US Filter M SERIES - 2000 Operation and Maintenance Manual

Type
Operation and Maintenance Manual
Operation and
Maintenance
Manual
for
M SERIES - 2000
Reverse Osmosis Systems
For Dialysis Applications
Serial # ________________________
8/97 MANOWNROSM REV F-
- i -
CAUTION The Food and Drug Administration has found this device (System) to be
substantially equivalent to devices (Systems) intended for use in specific
medical device applications. The System may not be modified,
reconfigured, or used in a medical application other than that for which it
is labeled. Modification of this system is a violation of federal law and will
void all warranties.
INDICATIONS
Use of the US Filter Reverse Osmosis System is indicated when water for use in the preparation of
dialysate fails to consistently meet the minimum requirements specified in the ANSI/AAMI approved
American National Standard for Hemodialysis Systems, RD5, 1981. The US Filter Reverse
Osmosis System may be used with any proportioning dialysis machine having a compatible water
demand.
CONTRAINDICATIONS
There are no absolute contraindications to reverse osmosis water treatment.
CAUTIONS AND WARNINGS
Procedures using the Reverse Osmosis System must be performed under the direction of a
physician.
Prior to operating or servicing this device, this manual must be read and understood. If something
is not clear, call for assistance before proceeding. Keep this and other associated manuals for
future reference and for new operators or qualified service personnel.
All electrical work should be performed by a qualified electrician in accordance with the latest
edition of the National Electrical Code, as well as local codes and regulations.
To avoid electrical shock hazard, do not remove covers or panels when power is supplied to the
device. Do not operate the device when covers or panels are removed.
WARNING A faulty pump motor or wiring can be a serious shock hazard if it or
surrounding water are accessible to human contact. To avoid this danger,
DO NOT remove any grounding wire from the system.
DISCLAIMER STATEMENT
This operation and maintenance manual is intended to be used with the supplier information
provided in the Appendix. These manuals should provide complete and accurate information to
meet your operating and/or service requirements based on the information available at the time of
publication. However, US Filter Corporation (“US Filter”) assumes no responsibility for the
technical content of the supplier literature.
This manual should be read fully and understood before installation, operation or maintenance of
the system is attempted. The information in this manual may not cover all operating details or
variations or provide for all conditions in connection with installation, operation and maintenance.
Should questions arise which are not answered specifically in this manual, contact your local
Culligan Dealer.
US Filter reserves the right to make engineering refinements that may not be reflected in these
manuals. The material in these manuals is for informational purposes and is subject to change
without notice and should not be construed as a commitment by US Filter. US Filter assumes no
responsibility for any errors that may appear in this document. This manual is believed to be
complete and accurate at the time of publication. In no event shall US Filter be liable for incidental
or consequential damages in connection with or arising from the use of this manual.
Copyright 1999 United States Filter Corporation
Printed in the United States of America. All rights reserved.
THIS BOOK OR PARTS THEREOF MAY NOT BE REPRODUCED IN ANY FORM
WITHOUT THE WRITTEN PERMISSION OF THE PUBLISHERS.
PROPRIETARY RIGHTS STATEMENT
This manual discloses information in which US Filter has proprietary rights. Neither receipt nor
possession of this manual confers or transfers any right to the client, and by its retention hereof,
the client acknowledges that it will not reproduce or cause to be reproduced, in whole or in part, any
such information except by written permission from US Filter. The client shall have the right to use
and disclose to its employees the information contained herein for the purpose of operating and
maintaining the US Filter equipment, and for no other purpose.
In the event the content of this manual is altered or section/items are omitted during a reproduction,
in whole or in part, and instructions or definitions within the reproduction result in personal injury to
those who follow the altered instructions, the burden of responsibility for personal injury falls solely
on the party who affects the reproduction.
MANUAL USER’S GUIDE
This manual describes the procedures necessary to install, operate, and maintain your US Filter
liquid treatment system. Please read this manual carefully before installing and operating your
equipment. The equipment warranty may be voided if installation or operation instructions are not
followed correctly.
- iii -
Warnings, Cautions, and Notes are used to attract attention to essential or critical information.
Warnings and Cautions will appear before the text associated with them, and notes can appear
either before or after associated text.
WARNING Warnings indicate condition, practices, or procedures which must be
observed to avoid personal injury or fatalities.
CAUTION Cautions indicate a situation that may cause damage or destruction of
equipment or may pose a long term health hazard.
NOTE: Notes are used to add information, state exceptions, and point out areas that
may be of greater interest or importance.
EQUIPMENT SUPPORT
US Filter continually strives to provide safe, efficient, trouble-free equipment using the optimum
technology for your application. If problems should develop, contact your local Culligan dealer.
MANOWNROSM INTRODUCTION
US Filter 1
1.0 INTRODUCTION
1.1 System Function...................................................................................................1-1
1.2 Quality Assurance/Quality Control......................................................................1-1
1.3 Functions of the System Components..................................................................1-2
2.0 SYSTEM INSTALLATION
2.1 Precautions Before Installation............................................................................2-1
2.2 Unpacking............................................................................................................2-1
2.3 System Location...................................................................................................2-1
2.4 Power Requirements............................................................................................2-1
2.5 Water Requirements.............................................................................................2-2
2.6 Permeate Line ......................................................................................................2-2
2.7 Accessory Sensors (Pretreatment Interlock and Tank Level)..............................2-2
2.8 Setting the Flush Cycle Time (Optional).............................................................2-3
2.9 Before Power Turn On.........................................................................................2-3
2.10 Installation of Reverse Osmosis Membrane(s)....................................................2-3
3.0 OPERATING INSTRUCTIONS
3.1 Controls and Indicators........................................................................................3-1
3.2 Initial Startup of Reverse Osmosis System..........................................................3-1
3.3 Initial Flushing of the System..............................................................................3-2
3.4 Normal Operations...............................................................................................3-2
3.5 Shutdown (Less Than 4 Days).............................................................................3-3
3.6 Prolonged Shutdown (Greater Than 4 Days).......................................................3-3
4.0 SYSTEM MAINTENANCE
4.1 General.................................................................................................................4-1
4.2 Sanitizing the Reverse Osmosis Membrane(s) ....................................................4-1
4.3 Maintenance Summary Log.................................................................................4-10
INTRODUCTION MANOWNROSM
US Filter
2
5.0 TROUBLESHOOTING
5.1 System Inoperative...............................................................................................5-1
5.2 Low System Pressure...........................................................................................5-1
5.3 Low Percent Recovery.........................................................................................5-1
5.4 High Percent Recovery ........................................................................................5-1
5.5 Low Salt Rejection...............................................................................................5-2
5.6 Low Calcium Rejection .......................................................................................5-2
5.7 System Keeps Shutting Off..................................................................................5-2
5.8 System Shutoff, Low Pressure.............................................................................5-2
5.9 System Shutoff, High Temperature......................................................................5-3
5.10 Low Permeate Flow Rate.....................................................................................5-3
5.11 Motor Overload Shutdown ..................................................................................5-3
5.12 Controller Malfunction ........................................................................................5-3
6.0 warranty
6.1 General Limited Warranty...................................................................................6-1
6.2 Water System Limited Warranty..........................................................................6-1
7.0 appendix...........................................................................................................................7-1
TABLE # DESCRIPTION PAGE
4-1 Sanitization Times and Concentrations for Polyamide.................................................4-2
Thin Film Composite Cartridges
4-2 Cleaning Solutions for Polyamide Thin Film................................................................4-8
Composite Membranes
4-3 Storage Solutions for Polyamide Thin Film Composite Cartridges 4-10
MANOWNROSM INTRODUCTION
US Filter 3
1. 0 INTRODUCTION
Congratulations on your selection of this US Filter Corporation water purification system. This
system is designed and manufactured to the highest standard of quality and fully tested and
inspected by competent personnel. In order to get maximum performance, we ask you to read
all of the following instructions before installing and operating this system. Any warranty offered
will be void unless directions are followed exactly.
This manual is divided into sections for easy reference. Review this manual thoroughly and
then return to section 2.0 System Installation for step-by-step installation instructions.
If you need technical assistance in operating or maintaining your unit, call your local US Filter
representative or should you require further assistance call US Filter Technical Support at the
phone number provided on the cover of this manual.
1.1 SYSTEM FUNCTION
The function of a Reverse Osmosis (RO) system is to separate contaminants from a
water supply using a Reverse Osmosis (RO) membrane. The system produces purified
water that is low in inorganic salts, organic matter and bacteria. The purified water can
be used as a direct feed to a distribution system or stored in a reservoir.
RO systems operate on the principle of passing water through the membrane under
high pressure. Most of the water passes through the membrane and is purified. A
portion of the water containing the impurities is not passed through the membrane and
is rejected to drain. Prefilters are used in many systems to reduce the amount of
particulate contamination reaching the membrane thus insuring the membrane does not
become clogged.
Please refer to general membrane specifications and System Specifications Sheet in the
Appendix for other parameters important to the operation of the system.
1.2 QUALITY ASSURANCE/QUALITY CONTROL
Quality Assurance and Quality Control are words that mean many things to many
people. All too often they are used as "catch" phrases that sound good but mean little.
At US Filter we have defined these phrases into a series of engineering and
manufacturing practices and control procedures to insure that you, our customer,
receive the finest product available in the world. We will continue to update these
practices and procedures in an effort to improve our systems and to insure that they
continue to reflect state-of-the-art technology and reliability.
Quality assurance is the cornerstone of our quality program. It begins with the selection
and qualification of components and materials. Before suppliers are selected, their
products must meet or exceed rigid standards set by our Quality Assurance
Department. From that point, our incoming material inspection insures only reliable
components will be used to manufacture systems. Numerous checks are performed as
the system passes through the manufacturing process. The completed system must
pass a series of tests before release to insure the system meets technical and
performance specifications. In the Appendix is a copy of our Quality Control form for
your system.
INTRODUCTION MANOWNROSM
US Filter
4
Quality audits do not improve products. They do, however, insure that a system meets
or exceeds specifications required for your system.
Customer feedback is also an important element of our program. Division heads meet
on a routine basis to review customer feedback and initiate actions that result in process
and product improvements.
1.3 FUNCTIONS OF THE SYSTEM COMPONENTS
A. Low Suction Pressure Switch
A low suction pressure switch is incorporated into the system for the protection
of the high pressure pump. If the inlet pressure on the pump suction falls below
0.69 bar (10 psig) for longer than three seconds the switch activates a safety
shutoff in the controller.
B. Reverse Osmosis Pump
The RO system contains a submersible pump driven by an AC motor. The pump
motor is turned on and off by a motor starter relay. The pump motor starter also
contains an overload relay.
C. High Pump Pressure Switch
A high pump pressure switch is utilized in the system to prevent damaging
pressures from reaching the RO membranes. This switch immediately shuts off
the RO pump if the pressure exceeds safety limits.
D. Pump Throttle Valve
A pump throttle valve is used to control the pressure and volume of water
supplied to the RO membranes.
E. Reverse Osmosis Membranes
The prefiltered water enters the RO housing and passes over the surface of the
RO membrane. A portion of the feedwater continues out of the system through
a reject needle valve. The remainder of the water permeates the membrane and
is directed to a distribution pipeline or to an accessory storage reservoir. Percent
conversion (the ratio of RO product water to feedwater) will vary depending on
water quality and the application.
F. Reject Throttle Valve
The water exiting the RO membranes is controlled by a reject throttle valve.
Closing the valve increases the pressure to the membranes and reduces the
amount of water to drain.
G. Reject Recirculation (Optional)
If the system is provided with the reject recirculation option, it permits part of the
reject water to be returned to the inlet of the high pressure pump. This may
allow the conservation of water depending on the water chemistry.
H. Product Recirculation
MANOWNROSM INTRODUCTION
US Filter 5
Product recirculation permits excess product water to be returned to the inlet of
the high pressure pump. This function is used when capacity exceeds demand,
conserving water and increasing percent rejection (water quality).
I. Autoflush (Optional)
In the TANK FEED and STANDBY modes, an optional solid state timer
automatically opens the reject solenoid valve for one minute every hour,
bypassing the throttle valve. This flush cycle is programmable for 1, 2, or 4
minutes per hour as determined by the fouling characteristics of the feedwater.
In the DIRECT FEED mode the autoflush function is inhibited.
J. Controller
The controller has the following capabilities:
1. Low pressure shutdown - The pump motor is shut down when the inlet
water pressure falls below 0.69 bar (10 psi). There is a 3 second delay to
prevent shutdown during momentary pressure losses.
2. High pressure shutdown - The pump motor is shut down when the
pressure exceeds a set point (determined by the choice of membranes)
to prevent damage to the pump or elements.
3. High temperature shutdown - The pump motor is shut down on high
temperature to protect the RO membranes from exposure to damaging
temperatures.
4. Low percent rejection alarm - The alarm function is activated when the
percent rejection falls below determined limits to warn the operator if the
product water quality no longer meets specifications.
5. Pump startup delay - There is a ten second delay after turning on the unit
before the pump starts.
6. Automatic flush - The system automatically flushes each hour when the
system is in the STANDBY mode or the TANK FEED mode.
7. Tank full shutdown - The system goes into STANDBY when a signal from
a level switch indicates that a storage tank is full (optional).
8. Interlock - This feature stops the system when the pretreatment
equipment is in the regeneration cycle and restarts the system
automatically when the regeneration cycle is completed.
K. Accessories
In order to comply with AAMI standards for audible alarm capability, an
accessory audible alarm is necessary. This alarm can be remotely mounted
providing monitoring in an area away from the unit for safety and convenience. If
the system is used to feed a storage reservoir, a level switch is necessary to
provide a signal to the controller while in the TANK FEED mode.
L. Water Quality Monitor
The controller is designed to measure the conductivity of the feedwater to the
system as well as the conductivity of the product water. It automatically
calculates the percent of impurities being rejected by the RO membranes. The
INTRODUCTION MANOWNROSM
US Filter
6
readings are on an LCD display. A more detailed description of this module can
be found in the Reverse Osmosis Controller, Operating Instructions in the
Appendix of this manual.
M. Alarms
The controller is equipped with a visual alarm for the following conditions:
Low Feed Pressure
High Pressure to Membranes
High Feed Temperature
Tank Full
Pretreatment Interlock
Low Percent Rejection
As these units are normally installed in a location away from patient areas for
noise containment, an accessory audible alarm is necessary to comply with
AAMI standards. The controller generates a signal to drive this alarm for the
following conditions:
Low Feed Pressure
High Pressure to Membranes
High Feed Temperature
Low Percent Rejection
MANOWNROSM SYSTEM INSTALLATION
US Filter
7
2. 0 SYSTEM INSTALLATION
2.1 PRECAUTIONS BEFORE INSTALLATION
The following are feedwater source requirements:
A. Inlet pressure of 0.69 to 6.9 bar (10 to 100 psig) at design feed flow.
B. Temperature of between 2°C (35°F) and 40°C (104°F). Recommended
operating temperature is 25°C (77°F).
C. Free chlorine concentration must be non-detectable on a continuous basis (<0.1
mg/l).
If the free chlorine level is greater than 0.1 mg/l, a carbon filter or bisulfite system
should be used as pretreatment to remove chlorine from the feedwater entering
the RO unit. Chlorine will cause irreversible damage to the membranes.
D. Langelier Saturation Index (LSI) of the reject must be negative at the system's
operating recovery to avoid fouling the membranes.
CAUTION pH must be between 4 and 11. Chlorine must be less than 0.1 mg/l.
NOTE: If feedwater does not meet the above requirements or there is some doubt,
contact your US Filter representative for recommendations.
2.2 UNPACKING
The RO systems are shipped completely assembled. Pressure switches, gauges,
control box, pump, valves and regulators are all pre-mounted on the main frame and
factory set to correct values. Carefully remove from the packing crate to avoid damage
to gauges, RO membranes, etc. Do not drop or handle with excessive roughness to
prevent damage to the electronic components and to prevent damage to the piping
joints.
2.3 SYSTEM LOCATION
A drain capable of handling the flow of reject water during system flushing should be
located as close to the system as possible. The reject line should be secured so that
during the high velocity flush the line is not dislodged from the drain.
2.4 POWER REQUIREMENTS
The power requirements for the system are given on the system I.D. plate just above
the controller enclosure. The system should be connected with a multiconductor,
grounded cable to a disconnect switch box. The switch box should be located within
approximately 10 feet of the system for the convenience of technicians servicing the
unit.
The system is provided with a motor starter to permit coordinated over-current
protection as defined by the National Electrical Code.
Ground the power cable to the ground connection in the motor starter housing box.
Branch circuit conductors supplying the motor shall have an amperage capacity of not
SYSTEM INSTALLATION MANOWNROSM
US Filter
8
less than 125% of the motor full load current rating as defined by the National Electric
Code.
2.5 WATER REQUIREMENTS
The inlet water supply must be adequate to provide the minimum flow requirement
shown on the System Specifications sheet in the Appendix. The inlet line should be
sized to deliver this flow with a minimum pressure of 0.69 bar (10 psig) at the RO unit.
A shutoff valve within 10 feet of the unit and a cleaning connection should be provided.
A sample valve to check feedwater quality is also recommended.
2.6 PERMEATE LINE
The permeate line should have sufficient internal diameter to minimize line pressure
drop. It is recommended that a small sample valve be placed in the product line to
facilitate sampling of the permeate water quality. A flow meter is supplied with the unit
to continuously measure the permeate flow rate. The flow meters supplied are fluid
calibrated to obtain accurate flow measurements. Any replacement flow meter should
also be fluid calibrated.
2.7 ACCESSORY SENSORS (PRETREATMENT INTERLOCK AND TANK LEVEL)
Provisions have been made for shutdown by accessory sensors, provided such sensors
are equivalent to a single pole single throw switch. The shutdown is such that normal
operation will resume automatically when the failure is removed.
A. Remove the lower panel from the control box and the attached integrated
controller circuit to expose the terminal connections.
B. See the Reverse Osmosis Controller, Operating Instructions in the Appendix to
connect PRETREATMENT INTERLOCK, HIGH TANK LEVEL and MID TANK
LEVEL sensors.
NOTE: If only a high tank level switch is used, a switch should be selected with a built
in “dead band” in order to prevent the RO pump from cycling. Using a switch
without a “dead band” could cause the motor starter to fail and can contribute
to low product water quality.
C. These terminals are normally closed. An open circuit is necessary to shut down
the system.
D. To connect the accessory, remove the shorting wire and connect a single pole
single throw normally closed sensor between the respective terminals. It is
recommended that shielded cable, properly grounded, be used to avoid radio
signal interference with these functions.
2.8 SETTING THE FLUSH CYCLE TIME (OPTIONAL)
See the see the Reverse Osmosis Controller, Operating Instructions in the Appendix.
MANOWNROSM SYSTEM INSTALLATION
US Filter
9
2.9 BEFORE POWER TURN ON
The controller should be installed in the controller housing before connecting electrical
power to the unit.
WARNING Personal injury and permanent damage to the controller can result if the
controller is installed with power on the unit.
Once the inlet water is plumbed to the system and the electrical power is connected, the
electrician should verify correct motor rotation. Please refer to the Pump Information
included in the Appendix.
CAUTION Verify that all retaining rings are in place on the RO membrane housings
before turning on the pump.
CAUTION Be sure the product water will not enter the user system during startup.
The RO system must be operated for four hours to flush out the
preservative solution before connecting to the user system.
Place the system in the DIRECT FEED mode and turn the POWER switch to ON.
2.10 INSTALLATION OF REVERSE OSMOSIS MEMBRANE(S)
RO units are normally shipped with the membranes installed and preserved in sodium
bisulfite solution. However, should the need arise, the following procedures may be
followed to install membranes:
A. See the Appendix for a diagram of the individual components of the membrane
assembly. Before sliding the membranes into the fiberglass housing, install the
proper product tube adapters on/in the product tube. Lightly lubricate the o-rings
with glycerin.
B. Load the RO membrane into the fiberglass housing from the direction in which
the feedwater will enter, inserting the end without the chevron seal first. Before
the chevron seal enters the housing lightly lubricate the chevron seal with
glycerin before it enters the housing and press the membrane to the end of the
housing.
C. Place an end plug into the RO housing end opposite from loading end. Using
snap ring pliers secure the end plug with the retaining ring. Press the membrane
into the end plug to seat. Hoses should be connected to end cap before
insertion into the housing.
WARNING Never engage or disengage retaining rings without wearing safety
glasses.
D. Install the adapter and end plug on the feed end and secure it with a snap ring.
CAUTION Never allow the RO membrane to dry out once it has been wet, or
irreversible damage will result.
MANOWNROSM OPERATING INSTRUCTIONS
US Filter
11
3. 0 OPERATING INSTRUCTIONS
3.1 CONTROLS AND INDICATORS
The controls and indicators used in the operation of the system are outlined in the
Appendix (Figure 1.1). The operator must be thoroughly familiar with the controls and
their respective functions for proper operation and in order to maintain the quality of the
permeate and eliminate unnecessary shutdowns.
3.2 INITIAL STARTUP OF REVERSE OSMOSIS SYSTEM
A. The operator should verify that appropriate pretreatment is in place and
operating satisfactorily. Inappropriate or inoperative pretreatment can damage
the membranes and result in water which does not meet AAMI recommended
standards.
B. Make all water, electrical, and drain connections according to installation
instructions.
C. Verify that the Key switch is in the OFF position (see the Reverse Osmosis
Controller, Operating Instructions in the Appendix). Turn on the inlet water and
the electrical power at the remote disconnect. Direct product water to the drain
until proper operation of the system is verified and product quality is known to be
within specifications.
CAUTION Verify that all RO housing retaining rings are in place and engaged.
D. Turn on the Key switch and press the power button. Place the controller in the
Direct Feed Mode.
E. Check for proper pump rotation. Slowly close the pump throttle valve (Figure
1.1) and record the pressure indicated on the PUMP PRESSURE gauge
upstream of the valve. Immediately turn the system off. DISCONNECT THE
POWER and switch any two electrical leads in the pump starter. Turn the
system on and again close the pump throttle valve and record the pump
discharge pressure. Do not run the system for more than a few seconds
with the pump discharge valve closed. The lead hookup that gave the
highest discharge pressure indicates the correct rotation. (Please refer to the
Pump Installation and Operating Instructions, included in the Appendix of this
manual for information concerning motor rotation and any other electrical
requirements.)
CAUTION Do not run the motor for more than 30 seconds backwards or irreversible
pump damage may result. The membrane pressure should not be
allowed to exceed 24.15 bar (350 psig) during the procedure.
F. Adjust the reject throttle valve (Figure 1.1 in the Appendix) to give the design
product flow indicated on the System Specifications sheet in the Appendix. DO
NOT exceed 24.15 bar (350 psig) membrane pressure.
OPERATING INSTRUCTIONS MANOWNROSM
US Filter
12
G. If your RO system is equipped with the reject recirculation option, set the reject
recirculation flow to zero, but DO NOT exceed pressure to membrane limits
(24.15 bar/350 psig). Open the reject valve further if necessary to avoid over
pressuring the system.
CAUTION Do not use reject recirculation option for the first 24 hours of operation
to avoid recirculating preservative chemicals which can permanently
damage the membranes.
H. Set the percent rejection alarm setting to 90%.
I. Check for plumbing leaks and make appropriate repairs as necessary. All
systems are leak checked at the factory but vibration during transit can cause
new leaks to develop.
J. Once the system is checked out and operating correctly, place the controller in
the desired mode. Read the next section on initial flushing before directing
product water to the RO water storage tank or user system.
3.3 INITIAL FLUSHING OF THE SYSTEM
Since the RO membrane is shipped moist with sodium bisulfite preservative solution,
initial flushing of the system is required. Before operating a new system, and also after
the replacement of a RO membrane, perform the following steps:
A. Place the outlet of the permeate at a convenient drain.
B. Turn on the feedwater and the system pump (as outlined in Operating
Instructions, Initial Startup of the Reverse Osmosis System section of this
manual) and let the system run for 24 hours. This will flush residual bisulfite
solution and any particulate impurities from the membranes, housings and
plumbing of the system.
C. Permeate quality will increase slowly for 1 hour. Monitor the inlet quality vs. the
permeate quality by observing the conductivity meter outputs.
NOTE: The system should have better than 90% rejection within 2 hours as the
membranes equilibrate. If the permeate quality does not improve, refer to
section 5.0 Troubleshooting.
D. After flushing, turn off the system and reconnect the permeate line. The system
is now ready to be put into a normal operating mode.
3.4 NORMAL OPERATIONS
The following procedure is for operating a U.S. Filter RO system that has been installed
and initial flushing has been completed. Refer to the Reverse Osmosis Controller,
Operating Instructions in the Appendix for control location and identification.
A. Initiate operation by pressing the POWER button.
B. If the controller is in the STANDBY or TANK FEED mode, allow the system to
complete the flush cycle before proceeding.
MANOWNROSM OPERATING INSTRUCTIONS
US Filter
13
C. Place the controller in the DIRECT FEED mode.
D. After five minutes of operation to allow the system to stabilize, verify and record
the operating data in the daily log sheet (sample in Appendix). Follow the
procedures outlined in the Maintenance Summary section of this manual.
E. Set the percent rejection alarm to the limits prescribed by the physician. The
alarm limits for the purposes of monitoring shall be calculated based on the
feedwater analysis and initial rejection characteristics and shall correspond to the
highest rejection coefficient at which contaminants reach unsafe limits. Consult
your US Filter representative for the determination of these limits. Under no
circumstances is it recommended that the alarm be set at less than 90%.
F. After performing checks and recording the results reset the MODE SELECTION
SWITCH to the desired location if necessary. If a storage tank is utilized, set the
MODE SELECTOR SWITCH to the TANK FEED position. If no storage tank is
used, set the MODE SELECTION SWITCH to the DIRECT FEED position. The
system is now ready to begin dialysis.
3.5 SHUTDOWN (LESS THAN 4 DAYS)
If the system is not to be run for up to 4 days, the system should be placed in the
STANDBY mode of operation. This will allow the system to return to operation at any
time with no adverse affects on the permeate or membranes.
CAUTION If the shutdown is due to a malfunction, be sure to disconnect the
electrical power and turn off the water supply before making repairs or
replacements.
3.6 PROLONGED SHUTDOWN (GREATER THAN 4 DAYS)
When the system is going to be inactive for more than 4 days (not in a STANDBY
mode), perform the following steps:
A. Turn off the feedwater source.
Bisulfite Sanitization (4 days to 1 month shutdown)
Drain water from the system and pump in a 0.5% sodium bisulfite solution
to prevent biological growth in the elements.
or
Formaldehyde Sanitization (shutdowns exceeding 1 month)
Drain water from the system and pump in a solution of 3% formaldehyde
to prevent biological growth and kill residual bacteria.
B. Remove the prefilter cartridges and discard. Clean the prefilter housing, leaving
the cleaned components to dry during the shutdown period.
C. Turn off the feedwater and electrical power until the system is to be placed into
operation again.
D. To resume system operation, first drain the solution from the RO housings,
install new prefilter cartridges, disconnect the permeate line from the reservoir
OPERATING INSTRUCTIONS MANOWNROSM
US Filter
14
and run to drain. Then turn on the feedwater and start the system, run the
permeate water to drain for 4 hours to flush the system. The product water
should be analyzed with the Fast Formalert or other formaldehyde test kit until
detection limits are reached. Reattach the permeate line and start the system as
outlined in the Operating Instructions, Normal Operations section of this manual.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39

US Filter M SERIES - 2000 Operation and Maintenance Manual

Type
Operation and Maintenance Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI