7
CONFIGURATION: “SEL.CFG” Use Gross/Net to enter the menu and step to each category,
Tare Recall to select parameters. ENT to return to menu selection.
Capacity is the combination of “1”, “2” and “3”.
Example: 1__100, 2___2 and 3__0.0 = 2,000 x 0.2 lb
Step Parameters Definition
1 5, 10, 15, 20…100, 120…1000
Number of divisions x100
100 = 10,000 divisions
2 1, 2, 5, 10, 20, 50, and 100.
Count by selection
10,000 divisions, count by 2 = 20,000
3 0, 0.0, 0.00, 0.000, and 0.0000 Decimal point selection
4 105P, 9 d (105% or 9 divisions) Overrange selection
5 1, 2, 3, 4, 5, 6, 8, 10, 12, 15……90 Digital filter selection (averaging)
6 off, 0.5, 1, 3, 5, 10 (divisions) Auto Zero Maintenance (AZM)
7 1.9, 5, 10, 20, FS (% of capacity)
Zero range selection
1.9% of 2,000 x 0.2 = 38.0 lb
7.1 off, on (ISM) Zero’s scale on power-up
8 off, 1, 3, 5,10 (divisions) Motion Band selection
9 lb, kg, con Units selection and convert
10 nt, Gtn, n.nt, n.Gtn.
Port 1 serial output selection
nt display only, Gtn is Gross Tare Net
and n.nt/n.Gtn inhibit negative gross
printing
11 off, co, de Off, Continuous, or Demand
12 7o, 7E, 8n 7- odd, 7- even or 8- none
13 12, 24, 48, 96 Baud rate selection
14 off, 1, 2, 3, 5, 10, 15 (seconds)
Delay between lines or continuous
output.
19 A, b
A : adds “STX” in continuous
b : No “STX” in continuous
20 nt, Gtn, n.nt, n.Gtn Port 2 serial output selection
21 off, co, de, Ln Off, Continuous, Demand, Network
22 7o, 7E, 8n 7- odd, 7- even or 8- none
23 12, 24, 48, 96 Baud rate selection
24 off, 1, 2, 3, 5, 10, 15 (seconds)
Delay between lines or continuous
output.
25 1 – 16 (RS485/RS422) Network address selection
30 off, on DIO Inputs