TRACE™GC Ultra Quick Reference
0 cm
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
ECD 109 mm
FPD 127 mm
PID Column 133 mm
PID Exit 144 mm
FID/NPD 94/97 mm
SPLITLESS 64 mm
SPLIT 40 mm
PTV 30 mm
COLUMN
INSERTION
DEPTHS
PDD 123 mm
NPD FID
This Quick Reference card shows the proper syringe needle lengths, column insertion depths, and
injector liner types for PTV, split/splitless, and on-column injectors. Column insertion depths for FID,
NPD, FPD, PID, ECD and PDD detectors are illustrated adjacent to the ruler. Detector and injector
insertion depths are measured from the bottom of the ferrule.
PTV Injection (a) Splitless Injection (c)
Syringe Needle Length 51 mm Syringe Needle Length 80 mm
Column Insertion Depth 30 mm Column Insertion Depth 64 mm
Split Injection (b) On-Column Injection (d)
Syringe Needle Length 51 mm Syringe Needle Length 80 mm
Column Insertion Depth 40 mm
INJECTORS NOT DRAWN TO SCALE
SPLIT INJECTION (b)
Syringe needle
length (51 mm)
Length of column
inside injector from
bottom of ferrule
(40 mm)
Glass liner
(5 mm or 3 mm ID)
PTV INJECTION (a)
Syringe needle
length (51 mm)
Length of column
inside injector from
bottom of ferrule
(30 mm)
Glass liner
(2 mm ID)
SPLITLESS INJECTION (c)
Syringe needle
length (80 mm)
Length of column
inside injector from
bottom of ferrule
(64 mm)
Tapered glass liner
(5 mm or 3 mm ID)
ON-COLUMN INJECTION (d)
Syringe needle
length (80 mm)
Capillary Column
© 2007-2009 Thermo Fisher Scientific Inc. All rights reserved.
Published by Thermo Fisher Scientific S.p.A. Milan Italy
Printed in ITALY
PN 317 092 40 Rev. I, April 2009