7228 OPERATOR’S MANUAL – SECTION 5
and allows the WECO connector to function as a
balanced input.
SW#6: Input Coupling
When the Input Coupling DIP switch is in the UP
position (default), the amplier can receive and
amplify both DC and AC signal. When this switch
is in the DOWN position, a 2-Hz high-pass lter on
the inputs prevents the transmission of DC signal.
NOTE: Both the Input Coupling and DC Servo DIP
switches must be placed in the DOWN position
when activating the DC Servo function. Placing the
Input Coupling DIP switch in the UP position and
the DC Servo DIP switch in the DOWN position will
cause the amplier to be placed in Standby/Stop
mode. To return the amplier to operational condi-
tions, set both the DC Servo and the Input Cou-
pling DIP switches in the DOWN position or place
the DC Servo Switch in the UP position.
suitable for your application. For more information
on Controlled-Current operation, including how to
determine and congure a custom compensation
network, see the Applications section.
NOTE:
If the Controlled Mode DIP switch set-
ting is changed while the amplier is enabled, the
amplier will temporarily be placed in Standby/
Stop mode. Press the Enable button to release the
amplier.
SW#8: Bi-level Mode
When the Bi-level Mode DIP switch is in the UP
position (default), the amplier will automatically
switch the power supply transformers between a
series and a parallel conguration, depending on
the present voltage requirements of the applica-
tion. When this switch is in the DOWN position,
the amplier will not switch automatically, but will
operate according to the Manual Bi-level Mode
DIP switch setting (SW #9). For more information
on how to set the Bi-level Mode DIP switch for
increased voltage or current, see the Applications
section.
SW#9: Manual Bi-level Mode
The Manual Bi-level Mode switch is enabled when
the Bi-level Mode switch (SW#8) is set to Manual
(DOWN position). Once enabled, the amplier
will operate with the power supply transformers in
a series conguration when the Manual Bi-level
Mode DIP switch is in the UP position (default).
When this switch is in the DOWN position, the am-
plier will operate with the power supply transform-
ers in a parallel conguration. For more informa-
tion on how to set the Manual Bi-level Mode DIP
switch for increased voltage or current, see the
Applications section.
SW#10: Power Supply Rail Mode
When the Power Supply Rail Mode DIP switch is in
the UP position (default), the amplier will operate
with the power supply rails in a series congura-
tion for high-voltage output. When this switch is in
the DOWN position, the amplier will operate with
the power supply rails in a parallel conguration for
high-current output. For more information on how
to set the Power Supply Rail Mode DIP switch for
In Controlled-Current Mode, the load is part of the ampli-
er circuit, and the relationship of the load to the amplier is
critical. For proper and safe operation in Controlled-Current
mode, you must observe the following guidelines:
1. Properly attach a load before operating the amplier.
2. DO NOT use a blocking capacitor. The load must have a
DC path.
3. Never leave the load open. If you feel the load must be
fused, which could lead to a potential open circuit, please
contact AE Techron Technical Support.
4. Make sure the load has some inductive component.
5. Provide appropriate compensation for the load.
6. If oscillation occurs, turn o the amplier immediately.
Failure to follow these guidelines may result
in damage to the amplier or load.
CAUTION
SW#7: Controlled Mode
When the Controlled Mode DIP switch is in the
UP position (default), the amplier will operate
in Controlled-Voltage mode, and the amplier’s
output voltage will be controlled by its input voltage
signal. When this switch is in the DOWN position,
the amplier will operate in Controlled-Current
mode, and the amplier’s output current will be
controlled by its input voltage signal.
IMPORTANT: Controlled-Current operation re-
quires the use of a compensation network, and the
7228’s default compensation network may not be