Thermo Fisher Scientific DS‑33 GeneScan User guide

Type
User guide
GeneScan DS33 Installation Standards with GeneScan 500
LIZ Size Standard
3130, 3730, and 310 Series Instruments
Catalog Number 4330397
Pub. No. 4363118 Rev. E
WARNING! Read the Safety Data Sheets (SDSs) and follow the handling instructions. Wear appropriate protective eyewear,
clothing, and gloves. Safety Data Sheets (SDSs) are available from thermosher.com/support.
Product description
GeneScan DS-33 Installation Standards with GeneScan 500 LIZ Size Standard are used to perform the install check. The GeneScan
DS-33 Installation Standard consists of pooled PCR products that are labeled with 6-FAM, VIC, NED, and PET dyes. To generate the
pooled products, CEPH individual 1347-02 genomic DNA has been amplied with 8 uorescently labeled PCR primer pairs that amplify
selected microsatellite loci.
Table 1 Average loci sizes for GeneScan DS33 Installation Standard (base pairs)
Locus
POP-4 polymer POP-7polymer
310 sizes (bp)[1] 3130 series sizes (bp)[1] 3730 series sizes (bp)[2]
D20S119 [FAM] 112 bp 118 bp 112 bp 117 bp 113 bp 119 bp
D9S1690 [FAM] 236 bp 238 bp 236 bp 238 bp 237 bp 239 bp
D5S644 [VIC] 84 bp 96 bp 84 bp 96 bp 85 bp 97 bp
D5S424 [VIC] 217 bp 219 bp 216 bp 218 bp 217 bp 219 bp
D9S288 [NED] 137 bp 145 bp 137 bp 145 bp 138 bp 146 bp
D6S289 [PET] 173 bp 175 bp 173 bp 175 bp 174 bp 176 bp
D15S117 [PET] 339 bp 341 bp 339 bp 341 bp 338 bp 340 bp
D18S462 [NED] 304 bp[3] 303 bp[3] 303 bp[3]
[1] Sizes were generated using the GS500(250)LIZ size standard definition.
[2] Sizes were generated using the GS500 (–35,–250)LIZ size standard definition.
[3] Homozygous locus.
Contents and storage
Contents Amount Storage[1]
GeneScan DS33 Installation Standard 4 tubes Store the kit at –25°C to –15°C until
ready to use. Upon thawing, store the
size standard tube at 2°C to 8°C.
Protect from light.
Refer to the expiration date on the
label. Do not use expired product.
GeneScan 500 LIZ Size Standard 1 tube
[1] See the expiration date on the package. Do not use expired product.
PRODUCT INFORMATION SHEET
For Research Use Only. Not for use in diagnostic procedures.
Prepare the installation reagent:
3130/3130
xl
Genetic Analyzer
1. Resuspend the contents of the installation standard and the
size standard tubes, then centrifuge briey to bring the
contents to the boom of the tubes.
2. Combine the following components in a microcentrifuge
tube:
Component Amount (16-capillary
array)[1]
GeneScan DS33 Installation
Standard
5 µL
GeneScan 500 LIZ Size Standard 5 µL
HiDi Formamide 190 µL
Total 200 µL
[1] Scale the volumes for 4-capillary instruments.
IMPORTANT! Use the installation reagent within 16 hours of
preparation.
3. Mix thoroughly, then centrifuge to bring the mixture to the
boom of the tube and eliminate air bubbles.
4. To denature the DNA fragments, incubate at 95°C for
5 minutes. Immediately place the mixture on ice for
³ 2 minutes.
5. Transfer installation reagent to the appropriate wells a 96-
well plate.
Option Description
16-capillary
instruments
Dispense 10 µL of standard into wells
A1 to H2.
4-capillary instruments Dispense 10 µL of standard into wells
A1 to D1.
6. Cover the plate with a 96-well septa (Cat. No. 4315933), then
centrifuge to bring the mixture to the boom of the tube and
eliminate air bubbles.
7. Assemble the plate with the Retainer and Base, then load on
the instrument.
8. See the instrument user guide for specics on seing up the
run.
Prepare the installation reagent: 3730xl DNA
Analyzer
Note: This installation kit is not for use on 96-capillary
instruments.
1. Resuspend the contents of the installation standard and the
size standard tubes, then centrifuge briey to bring the
contents to the boom of the tubes.
2. Combine the following components in a microcentrifuge
tube:
Component Amount (48-capillary
array)
GeneScan Installation Standard 25 µL
GeneScan 500 LIZ Size Standard 25 µL
HiDi Formamide 460 µL
Total 510 µL
IMPORTANT! Use the installation reagent within 16 hours of
preparation.
3. Mix thoroughly, then centrifuge to bring the mixture to the
boom of the tube and eliminate air bubbles.
4. To denature the DNA fragments, incubate at 95°C for
5 minutes. Immediately place the mixture on ice for
³ 2 minutes.
5. Transfer installation reagent to the appropriate wells a 96- or
384-well plate.
Option Description
96-well plate Dispense 10 µL of standard into every other
column (A1-H1, A3-H3,...).
384well plate Dispense 5 µL of standard into the
corresponding wells for a single 48-capillary
injection (A1, C1, E1, G1, I1, K1, M1, O1, A5,
C5,....).
6. Cover the plate with a 96-well septa (Cat. No. 4315933), then
centrifuge to bring the mixture to the boom of the tube and
eliminate air bubbles.
7. Assemble the plate with the Retainer and Base, then load on
the instrument.
8. See the instrument user guide for specics on seing up the
run.
Run the install standard with the Gene Mapper Software Run
Module and select Dye Set G5.
2
GeneScan
DS33 Installation Standards with GeneScan
500 LIZ
Size Standard Product Information Sheet
Prepare the installation reagent: 310 Genetic
Analyzer
1. Resuspend the contents of the installation standard and the
size standard tubes, then centrifuge briey to bring the
contents to the boom of the tubes.
2. Combine the following components in a microcentrifuge
tube:
Component Amount[1]
GeneScan Installation
Standard DS33
1.0 µL
GeneScan 500 LIZ Size
Standard
0.5 µL
HiDi Formamide 12.5 µL
Total 14.0 µL
[1] The volumes are per sample. Scale up for the number of samples.
IMPORTANT! Use the installation reagent within 16 hours of
preparation.
3. Mix thoroughly, then centrifuge to bring the mixture to the
boom of the tube and eliminate air bubbles.
4. To denature the DNA fragments, incubate at 95°C for
5 minutes. Immediately place the mixture on ice for
³ 2 minutes.
5. See the instrument user guide for specics on seing up the
run.
Run the install standard with the appropriate run module (for
more information, see G5v2 Module for Use with Dye Set 33 (DS‑33)
User Bulletin, Pub. No. 4339367):
GS STR POP4 (1 mL) G5
GS STR POP4 (1 mL) G5 V2
Running samples
See the instrument user guide or geing started guide for
instructions on running samples.
Customer and technical support
Visit thermosher.com/support for the latest service and support
information.
Worldwide contact telephone numbers
Product support information
Product FAQs
Software, patches, and updates
Training for many applications and instruments
Order and web support
Product documentation
User guides, manuals, and protocols
Certicates of Analysis
Safety Data Sheets (SDSs; also known as MSDSs)
Note: For SDSs for reagents and chemicals from other
manufacturers, contact the manufacturer.
Limited product warranty
Life Technologies Corporation and/or its aliate(s) warrant their
products as set forth in the Life Technologies' General Terms and
Conditions of Sale at www.thermosher.com/us/en/home/global/
terms-and-conditions.html. If you have any questions, please
contact Life Technologies at www.thermosher.com/support.
Thermo Fisher Scientific Baltics UAB | V.A. Graiciuno 8, LT-02241 | Vilnius, Lithuania
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT,
PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Revision history: Pub. No. 4363118
Revision Date Description
E 09 May 2019 Update allele sizing information; change manufacturing address to Vilnius; minor style and
format updates.
D 21 December 2017 Pub. No. 4363118 was incorrectly changed to content for DS-33 GeneScan Installation
Standards with 600 LIZ Size Standard v2.0 (Cat. No. 4376911). Update to replace content for
GeneScan DS33 Installation Standards with GeneScan 500 LIZ Size Standard (Cat. No.
4330397).
C 11 Aug 2009 Baseline for this revision
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all
applicable Limited Use Label Licenses.
©2019 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified.