Covidien Nellcor PM10N User manual

Type
User manual

This manual is also suitable for

Service Manual
Nellcor
Portable SpO2 Patient Monitoring System
TM
COVIDIEN, COVIDIEN with logo, and Covidien logo and Positive Results for Life are U.S. and internationally registered
trademarks of Covidien AG. ™* brands are trademarks of their respective owner. Other brands are trademarks of a Covidien
company.
Service Manual i
Table of Contents
1 Introduction
1.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2.1 Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.2 Explosion, Shock, and Toxicity Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2.3 Service Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.4 Monitoring System Operation and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1.2.5 Patient Monitoring and Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
1.2.6 Monitoring System Readings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
1.2.7 Sensors, Cables, and Other Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
1.2.8 Electromagnetic Interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
1.2.9 Connections with Other Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7
1.2.10 Monitoring System Storage, Transport, and Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
1.3 Obtaining Technical Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.3.1 Technical Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
1.3.2 Related Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
1.4 Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9
2 Service Menu Options
2.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Safety Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.3 Service Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.3.1 Power On Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.3.2 Permission to Mute Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
2.3.3 Alarm Silence Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.3.4 Power Saving Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.3.5 Battery Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.3.6 Date/Time Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.3.7 Communication Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
2.3.8 Alarm Priority Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
2.3.9 Languages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
2.3.10 Dynamic Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-5
2.3.11 Enable PCSync Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.3.12 About Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
2.3.13 Covidien Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
3 Data Management
3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2 External Data Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.2.1 Monitoring History Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
ii Service Manual
3.3 Firmware Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
4 Modification and Testing
4.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3 System Performance Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3.1 Power-On Self-Test (POST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
4.3.2 Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4.3.3 Patient Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
4.3.4 Homecare Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.3.5 Sleep Study Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
4.3.6 Dynamic Passwords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
4.3.7 Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
4.4 Operational and Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
4.4.1 General Operation Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-10
4.4.2 Functional Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-26
4.5 Verification Check Sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
5 Troubleshooting
5.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2 Troubleshooting Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
5.2.1 Error Conditions by Category . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.2 System Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6
5.3 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
6 Repair
6.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.2 Spare Parts and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
6.3 Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6.4 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
6.4.1 Remove the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.4.2 Replace the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
6.5 Disassembly and Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.5.1 Front and Rear Assembly Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
6.5.2 NELL1SR Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.5.3 Main Board Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
6.5.4 Coin Cell Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-11
6.5.5 LCD Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-12
6.5.6 PI Cable and Cable Housing Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
Service Manual iii
List of Tables
Table1-1.Safety Symbol Definitions...............................................................................1-2
Table2-1.Service Menu Settings......................................................................................2-2
Table3-1.Monitoring Status Codes.................................................................................3-6
Table4-1.Required Test Equipment ...............................................................................4-2
Table4-2.Patient Modes .....................................................................................................4-5
Table4-3.Functional Tests with SRC-MAX................................................................. 4-27
Table5-1.Error Conditions and Resolutions ................................................................5-2
Table5-2.System Error Codes ...........................................................................................5-6
Table6-1.Spare Parts List by Callout Number.............................................................6-2
Table6-2.Monitoring System Accessories....................................................................6-4
Page Left Intentionally Blank
iv Service Manual
Service Manual v
List of Figures
Figure2-1. Service Menu ................................................................................................................. 2-3
Figure3-1. Mini-USB Port ................................................................................................................ 3-3
Figure3-2. Transfer Data Type ...................................................................................................... 3-4
Figure3-3. Transfer Data by USB .................................................................................................. 3-4
Figure3-4. Sample Monitoring History Printout .................................................................... 3-5
Figure3-5. Sample Bridge Driver Installer Window .............................................................. 3-7
Figure3-6. Sample New Hardware Wizard Screen ................................................................ 3-7
Figure3-7. Device Manager Button, Hardware Tab .............................................................. 3-8
Figure3-8. Sample Hardware List in Device Manager Window ........................................ 3-9
Figure3-9. Sample Initial USB to UART Bridge Properties Window .............................. 3-10
Figure3-10. Baud Rate List, Port Settings Tab ....................................................................... 3-11
Figure3-11. Firmware Upgrade Mode ..................................................................................... 3-12
Figure3-12. Firmware Upgrade, PCSync™* Utility ............................................................... 3-12
Figure3-13. Firmware Upgrade Process ................................................................................. 3-13
Figure3-14. POST Screen and Firmware Version ................................................................. 3-14
Figure4-1. Power-On Self-Test Sequence ................................................................................ 4-3
Figure4-2. Change Patient Mode Menu ................................................................................... 4-4
Figure4-3. Homecare Mode Menu Item ................................................................................... 4-5
Figure4-4. Homecare Mode Monitoring Screen .................................................................... 4-6
Figure4-5. Sleep Study Mode Menu Item ................................................................................ 4-7
Figure4-6. Service Menu ................................................................................................................. 4-8
Figure4-7. Dynamic Password Menu (Homecare to Standard Example) ...................... 4-8
Figure4-8. Password Set or Reset (Homecare to Standard Example) ............................ 4-9
Figure4-9. Date/Time Settings ..................................................................................................... 4-9
Figure4-10. Sensor Port ................................................................................................................ 4-10
Figure4-11. Sensor Port ................................................................................................................ 4-11
Figure4-12. Low SpO2 Alarm Limit of 99% ............................................................................ 4-12
Figure4-13. Low Pulse Alarm Limit of 160BPM .................................................................... 4-13
Figure4-14. Alarm Silence Duration Setting of 30 Seconds ............................................ 4-14
Figure4-15. Permission to Mute Alarm ................................................................................... 4-15
Figure4-16. Confirmation for Muted Alarm ........................................................................... 4-15
Figure4-17. Alarm Volume Default Setting of 2 ................................................................... 4-17
Figure4-18. Key Beep Volume Default Setting of 0 ............................................................ 4-18
Figure4-19. Pulse Volume Default Setting of 0 .................................................................... 4-19
Figure4-20. Brightness Setting .................................................................................................. 4-20
Figure4-21. “Spot Reading Saved” Message ......................................................................... 4-21
Figure4-22. Monitoring History Spot Data ............................................................................ 4-21
Figure4-23. Sensor Disconnect Alarm Priority Setting ...................................................... 4-22
Figure4-24. Sensor Off Alarm ..................................................................................................... 4-23
Figure4-25. Screen Saver Time Setting ................................................................................... 4-24
Figure4-26. Auto Power Off Time Setting .............................................................................. 4-26
Figure4-27. SRC-MAX Oximetry Tester ................................................................................... 4-28
Figure4-28. SRC-MAX Tester-Generated Waveform .......................................................... 4-29
vi Service Manual
Figure4-29. SRC-MAX Increase to 200 BPM .......................................................................... 4-30
Figure4-30. SRC-MAX Decrease to 60 BPM ........................................................................... 4-30
Figure4-31. SRC-MAX %SpO2 Increase to 90 ....................................................................... 4-31
Figure4-32. SRC-MAX %SpO2 Decrease to 75 ..................................................................... 4-32
Figure4-33. SRC-MAX High Modulation ................................................................................. 4-33
Figure4-34. BPM of 200 with High Modulation ................................................................... 4-34
Figure4-35. BPM of 60 with High Modulation ...................................................................... 4-35
Figure4-36. %SpO2 of 90 with High Modulation ................................................................ 4-36
Figure4-37. %SpO2 of 75 with High Modulation ................................................................ 4-37
Figure4-38. %SpO2 of 75 with Low Modulation ................................................................. 4-38
Figure4-39. High Light Condition ............................................................................................. 4-39
Figure4-40. BPM of 200 with High Light Condition ........................................................... 4-40
Figure4-41. BPM of 60 with High Light Condition .............................................................. 4-41
Figure4-42. %SpO2 of 90 with High Light Condition ........................................................ 4-42
Figure4-43. %SpO2 of 75 with High Light Condition ........................................................ 4-43
Figure4-44. High Modulation and High Light Condition ................................................. 4-44
Figure5-1. Return Packaging ......................................................................................................... 5-9
Figure6-1. Exploded View ............................................................................................................... 6-2
Figure6-2. Standard Cover (3 Shown) and Ambulatory Cover ......................................... 6-4
Figure6-3. Battery Replacement .................................................................................................. 6-5
Figure6-4. Front and Rear Assembly Replacement ............................................................... 6-7
Figure6-5. NELL1SR Board Replacement .................................................................................. 6-9
Figure6-6. Main Board Replacement ....................................................................................... 6-10
Figure6-7. Coin Cell Battery Replacement ............................................................................. 6-11
Figure6-8. LCD Replacement ..................................................................................................... 6-12
Figure6-9. PI Cable and Cable Housing Replacement ...................................................... 6-13
1-1
1 Introduction
1.1 Overview
This manual contains information for servicing the Nellcor™ portable SpO2 patient monitoring
system.
This manual applies to the following product:
Note:
Before use, carefully read this manual, the Operator’s Manual, accessory Instructions for Use, and all
precautionary information and specifications.
Reference the Operator’s Manual for the following information:
Intended Use statement
Operations-related warnings and cautions
Overviews of the display and operating buttons
Descriptions of product and packaging symbols
Installation and operation instructions
Alarms management
Preventive maintenance
Performance considerations
Accessories
Theory of operations
Product specifications
Clinical studies
PM10N
Introduction
1-2 Service Manual
1.2 Safety Information
This section contains important safety information related to general use of the Nellcor™ portable
SpO2 patient monitoring system. Other important safety information appears throughout the
manual. The Nellcor™ portable SpO2 patient monitoring system is referred to as the “monitoring
system” throughout this manual.
1.2.1 Safety Symbols
1.2.2 Explosion, Shock, and Toxicity Hazards
WARNING:
Explosion hazard — Do not use the monitoring system in the presence of flammable anesthetics.
WARNING:
Shock hazard—Do not pour or spill liquids onto the monitoring system.
WARNING:
Shock hazard—Firmly close the battery cover to prevent moisture from entering the monitoring
system.
WARNING:
The LCD panel (display) contains toxic chemicals. Do not touch broken LCD panels. Physical
contact with a broken LCD panel can result in transmission or ingestion of toxic substances.
Table1-1.Safety Symbol Definitions
Symbol Definition
WARNING
Alerts users to potential serious outcomes (death, injury, or adverse events) to the
patient, user, or environment.
Caution
Identifies conditions or practices that could result in damage to the equipment or
other property.
Note
Provides additional guidelines or information.
Safety Information
Service Manual 1-3
1.2.3 Service Procedures
WARNING:
To avoid possible injury, do not attempt to service the monitoring system if there are any signs of
burning or smoking coming from the monitoring system.
WARNING:
Before attempting to service the monitoring system, disconnect it from the patient to avoid
possible injury to the patient.
WARNING:
Before attempting to disassemble the monitoring system, remove the batteries to prevent
possible injury.
WARNING:
Ensure that conductive portions of the electrodes, leads, and cable do not come into contact with
any other conductive parts.
WARNING:
High voltage is generated by the LCD backlight driver. Exercise caution when operating the
monitoring system with covers open.
WARNING:
Extreme care must be taken in modifying default or other settings to ensure they are appropriate
to the intended use.
WARNING:
Make sure to complete all performance tests in Chapter 4, Modification and Testing before placing
the monitoring system into operation after repair or maintenance. Failure to perform all tests
could result in erroneous monitoring system readings.
WARNING:
Any connections between this monitoring system and other devices must comply with applicable
medical systems safety standards such as IEC 60601-1. Failure to do so could result in unsafe
leakage current and grounding conditions.
WARNING:
To ensure accurate performance and prevent device failure, do not expose the monitoring system
to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate
performance or device failure. Reference the Operator’s Manual for fluid ingress specifications.
Introduction
1-4 Service Manual
1.2.4 Monitoring System Operation and Service
WARNING:
Inspect the monitoring system and all accessories before use to ensure there are no signs of
physical damage or improper function. Do not use if damaged.
WARNING:
To ensure accurate performance and prevent device failure, do not expose the monitoring system
to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate
performance or device failure. Do not immerse in water, solvents, or cleaning solutions, since the
monitoring system and pulse oximetry sensors and connectors are not waterproof.
WARNING:
Do not sterilize the monitoring system by irradiation, steam, or ethylene oxide.
WARNING:
The monitoring system should not be used adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, observe the monitoring system to verify normal operation in
the desired configuration.
WARNING:
The only user-serviceable parts inside the monitoring system are the four AA batteries. While users
can open the battery cover to change the batteries, only qualified service personnel should
remove the cover or access internal components for any other reason. Users should not modify
any components of the monitoring system.
WARNING:
Do not spray, pour, or spill any liquid on the monitoring system, its accessories, connectors,
switches, or openings in the casing, since this may cause damage to the monitoring system. Never
place fluids on the monitoring system. If fluid spills on the monitoring system, remove batteries,
wipe all components dry immediately, and have the monitoring system serviced to ensure no
hazard exists.
WARNING:
Do not damage the batteries by applying pressure. Do not throw, hit, drop, or impact the batteries.
WARNING:
Keep the monitoring system and batteries out of the reach of children.
Caution:
The monitoring system may not operate properly if it is operated or stored at conditions outside
the ranges stated in this manual, or if it is subjected to excessive shock or dropping.
Safety Information
Service Manual 1-5
1.2.5 Patient Monitoring and Safety
WARNING:
Always disconnect and remove the monitoring system and sensors during magnetic resonance
imaging (MRI) scanning. Attempting to use the monitoring system during an MRI procedure could
cause burns or adversely affect the MRI image or the monitoring system's accuracy.
WARNING:
Keep patients under close surveillance when monitoring. It is possible, although unlikely, that
radiated electromagnetic signals from sources external to the patient and the monitoring system
can cause inaccurate measurement readings.
WARNING:
As with all medical equipment, carefully route patient cabling to reduce the possibility of patient
entanglement or strangulation.
WARNING:
Do not lift or carry the monitoring system by the pulse oximetry sensor or pulse oximetry interface
cable. The cable may disconnect and cause the monitoring system to drop on a patient or cause
damage to monitoring system surfaces.
1.2.6 Monitoring System Readings
WARNING:
The monitoring system may remain attached to the patient during defibrillation or during use of
an electrosurgical unit; however, the monitoring system is not defibrillator-proof, and readings
may be inaccurate during defibrillation and shortly thereafter.
WARNING:
Check the patient's vital signs by alternate means should there be any doubt about the accuracy
of any measurement. Request a qualified service technician confirm the monitoring system is
functioning correctly.
WARNING:
For best product performance and measurement accuracy, use only accessories supplied or
recommended by Covidien. Use accessories according to their respective Instructions for Use.
Introduction
1-6 Service Manual
1.2.7 Sensors, Cables, and Other Accessories
WARNING:
Before use, carefully read the pulse oximetry sensor Instructions for Use, including all warnings,
cautions, and instructions.
WARNING:
Use only the Covidien-approved pulse oximetry sensors, interface cables, and accessories. Use of
other sensors, cables, and accessories can result in inaccurate readings and increased monitoring
system emissions.
WARNING:
Do not use any other cables to extend the length of the Covidien-approved interface cable.
Increasing the length will degrade signal quality and may lead to inaccurate measurements.
WARNING:
To prevent damage, avoid undue bending of the sensor cable.
WARNING:
The sensor disconnect error message and associated alarm indicate the pulse oximetry sensor is
either disconnected or has faulty wiring. Check the connection and, if necessary, replace the
sensor, the pulse oximetry cable, or both.
1.2.8 Electromagnetic Interference
WARNING:
Any radio frequency transmitting equipment or other nearby sources of electrical noise may result
in disruption of the monitoring system.
WARNING:
The monitoring system is designed for use in environments in which the signal can be obscured by
electromagnetic interference. During such interference, measurements may seem inappropriate
or the monitoring system may not seem to operate correctly.
WARNING:
Large equipment using a switching relay for its power on/off may affect monitoring system
operation. Do not operate the monitoring system in such environments.
Safety Information
Service Manual 1-7
Caution:
This device has been tested and found to comply with the limits for medical devices related to IEC
60601-1-2: 2007 and IEC 60601-1-2:2014. These limits are designed to provide reasonable
protection against harmful interference in a typical medical installation.
Caution:
This monitoring system generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful interference to other
devices in the vicinity. If interference is suspected, move pulse oximetry cables away from the
susceptible device.
Caution:
Be aware of possible interference from sources of electromagnetic interference, such as cellular
phones, radio transmitters, motors, telephones, lamps, electrosurgical units, defibrillators, and
other medical devices. If pulse oximetry readings are not as expected for the patient’s condition,
remove the sources of possible interference.
1.2.9 Connections with Other Equipment
Caution:
Accessory equipment connected to the monitoring system's data interface must be certified
according to IEC 60950-1 for data-processing equipment. All combinations of equipment must be
in compliance with IEC 60601-1 Requirements for Medical Electrical Systems. Anyone who
connects additional equipment to the signal input or signal output port configures a medical
system and is therefore responsible for ensuring the system complies with the requirements of IEC
60601-1, IEC 60601-1-2:2007, and IEC 60601-1-2:2014.
Caution:
When connecting the monitoring system to any instrument, verify proper operation before clinical
use.
Caution:
Anyone who connects a PC to the data output port configures a medical system and is therefore
responsible for ensuring that the system complies with the requirements of IEC 60601-1-1 and the
electromagnetic compatibility IEC 60601-1-2.
1.2.10 Monitoring System Storage, Transport, and Disposal
Caution:
Remove the batteries from the monitoring system before placing it in storage or when not using
it for a long period.
Introduction
1-8 Service Manual
Caution:
Do not short-circuit the batteries, as they may generate heat. To avoid short-circuiting, do not let
the batteries come in contact with metal objects at any time, especially during transport.
Caution:
Follow local government ordinances and recycling instructions regarding disposal or recycling of
the monitoring system and its components, including batteries and accessories.
1.3 Obtaining Technical Assistance
1.3.1 Technical Services
For technical information and assistance, contact Covidien or a local Covidien representative.
When calling Covidien or a local Covidien representative, have the monitoring system serial
number available. The serial number label is located on the bottom of the monitoring system.
Provide the firmware version number displayed during the power-on self-test (POST).
1.3.2 Related Documents
Nellcor™ Portable SpO2 Patient Monitoring System Operator’s Manual —
Provides basic information
for operating the monitoring system and troubleshooting errors or malfunctions.
Nellcor™ Pulse Oximetry Sensor Instructions for Use — Guides sensor selection and usage. Before
attaching any of the various Covidien-approved pulse oximetry sensors to the monitoring system,
refer to the individual Instructions for Use.
Nellcor™ Oxygen Saturation Accuracy Specification Grid — Provides sensor-specific guidance related
to desired SpO2 saturation accuracy measurements. Available online with the product manuals for the
monitoring system at www.covidien.com.
Covidien Technical Services: Patient Monitoring
15 Hampshire Street
Mansfield, MA 02048 USA
1.800.635.5267, 1.925.463.4635,
or contact a local Covidien representative
www.covidien.com
Warranty Information
Service Manual 1-9
1.4 Warranty Information
The information contained in this document is subject to change without notice. Covidien makes
no warranty of any kind with regard to this material, including, but not limited to, the implied
warranties of merchantability and fitness for a particular purpose. Covidien shall not be liable for
errors contained herein or for incidental or consequential damages in connection with the
furnishing, performance, or use of this material.
Introduction
Page Left Intentionally Blank
1-10 Service Manual
2-1
2 Service Menu Options
2.1 Overview
This chapter describes Service Menu settings for the Nellcor™ portable SpO2 patient monitoring
system. A password is required to access the Service Menu. Contact Technical Services
(reference Technical Services, page 1-8) for the default password.
2.2 Safety Reminders
WARNING:
For best product performance and measurement accuracy, use only accessories supplied or
recommended by Covidien. Use accessories according to their respective Instructions for Use.
WARNING:
The sensor disconnect error message and associated alarm indicate the pulse oximetry sensor is
either disconnected or has faulty wiring. Check the connection and, if necessary, replace the
sensor, the pulse oximetry cable, or both.
WARNING:
Do not use damaged pulse oximetry sensors. Do not use with exposed optical components. Do
not immerse completely in water, solvents, or cleaning solutions, since pulse oximetry sensors
and connectors are not waterproof. Do not sterilize by irradiation, steam or ethylene oxide. Refer
to the cleaning instructions in the Instructions for Use for reusable sensors.
Caution:
Do not attach any cable intended for computer use to the sensor port connector.
2.3 Service Menu Overview
WARNING:
Only qualified service personnel should open the monitoring system housing, remove and
replace parts, or make adjustments.
Service Menu Options
2-2 Service Manual
The monitoring system contains a Service Menu that is accessible only by password. Contact
Technical Services (reference Technical Services, page 1-8) for the default passwords for all
restricted menus and functions.
The monitoring system ships with the Service Menu factory default settings listed in Table 2-1. For
a list of the factory default settings for other menus, reference the Operator’s Manual.
Table2-1.Service Menu Settings
Parameter Ranges/selection
Factory default
Adult Pediatric Neonatal
Power On Settings Factory Defaults, Institutional Defaults, Last Setting Factory Defaults
Permission to Mute Alarm No, Yes No
Alarm Silence Duration 30, 60, 90, 120 seconds 120 seconds
Power Saving Settings Screen Saver Time:
Never, 1 - 10 minutes
3 minutes
Screen Saver Brightness:
0% - 50% in 10% increments
20%
Auto Power Off Time:
Never, 1 - 10 minutes
3 minutes
Battery Type Lithium, Alkaline Lithium
Date/Time Settings Date Format:
YY/MM/DD, DD/MM/YY, MM/DD/YY
YY/MM/DD
Communication Settings Serial Connectivity Settings:
ASCII 19200, ASCII 115200,
SPDout 19200, SPDout 115200
ASCII, 19200
Alarm Priority Settings Low, Medium, High:
- Sensor Disconnect
- Sensor Off
- Sensor Failure
Low
Low, Medium, High:
- SpO2 High
- Pulse Rate High
Medium
Medium, High:
- SpO2 Low
- Pulse Rate Low
Medium
Languages Chinese, Czech, Danish, Dutch, English, Finnish,
French, German, Greek, Hungarian, Italian,
Japanese, Korean, Norwegian, Polish, Portuguese,
Russian, Slovakian, Spanish, Swedish, Turkish
English
Dynamic Password Passwords can be changed for entry into Homecare
Mode and Sleep Study Mode and back to Standard
Mode
(Contact Technical Services)
Enable PCSync Mode On, Off Off
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112

Covidien Nellcor PM10N User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI