Agilent Technologies G3180B User manual

Category
Water pumps
Type
User manual

Agilent Technologies G3180B is a Two-Way Splitter Kit with Makeup Gas designed to interface with capillary columns on an Agilent 6890 gas chromatograph (GC). It enables the division of column effluent between two detectors, which can operate at different pressures. The kit includes metal column ferrules to eliminate air leakage and a microfluidic plate for low dead volume connections. The constant pressure operation and the ability to calculate chromatographic parameters beforehand simplify the setup and optimization process.

Agilent Technologies G3180B is a Two-Way Splitter Kit with Makeup Gas designed to interface with capillary columns on an Agilent 6890 gas chromatograph (GC). It enables the division of column effluent between two detectors, which can operate at different pressures. The kit includes metal column ferrules to eliminate air leakage and a microfluidic plate for low dead volume connections. The constant pressure operation and the ability to calculate chromatographic parameters beforehand simplify the setup and optimization process.

Agilent Technologies
Agilent G3180B
Two-Way Splitter Kit
With Makeup Gas
Installation and Operation
Guide
2 Installation and Operation Guide
Notices
©Agilent Technologies, Inc. 2006
No part of this manual may be reproduced in
any form or by any means (including elec-
tronic storage and retrieval or translation
into a foreign language) without prior agree-
ment and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
G3180-90120
Supercedes G3180-90110
Edition
First edition, April 2006
Printed in USA
Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA
Warranty
The material contained in this
document is provided “as is,” and
is subject to being changed, with-
out notice, in future editions. Fur-
ther, to the maximum extent
permitted by applicable law,
Agilent disclaims all warranties,
either express or implied, with
regard to this manual and any
information contained herein,
including but not limited to the
implied warranties of merchant-
ability and fitness for a particular
purpose. Agilent shall not be
liable for errors or for incidental
or consequential damages in
connection with the furnishing,
use, or performance of this
document or of any information
contained herein. Should Agilent
and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the
separate agreement shall control.
Safety Notices
CAUTION
A CAUTION notice denotes a haz-
ard. It calls attention to an operat-
ing procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated condi-
tions are fully understood and
met.
Acknowledgement
Microsoft® is a U.S. registered trademark
of Microsoft Corporation.
Installation and Operation Guide 3
In this Guide. . .
This Installation and Operation Guide contains information for installing and
using an effluent splitter on an Agilent 6890 gas chromatograph (GC). The
G3180 splitter is intended for use with capillary columns and uses makeup gas
to maintain adequate flows throughout the system.
1 Introduction
This chapter describes how the splitter works, the GC and software
requirements of the system and the contents of the installation kit.
2 Hardware Installation
See this chapter for a detailed procedure for installing the splitter hardware
and connecting the makeup gas supply.
3 Splitter Configurations
The split ratio (how the column effluent divides between the two detectors) is
governed by two restrictors, which are lengths of deactivated fused silica
tubing. This chapter presents a set of precalculated “typical” configurations. If
desired, you can create a custom configuration to meet specific needs. The
chapter describes a set of software tools, included in the kit, to assist you in
designing such configurations. Finally, installation of the column and
restrictors is covered.
4Operation
This chapter contains a worked-out custom configuration, plus a few special
topics.
4 Installation and Operation Guide
Installation and Operation Guide 5
Contents
1 Introduction
Overview 8
How It Works 9
Details 10
Metal ferrules 10
Microfluidic plate 10
Constant pressure operation 10
Calculation of chromatographic parameters 11
GC Requirements 12
Other Requirements 12
Parts Supplied 13
Part Identification 14
Parts Not Supplied 15
Tools Required 15
2 Hardware Installation
Prepare the GC 18
Install the Column Clips 20
Install the Bracket and Splitter 21
Connect the Makeup Gas Supply 24
To supply the makeup gas froma PCM 24
To supply the makeup gas from an Auxiliary Pressure
controller 24
6 Installation and Operation Guide
3 Splitter Configurations
Typical Configurations 26
Splitting to an MSD 28
Custom Configurations 29
Restrictor id and length 32
Maximum and minimum flows 33
Column outlet pressure 34
Inlet pressure 34
Restrictor and Column Installation 35
Install the column 35
Connect the splitter 35
Disconnect tubing from the splitter 37
4Operation
An Example 40
Column flow 40
Select restrictors 42
Calculate column flow 43
Calculate ECD restrictor flow 44
Calculate MSD restrictor flow 45
Changing Columns Without Venting the MSD 46
Backflushing the Column 47
7
Agilent G3180B Splitter Kit
Installation and Operation Guide
Agilent Technologies
1
Introduction
Overview 8
How It Works 9
Details 10
Metal ferrules 10
Microfluidic plate 10
Constant pressure operation 10
Calculation of chromatographic parameters 11
GC Requirements 12
Other Requirements 12
Parts Supplied 13
Part Identification 14
Parts Not Supplied 15
Tools Required 15
This manual covers the installation and operation of the G3180B effluent
splitter with makeup gas kit on the Agilent 6890 series gas chromatograph
(GC).
8 Installation and Operation Guide
1Introduction
Overview
Splitter installation is done in three steps:
1Hardware installation. This gets the hardware installed and the gas flows
connected.
2Restrictor configuration. You can choose to use a typical, precalculated
configuration or create a custom one using software tools supplied on a CD.
3Restrictor and column installation. Using the results of step 2, cut the
appropriate lengths of the appropriate diameter tubing for the restrictors.
Install the restrictors and the analytical column.
Introduction 1
Installation and Operation Guide 9
How It Works
The splitter divides the effluent from a column between two different
detectors. The detectors can be operating at different pressures, that is, any
mix of the following can be used:
Atmospheric pressure
FID (flame ionization detector)
TCD (thermal conductivity detector)
NPD (nitrogen phosphorus detector)
ECD (electron capture detector)
FPD (flame photometric detector)
Below atmospheric pressure
MSD (mass selective detector)
Above atmospheric pressure
AED (atomic emission detector)
The split ratio is determined by the length and diameter of tubing connecting
the splitter to the detectors. Tubing dimensions may be determined from
Table 2 on page 26 in this manual or from a spreadsheet calculator that is
included for calculating tubing dimensions for special situations.
Figure 1 shows the plumbing configuration for the G3180B splitter.
Figure 1 Splitter plumbing
Restrictor 1 Restrictor 2
Constant pressure
makeup gas supply
Column
Splitter
GC oven
Det 1 Det 2Inlet
10 Installation and Operation Guide
1Introduction
The column flow mixes with the makeup flow in the splitter. This mixture then
flows through lengths of uncoated, deactivated, fused-silica tubing to each
detector. These tubes act as flow restrictors. While the flow through each
restrictor changes with oven temperature, the ratio of the two flows at any
temperature is the same.
Details
The G3180B kit addresses several limitations of previous approaches to
splitting column effluent between two detectors:
Metal ferrules
The splitter uses metal column ferrules, which eliminate air leakage into the
sample stream. Unlike polyimide, metal ferrules do not loosen upon thermal
cycling of the oven. They also do not outgas contaminants or shed particles
(like graphite) that can result in chromatographic problems.
Microfluidic plate
The splitting hardware is based on microfluidic plate technology. This allows
very low dead volume connections between the column end and the two
detector restrictor tubes. The thin metal plate has fast thermal response and is
mounted solidly on the oven wall for ease of use. The interior plate surfaces
are deactivated to prevent adsorption by active compounds.
Constant pressure operation
The splitter uses a source of makeup gas supplied by electronic pneumatics
control (EPC). This maintains the splitter at a known and constant pressure.
Constant pressure allows easier splitting to vacuum detectors like the MSD. It
simplifies choice of splitter parameters, allowing all aspects of the
chromatographic setup to be calculated. Constant pressure makeup allows the
column to be run in constant flow mode while still maintaining a constant
split ratio between two detectors of different operating pressures such as the
FPD and the MSD. Because the EPC pressure can be time programmed, useful
