Hach SP-510 Instrument Manual

Type
Instrument Manual
Cat. No. 54100-18
SP-510
HARDNESS MONITOR
INSTRUMENT MANUAL
CAT. NO. 54100-18
SP-510™
HARDNESS MONITOR
INSTRUMENT MANUAL
© Hach Company, 1999. All rights reserved. Printed in the U.S.A. dd/dp 7/99 1ed
Rev. 1, 10/99
2
TRADEMARKS OF HACH COMPANY
3
TABLE OF CONTENTS
CERTIFICATION............................................................................................................................................................ 5
SAFETY PRECAUTIONS.............................................................................................................................................. 7
SPECIFICATIONS.......................................................................................................................................................... 9
SECTION 1 GENERAL INFORMATION......................................................................................................... 13
1.1 Instrument Description............................................................................................................................................. 13
1.2 Principle of Operation.............................................................................................................................................. 14
INSTALLATION....................................................................................................................................................... 15
SECTION 2 INSTALLATION............................................................................................................................... 17
2.1 Instrument Environmental Considerations............................................................................................................... 17
2.2 Unpacking the Instrument........................................................................................................................................ 17
2.3 Wall Mounting the Instrument................................................................................................................................. 18
2.4 Plumbing Connections ............................................................................................................................................. 21
2.5 Installing the Sample Line........................................................................................................................................ 22
2.6 Optional Air Purge................................................................................................................................................... 23
2.7 Sample Conditioning................................................................................................................................................ 23
2.7.1 Assembling the Sample Conditioning Kit...................................................................................................... 23
2.7.2 Using the Sample Conditioning Kit................................................................................................................ 24
2.8 Electrical Connections.............................................................................................................................................. 26
2.8.1 Power Connections......................................................................................................................................... 26
2.8.2 Wiring the Instrument..................................................................................................................................... 28
2.8.3 Voltage Selection for Alternate Voltage Operation........................................................................................ 28
2.8.4 Alarm Connections......................................................................................................................................... 29
2.9 Installing Pump/Valve Pinch Plate........................................................................................................................... 31
START UP................................................................................................................................................................... 33
SECTION 3 STARTING THE INSTRUMENT................................................................................................ 35
3.1 Introduction.............................................................................................................................................................. 35
3.2 Installing the Reagents............................................................................................................................................. 35
3.2.1 Priming the Reagents (optional)..................................................................................................................... 36
3.3 Installing the Stir Bar ............................................................................................................................................... 36
3.4 Supplying Sample .................................................................................................................................................... 37
3.5 Supplying Power...................................................................................................................................................... 37
3.6 Two-point Standardization....................................................................................................................................... 38
4
TABLE OF CONTENTS, continued
SECTION 4 OPERATION ...................................................................................................................................... 39
4.1 Instrument Operation................................................................................................................................................ 39
MAINTENANCE....................................................................................................................................................... 41
SECTION 5 MAINTENANCE............................................................................................................................... 43
5.1 Scheduled Maintenance............................................................................................................................................ 43
5.1.1 Replenishing the Reagents.............................................................................................................................. 43
5.1.2 Replacing the Pump Tubes.............................................................................................................................. 43
5.1.3 Tubing Replacement ....................................................................................................................................... 44
5.2 Unscheduled Maintenance........................................................................................................................................ 46
5.2.1 Changing Instrument Alarm Point .................................................................................................................. 46
5.2.2 Fuse Replacement .......................................................................................................................................... 46
5.2.3 Cleaning the Instrument Enclosure................................................................................................................. 47
5.2.4 Replacing the Sample Conditioning Filter...................................................................................................... 47
5.2.5 Reagent Spill Clean Up Instructions............................................................................................................... 47
SECTION 6 TROUBLESHOOTING.................................................................................................................... 49
GENERAL INFORMATION.................................................................................................................................. 51
REPLACEMENT ITEMS .............................................................................................................................................. 53
HOW TO ORDER.......................................................................................................................................................... 54
REPAIR SERVICE......................................................................................................................................................... 55
WARRANTY ................................................................................................................................................................. 56
INDEX............................................................................................................................................................................ 57
5
CERTIFICATION
Hach Company certifies this instrument was tested thoroughly, inspected, and
found to meet its published specifications when it was shipped from the factory.
The SP-510™ Hardness Monitor has been tested and is certified as indicated to
the following instrumentation standards:
Product Safety
The SP-510™ was tested for safety as follows:
UL 3101-1 (ETL Listing # H0492805390)
CSA C22.2 No. 1010.1 (ETLc Certification # H0492805390)
Certified by Hach to EN 61010-1 (IEC1010-1) per 73/23/EEC, supporting test
records by Intertek Testing Services.
Immunity
The SP-510™ Hardness Monitor was tested for EMC as follows:
EN 61326:1998 (EMC Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use) per 89/336/EEC EMC:
Supporting test records by Hach Company, certified compliance by
Hach Company.
Standards include:
IEC 1000-4-2:1995 (EN 61000-4-2:1995) Electro-Static Discharge Immunity
(Criteria B)
IEC 1000-4-3:1995 (EN 61000-4-3:1996) Radiated RF Electro-Magnetic
Field Immunity (Criteria A)
IEC 1000-4-4:1995 (EN 61000-4-5:1995) Electrical Fast Transients/Burst
(Criteria B)
IEC 1000-4-5:1995 (EN 61000-4-5:1995) Surge (Criteria B)
IEC 1000-4-6:1996 (EN 61000-4-6:1996) Conducted Disturbances Induced
by RF Fields (Criteria A)
IEC 1000-4-11:1994 (EN 61000-4-11:1994) Voltage Dip/Short Interruptions
(Criteria B)
Additional Immunity Standard/s include:
ENV 50204:1996 Radiated Electro-Magnetic Field from Digital Telephones
(Criteria A)
Emissions
The SP-510™ Hardness Monitor was tested for Radio Frequency Emissions
as follows:
Per 89/336/EEC EMC: EN 61326:1998 (Electrical Equipment for measurement,
control and laboratory use-EMC requirements) Class "TBD" emission limits.
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.
Standards include:
EN 61000-3-2 Harmonic Disturbances Caused by Electrical Equipment
EN 61000-3-3 Voltage Fluctuation (Flicker) Disturbances Caused by
Electrical Equipment
6
CERTIFICATION, continued
Additional Emissions Standard/s include:
EN 55011 (CISPR 11), Class "TBD" (Test date 4-21-99) emission limits
Canadian Interference-causing Equipment Regulation, IECS-003, Class A
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.
This Class A digital apparatus meets all requirements of the Canadian
Interference- Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement
sur le matériel brouilleur du Canada.
FCC PART 15, Class "A" Limits
Supporting test records by Hewlett Packard, Fort Collins, Colorado Hardware Test
Center (A2LA # 0905-01) and certified compliance by Hach Company.
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions:
(1) this device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may cause
undesired operation.
Changes or modifications to this unit not expressly approved by the party
responsible for compliance could void the user's authority to operate the
equipment.
This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with
the instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference, in which case the user will be required to correct the interference at
his own expense. The following techniques of reducing the interference problems
are applied easily.
1. Disconnect the SP-510™ Harness Monitor from its power source to verify
that it is or is not the source of the interference.
2. If the SP-510™ Harness Monitor is connected into the same outlet as the
device with which it is interfering, try another outlet.
3. Move the SP-510™ Harness Monitor away from the device receiving the
interference.
4. Reposition the receiving antenna for the device receiving the interference.
5. Try combinations of the above.
7
SAFETY PRECAUTIONS
This symbol, if noted on the product, indicates the need for protective
eye wear.
Section 2.1 Instrument Environmental Considerations on page 17
Section 2.4 Plumbing Connections on page 21
Section 2.6 Optional Air Purge on page 23
Section 2.8 Electrical Connections on page 26
Section 2.8.1 Power Connections on page 26
Section 2.8.2 Wiring the Instrument on page 28
Section 2.8.3 Voltage Selection for Alternate Voltage Operation on
page 28
Section 2.8.4 Alarm Connections on page 29
SECTION 3 STARTING THE INSTRUMENT on page 35
SECTION 5 MAINTENANCE on page 43
8
SAFETY PRECAUTIONS, continued
Section 5.1 Scheduled Maintenance on page 43
Section 5.2 Unscheduled Maintenance on page 46
Section 5.2.2 Fuse Replacement on page 46
9
SPECIFICATIONS
GENERAL SPECIFICATIONS
Hard/Soft Display: Two LEDS with one indicating hard water and the second
one indicating soft water
Time Delay: A soft-to-hard water alarm transition requires two successive
readings above the trip point. A hard-to-soft water alarm clearing transition
requires one reading below the trip point.
Trip Points:
1.0, 2.0, 5.0, 10.0, 20.0, 50.0, and 100.0 mg/L hardness measured as CaCO
3
Reagents Required: Buffer and indicator for the desired trip point
Reagent Bottle Volume: 500 mL
Reagent Shelf Life: 1year
Reagent Consumption Rate: 500 mL of each reagent every two months
Sampling Rate:
New sample every 1.9 minutes ± 5% at 60 Hz; new sample every 2.3 minutes
±5%at50Hz
Controls: On/Off Switch, ALARM OFF, ZERO CAL, HARD CAL
Service Requirements:
Restandardization every 2 months. Replace pump tubes every 6 months.
Trip Point Buffer Cat. No. Indicator Cat. No.
1 mg/L 27685-49 27690-49
2 mg/L 27685-49 27691-49
5 mg/L 27685-49 27692-49
10 mg/L 27686-49 27692-49
20 mg/L 27687-49 27692-49
50 mg/L 27688-49 27692-49
100 mg/L 27689-49 27692-49
10
SPECIFICATIONS, continued
PERFORMANCE SPECIFICATIONS
Repeatability:
±10% of trip value on 1 and 2 mg/L ranges; ±4% of trip value on
remaining ranges.
ENVIRONMENTAL SPECIFICATIONS
Operating Humidity Range: 5 to 95%, noncondensing
Storage Temperature Range: -40 °C to +60 °C (-40 °F to +140 °F)
Operating Temperature Range: Cto4C(32°Fto10F)
Influence of Temperature On Trip Point:
ELECTRICAL SPECIFICATIONS
Power Requirements: 115/230 Vac, (switch selectable inside instrument)
70 VA, 50/60 Hz, 1.25 Amp Fuse (see Section 2.8.3 on page 28).
Installation Category: II
Electrical Connection:
Two 3-wire barrier terminal blocks. Wire range 12
18 AWG. Holes for two
1/2" conduit fittings are provided.
Hard Water Alarm Relay: SPDT relay, actuated when hard water indicator is
on. (Alarm may be disabled from front keypad.)
Contact ratings: 5 A resistive at 100-240 Vac
Nominal Alarm Point Minimum Trip Value Maximum Trip Value
1mg/L 0.75mg/L 1.25mg/L
2 mg/L 1.5 mg/L 2.5 mg/L
5mg/L 3.75mg/L 6.25mg/L
10 mg/L 7.5 mg/L 12.5 mg/L
20 mg/L 15.0 mg/L 25.0 mg/L
50 mg/L 37.5 mg/L 62.5 mg/L
100 mg/L 75.0 mg/L 125.0 mg/L
1 mg/L –0.03 mg/L/°C
2 mg/L –0.03 mg/L/°C
5 mg/L –0.06 mg/L/°C
10 mg/L –0.08 mg/L/°C
20 mg/L –0.09 mg/L/°C
50 mg/L –0.29 mg/L/°C
100 mg/L –0.60 mg/L/°C
SPECIFICATIONS, continued
11
OPTICAL SPECIFICATIONS
Light Source: Class 1 LED (Light Emitting Diode) with peak wavelength of
610 nm. Estimated 50,000 hours minimum life.
Detector: Silicon photo detector
Optical Path Length: 0.35"
MECHANICAL SPECIFICATIONS
Enclosure: Environmental Rating IP62
Dimensions: Approximately 42 x 31.5 x 18 cm (16.5 x 12.5 x 7") (refer to the
transmittal drawing for exact space requirements).
Case Construction: IP62 instrument enclosure with gasketed door
Mounting: Wall-mounted
Shipping Weight: 11.3 kg (25 lbs)
HYDRAULIC (SAMPLING) SPECIFICATIONS
Sample Flow Rate to Sample Conditioning: 50 to 500 mL/minute flow
rate required.
Inlet Pressure to Instrument: 1 to 5 psig, 1.5 psig is optimum. Exceeding 5 psig
can cause sample tubing failure.
Inlet Pressure to Sample Conditioning: 1.5psigto75psig(withthesample
tube level with the bottom of the instrument)
Inlet Fitting:
At instrument, ¼" OD polyethylene tubing with quick-disconnect fitting
Drain Fitting: Hose barb for ½" ID flexible tubing
Sample Temperature Range: Cto4C(41°Fto10F)
Sample Conditioning: use provided sample conditioning
12
13
SECTION 1 GENERAL INFORMATION
1.1 Instrument Description
The Model SP-510™ Hardness Monitor is designed to monitor water softener
effluents continuously to detect hardness breakthrough due to softener exhaustion.
It enables commercial and industrial water softener operators to establish
automatic control of their systems by initiating regeneration sequences with the
instrument's alarm circuit. The Hardness Monitor is also suited for other
applications requiring the monitoring of hardness.
By selecting the appropriate hardness indicator and buffer reagents, the
monitor will alarm at 1, 2, 5, 10, 20, 50, or 100 mg/L hardness measured as
CaCO
3
. When the preselected alarm point is exceeded, the alarm relay responds,
closing the normally open contacts and opening the normally closed contacts.
These dry contacts can be used to actuate annunciators and/or initiate softener
regeneration. Control panel indicators provide “hard” or “soft” sample status.
The SP-510™ Hardness Monitor enclosure is environmentally rated for IP62 per
IEC 529. The enclosure is dust-tight, and drip-resistant but is not designed for
outdoor use.
The instrument is designed so electronic components are isolated from the
hydraulic components. Windows allow the operator to observe the display
indicators and reagent supply without opening the instrument enclosure.
Figure 1 Model SP-510™ Hardness Monitor
14
SECTION 1, continued
1.2 Principle of Operation
The Hardness Monitor uses a buffered colorimetric hardness indicator solution
that is mixed with the water sample as it flows through the instrument. While in
the colorimeter sample cell, light is transmitted through the solution to a detector,
which provides the colorimetric measurement.
Note: The Class 1 Light Emitting
diode(LED) used as thelightsource
for the photometric measurement
does not present any potential for
eye damage.
Indicator solutions are formulated to provide 50% transmittance (%T) to the
detector at their nominal alarm point values. Below 50%, the color of
sample/reagent mixture is blue. As hardness increases, the color becomes red.
Because the light source for the colorimetric measurement is an orange light-
emitting diode, a blue solution blocks light to the detector while a red solution
allows light to be transmitted unattenuated through the measurement cell to the
detector. Maximum light transmittance through the cell occurs when all of the
indicating reagent available has reacted with sample hardness and any further
increasing levels of hardness have no effect on solution color.
A linear, peristaltic pump/valve controls the flow of incoming sample and injects
metered volumes of the buffer and indicator reagents on a sequenced cycle. The
pump/valve uses a motor-driven cam to operate pinch blocks that squeeze special
thick-walled tubing against a fixed plate, the sequence operates as follows:
1. The sample inlet line opens and remains open to allow sufficient sample
flow to purge the inlet line and measure cell. While the inlet sample is
flowing, the reagent inlet valve and pinch blocks are opened, filling the tubing
with reagent.
2. The sample inlet line is then pinched shut, leaving a fresh sample in the
measuring cell. The cell volume is controlled by an overflow weir.
3. Reagent inlet valve block is then closed.
4. Reagent outlet valves open and the pinch block squeezes the tubing, forcing
reagent into the measuring cell. A magnetic stirrer mounted beneath the
measuring cell mixes sample and reagent in the cell. Sufficient time is allowed
for sample and reagents to react.
5. An optical interrupter circuit on the drive motor shaft signals the monitor
electronics to take a measurement.
This cycle is repeated approximately every 2 minutes.
15
DANGER
Some of the tasks in this section of the manual have safety issues associated with them. Because the potential for injury to
individuals and equipment exists when these safety issues are not addressed, Hach Company strongly recommends that
qualified personnel conduct the installation, and that all installation personnel review the associated instructions carefully.
PELIGRO
Algunas de las tareas comprendidas en esta sección del manual pueden ocasionar daños a las personas y al material si no
observan la medidas de suguridad. Hach Company recomienda encarecidamente que el material sea instalado por un personal
cualificado y que el personal encargado de la instalación lea atentamente estas instrucciones.
PERIGO
A execução de algumas tarefas previstas nesta secçao do manual pode causar ferimentos às pessoas ou estragos no
equipamento se não forem observadas precauções de suranca. A Hach Company recomenda vivamente que o equipmento seja
instalado por pessoal qualificado e que todas as pessoas afectadas à sua instalação leiam atentamente estas instruções.
DANGER
Certaines tâches dans ce chapitre du mode d'emploi peuvent causer des blessures aux personnes et endommager le matériel si
les consignes de sécurité ne sont pas suivies. Hach Company recommande vivement que l'installation soit faite par du personnel
qualifié et que toutes les personnes effectuant l'installation lisent attentivement ces instructions.
GEFAHR
Einige der in diesem Abschnitt der Betriebsanleitung beschriebenen Arbeiten können bei Nichtbeachtung der
Sicherheitsvorschriften zu Verletzungen von Personen oder Schäden am Gerät führen. Es wird dringend empfohlen, die
Installation ausschließlich von qualifiziertem Personal duchführen zu lassen; mit der Installation befaßte Personen sollten
diese Anweisungen aufmerksam lesen.
INSTALLATION
16
17
SECTION 2 INSTALLATION
2.1 Instrument Environmental Considerations
The instrument enclosure has been tested to IEC 529 IP 62 standards and is
designed for general-duty, indoor installation. Ambient temperatures may fall
within the range of to 40 °C (18° to 104 °F) but they should be nearly constant.
Do not mount the instrument in direct sunlight; shield from dripping water.
2.2 Unpacking the Instrument
Remove the Hardness Monitor from its shipping carton and inspect it for any
damage that may have occurred during shipment. Verify that the Installation Kit,
Cat. No. 55164-00, and Maintenance Kit, Cat. No. 55165-00, are included in
addition to reagents.
If damage is evident or the shipment is incomplete, please contact the Customer
Service Department, Hach Company, Loveland, Colorado for instructions. The
toll-free number is 800-227-4224. Contents of the Installation and Maintenance
kits are listed in Replacement Parts.
18
SECTION 2, continued
2.3 Wall Mounting the Instrument
The instrument case is designed for wall mounting. Refer to Figure 2, Figure 3,and
Figure 4 for dimensions and other installation information. Use ¼-inch screws for
mounting. Mount the instrument as close to the sampling point as practical to ensure
complete purging of the sample line each cycle. Leave adequate clearance at the
sides and bottom of the instrument case for plumbing and wiring connections.
Figure 2 SP-510 Instrument Dimensions (1 of 3)
SECTION 2, continued
19
Figure 3 SP-510 Instrument Dimensions (2 of 3)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61

Hach SP-510 Instrument Manual

Type
Instrument Manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI