Norbar 34315 is a Torque Tool Tester that can be used to measure the torque of various types of tools, including track, dial & electronic, impulse tool, clutch tool, and stall tool. It has a range of features that make it suitable for a variety of applications, including:

  • Multiple measurement modes: The Norbar 34315 can be used to measure torque in a variety of modes, including track, dial & electronic, impulse tool, clutch tool, and stall tool. This makes it suitable for a wide range of applications.
  • High accuracy: The Norbar 34315 is highly accurate, with an accuracy of ±0.5% of reading. This makes it suitable for use in applications where accurate torque measurement is critical.

Norbar 34315 is a Torque Tool Tester that can be used to measure the torque of various types of tools, including track, dial & electronic, impulse tool, clutch tool, and stall tool. It has a range of features that make it suitable for a variety of applications, including:

  • Multiple measurement modes: The Norbar 34315 can be used to measure torque in a variety of modes, including track, dial & electronic, impulse tool, clutch tool, and stall tool. This makes it suitable for a wide range of applications.
  • High accuracy: The Norbar 34315 is highly accurate, with an accuracy of ±0.5% of reading. This makes it suitable for use in applications where accurate torque measurement is critical.
www.norbar.com © Norbar Torque Tools 2013
INITIAL USE
Ensure that TD selection switch is in correct position (TD#1, TD#2, TD#3).
Switch on at side panel.
Track screen displayed. Press for relevant method of measurement.
Press EXIT for setup.
NOTE: It is essential that a new TTT is charged for 3.3 hours.
For AC power operation, connect power adapter.
For further information on operation & applications see
TTT Operators Manual (Part number 34308).
BUTTON FUNCTIONS
Button
Function
Button
Function
Track
Stall Tool
Dial & Electronic
Unit selection
Click and Cam
PRINT / RESET
Impulse Tool
PRINT / X PRINT
Clutch Tool
Exit
MODES
Mode
Filter setting
How it works
Track
500 Hz
Follows signal.
Dial & Electronic
500 Hz
Hold highest signal
until reset by user.
Impulse Tool
500 Hz
Clutch Tool
500 Hz
Stall Tool
500 Hz
Click & Cam
500 Hz
Hold 1st signal peak for
set time, then resets.
QUICK REFERENCE GUIDE
TORQUE TOOL TESTER
Part Number 34315 | Issue 3 | Original Instructions (English)
www.norbar.com © Norbar Torque Tools 2013
NOTE: The filter setting for each mode can be set independently. To change the filter
setting , select ‘SETTINGS’ and then ‘MODE FREQUENCY’. Choose the
measurement mode that you want to set and then select a frequency. When
selecting ‘OTHER FREQUENCY’ the user can enter a filter setting that does not
appear on the selection list.
SETTINGS
Feature
Settings
Language
11 Language choices
Password
Any 6 characters. (Factory set to 000000).
Frequency
Variable, 100Hz to 2500Hz.
Thresholds / Auto
Reset Hold Time
1 / 2 / 3 / 4 seconds.
Thresholds / First Peak
Sensitivity
HIGH (2.5% of reading) / MEDIUM (5% of reading) / LOW (10% of reading)
Units
N∙m, dN∙m, cN∙m, lbf∙ft, lbf∙in, ozf∙in, ft∙lb, in∙lb, in∙oz, kgf∙m, kgf∙cm, gf∙m ,gf∙cm, USER
Power Down Time
1 TO 99 minutes. Set to 0 (zero) to disable.
Limits
Variable, set by user in set up menu
TROUBLESHOOTER’S GUIDE
1. Zero does not function in track mode
Transducer overstrained. Return to Norbar.
2. Measurement modes do not function correctly.
Ensure that ‘TRIGGER FROM’ setting is not too high.
3. Battery only powers TTT for a short time.
Battery pack may need replacing (see manual)
4. Serial data output is not communicating with other equipment.
Check that control word on TTT and data equipment match.
Check baud rate is set to the same as data equipment.
See manual.
5. Serial data output is overwritten.
Printer may need line feed enable via menus SETUP SETTINGS - SERIAL PORT
6. Serial data output too fast.
Printer too slow. Slow down TTT by changing line delay using menus listed above.
7. Display shows ‘SMART TD NOT INITIALISED’.
i) Transducer lead has a broken connection.
ii) Unmodified transducer plugged in.
iii) SMART transducer lost memory; return to Norbar.
8. Menu 82. ‘CLOCK NOT INITIALISED’ displayed when power on.
Back up battery has failed. Replace or return to Norbar (see manual).
9. Password lost.
Contact Norbar quoting coded number on password menu.
NORBAR TORQUE TOOLS LTD, Wildmere Road, Banbury, Oxfordshire, OX16 3JU, UNITED KINGDOM
Tel: + 44 (0) 1295 270333
Email: enquir[email protected]
  • Page 1 1
  • Page 2 2

norbar 34315 Reference guide

Type
Reference guide

Norbar 34315 is a Torque Tool Tester that can be used to measure the torque of various types of tools, including track, dial & electronic, impulse tool, clutch tool, and stall tool. It has a range of features that make it suitable for a variety of applications, including:

  • Multiple measurement modes: The Norbar 34315 can be used to measure torque in a variety of modes, including track, dial & electronic, impulse tool, clutch tool, and stall tool. This makes it suitable for a wide range of applications.
  • High accuracy: The Norbar 34315 is highly accurate, with an accuracy of ±0.5% of reading. This makes it suitable for use in applications where accurate torque measurement is critical.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI