Connecting External
Platform to
the HY10
Indicator via WiFi
Using the RS232/RS485 Converter for WiFi
USER MANUAL
2
MARCH 2021
3
CONTENTS
1. INTENDED USE ..................................................................................................... 4
2. SUPPORTED TERMINALS .................................................................................... 4
3. INSTALLATION ...................................................................................................... 5
3.1 Converter-Related Activities .............................................................................. 5
3.2 Scale-Related Activities .................................................................................... 7
4
1. INTENDED USE
The manual explains how to connect an external platform using the
RS232/RS485 converter for WiFi.
2. SUPPORTED TERMINALS
For the purpose of this user manual, the HY10 indicator was used. The
configuration on other scales will look the same.
5
3. INSTALLATION
3.1 Converter-Related Activities
Once the device is connected to the power supply, communicate with it and
set the respective operating parameters. Use a computer or other device with a
wireless network adapter. By default, the converter works in Access Point mode
without an access password, which means that in order to connect to it, you have to
find it in the list of available network points and press "Connect" button.
6
Next, open any web browser and type the following value in the address bar:
192.168.1.1. It is the IP address of the converter.
In the next step, the most important information is marked red. Change the language
to English, go to the "WiFi Set" tab and set the type of converter network security and
a password.
Press „Save” button. To save introduced modifications, restart the converter. To do
that, go to Reset&Restart tab and press "Restart" button.
7
Press „Save” button. To save introduced modifications, restart the converter. To do
that, go to Reset&Restart tab and press "Restart" button.
8
3.2 Scale-Related Activities
On the scale enter Parameters-> Communication-> WiFi->Available networks.
Select converter network and enter previously set password.
Once you have successfully connected to the converter's network, log in using the
service password and go to Parameters-> Global-> Number of Platforms menu.
Increase the value by 1. Next, go to Parameters-> Global-> Weighing module type
and select respective weighing module, e.g. MW-MH. Restart the device. After
restart, log in again with the service password and go to Parameters-> Factory
parameters-> Platform (that you want to add)-> Misc.-> Pue Y: Communication and
set the parameters as presented below:
9
10
/