GW Instek tware User manual

Type
User manual
HowtouseARBEditingSoftwarwe?
A.Connection
1.PleaseexecutetheDeviceManager
(Route:Windows>computer>DeviceManager)
2.Checkthecomports
3.OpentheArbWaveformEditingSoftware
4.Settheinterfacethatyouuse
(GPIB,USB,LAN)
5.Checkthecomportagain
(Route:Setup>ConfigI/O>Option)
(EX:MyAFGCDCisCOM7)
B.Edit
1.MaxEditLengthmeansthemaxwaveformlength
2.AfterStep1,Thepopupwindowwillshowtheparameter.
Youcansetthesamplerateandthehorizontallength.
Ifyousetthehorizontallengthto10k,itwillzoominto10kpoints.(ex.)
Butthemaxlengthisstill1M.
3.Thebasicwaveform.
Ifyouwanttocreateabasicwaveform,everysettingswillchangebykey‐
in.
SampleRate(Hz)/points=Frequency,200MHz/1k=200kHz
Butitmeansthesinewaveis200knottheoutput.


4.Setthewaveformtoinstrument
(1)Fromwave:Itmeanswhichpartofwaveformthatyouwanttoselect.
(2)Toinstrument:Itmeansthestartpointofinstrument.
(3)Output:Itmeanstheoutputrange.Thefrequencyisalsodecidedby
thepoints.SampleRate(Hz)/points=Frequency
EX.Setting
(1)Fromwave:
Start:200,length:500
(2)Output:
Start:0,length:700
0~699=700Points
Frequency=SampleRate/Points=>200MHz/700Points=285.71 