Extech Instruments SC600 User manual

Category
Kitchen scales
Type
User manual

Extech Instruments SC600 is a versatile electronic counting scale/balance, equipped with a high-precision load cell transducer for accurate weighing and counting applications. Its large measurement capacity, high resolution display, and multiple weighing units (grams and ounces) make it suitable for various tasks. The SC600 features a tare function for determining the net weight of objects, a self-calibration utility for maintaining accuracy, and RS-232 PC interface for data transfer.

Extech Instruments SC600 is a versatile electronic counting scale/balance, equipped with a high-precision load cell transducer for accurate weighing and counting applications. Its large measurement capacity, high resolution display, and multiple weighing units (grams and ounces) make it suitable for various tasks. The SC600 features a tare function for determining the net weight of objects, a self-calibration utility for maintaining accuracy, and RS-232 PC interface for data transfer.


USERGUIDE
ElectronicCountingScale/Balance
ModelSC600

2SC600EUENv1.34/14
Introduction
ThankyouforselectingtheExtechInstrumentsModelSC600ElectronicCountingScale/Balance.
The SC600 is microprocessor controlled and uses a load cell transducer for high accuracy and
optimumperformance.TheSC600offersalargemeasurementcapacity,highresolution,andlarge
easy to read display. The slide switch on the back of the scale allows for the selection of
measurementunits(gramorounce).
A selfcalibration utility is builtin and the calibration data is stored internally even upon power
down. The RS232 PC interface allows for the transfer of measurement data between the SC600
andaPC.
The SC600 includes ‘bubble’ level, adjustable rubber feetfor leveling, and heavyduty ABSplastic
housing.AnACadaptormaybeusedtopowertheunitortheunitmaybeusedonbatterypower.
Thisdeviceisshippedfullytestedandcalibratedand,withproperuse,willprovideyearsofreliable
service. Please visit our website (www.extech.com) to check for the latest version of this User
Guide,ProductUpdates,andCustomerSupport.
3SC600EUENv1.34/14
MeterDescription
1. 5digitmultifunctionLCDDisplay
2. ‘Bubble’Leveler
3. WeightPlatform
4. ACAdaptorjack(9V/300mA)
5. 3.5mmRS232PCinterfacejack
6. UnitofMeasureselectswitch(gramsorounces)
7. PowerONOFFswitch
8. TAREbutton
9. CAL(Calibration)button
Batterycompartmentandadjustable
levelinglegslocatedonbottomofunit.
1
2
3
4
5
6
7
8
9
4SC600EUENv1.34/14
Operation
PreparingtheSC600
PlacetheSC600onafirm,levelsurface.Adjusttherubberpadlegsonthebottomoftheunitto
level; ensure that the SC600 is level before use. The ‘bubble’ level located on the top of the
SC600canbeusedasanaccuratelevelguide.
Remove the protective tape
from the weight platform area and install the weight platform as
showninthediagramusingthe4screws(supplied).
5SC600EUENv1.34/14
Poweringthemeter
Installsix(6)1.5VAAbatteriesintothebatterycompartmentlocatedonthebottomoftheunit.
TheACAdaptorcanalsobeusedtopowertheunit.TheACadaptorjackislocatedontherearof
theSC600.
PowertheunitONorOFFusingtheONOFFswitch
locatedontherightsideoftheSC600.
The meter will perform a short selftest during which the display will count down to zero; the
unitwillswitchONinitsnormaloperating mode.
Pleaseallow30minutesofwarmupandstabilizationtimebeforeusingtheSC600.
Selectingtheunitsofmeasure(gramsorounces )
Use the recessedselector switch on the rearof the unit to select the desired unit of measure
gramsorounces.TheLCDdisplaywillreflecttheselectedunitofmeasure.
Scale(Weighing)Operation
FollowtheinstructionsaboveforSC600preparationand,whentheunitisready,gentlyplacean
objectontheweightplatformandreadtheweightmeasurementontheLCDdisplay.Ensurethat
the weight does not exceed the maximum capacity of the SC600 as published in the
specificationsofthisUserGuide.Iftheweightofanobjectexceedsthemeasurementcapacityof
the SC600, the displaywill show all dashes on the upper part of the LCD display (‐ ‐ ‐ ‐ ).For
underrangemeasurementsthedisplayshowsdashesonthelowerpartof thedisplay(____).
TAREFunction
1. After weighing the first item, press the TARE button; the display will reset to zero
automaticallyandthedisplaywillshowtheTAREicon.
2. Removethefirstitemfromtheplatformandplacetheseconditemontheplatform.
3. TheSC600willnowdisplayonlythe
weightoftheseconditem.
4. TheTAREdisplayiconwillswitchOFFthenexttimethattheSC600isswitchedOFF.
TareConsiderations:Ifthe display shows a negativeweight(minus sign shown)oralldashes
onthelowerareaoftheLCD_____(weight
valuebelowzero)anerrormayhaveoccurred
andtheTareprocedureshouldberepeated.
6SC600EUENv1.34/14
CounterOperation
Foracollectionofobjectsorpieces(eachpiecehavingthesameweightasthenext),theSC600
can calculate the number of these objects in a collection. A group of pieces can be used as a
sampleexampleingroupsof10,20,50,or100pieces.
1. Prepare a
group of pieces (ina groupsize of10, 20, 50,or 100pieces). This will be the
‘standard’ that the SC600 uses to count subsequent collections of pieces. Do not place
anypiecesonthescaleyet.
2. WiththeSC600poweredOFF, pressandholdtheCALbutton
whileswitchingtheSC600
ON
3. HoldtheCALbuttonuntilonebeepisheard
4. Thedisplaywillcountdown to zero andthebottomofthedisplaywillshowtheCOUNT
andPCS(pieces)icons
5. The display will scroll through the group sizes (10, 20, 50, and 100)
one group size per
second
6. Placethegroupofpieces(fromstep1)ontotheweightplatform
7. PresstheCALbuttonwhentheappropriategroupsize(10,20,50,and100)isdisplaying.
Thedisplaywillflashtwiceandfreezethegroupsizenumber.
8. Removethegroup
ofpiecesfromtheplatform.TheSC600isnowreadyforcounting.
9. Place anunknown number of pieces (of thesame weight asthe ones from step 1) and
theSC60willdisplaythenumberofpiecespresentlyontheplatform.
ChangetheCountSampleNumber
1. RemoveallitemsfromtheSC600platform
2. Press the CAL button once; the display will scroll the piece group sizes (10, 20, 50, and
100)oncepersecond
3. Place a group of pieces (10, 20, 50, or 100) onto the weight platform and
when the
numberontheSC600displaymatchesthenumberofpiecesontheplatformpresstheCAL
button
4. Thedisplaywillflashtwiceandfreezethecountnumber
5. Removethegroupofpiecesfromtheplatform.TheSC600isnowreadyforcounting.
6. Placeanunknown
numberofpieces(ofthesameweightastheonesfromstep3)andthe
SC60willdisplaythenumberofpiecespresentlyontheplatform.
CounterOperationConsiderations
Theonepieceweightshouldbelargerthan0.1gotherwiseerrorsmayoccurincounting.The
accuracyofthecounterdependsonthenumberofunitscounted.Itisstrongly recommended
thatmoresamples,forexample50unitsormore,becountedforthebestaccuracy.
7SC600EUENv1.34/14
Calibration
PreparingtheCalibration
Standardweightsarerequiredforcalibration(notsupplied).A200gramora400gramstandard
canbeused.
WarmuptheSC600foratleast30minutesbeforestartingthecalibration.
IMPORTANTCALIBRATIONNOTE:
Before calibrating the SC600 determine the expected weight measurements that will be
encounteredduringnormal
use.Forexpectedweightmeasurementsupto200gramsplease
performa200gramcalibration.Forexpectedweightsabove 200gramspleasecalibratetothe
400gramstandard.
400gramcalibration:
1. AftertheSC600iswarmedup,gentlyplacethe400gramstandardweightonthecenterof
the
weightplatform.Iftheweightdisplaydoesnotshowexactly400gramsthencontinue
with the procedure, otherwise calibration is not required and the procedure can end at
thispoint
2. PresstheCALbuttononceandthedisplaywillshow‘200grams’
3. PresstheTAREbuttononceandthe
displaywillshow‘400grams’
4. PresstheCALbuttononceagainandthedisplayvalue(400g)willflashsix(6)times.
5. Calibration is now complete. Calibration data will be stored internally in nonvolatile
memory meaning that calibration data will be saved even when the SC600 is powered
OFF.
200gramcalibration:
1. AftertheSC600iswarmedup,gentlyplacethe200gramstandardweightonthecenterof
theweightplatform.Iftheweightdisplaydoesnotshowexactly200gramsthencontinue
with the procedure, otherwise calibration is not required and the procedure can end at
this
point
2. PresstheCALbuttononceandthedisplaywillshow‘200grams’
3. PresstheCALbuttononceagainandthedisplayvalue(200g)willflashsix(6)times.
4. Calibration is now complete. Calibration data will be stored internally in nonvolatile
memory meaning that calibration data
will be saved even when the SC600 is powered
OFF.

8SC600EUENv1.34/14
RS232PCInterface
The SC600can stream data to a PC viathe rear 3.5mm RS232 Output jack, theoptional 407001
USB kit (RS232 to USB cable and driver CD) along with the 407001 software (available free at
www.extech.com)arerequired.
Once the equipment listed above is obtained, set the 407001USB to channel
number 2 and
connectittotheSC600.Openthesoftwareapplicationandfollowtheonscreenprompts.
BatteryReplacement
WhenthelowbatteryiconappearsontheLCD(LO),thebatteriesmustbereplaced.
ThebatterycompartmentislocatedonthebottomoftheSC600.
Openthebatterycompartmentusingthebatterycompartmentlatches.
Replacethebatteriesobservingcorrectpolarityandreplacethecompartmentcarefully.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
As consumers, users are legally required to take used batteries to appropriate
collection sites, the retail store where the batteries were purchased, or wherever
batteriesaresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuseris
obligatedtotakeend
oflife devices to a designated collection point for the disposal of electrical and electronic
equipment.
ACPowerAdaptor
TheACadaptorcanbeusedtopowertheSC600froma110/220/240VACpowersource.The
adaptorcapacityis300mA.TheACadaptorplugsintotheSC600viathejackattherearofthe
instrument.
9SC600EUENv1.34/14
Specifications
Display 5digitmultifunctionLCD
Digitheight:20.9mm(0.8”)
RangesandResolution 0.10to600g(resolution:0.01g)
0.005to9.9995oz.(resolution:0.0005oz.)
10.000to21.164oz.(resolution:0.001oz.)
Unitofmeasure Selectableviarearswitch(grams/ounces)
Accuracy ±(0.05%+0.04g);within2hoursofcalibration
Accuracy applies at 23
o
C ±5
o
C (specifications tested under the
following RF environmental conditions: field strength < 3V/M and
frequency<30MHzonly)
Samplingtime Approx.1measurementpersecond
TareControl 600gmaximum(approx.)
Transducertype Loadcell
UserCalibration Using200gor400gstandardweights(notsupplied)
Dataoutput RS232PCinterfacejackon
rearofinstrument
OptionalPCinterfacecableandsoftwarePN407001
OptionalUSBadaptorPN407001USB
Overrangeindication Upperdashes‐‐‐‐‐‐appearontheLCD
Underrangeindication Lowerdashes____appearontheLCD
Lowbatteryindication ‘LO’BatteryiconappearsontheLCD
Power
supply 6 x 1.5V AA batteries or power adaptor 9V/300mA (supplied with
multiplugs)
Powerconsumption 17mADCapproximately
OperatingTemperature 0to50
o
C(32to122
o
F)
OperatingHumidity <80%RH
Dimensions Base:250x190x70mm(9.8x7.5x2.8”)
Platformdiameter:120mm(4.7”)
Weight 1kg(2.2lbs.)
Copyright©2014FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
www.extech.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9

Extech Instruments SC600 User manual

Category
Kitchen scales
Type
User manual

Extech Instruments SC600 is a versatile electronic counting scale/balance, equipped with a high-precision load cell transducer for accurate weighing and counting applications. Its large measurement capacity, high resolution display, and multiple weighing units (grams and ounces) make it suitable for various tasks. The SC600 features a tare function for determining the net weight of objects, a self-calibration utility for maintaining accuracy, and RS-232 PC interface for data transfer.