Air Liquide Presence PR Instructions For Use Manual

Type
Instructions For Use Manual
www.airliquidehealthcare.com.au
Presence
TM
PR, PRC & PRE
Instructions for Use
Content
Inspection checks to be performed by the user before initial use
Operational Information
2.1 Connecting to the barbed outlet
2.2 Connecting to the bleeved outlet
Troubleshooting guide
2
The Integrated Regulator-Valve is manufactured by Air Liquide Medical Systems,
6 rue Georges Besse 92182 Antony France and is supplied as part of the Presence™ range
of products in Australia by:
Air Liquide Healthcare Pty. Ltd.
5/476 Gardeners Road – Alexandria NSW 2015
1300 36 02 02
www.airliquidehealthcare.com.au
Disclaimer of Liability:
Whilst proper care has been taken in the preparation of this document, no liability for injury or damage resulting from
its use can be accepted.
1
2
3
Presence™ PR, PRC and PRE are ready to use medical oxygen cylinders complete with
an integrated regulator-valve and flow meter which are intended to supply regulated medical
oxygen to a patient or another medical device. They deliver medical oxygen either through a
flow regulated barbed outlet or a fixed flow sleeved outlet.
Presence
TM
Presence
TM
Presence
TM
PRE PR PRC
Key features:
3 Easy to use and ergonomic for
a quick and easy set-up
3 Eliminates the cost of additional
equipment (separate regulators
and flowmeters)
3 Unique interlocking feature to
prevent accidental valve opening
3 Dedicated carry handle* and
bed/wall/trolley system
minimising WHS issues
3 Available in small sizes for
ambulatory needs and large size
for backup requirements
* Not available with the PRE
3
1. Inspection checks to be performed by the user
before initial use
a. Check that a clear plastic heat shrink wrap with Air
Liquide Healthcare logo is intact, and fully covers the
shroud (refer Fig 1). Check that the cylinder is clean and
free from oil and grease.
b. Check outlet protective caps are in place prior to
removing wrap. When possible always replace the caps
when not in use (refer Fig 1).
c. Check that the valve is in the red “O” OFF position (refer
Fig 2). View the front of the valve assembly and identify
all the components (the barbed and sleeved outlets and
the pressure gauge).
d. Check that the pressure gauge shows the cylinder is
fully charged and check that the pressure indicated on
the gauge matches the pressure on the product label. If
the cylinder has already been used, check the
pressure gauge to ensure there is sufficient medical
oxygen left. The coloured sections are an indication of
the available contents:
• Green: Sufficient contents
• Yellow: Contents are getting low
Red: Contents are low. Discontinue use and
replace with a full one
e. Check the product label to ensure that the gas matches
the application. Read the operating instructions
provided on the label fitted to the side of the shroud
(refer Fig 3).
If any deviation is observed do not use the cylinder
and contact your nearest Air Liquide Healthcare
office on 1300 360 202.
Figure 1 : Heat shrink wrap
and protective caps
Figure 2: Outlets and
pressure gauge
Figure 3: Operating
instructions and product
label
4
a
b
c
d
e
2. Operational information
a. Remove the cap and connect the oxygen tubing to the
barbed outlet (refer Fig 4).
b. Turn the ON/OFF valve (outer ring) anti clockwise to
show the green “I” in the portal on top of the valve (refer
Fig 5).
c. Select the flow rate, “1”, “1,5”, ‘2”, “3”, “4”, “6”, “9”, “12” or
“15” litres per minute (“l/min”) by turning the flow selector
knob at the top of the valve (inner section) clockwise. A
“click” will be heard when each flow rate is selected until
the desired flow rate is reached (refer Fig 6).
d. Check that medical oxygen is flowing from the barbed
outlet before administering to patient.
e. When therapy has concluded, turn off the flow by
turning the flow selector knob in an anti clockwise
direction until it no longer moves and the green “I” is
displayed in the portal.
f. Turn the valve clockwise showing the red “O” in the
portal.
g. When required remove the oxygen tube from the barbed
outlet and replace the cap.
2.1 Connecting to the barbed outlet
Ensure that the flow rate position is not in-between
any available flow rate position as this may interrupt
the supply of medical oxygen.
Figure 4: Tubing connected
to the barbed outlet
Figure 5: Valve in the green
“I” ON position
Figure 6: Flow selector
knob turned to the selected
flow rate
5
a
b
c
2.2 Connecting to the sleeved outlet:
a. Remove the cap and connect the desired equipment to
the sleeved outlet (i.e. hose or suction equipment) and
ensure the fitting is sufficiently hand tightened (refer Fig
7 & 8).
b. Turn the ON/OFF Valve anti clockwise to show the
green “I” in the portal on top of the valve (refer Fig 5).
c. Use the attached equipment as required for proper
dispensation of medical oxygen. The outlet pressure
flow rate of the sleeved index outlet is at least 50 l/min
whilst the contents are within the “yellow” and “green”
coloured sections of the pressure gauge.
d. When therapy has concluded, turn off the attached
equipment.
e. Turn the ON/OFF valve clockwise showing the red “O”
in the portal.
f. When required remove the attached equipment from
the sleeved outlet and replace the cap.
g. When attaching a suction unit to a PRE cylinder, the use
of a small extension piece will be required (refer Fig 9).
Extension pieces should not be fitted to PRC and
PR cylinders as this will cause the cylinder to become
unstable.
h. When a venturi suction unit (or similar) is attached to the
sleeve outlet, the pressure gauge on the Presence™ PR,
PRC and PRE may be obscured. A close up view of the
pressure gauge (such as looking under and around the
device) may be required to ensure a clear viewing.
Figure 7: Hose connected to
the sleeved outlet
Figure 8: Venturi Suction unit
connected to the Sleeved
Outlet
Figure 9: Connection to
the PRE Sleeved Outlet
using extension piece
Ensure that the extension piece is removed from
the cylinder prior to returning the empty cylinder to
Air Liquide Healthcare.
6
a
g
a
3. Troubleshooting guide
Fault/concern
No flow of gas
Cannot
connect to the
Outlet
Hissing noise
Possible cause
Cylinder is empty
Outlet nozzle may
be blocked
Flow selector
in-between settings
Blocked port in flow
selector
Hose diameter is
wrong size
Shroud is twisted
Damaged sleeved
outlet
Gas leak
Response to be taken
Check that the flow selector knob or any attached device has not been
left ON. If contents have been depleted then place unit in the empty
cylinder storage area and replace with full one.
If it’s a first time use, then check the pressure gauge and if the gauge
is in the red section immediately contact customer service and
quarantine the unit for investigation by an Air Liquide Healthcare
representative.
DO NOT attempt to clear a blocked nozzle. Immediately contact
customer service and quarantine the unit for investigation by an Air
Liquide Healthcare representative.
View the top of the flow selector knob to ensure the flow setting
is visible through the viewing port. If the number is not clearly visible
then rotate the knob till the required flow is clearly visible and clicks
into place.
Immediately contact customer service and quarantine the unit for
investigation by an Air Liquide Healthcare representative.
Check that the tubing fits firmly over the barbed outlet. If not then
replace the tubing.
View the front of the unit and if the shroud is twisted contact
customer service and quarantine the unit. Obtain another unit for use.
If the unit is required immediately, then twist the shroud back in-line
and use as normal. At the completion of use, contact customer
service and quarantine the unit for investigation by an Air Liquide
Healthcare representative.
Check that the thread is not damaged to either the sleeved outlet or
the device it’s being connected to.
Check that the outer rim of the sleeved outlet has not been
damaged and is circular in shape. Check the hand wheel engages
properly.
If sleeved outlet is damaged then contact customer service and
quarantine the unit for investigation by an Air Liquide Healthcare
representative.
First check to make sure that the flow selector knob has not been left ON.
Do not attempt to stop a leak. Turn the valve OFF and immediately
contact customer service and quarantine the unit for investigation by
an Air Liquide Healthcare representative.
• Read the product label and SDS before using.
• Do not use grease or oil or any oil-based product.
Do not use or store near a naked flame, a source of
ignition or combustible material.
Do not smoke nearby and use appropriate signage.
Use and store in a well ventilated area and store in a
secure and upright position.
Do not connect any device to the fill port located at the
rear of the integrated regulator-valve. The metal cap
shall NOT be removed under any circumstance. Refer
to the warning label located near the fill port.
Only attach fittings and accessories to the integrated
regulator-valve that are compatible with medical oxygen
Do not store Presence™ units in any area where
insect infestation may occur. Do not place objects
inside the nozzles and ensure proper protection
of the outlet nozzles when not in use by using the
plastic caps as supplied.
Flow inaccuracies can be caused by back-pressure
downstream of the integrated regulator-valve.
Any repair, service or modification shall only be
carried out by an authorised agent.
SAFETY INFORMATION
7
www.airliquide.com
OUR TEAM IS HERE TO HELP.
FOR ANY ENQUIRIES PLEASE CONTACT US ON:
DBRO-SMTG-0270-2
NSW
Unit 5/476 Gardeners Road
Alexandria NSW 2015
Tel: 02 9364 7474
eFax*: 02 9364 7476
QLD
25 Campbell Street
Bowen Hills QLD 4006
Tel: 07 3250 3900
eFax*: 02 9364 7497
SA
95 Fullarton Road
Kent Town SA 5067
Tel: 08 8331 1648
eFax*: 02 9364 7477
VIC & TAS
40 Bunnett Street
North Sunshine VIC 3020
Tel: 03 9300 5644
eFax*: 02 9364 7482
WA & NT
51 Hampden Rd
Nedlands 6009
Tel: 08 6389 0707
Fax: 08 6389 1199
*All eFax area codes begin with (02).
World leader in gases, technologies and services for Industry and Health, Air Liquide is present in 80
countries with approximately 65,000 employees and serves more than 3 million customers and patients.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Air Liquide Presence PR Instructions For Use Manual

Type
Instructions For Use Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI