NUVE FN 400 User manual

Type
User manual

This manual is also suitable for

NUVE FN 400 P is a technologically advanced dry air sterilizer and oven designed to meet the demanding requirements of laboratories and research facilities. With its precise temperature control, air forced ventilation, and programmable microprocessor, the FN 400 P offers exceptional performance for a wide range of applications, including sterilization, drying, and heating.

The FN 400 P features a spacious 44-liter useful volume, allowing you to accommodate larger samples or multiple items simultaneously. Its electro-acid coated aluminum interior and stainless steel exterior ensure durability and resistance to corrosion, making it suitable for even the most challenging environments.

NUVE FN 400 P is a technologically advanced dry air sterilizer and oven designed to meet the demanding requirements of laboratories and research facilities. With its precise temperature control, air forced ventilation, and programmable microprocessor, the FN 400 P offers exceptional performance for a wide range of applications, including sterilization, drying, and heating.

The FN 400 P features a spacious 44-liter useful volume, allowing you to accommodate larger samples or multiple items simultaneously. Its electro-acid coated aluminum interior and stainless steel exterior ensure durability and resistance to corrosion, making it suitable for even the most challenging environments.

1
NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET
A.Ş.
FN 300 - FN 400 - FN 500
FN 400 P – FN 500 P
DRY AIR STERILIZERS/OVENS
USER’S MANUAL
NÜVE SANAYİ MALZEMELERİ İMALAT VE TİCARET A.Ş.
Z14.K 25 002 Rev. No:01 Rev. Date: 09/2002
2
NÜVE SANAYI MALZEMELERI
EUROPEAN REPRESENTATIVE:
İMALAT VE TICARET A.Ş.
Esenboğa Yolu 22 km Mr. Afxentios Pimenidis
Akyurt - ANKARA / TURKEY Karpou 19
Tel : (00 90 312) 399 28 30 ( 3 lines ) 116 31 – N. Kosmos GREECE
Fax : (00 90 312) 399 21 97 Tel: 00 30 10 9028051-4
e-mail: nuveex[email protected] Fax: 00 30 10 9029855
WARRANTY CERTIFICATE
1. Nüve warrants that the equipment delivered is free from defects in material and
workmanship. This warranty is provided for a period of one year. The warranty period
starts from the delivery date.
2. Warranty does not apply to parts normally consumed in operation, current
maintenance and the adjustments described in the operating instructions provided
with the equipment.
3. Nüve does not accept any liability in the case where the goods are not used in
accordance with their proper intent.
4. The warranty may not be claimed for damages occurred during the shipment, for
damages resulting from improper handling or use, the defects in maintenance,
negligence, bad functioning of auxiliary equipment, in the case of force majeure or
accident and incorrect power supply.
5. In the event of failure, Nüve shall be under no liability for any injury, or any loss or
damage as the result of the failure other than the guarantee conditions.
BEFORE OPERATING THE INSTRUMENT THIS MANUAL SHOULD BE READ
CAREFULLY.
INFORMATION CONTAINED IN THIS DOCUMENT IS THE PROPERTY OF NÜVE. IT
MAY NOT BE DUPLICATED OR DISTRIBUTED WITHOUT HIS PERMISSION.
THE VALIDITY OF THE GUARANTEE IS SUBJECT TO THE OBSERVATION OF THE
INSTRUCTIONS AND PRECAUTIONS DESCRIBED IN THIS MANUAL.
3
CONTENTS
Page
1. USE AND FUNCTION
4
2. TECHNICAL SPECIFICATIONS
4
2.1. Technical Specifications Table 4
2.2. General View 5
3. INSTALLATION PROCEDURE
3.1. Lifting and Transport 6
3.2. Unpacking 6
3.3. Environmental Conditions
3.4. Mains Supply
3.5. Positioning
3.6. Before Starting the Operation
6
6
6
7
4. OPERATING THE DRY AIR STERILIZER
4.1. Switching On
4.2. Control Thermostat
4.3. Programming
4.4. End of the Program
5. PERIODICAL MAINTENANCE AND CLEANING
5.1. Periodical Maintenance
5.2. Cleaning
6. TROUBLESHOOTING
7
8
9
9
10
10
10
7. ELECTRICAL CIRCUIT DIAGRAMS
7.1. FN 300 11
7.2. FN 400
7.3. FN 500
7.4. FN 400 P
7.5. FN 500 P
12
13
14
15
4
1. USE AND FUNCTION
The FN series dry air sterilizers/ovens are designed to sterilize, heat and dry samples in
laboratories. They offer excellent sterilization, drying and heating conditions.
They maintain temperatures between 70 °C and 250 °C and keep the temperature stable
within the given tolerances.
The FN series dry air sterilizers/ovens provide homogeneous temperature distribution by
means of the sheet heaters placed onto the outer surface of the useful volume.
The insulation of the dry air sterilizers/ovens not only provides economical working
conditions but also helps to achieve the excellent homogeneity.
The FN series dry air sterilizers/ovens provide precise operation conditions with the PID
microprocessor controlled thermostat and with the timer, they also ensure safe operation
conditions with the safety thermostat.
The FN series dry air sterilizer/ovens are manufactured according to the following
standards,
EN 61010-1, EN 60601-1, EN 60601-1-2, EN 1441, EN 980
Do not operate the instrument for purposes other than main purpose.
The instrument is only to be used by authorized people after the users manual has
been read carefully. Only technical personnel can handle the product in case of a
breakdown.
5
2. TECHNICAL SPECIFICATIONS
2.1. Technical Specifications Table
FN 300 FN 400 FN 400 P FN 500 FN 500 P
Temperature range 70°C / 250°C
Temperature sensor Fe-Const
Air forced ventilation Optional Optional Standard Optional Standard
Control system PID program mable microprocessor
Temperature set and 1°C
display sensitivity
Temperature variation ±2°C
(110°C-190°C)
Temperature fluctuation ±1°C
Timer 1-999 min. + Hold position
Number of shelves(max/min) 2/6 2/7 2/7 2/10 2/10
Safety thermostat Gas expansion thermostat (50°C- 300°C)
Power consumption 500 W 800 W 750 W 1600 W 1500 W
Power Supply 230 V, 50 Hz.
Useful volume 22 liters 44 liters 42 liters 120 liters 110 liters
Internal material Electro-acid Electro-acid Stainless Electro-acid Stainless
coated aluminum coated aluminum Steel coated aluminum Steel
External material Epoxy-polyester painted steel
Internal dim.s (wxdxh) mm 300x240x300 420x320x360 420x320x360 500x490x500 500x490x500
External dim.s (wxdxh) mm 555x380x460 705x470x540 705x470x540 790x630x670 790x630x670
Net/Packed weight 22/25 35/39 35/39 49/59 49/59
Packing dims.(wxdxh)mm 640x480x530 810x550x615 810x550x615 890x720x730 890x720x730
6
2.2. General View
1. Control panel 6. Shelf
2. Safety thermostat 7. Shelf carrier holder
3. On / Off switch 8. Shelf carrier
4. Supply inlet and the fuses 9. Ventilation hole clips
5. Chamber gasket
7
3. INSTALLATION PROCEDURE
3.1. Lifting And Transport
Because of the heavy weight of the dry air sterilizer/oven, all lifting and transport must be
carried out using proper handling equipment. The dry air sterilizer/oven must be
supported from underneath and never be turned over.
3.2. Unpacking
Unpack the wooden packing.
Remove the nylon packing of the dry air sterilizer/oven and of the shelves.
Provided with the unit are below,
! Users Manual
! Electrical cables
! 2 pieces of shelves
3.3. Environmental Conditions
Please pay special attention to the followings,
! Indoor use only
! Temperature from 5°C to 40 ° C
! Humidity level % 80 up to 22° C
! Maximum height 2000 m.
3.4. Mains Supply
The dry air sterilizer/oven requires 230 V ± 10%, 50 Hz.
Please make sure that the supplied mains matches the required power ratings. If no,
provide an extra line to support.
Always connect the dry air sterilizer/oven to properly earthed sockets.
A supply fitted with a circuit breaker should be used for protection against short
circuit in case of an isolation fault.
3.5. Positioning
Check that no damage has occurred during transport.
Lift the dry air sterilizer/oven underneath and carry it to its place carefully.
Balance the dry air sterilizer/oven on the four pedestals. If necessary, provide stable
standing by adjusting the pedestal heights.
Place the shelf carriers and then the shelves.
8
Check the followings,
! the proposed site is suitable for users,
! the operator can follow up the unit even he deals with something else.
! the dry air sterilizer/oven does not occupy the utilisation space of others or damage
them.
3.6. Before Starting The Operation
Attention to the followings,
Do not heat the samples which can liquefy or expand during the operation in sealed
containers.
Determine of the boiling points of the samples and set the temperature accordingly.
Make sure that the samples do not overflow from their containers when heated.
Do not operate the dry air sterilizer/oven at the temperatures which do harm to the
samples or cause them to deform.
Make sure that the ventilation hole is open to discharge the gas or vapour which
occurs during the operation.
Make sure that the vapour occurred during the operation is not harmful to humans or
flammable or explosive.
IMPORTANT! The safety thermostat set value should always be higher than the working
temperature.
4. OPERATING THE STERILIZER
4.1. Switching On
Push On/Off switch.
See that the microprocessor activates.
Read section 4.2. to learn the functions of the microprocessor.
Set the values ( see section 4.3) and start the operation.
9
4.2. Control Thermostat
1. Temperature Display: This display shows the chamber temperature during the
operation and the set temperature value during programming. The error codes are
also shown on this display.
2. Time Display: This display shows the elapsed time during the operation and the set
value during programming.
3. Alarm Mute Key: This key is pushed to interrupt the alarm which goes off when the
program ends and if any failure occurs during the operation.
4. Start / Stop Key: This is key which is pushed to start the program or to stop the
running program.
5. Temperature Set Key: This key is pushed to set the temperature.
6. Temperature Value Increase/Decrease Keys: These key are pushed to increase or
decrease the values on the displays.
7. Time Set Key: This key is pushed to set the time. (01-999 minutes and Hold position)
8. Time Value Increase/Decrease Keys: These key are pushed to increase or
decrease the values on the displays.
9. Heat Led: This led indicates that the chamber is heated.
10
10. Alarm Led: It lights up when the program ends and if any failure occurs during the
operation.
4.3.Programming
Push the temperature set key.
Set the temperature by pushing the value
increase/decrease keys.
Push the temperature set key again to save the
temperature value.
Push the time set key.
Set the time values by pushing the value
increase/decrease keys (01 min. to 999 min. or
Hold)
Push the time set key again to save the value.
Push the start/stop key to start the program.
4.4. End of the Program
See that the program is over.
Take the samples out.
Wipe the chamber surface to clean if needed.
You may leave the dry air sterilizer / oven on stand-by position or switch it off.
11
5. CLEANING AND MAINTENANCE
5.1. Periodical Maintenance
There is no periodical maintenance.
5.2. Cleaning
Make all cleaning after you have disconnected electrical cable from the dry air
sterilizer/oven and at room temperature.
You may use a piece of damp cloth against dust and dirt.
You may use mild liquid soap against difficult dust and dirt.
Protect the dry air sterilizer/oven against rust. Please remove possible small rust
before it affects the whole surface.
Please be aware of the undesirable effects of the chemicals and be careful while
applying them.
6. TROUBLESHOOTING
If the dry air sterilizer does not operate, check the following
it is switched on,
it is plugged in properly or connected properly to the electric supply,
the plug is not defective,
the installation of the supply is not defective,
there is mains supply,
The fuses are sound.
Errors Recognised By The Microprocessor Control System
OFL: 1. The chamber temperature exceeds 255°C.
2. The temperature sensor(Fe-Const) endings are broken.
Switch off the dry air sterilizer/oven and call for an technical help.
EOF: This error code is displayed when power is cut and the temperature in the
chamber goes out of the ±2 °C alarm limit.
The sterilization operation is interrupted. You may re-start the dry air
sterilizer/oven by pushing the start/stop key.
PLEASE CONTACT TO AN AUTHORIZED NUVE AGENT TO SEEK TECHNICAL
HELP IF AN ERROR OCCURS.
12
7. ELECTRICAL CIRCUIT DIAGRAMS
7.1. FN 300
13
7.2. FN 400
14
7.3. FN 500
15
7.4. FN 400 P
16
7.5. FN 500 P
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16

NUVE FN 400 User manual

Type
User manual
This manual is also suitable for

NUVE FN 400 P is a technologically advanced dry air sterilizer and oven designed to meet the demanding requirements of laboratories and research facilities. With its precise temperature control, air forced ventilation, and programmable microprocessor, the FN 400 P offers exceptional performance for a wide range of applications, including sterilization, drying, and heating.

The FN 400 P features a spacious 44-liter useful volume, allowing you to accommodate larger samples or multiple items simultaneously. Its electro-acid coated aluminum interior and stainless steel exterior ensure durability and resistance to corrosion, making it suitable for even the most challenging environments.