Application:
The temperature sensors TG8 are designed for measuring temperatures of gaseous and liquid
substances. The temperature range for application of the sensor is – 50°C to 125 °C and it must not be
exceeded even for a short term. The sensors may be used for all control systems compatible with the
DS 18B20 temperature sensor. The temperature sensors are designed for universal application. The
sensors in combination with JTG8 thermowell can be used for measuring temperatures in piping and
simultaneously as pressure equipment in terms of Government Order No. 26/2003 Coll. as amended, in
which the Pressure Equipment Directive 97/23/ES is implemented. The sensors are intended for
operation in chemically non-aggressive environments.
Sensor description:
The sensor consists of a metallic housing with the sensing element inside and a supply cable. The sensor
housing is made of stainless steel grade 17240. The sensors are connected as two-wire probes. The
supply cable has external silicone insulation and is shielded. The shielding is not connected with the
housing or with the temperature element. The length of housing can be selected from 40 to 200 mm as a
standard.
Accuracy class of element *
± 0,5 °C in the range –10 to 85 °C
± 2 °C in the range -50 to 125 °C
Temperature element wiring
IP 67 according to ČSN EN 60 529
0,5
7 sec (in flowing water 0,2 m.s
-1
)
500 Vef according to ČSN EN 61010-1 Art. 6.8.4.
200 M at 500VDC, 25 ± 3 °C
shielded silicone 2 x 0,34 mm
2
0,105 / 1 m at a temperature of 25 °C
External pressure endurance
without thermowell 2,5 MPa
with thermowell 6,3 MPa
Maximum water flow speed when
measuring temperature in piping with
JTG8 thermowell
Thermowell length up to 60 mm 2 m.s
-1
Thermowell length 60 to 100mm 1,5 m.s
-1
Thermowell length 100 to 160mm 1,0 m.s
-1
Maximum speed of water and water
vapour flow when measuring temperature
in piping with JTG8 thermowell
Thermowell length up to 60 mm 20 m.s
-1
Thermowell length 60 to 100mm 15 m.s
-1
Thermowell length 100 to 160mm 10 m.s
-1
Class of electrical equipment