SECTION 4—OPERATING INSTRUCTIONS
Pneumatic Oxygen Conserver 20 Part No 1160840
WARNING
NOOXYGENisdeliveredinbetweensettings.Toobtain
desiredflow,theindicatingpointerMUSTalignwitha
specificnumberonthedial.
Toavoidinjurytopatient,ALWAYSconfirmprescribed
settingbeforeadministeringtopatientandmonitorflowona
frequentbasis.
10. Turntheselectorswitchuntilindicatingpointerisaligned
withtheprescribedsetting.FIGURE4.2onpage21.
NOTE:Thecontinuous flowsetting(CF)isdesignedforemergency
useonly.Theamountofoxygendeliveredwhenusingtheoxygen
conservingregulatorincontinuousflowmodeispresetto2L/min.
(litersperminute).Rememberthatinthismodetheoxygenwillbe
consumedatamuchfasterrate.Returntoanothersourcebefore
runningoutofoxygen.FIGURE4.2onpage21.
WARNING
Whenpointeroncontentsgaugeentersredzone,itis
recommendedtoreplaceorrefillthecylinder.
11. Breathethroughthenoseandfeelapulseofoxygenasyou
inhale.
NOTE:Whenonapulsesetting,thereisfloworapulseonlyatthe
beginningofeachbreath.Ifyoudonotfeelthepulseatthebeginning
ofeachbreath,checkthesetting.Ifthereisstillnopulse,turndialto
2L/min.continuoussetting.FIGURE4.2onpage21.
NOTE:Becauseeachpatient’sbreathingpatternisdifferentandthe
environmentvaries,itmaybedifficulttofeelsomelowsetting
pulses.