Hydrim L110W P1 & P2 Cycle with Revision 2.02 Software
1. Preparation drain pump runs
When the P1 or P2 cycle button is pressed the drain pump turns ON. The Hydrim will drain
the wash chamber and the sidebag. If the previous cycle was interrupted when the cycle is
started the drain pump will run for 80 seconds. If the previous cycle was a complete cycle
the drain pump will run for 20 seconds. If the unit fails to drain an E4 code will appear on
the display.
2. Sidebag fills with hot water target temperature 50°C/122°F
Turn ON Y7 hot water valve. Start timer whenever valve is open to calculate the Water
Debit Constant and Error Timer for filling sidebag. If water does not close pressure switch
B1 within 130 seconds declare “no hot water” and use just cold water. If pressure switch
does not close after another 130 seconds Hydrim fails with an E2 code. When the pressure
switch closes calculate the Water Debit Constant time for filling up 2.7 liters.
3. Fill wash chamber target temperature 28°C/82.4°F
When sidebag is full and pressure switch B1 activates (blue arm moves out) turning ON Y6
sidebag drain valve to allow sidebag to drain into the chamber. Pressure switch will
deactivate when sidebag starts draining into chamber. After approximately 45 seconds turn
ON circulation pump M1 and Y7 hot water valve. The on time for the hot water inlet valve
depends on the “Water Debit Constant” that was calculated previously. The chamber
capacity is 4 liters, 2.7 liters from the sidebag and 1.3 liters from the hot water inlet valve,
resulting valve Y6 is on for approximately 75 seconds and valve Y7 is on for 1.3/Water
Debit Constant. When water inlet valves timer expires add 10 seconds to the sidebag drain
valve Y6 before closing. Failure to fill chamber will cause an E2 code.
4. Circulation, heating and dosing
Start circulation pump M1, dosing pump M4 and turn ON water heater R1 if water
temperature in chamber is below target temperature. Start heating and dosing error timers.
Dosing pump runs for 7 seconds. If flow switch does not detect dosing before the dosing
error timer ends (30 seconds), the Hydrim fails with an ED code. Once selected dosing time
is achieved the dosing pump will turn OFF.
5. Circulation and heating target temperature 50°C/122°F
Circulation pump continues to run and water heater remains turned ON. If the heating error
timer ends and the chamber water temperature is below the target temperature, the Hydrim
fails with an E1 code.
6. Circulation, heating and dosing target temperature 50°C/122°F
Circulation pump continues to run, and water heater remains turned ON. When water
temperature in chamber reaches target temperature start dosing error timer. The dosing
pump runs for 7 seconds. If flow switch does not detect dosing before the dosing error timer
ends (30 seconds), the Hydrim fails with an ED code. Once selected dosing time is
achieved the dosing pump will turn OFF.
7. Circulation
Circulation pump M1 remains ON for 300 seconds/5 minutes.
8. Circulation and evacuation 10 seconds
Circulation pump M1 remains ON. Turn ON drain pump M2.
9. Evacuation 20 seconds
Circulation pump M1 turns OFF. Drain pump M2 remains ON for 20 seconds.
10. Fill wash chamber target temperature 40°C/104°F without filling sidebag
Turn ON valve Y6 and valve Y7 target temperature 40°C/104°F. The ON time for the water
inlet valve depends on the “Water Debit Constant” that was calculated previously. Turn
valve Y7 ON for 3/Water Debit Constant and Y6 for 3/Water Debit Constant + 10 seconds.
When water inlet valves timer expires add 10 seconds to the sidebag drain valve Y6.