operations like backflushing unwanted heavy materials from the column and
changing columns in MSD systems without venting are possible.
Introduction 1
Installation and Operation Guide 11
Calculation of chromatographic parameters
Because the pressure at the split point is known and constant, the
chromatographic parameters can be calculated before setup. This is especially
useful with GC/MSD setups, where there are limitations on the flow rates of
carrier gas allowed into the MSD. If a method that was originally developed on
an MSD is converted to a splitter setup, a new inlet pressure can be calculated
to produce retention times very similar to the original method.
12 Installation and Operation Guide
1Introduction
GC Requirements
The splitter mounts in an Agilent 6890 series GC.
The splitter requires an electronically controlled pressure source such as the
Three Channel Pressure controller (6890 option 205, 301, or 308) or a
Pneumatics Control Module (PCM).
Other Requirements
The calculator requires Microsof®t Excel 97 (or later), which is not supplied
with this kit.
Introduction 1
Installation and Operation Guide 13
Parts Supplied
The G3180B kit contains the following parts (Table 1).
Table 1 Parts supplied
Part number Description Quantity
0100-0124 Union, stainless steel, 1/16-inch tubing 2
0100-0241 Union, stainless steel, 1/8 to 1/16-inch reducing 1
G1580-00130 Valve box blanking plate 1
G1530-01340 Capillary column spring clips 4
0515-0374 Screw, M3 × 10 mm 7
G2855-60140 Oven bracket assembly 1
G2855-60560 T-screw oven bracket retainer 2
G2855-80022 Manual and calculator CD 1
G3180-90120 Manual, G3180B
0100-2354 Tubing, stainless steel, 1/16-inch od × 0.01-inch id, 1 m 1
G3180-61500 Compact splitter with makeup gas assembly 1
G2855-60150 Supplies and spares kit 1
14 Installation and Operation Guide
1Introduction
Part Identification
Most of the kit parts are easily recognized. The unique ones are identified in
Figure 2.
Figure 2 Part identification
Capillary column
spring clips
Valve box
Oven bracket
Compact splitter
blanking plate
assembly
with makeup gas
assembly
This assembly is shipped in a
plastic bag to keep contaminants
out of the tubing and the fittings.
Do not open the bag until you are
ready to install the splitter.
Introduction 1
Installation and Operation Guide 15
Parts Not Supplied
Brown-dot frit (19231-60610)
Tools Required
Side cutter, large
Open-end wrenches
16 Installation and Operation Guide
1Introduction
17
Agilent G3180B Splitter Kit
Installation and Operation Guide
Agilent Technologies
2
Hardware Installation
Prepare the GC 18
Install the Column Clips 20
Install the Bracket and Splitter 21
Connect the Makeup Gas Supply 24
This chapter describes the procedure for installing the splitter hardware and
connecting the makeup gas supply.
18 Installation and Operation Guide
2Hardware Installation
Prepare the GC
1Raise the GC top cover to expose the oven top.
2Remove the valve box cutout using a side cutter (Figure 3).
WARNING Turn the power off and disconnect the power cord before proceeding.
Figure 3 Remove the valve box cutout
Cut tabs
around edges
and remove
Hardware Installation 2
Installation and Operation Guide 19
3This exposes a layer of soft insulation. Remove it to expose the hard oven
insulation. Remove the precut insulation piece at the location shown in
Figure 4.
4Replace the soft insulation. Install the valve box blanking plate, using one
screw at the front and one at the rear to secure it. See Figure 5.
Figure 4 Remove the insulation cutout
Figure 5 Install valve box blanking plate
Remove
this
cutout
Two screws
20 Installation and Operation Guide
2Hardware Installation
Install the Column Clips
Install the four column clips on the oven shroud (Figure 6).
Figure 6 Oven clips
Oven
shroud
Clip
(1 of 4)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52

Agilent Technologies G3180B User manual

Category
Water pumps
Type
User manual

Agilent Technologies G3180B is a Two-Way Splitter Kit with Makeup Gas designed to interface with capillary columns on an Agilent 6890 gas chromatograph (GC). It enables the division of column effluent between two detectors, which can operate at different pressures. The kit includes metal column ferrules to eliminate air leakage and a microfluidic plate for low dead volume connections. The constant pressure operation and the ability to calculate chromatographic parameters beforehand simplify the setup and optimization process.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI