SciCan HYDRIM L110w G4 User manual

Type
User manual
HYDRIM
®
L110w G4
INSTRUMENT WASHER
HYDRIM L110w G4 Operator’s Manual 95-113645 CA ML R2 Copyright 2013 SciCan Ltd. All rights reserved.
Operator’s Manual
Manuel de l’utilisateur
Table of Contents
1. Introduction ......................1
2. Important Information ..............2
3. Disclaimers ......................3
4. Installation Instructions.............4
5. Instructions for Use ...............9
5.1 Instrument Reprocessing
Recommendations ..................9
5.2 Cycle Description Chart .............10
5.3 Hints and tips for best cleaning results . 11
5.4 Recommended load congurations ....12
5.5 Baskets and Accessories............14
5.6 Touchscreen Overview .............15
5.7 Set up menu overview ..............16
5.8 - 5.22 Setup menu options ...........17
5.23 Connecting to a network ............22
5.24 Connecting to a wireless network .....22
5.25 Activating the HYDRIM L110w G4.....23
5.26 Interrupting / aborting a cycle ........24
5.27 Additional user
and setup menu functions ...........25
5.28 Opening the Door in
Case of Power Failure ..............26
5.29 Storing and Retrieving
Cycle Information ..................27
5.30 Printer / Data Logger Setup ..........29
6. Care and Maintenance ...........30
6.1 Replacing the HIP™
Cleaning Solution ................. 30
6.2 Relling the Water Softener
Salt Reservoir...................... 32
6.3 Cleaning the HYDRIM L110w G4 .....32
6.4 Filter and Wash Arm Maintenance ....33
6.5 Preparing the unit for shipping .......33
6.6 Annual maintenance ...............34
7. Troubleshooting .................36
8. Specications ..................38
9. Serial Number Plate .............38
10. Warranty ......................39
HYDRIM and STATIM are registered trademarks of SciCan Ltd. BRAVO,
HIP, and SysTM are a trademarks of SciCan Ltd. All other trademarks referred
to in this manual are the property of their respective owners.
For all service and repair inquiries:
In Canada 1-800-870-7777
United States: 1-800-572-1211
Germany:
+49 (0)7561 98343 - 0
International: (416) 446-4500
Manufactured by:
SciCan Ltd.
1440 Don Mills Road,
Toronto ON M3B 3P9
CANADA
Phone: (416) 445-1600
Fax: (416) 445-2727
Toll free: 1-800-667-7733
EU Representative
SciCan GmbH
Wangener Strasse 78
88299 Leutkirch GERMANY
Tel.: +49 (0)7561 98343 - 0
Fax: +49 (0)7561 98343 - 699
SciCan Inc.
701 Technology Drive
Canonsburg, PA 15317 USA
Phone: +1 724 820 1600
Fax: +1 724 820 1479
Toll free: 1-800-572-1211
SciCan Medtech
Alpenstrasse 16
CH-6300 ZUG SWITZERLAND
Phone: +41 (0) 41 727 7027
Fax: +41 (0) 41 727 7029
Thank you for selecting the HYDRIM L110w G4 instrument washer. The HYDRIM
complements the STATIM
®
and BRAVO™ autoclaves by quickly and hygienically
preparing soiled instruments for sterilization. To ensure years of safe, trouble-free service,
carefully review this Operator’s Manual before operating the unit. The HYDRIM L110w G4
is suitable only for the applications listed in this manual. Using the HYDRIM for other
purposes may be dangerous. The manufacturer cannot be held responsible for damage
caused by improper use.
Operational, maintenance and replacement instructions must be followed for the product
to perform as designed. All trademarks referred to in this manual are the property of their
respective owners. Contents of this manual are subject to change without notice to reect
changes and improvements to the HYDRIM product.
The HYDRIM L110w G4 should only be installed and serviced by a qualied contractor
as it is an Installation Category 2 device. Before connecting the unit, the installer should
check that the voltage and frequency of the electrical supply correspond with the details
in the installation instructions. The unit must only be plugged into a grounded outlet.
Please note that this unit is completely isolated from the electrical supply only when it is
unplugged or the circuit breaker is turned OFF. This must be done before any repair work
is carried out.
The HYDRIM L110w G4 uses HIP™ Cleaning Solution exclusively. The unit will not
perform as described if other detergents / cleaning products are used. Other detergents /
cleaning products may damage the unit, the instruments, cause excessive foaming, and
void the warranty. Do not use solvents in this unit. Do not allow small children or
unauthorized personnel access to this unit or its controls.
1. Introduction
1
2. Important Information
The following symbols appear in this book.
Caution, there is a potential hazard to the operator.
This situation may lead to a mechanical failure.
This is important information.
2
Do not permit any person other than authorized personnel to supply parts for service or
maintenance for your HYDRIM. SciCan is not liable for incidental, special or consequential
damages caused by any maintenance or services performed on the HYDRIM by unauthorized
personnel, or for the use of equipment or parts manufactured by a third party, including lost prots,
any commercial loss, economic loss, or loss arising from personal injury.
Never remove any unit panels and never insert objects through holes or openings in the cabinetry.
Doing so may damage the unit and/or pose a hazard to the operator.
If the unit is used in a manner other than that specied, the protection provided by the equipment
may be impaired.
Cleaning solutions may irritate. Avoid contact with eyes, skin and mouth.
Never sit, stand or lean on the open door. The unit may tip forward causing injury.
Always turn the unit OFF before adding softener salt, adding solutions or performing routine
maintenance to the unit.
The owner shall not allow any personnel other than the trained and authorized personnel
to operate the unit.
SciCan recommends leaving the unit turned ON at all times, in particular over times when no staff
is in the ofce or when no water shut-off mechanism is installed. Leaving the unit turned ON will
activate the drain pump if water is leaking into the chamber as a result of any faulty circumstance.
For further information or questions about the HYDRIM L110w G4, contact your authorized dealer.
3. Disclaimers
3
Installation:
Installation should be undertaken by a
SciCan approved technician. The use of
an unapproved installer may invalidate
the warranty.
If the HYDRIM is installed in a Sterilization
Center, the manufacturer of the Sterilization
Center should allow a 10mm / 1/2” space
at the top, back and both sides of the HYDRIM.
This will facilitate installation, and service
access to the HYDRIM.
Do not move the HYDRIM into place by
maneuvering the open wash chamber door.
This may cause to door to become misaligned
and leak.
During installation, all consumables should
have been added to the machine as appropriate.
It is important that this be done before starting
the machine.
4. Installation Instructions
To open the wash chamber door, plug in the unit,
turn it on and press the unlock symbol.
Alternatively, remove the screw in the kickplate
and pull the ring. See section 5.28.
Tools and supplies required for installation:
Slot screwdriver
Channel locks
Specications:
Voltage: 208 240 V
Frequency: 60 Hz
Rated load: 2.5 kW
Circuit
breaker: 15 A
per phase
Height: 850 mm / 33.5”
Width: 598 mm / 23.54”
Depth: 600 mm / 23.75”
Depth with
door open: 1200 mm / 47”
Weight: 80 kg / 176 lbs
Max. Running Noise: 78 dBA
Utility Hook-ups
For safety reasons, do not install
the HYDRIM without the top cover.
4
Levelling the HYDRIM:
The unit is standing on three supports: rollers (wheels) at the
back and two legs at the front.
1. Remove the front kickplate. Push the HYDRIM
into place while lifting the strap at the front to allow the unit
to roll on the rollers.
2. Adjust the front legs as required until the
HYDRIM is level.
3. The rear two legs are used only if the oor is uneven or
cannot provide support to the rollers.
4. Tuck the strap under the machine before
replacing the kickplate, so as not to interfere
with the exhaust air vents.
RO water
inlet hose
hot water
inlet hose
power
connector
ethernet
connector
printer
connector
drain hose
fuse holder
cold water
inlet hose
Rear of Unit
Hose / Length / Max. Distance Water Shut-off
Cord Diameter Distance from Pressure valve
inlet / drain
Hot Inlet 1.9 m / 6 ft. 1.5 m / 5 ft. 1-10 bar / Yes
3/4” 14.5-145 psi
Cold Inlet 1.9 m / 6 ft. 1.5 m / 5 ft. 1-10 bar / Yes
3/4” 14.5-145 psi
R/O Inlet 1.9 m / 6 ft. 1.5 m / 5 ft. 1-10 bar / Yes
3/4” 14.5-145 psi
Drain 1.5 m / 5 ft.
3/4”
Electrical 1.8 m
AWG 14-3
4. Installation Instructions
Connecting the water inlet hoses:
Connect the hot and cold water inlet hoses to the hot and cold water supply.
Drain requirements:
Connect the drain hose to the drain outlet. The drain hose can be attached to existing drain
lines using a 3.5 cm / 1.5” or larger standpipe / P-trap combination. If the hose is connected
directly to the drain line, ttings and adapters should not reduce water ow. The drain hose
should be attached to the main drain at a point no more than 1 metre /3 ft. above the base of
the HYDRIM. A oor drain is acceptable (check local codes).
Electrical requirements:
Connect the electrical cord to the power supply. The HYDRIM must have a dedicated circuit
equipped with a 15A fuse. The outlet must be grounded.
Additional Information:
The HYDRIM unit is heavy. Use a forklift or dolly and exercise caution when moving it.
The HYDRIM should be serviced on site.
The HYDRIM is equipped with an air gap / anti-suction device to prevent backow of dirty
water into the water supply. No other air gap device is necessary.
If you need to extend the water inlet and drain hoses, ensure that you use commercial grade
plumbing hose.
The maximum length of the drain hose is 3.3 m / 13 ft.
Installing the water softener salt:
Unscrew the salt container lid and pour 1 litre / 1 quart of water into
the water softener. Add 1 kg / 2.2 lbs of water softening salt in the
same manner. Screw the salt container cap on tightly.
Handling and installing the HIP™ cleaning solution
The HIP™ cleaning solution is used in dilute form as a detergent for the cleaning
of surgical instruments in the HYDRIM range of washers and washer disinfectors. It is a
colourless and almost odourless liquid which is completely soluble in water.
The pH range of the concentrated solution is between 9.3 to 9.8 and the solution is
therefore alkaline. It is therefore advised that certain precautionary measures are observed
when handling and during the loading of the solution into the machine, and when removing
and disposing of the empty box.
5
4. Installation Instructions
Recommended precautionary measures with regard to safe handling of HIP™ cleaning solution.
Always wear undamaged gloves made of natural or butyl rubber, nitrile, or neoprene
(Surgical gloves are ideal).
Always wear eye protection.
Protect exposed skin on arms.
If solution is spilled on clothing, remove the item of clothing and wash affected skin
with plenty of water, in accordance with the Material Safety Data Sheet requirements.
Use good industrial hygiene practices in handling this material. When handling,
do not eat or drink.
After handling the product, remove and dispose of gloves and wash hands prior to
removing eye protection.
Do not use the personal protective equipment you use for clinical activities to handle
HIP™ products.
Always ensure that personal protective equipment, if contaminated with HIP™
cleaning solution, is cleaned, or disposed of as appropriate.
For comprehensive data on HIP™ cleaning solution, please visit SciCan.com to
view or download a PDF of the Material Safety Data Sheet.
To install the HIP™ cleaning solution box, follow these steps:
salt container
Installing Cleaning Solution:
Install HIP™ Cleaning Solution as follows:
Technician Installation Test: Turn on the shut-off valves. Run a cycle, checking for leaks in
the plumbing connections.
6
Water quality
The quality of the water being used in the HYDRIM to clean the instruments is very critical
to achieving satisfactory cleaning results and to protecting the instruments and the internal
parts of the unit from deterioration.
Drinking water typically contains many dissolved solids. The amount of dissolved solids
greatly depends on the local natural geological conditions and they can cause stains, spots
and corrosion on instruments and on the internal parts of the HYDRIM. Among others, Iron,
Manganese, Chloride, and Calcium Carbonate (CaCO
3
) are the dominant dissolved solids
that affect the cleaning results when using a washer.
Iron and Manganese can cause orange and brown or black stains on the instruments and
on the internal parts of the HYDRIM. Chloride is usually responsible for causing stains,
spotting, pitting and scaling. If the instruments or the inner parts of the HYDRIM L110w G4,
such as the chamber, show any of these spots or stains, a water test may be required to
determine the cause. The installation of a water treatment system may be required to
reduce the amount of dissolved solids in the water and to improve the cleaning
performance of the HYDRIM.
Calcium Carbonate is the principle cause of water hardness and leaves white spots or
scales. The HYDRIM is equipped with a built-in water softening system that must be
adjusted according to the local water hardness. Please see the next section of this manual
for recommended regeneration settings.
Before using the HYDRIM, SciCan recommends testing the water and recording
the results for water hardness, pH value, and water conductivity for future reference.
IMPORTANT: The HYDRIM's water softening system reduces the water hardness by
taking out Calcium Carbonate. If your water testing results show that your water hardness
is outside the unit's range of adjustment, or if other dissolved solids in the water cause
stains or deposits on the instruments or chamber, an external water treatment system may
be required.
4. Installation Instructions
7
4. Installation Instructions
Testing water quality
The HYDRIM L110w G4 is equipped with
a built-in water softening system that must
be adjusted according to the local water
hardness. To read local water hardness,
proceed as follows:
1. The water test kit included with your
HYDRIM contains three water
hardness test strips in bags. Take a
water sample from the location where
the machine will be installed.
2. Open one of the bags, remove the
test strip and dip it into the water.
3. Compare the color of the strip with
the chart on the back of the bag
Determine the water hardness according
to the chart on the water test kit envelope.
4. Power the unit on and select the
Settings key from the main menu.
5. Go to the Setup Menu and select
“Set Regeneration”.
6. Using the up and down arrows, set the
water softener regeneration level
according to the water hardness table
in this section. If your water hardness
falls between two settings, select the
higher setting.
7. Unscrew the water softener container
lid from the bottom right of the chamber
and pour at least 1.0 litres (0.26 gal) of water
into the water softener container. Then fill the
salt container to the top (maximum 1 kg/
2.2 lbs.) with salt. Close by screwing the lid
tightly back into place and run a complete
cycle with no instruments. An improper seal can lead to corrosion.
*Please note: The water test strip is only accurate up to 250 ppm. If the reading on the test strip exceeds
250 ppm and/or if the location in which the HYDRIM is installed has known water quality problems, having
a more detailed and accurate water test done by a test lab is strongly recommended.
Water Hardness Conversion and salt regeneration levels
°dH
US
GPG
PPM
(mg CaCO3 / Litre)
Regen.
1 1.0 18
2 2.1 36
3 3.1 54
Very Soft
4 4.2 71
5 5.2 89
5.6 5.8 100
1
6 6.3 107
6.2 6.4 110
7 7.3 125
8 8.3 143
2
8.4 8.8 150
9 9.4 161
10 10.4 178
10.1 10.5 180
11 11.5 196
3
11.2 11.7 200
11.8 12.3 210
Typically
No Treatment
Necessary
(values from 18-143)
External Treatment
Required
(values >535)
Additional
Water
Treatment
Required
May Require External Treatment
(values from 150-535)
12 12.5 214
13 13.6 232
4
14 14.6 250*
15 15.6 268
16 16.7 286
5
16.8 17.5 300
17 17.7 303
18 18.8 321
19 19.8 339
6
19.6 20.5 350
20 20.9 357
20.2 21.0 360
21 21.9 375
22 22.9 393
22.4 23.4 400
23 24.0 411
24 25.0 428
25 26.1 446
7
25.2 26.3 450
26 27.1 464
27 28.2 482
28 29.2 500
28.6 29.8 510
29 30.2 518
30 31. 3 535
30.3
31.6
540
8
8
5.1 Instrument Reprocessing Recommendations
The HYDRIM L110w G4 is designed to clean general dental and medical instruments.
Terminal sterilization MUST follow processing in the HYDRIM L110w G4.
Prior to placing instruments in the HYDRIM, consult the instrument manufacturer’s reprocessing
instructions.
Examples of medical and dental instruments suitable for cleaning in the HYDRIM L110w G4
include:
• Stainless steel scalers, spatulas and hemostats.
Stainless steel retractors, needle holders, skin hooks, tissue forceps, scissors, curettes, and
other stainless steel instruments free of deep cavities.
For best results, instruments processed in the HYDRIM L110w G4 should
have the following properties:
Heat resistance to a temperature of up to 70°C / 158°F
Corrosion resistance in the presence of heat and alkalinity.
Burs, microkeratomes, phaco tubing, beroptics and electrical equipment should not be processed
in the HYDRIM.
Nickel-plated instruments, or instruments containing aluminum or colour anodized aluminum
as well as single use instruments are not suited for processing in the HYDRIM.
HIP™ (HYDRIM Cleaning Solution with Instrument Protection) has been carefully formulated to
provide outstanding material compatibility with a wide range of metals and coatings. However,
some manufacturers’ instruments or instrument coatings may not be compatible.
Please consult the instrument manufacturer for their recommended cleaning procedures.
In general, it is not recommended to process carbon steel and chrome plated instruments
in the HYDRIM.
All hinged instruments should be placed in open position in open baskets or in the hinged
instrument rack (SciCan part number 01-110409S) in P3 Heavy-Duty cycle.
When processing small or light objects in the HYDRIM use the basket with lid (SciCan part number
01-109966S).
Only the outer surfaces of dental handpieces can be cleaned in the HYDRIM L110w G4 with
the handpiece holder (9 handpieces, PN 01-113113S or 21 Handpieces, PN 01-113112S).
Set-able materials, such as cements, amalgam, and composites need to be removed from the
instruments after use at chairside. The HYDRIM will not be able to clean them off the instruments.
.
5. Instructions for Use
9
5. Instructions for Use
5.2 Cycle Description Chart
10
* This is not a wash cycle. Always run a wash cycle following the rinse & hold cycle.
** Cycle times depend on the temperature and pressure of incoming water.
*** Rinse times are adjustable up to 10 minutes by an authorized technician.
Cycle Prewash Wash
Intermediate
Rinse
Rinse Dry
Total Time
**
w/o Drying
Water
Consumption
P0 – Machine
Cleaning Cycle
No initial draining.
<30ºC/86°F
(cold)
3 minutes
N/A N/A
<30ºC (cold)
2 minutes
N/A
7
minutes
16 L
4.23 Gal
P1 – Rinse and
Hold Cycle*
Use to prevent
soil from drying on
instruments when they
will not be washed
within one hour.
<30ºC/86°F
(cold)
3 minutes
N/A N/A
60ºC/140°F
1 minute
N/A
17
minutes
16 L
4.23 Gal
P2 – Regular
Cycle
Use for moderately
soiled loose
instruments.
N/A
50ºC/122°F
5-15 minutes
(default 5
minutes)
<60ºC/140°F
1 minute
60ºC/140°F***
1-10 minutes
(default 1 minute)
1-20
minutes
(default 10
minutes)
25
minutes
24L
6.34 Gal
P3 – Heavy Duty
Cycle
Use for heavily soiled
instruments and
cassettes.
<30ºC/86°F
(cold)
3 minutes
50ºC/122°F
5-15 minutes
(default 9
minutes)
<60ºC/140°F
1 minute
60ºC/140°F
1 minute
1-20 minutes
(default 10
minutes)
34
minutes
32L
8.45 Gal
5.3 Hints and tips for best cleaning results
Remove all cement, composite material and amalgam from instruments at chairside prior to
processing in the Hydrim.
Use only accessories, such as baskets, cassettes, trolleys and racks to hold instruments
that are designed for the HYDRIM L110w G4.
Use accessories for their intended purpose only. For example, do not place hinged
instruments in cassettes. Use the hinged instrument rack instead.
Disassemble instruments if possible.
Do not overlap instruments. Each instrument must be kept separate.
All instruments must be placed in the instrument washer in such a way as to allow good
rinsing of all surfaces. Any surface that is shielded from water and detergent cannot
be cleaned.
Open hinged instruments widely when placed in baskets or hinged instrument rack
(PN 01-110409S). Closed hinged instruments cannot be cleaned.
Place suction tubes in the rack for vertical instruments and ensure that water can flow
unrestricted in and out of the tubes. Do not place suction tubes horizontally in a basket
or in a cassette. For instruments containing lumens, such as dental handpieces, only the
external surfaces are cleaned.
Place trays and other concave shaped items at an angle to ensure water can run off
surfaces easily.
Do not overload the HYDRIM.
Verify the Hydrim setup:
Ensure that the washer spray arms are free of any blockage and can rotate freely.
Ensure that the mesh filters in the chamber drain are free from debris.
To achieve best cleaning results, instruments should be processed in the HYDRIM as
soon as possible. If immediate cleaning is not possible, it is recommended to process
the instruments using the P1 Rinse and Hold program immediately after use on
the patient.
Visually inspect instruments for cleanliness and integrity (corrosion, material changes)
after cleaning and repair or replace instruments if necessary.
If process enforced is activated, the operator must confirm that the instruments are
visually clean after a complete cycle is finished.
11
5. Instructions for Use
5. Instructions for Use
5.4 Recommended load congurations
The intended purpose of the L110w G4 washer is a clinical general instrument washer,
predominantly aimed at the general dental and medical markets. There will be many
combinations of instruments that can be processed in the L110w G4 unit depending on:
Size of practice
Types of procedures undertaken at any one time
Specialization of practice
Instrument management protocols
It is difcult to illustrate all the possible combinations av ailable, so below we have given
examples of the most common typical basket loadings, the recommended loading
of general instruments, and the worst case loading, for which the unit has been type
tested and validated.
Handle all instruments with care to prevent personal injuries through punctures for
example. Follow local health and safety regulations for preventing accidents, bodily
harm and personal injuries.
Observe the instrument manufacturer’s recommendations for cleaning and care.
When loading the unit, arrange loose instruments so that they do not touch one
another. Instruments with concave surfaces or cavities should be placed so that water
can drain freely.
12
Sample loading:
Example 1 – solid straight & hinged instruments
Example 2 – solid straight instruments
5. Instructions for Use
13
Example 7instrument trays
Example 6instrument sets in cassettes – half load
Example 5instrument sets in cassettes – full load
Example 4small & disassembled instruments
For example: Two piece mirrors, scalpel knife handles, etc.
For example: Bone fi les, pluggers, curettes, excavators, etc.
For example: Bone fi les, pluggers, curettes, excavators, etc.
Example 3hinged instruments, solid instruments and open hollow instruments
For example:
Scalers, curettes,
dental elevators,
single part mirrors,
suction tubes etc.
For example:
Rongeurs, forceps,
scissors etc.
The following are optional:
STATIM 5000 basket.
Part-number
01-107241
Long Basket.
Part-number
01-108232
Full Size Basket.
Part-number
01-111598
14
5. Instructions for Use
Rack for four 11” x 14”
cassettes, for upper
or lower trolley. Full size (1/1).
Part-number
01-109965S
Hinged instrument rack.
Part-number
01-110409S
STATIM 2000 basket.
Part-number
01-107240
Rack for ve 8” x11”
cassettes, for upper
or lower trolley. Full size (1/1).
Part-number
01-109963S
Rack for ve medium
size cassettes, for upper
or lower trolley. Half-size (1/2).
Part-number
01-109964S
Basket with lid.
Quarter size (1/4).
Part-number
01-109966S
Hygiene basket for
loose instruments.
Quarter size (1/4).
Part-number
01-109967S
5.5 Baskets and Accessories
The following are included with the HYDRIM:
5. Instructions for Use
Settings
Unlock door
Water softener,
detergent level,
and door status
Network connectivity
(green when active)
Programs
USB status
(green when active)
5.6 Touchscreen overview
15
5. Instructions for Use
User Technician Setup
5.7 Set up menu overview
Set EOL CR/LF
Set Regeneration
LCD Contrast
Theme
Remote Screen
Instructions
Instructions Delay
Remote Access
Prepare for Shipping
Baud RateLanguage
Country
Date/Time
Unit No.
Network Setup
Screen Saver
Temperature C/F
Set Button Beep
Beep Volume
Set Printer
Cleaning
Set Drying Time
Set Wash Time
Cycle Count
Process Enforced
User
16
5.8 Setting the language
The messages displayed by your HYDRIM can be presented in a number of different
languages. To change the current language, follow these steps:
1.
2. Scroll to
Language
and select.
3. From the LANGUAGE screen, press to scroll through the list of languages.
When you have found the desired language, press
to save your selection and return
to the Setup menu.
5.9 Setting the country
1.
2. Scroll to
Country
and select.
3. Using the keypad, type the name of the country and press
EN
to select. Press
to save and return to the Setup menu.
5.10 Setting the time
1.
2. Scroll to
Date / Time
and select TIME.
3. From the TIME screen, use the keypad to set the time. Press
EN
to save and
to return to the Setup menu.
NOTE: If the HYDRIM is connected to a network, it is important to also enter the correct
Time Zone. Enter the Time submenu, select Time Zone and scroll and select your local
time zone.
4. To change your unit to display 12-hour time format (24-hour time format is the default
setting), go to the Setup menu and use
to scroll to TIME 12/24, select it and
toggle to 12. Press
to save and return to the Setup menu.
5. To activate daylight savings time (DST), go to the Setup menu and use
to scroll
to DST ON/OFF and select. Use
to toggle DST ON or OFF and press the
to save and return to the Setup menu. If connected to the network, the time will
automatically adjust to DST.
17
5. Instructions for Use
5.11 Setting the date
1.
2. Scroll to
Date / Time
and select DATE.
3. From the DATE screen, use the keypad to set the date. Press
EN
to save and
to return to the Setup menu.
4. To change the format in which the date appears, return to the Setup menu and use
to scroll to DATE FORMAT. Select it, and follow the prompts to have
the date displayed in the desired format. Press
to save and return to
the Setup menu.
5.12 Assigning unit identi er number
1.
2. Scroll to
Unit No
and select.
3. Using the keypad, select a maximum of 3 digits to be used as the unit’s identifi er
number. Press
EN
to save and to return to the Setup menu.
5.13 Adjusting the screensaver delay
To change the length of time before the screensaver is activated, follow these steps:
1.
2. Scroll to
Screensaver
and select.
3. Use to scroll through your time options. When you have found the amount of
time you require, press it. Press
to save and return to the Setup menu.
5.14 Adjusting the temperature display
1.
2. Scroll to
Temperature C/F
and select.
3. Use to choose between having information displayed in degrees Celsius or
Fahrenheit. Press
to save and return to the Setup menu.
18
5. Instructions for Use
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81