Rosemount 890 UV Analyzer-Rev A Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual
Rosemount Analytical
M
ODEL
890
UV A
NALYZER
I
NSTRUCTION
M
ANUAL
748460-A
N
OTICE
The information contained in this document is subject to change without notice.
Manual Part Number 748460-A
January 2000
Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
Teflon® and Viton® are registered trademarks of E.I. duPont de Nemours and Co., Inc.
Suprasil II® is a registered trademark of Heraeus Amersil Inc.
Pyrex® is a registered trademark of Corning Glass Works.
SNOOP® is a registered trademark of NUPRO Co.
C
ONTENTS
I
748460-A
Rosemount Analytical
January 2000
P
REFACE
Safety Summary.................................................................................................P-1
Specifications - General......................................................................................P-4
Specifications - Sample......................................................................................P-4
Specifications - Physical.....................................................................................P-5
Specifications - Options......................................................................................P-5
Customer Service, Technical Assistance and Field Service...............................P-6
Returning Parts to the Factory............................................................................P-6
Training ......................................................................................................P-6
Documentation....................................................................................................P-6
Compliances ......................................................................................................P-7
S
ECTION
1. I
NTRODUCTION
1.1 General Description..................................................................................1-1
1.2 Available Options......................................................................................1-3
S
ECTION
2. U
NPACKING AND
I
NSTALLATION
2.1 Check for Shipping Damage.....................................................................2-1
2.2 Location....................................................................................................2-1
2.3 Voltage Requirements..............................................................................2-1
2.4 Electrical Connections..............................................................................2-2
2.4.1 Line Power Connections.............................................................2-2
2.4.2 Recorder Connections ................................................................2-3
2.5 SAMPLE INLET/OUTLET CONNECTIONS..............................................2-4
2.6 CALIBRATION GAS REQUIREMENTS ...................................................2-5
2.7 SAMPLE HANDLING SYSTEM................................................................2-5
2.8 LEAK TEST PROCEDURE.......................................................................2-5
2.9 SAMPLE FLOW RATE.............................................................................2-6
2.10 OPTION BOARDS....................................................................................2-6
2.10.1 Alarm Connections....................................................................2-7
2.10.2 Calibration and Gas Control Connections.................................2-7
2.10.3 Auto Zero/Span Connections....................................................2-8
C
ONTENTS
II
748460-ARosemount Analytical
January 2000
2.10.4 Remote Input/Output Connections ........................................... 2-9
2.11 Ordering Option Kits............................................................................... 2-10
S
ECTION
3. I
NITIAL
S
TARTUP AND
C
ALIBRATION
3.1 Leak Test ................................................................................................. 3-1
3.2 Power Verification.................................................................................... 3-1
3.3 Software/Countdown................................................................................ 3-2
3.4 Front Panel Indicators and Controls......................................................... 3-2
3.4.1 Display ..................................................................................... 3-2
3.4.2 Function Keys............................................................................. 3-2
3.4.3 User Programmable Keys .......................................................... 3-4
3.4.4 Run Mode Display...................................................................... 3-5
3.4.5 General Display Information....................................................... 3-5
3.5 Accessing Mode Displays ........................................................................ 3-6
3.6 Security Code........................................................................................... 3-8
3.7 Range Parameters................................................................................... 3-9
3.8 Analyzer Diagnostics................................................................................ 3-11
3.9 Zero Calibration........................................................................................ 3-11
3.10 Zero Calibration for the Analyzer with the Calibration Gas
Control Option .............................................................................. 3-13
3.11 Span Calibration....................................................................................... 3-13
3.12 Span Calibration for the Analyzer with the Calibration Gas
Control Option .............................................................................. 3-14
3.13 Linearization............................................................................................. 3-14
3.14 Alarm ...................................................................................................... 3-19
3.14.1 STATUS Display....................................................................... 3-21
3.15 Current Output ......................................................................................... 3-22
3.16 Auto Zero/Span Option ............................................................................ 3-22
3.17 Remote Range input/Output Option......................................................... 3-26
3.18 Interference Balance................................................................................ 3-28
S
ECTION
4. R
OUTINE
O
PERATION AND
T
HEORY
4.1 Routine Operation.................................................................................... 4-1
4.2 Recommended Calibration Frequency..................................................... 4-1
4.3 Shutdown................................................................................................. 4-1
4.4 Detection System Theory......................................................................... 4-1
S
ECTION
5. T
ROUBLESHOOTING
5.1 Error Code Summary............................................................................... 5-1
5.2 Iris Balance Adjustment ........................................................................... 5-2
5.3 Voltage Checks........................................................................................ 5-2
5.4 Digital GAIN Adjustment .......................................................................... 5-2
5.5 Case Heater............................................................................................. 5-3
5.6 ERL Error Message.................................................................................. 5-3
C
ONTENTS
iii
January 2000 Rosemount Analytical
748460-A
S
ECTION
6. R
OUTINE
S
ERVICING
6.1 Cell Removal, Cleaning and Replacement ...............................................6-1
6.2 Lamp Replacement...................................................................................6-3
6.2.1 Lamp Realignment......................................................................6-7
6.3 Cleaning Optical Components..................................................................6-9
6.3.1 Spectrally Selective Mirrors.........................................................6-9
6.3.2 Beam Splitter/Focusing Mirrors...................................................6-9
6.3.3 Source Envelope.........................................................................6-9
6.3.4 End Caps....................................................................................6-9
6.4 Electronic Circuitry....................................................................................6-10
6.4.1 Power Supply Board ...................................................................6-10
6.4.2 Signal Board ...............................................................................6-10
6.4.3 Preamplifier Board ......................................................................6-10
6.4.4 Adapter Board.............................................................................6-10
6.4.5 Micro Board.................................................................................6-11
6.4.6 Case Heater Temperature Control..............................................6-11
6.4.7 Dual Alarm/Calibration Gas Control Board (Option)....................6-11
6.4.8 Isolated Remote Range I/O Board (Option)................................6-11
6.4.9 Auto Zero/Span Board (Option) ..................................................6-11
S
ECTION
7. R
EPLACEMENT
P
ARTS
7.1 CIRCUIT BOARD REPLACEMENT POLICY............................................7-1
7.2 SELECTED REPLACEMENT PARTS......................................................7-1
7.3 LAMP REPLACEMENT............................................................................7-2
D
ATA
S
HEET
G
ENERAL
P
RECAUTIONS
F
OR
S
TORING AND
H
ANDLING
H
IGH
P
RESSURE
G
AS
C
YLINDERS
W
ARRANTY
F
IELD
S
ERVICE AND
R
EPAIR
F
ACILITIES
C
ONTENTS
IV
748460-ARosemount Analytical
January 2000
F
IGURES
1-1 Model 890 Analyzer................................................................................. 1-1
1-2 Model 890 Optical Bench......................................................................... 1-2
2-1 Power Supply Board................................................................................ 2-2
2-2 Cable Gland Connection.......................................................................... 2-3
2-3 Calibration Gas Control and Alarm Connections...................................... 2-7
2-4 Auto Zero/Span Connections................................................................... 2-8
2-5 Remote Input/Output Connections........................................................... 2-9
3-1 Model 890 Adjustments ........................................................................... 3-1
3-2 Model 890 Keypad................................................................................... 3-2
3-3 Run Mode Display.................................................................................... 3-5
3-4 Logic Flow Chart...................................................................................... 3-7
3-5 Security Mode.......................................................................................... 3-8
3-6 Range Mode ............................................................................................ 3-9
3-7 Diagnostics Mode .................................................................................... 3-12
3-8 Linearization Mode................................................................................... 3-15
3-9 Typical Application Linearization Curve ................................................... 3-16
3-10 Concentration Curve................................................................................ 3-18
3-11 Curve, Normalized................................................................................... 3-18
3-12 Alarm Option............................................................................................ 3-20
3-13 Status Display.......................................................................................... 3-21
3-14 Current Output Mode............................................................................... 3-22
3-15 Auto Zero/Span........................................................................................ 3-25
3-16 Remote Input/Output................................................................................ 3-26
4-1 Model 890 Timing Diagram...................................................................... 4-2
6-1 Optical Bench........................................................................................... 6-2
6-2 Sample Cell Assembly............................................................................. 6-3
6-3. Collector Block (Exploded View).............................................................. 6-5
6-4 Detector Block (Exploded View)............................................................... 6-6
6-5 Lamp Assembly 655000 .......................................................................... 6-7
6-6. Lampl Alignment...................................................................................... 6-8
7-1 Model 890 Component Locations ............................................................ 7-3
7-2 Optical Bench - Sensor Locations............................................................ 7-4
7-3 UV Lamp Life vs. Intensity....................................................................... 7-4
T
ABLES
3-1 Linearization Coefficients, Standard Ranges........................................... 3-16
3-2 Remote Range I/O Bit Designation.......................................................... 3-27
3-3 Remote Range I/O Binary and Decimal Bit Coding.................................. 3-27
5-1 Error Code Summary............................................................................... 5-1
6-1 Jumper Configurations for Options .......................................................... 6-12
C
ONTENTS
v
January 2000 Rosemount Analytical
748460-A
D
RAWINGS
(
LOCATED IN REAR OF MANUAL
)
623782 Schematic Diagram, Micro Board
624127 Schematic Diagram, Adaptor Board
624204 Schematic Diagram, Dual Alarm/Fail Safe Alarm
624251 Schematic Diagram, Remote Control
624599 Scheamtic Diagram, Auto/Zero Span
652687 Schematic Diagram, Signal Board SO
2
652715 Diagram, Electrical Interconnect SO
2
652807 Schematic Diagram, Power Supply Board
652857 Schematic Diagram, Preamplifier Board SO
2
654853 Installation Drawing, Model 890
656137 Schematic Diagram, Signal Board Cl
2
656138 Schematic Diagram, Preamplifier Board Cl
2
656911 Diagram, Electrical Interconnect Cl
2
C
ONTENTS
VI
748460-ARosemount Analytical
January 2000
N
OTES
P
REFACE
P-1
January 2000 Rosemount Analytical
748460-A
S
AFETY
S
UMMARY
To avoid explosion, loss of life, personal injury and damage to this equipment and on-site
property, all personnel authorized to install, operate and service the Model 890 Analyzer
should be thoroughly familiar with and strictly follow the instructions in this manual. Save
these instructions.
If this equipment is used in a manner not specified in these instructions, protective
systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe
personal injury, death, or substantial property damage if the warning is ignored.
WARNING is used to indicate the presence of a hazard which can cause severe
personal injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation or maintenance information which is
important but not hazard-related.
Do not operate without doors and covers secure. Servicing requires access to
live parts which can cause death or serious injury. Refer servicing to qualified
personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
Alarm and zero/span switching relay contacts wired to separate power sources
must be disconnected before servicing.
This instrument is shipped from the factory set up to operate on 115 volt, 50/60
Hz electric power. For operation on 230 volt, 50/60 Hz power, see Section 2.3 for
modifications.
WARNING: ELECTRICAL SHOCK HAZARD
P
REFACE
Rosemount Analytical January 2000
748460-A
P-2
This analyzer is of a type capable of analysis of sample gases which may be
flammable. If used for analysis of such gases, the instrument must be protected
by a continuous dilution purge system in accordance with Standard ANSI/NFPA
496-1989, Chapter 8.
If explosive gases are introduced into this analyzer, the sample containment
system must be carefully leak-checked upon installation and before initial
startup, during routine maintenance and any time the integrity of the sample
containment system is broken, to ensure the system is in leak-proof condition.
Leak-check instructions are provided in Section 2.8.
Internal leaks resulting from failure to observe these precautions could result in
an explosion causing death, personal injury or property damage.
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair.
Ultraviolet light from the source lamp can cause permanent eye damage. Do not
look at the UV source for prolonged periods. Use of UV filtering glasses is
recommended.
This analyzer requires periodic calibration with known zero and standard gases.
See General Precautions for Handling and Storing High Pressure Cylinders, at
the rear of this manual.
WARNING: POSSIBLE EXPLOSION HAZARD
WARNING: PARTS INTEGRITY
WARNING: INTERNAL ULTRAVIOLET LIGHT HAZARD
WARNING: HIGH PRESSURE GAS CYLINDERS
P
REFACE
January 2000 Rosemount Analytical
748460-A
P-3
This instrument measures toxic gases. Ensure gas lines are leak-free and
properly vented. Inhalation of toxic gases is highly dangerous and could result
in death.
Also, exhaust gas from this instrument is toxic and equally dangerous. Exhaust
must be connected either to its original source or an appropriate outside vent
using ¼-inch (6mm) tubing minimum.
This instrument’s internal pullout chassis is equipped with a safety stop latch
located on the left side of the chassis.
When extracting the chassis, verify that the safety latch is in its proper (counter-
clockwise) orientation.
If access to the rear of the chassis is required, the safety stop may be
overridden by lifting the latch; however, further extraction must be done very
carefully to insure the chassis does not fall out of its enclosure.
If the instrument is located on top of a table or bench near the edge, and the
chassis is extracted, it must be supported to prevent toppling.
Failure to observe these precautions could result in personal injury and/or
damage to the product.
WARNING: TOXIC GAS HAZARD
CAUTION: TOPPLING HAZARD
P
REFACE
Rosemount Analytical January 2000
748460-A
P-4
S
PECIFICATIONS
- G
ENERAL
R
ANGE
(S
TD
) (
FULLSCALE
)
0 to 50, 0 to 5000 ppm SO
2
at atmospheric pressure
0 to 100 to 0 to 5000 ppm Cl
2
at atmospheric pressure
O
PERATING
T
EMPERATURE
SO
2
applications: 32°F to 104°F (0°C to 40°C)
Cl
2
applications: 59°F to 95°F (15°C to 35°C)
R
EPEATABILITY
1% of fullscale
Z
ERO
D
RIFT
1
SO
2
: ±2% of fullscale per week
Cl
2
: ±2% of fullscale per 24 hours
S
PAN
D
RIFT
1
SO
2
: ±2% of fullscale per week
Cl
2
: ±2% of fullscale per 24 hours
N
OISE
1% of fullscale
R
ESPONSE
T
IME
(E
LECTRONIC
)
Variable, 90% of fullscale in 0.5 sec. to 20 sec, field
selectable (application dependent)
S
ENSITIVITY
SO
2
: 0.1 ppm
Cl
2
: 0.2 ppm
I
NTERFERENT
R
EJECTION
Discrimination ratio for NO2 is 1000:1 for SO
2
applications
A
NALOG
O
UTPUT
Standard: 0 to 5 VDC and 0 to 20 mA/4 to 20 mA DC,
isolated (maximum load 700 ohms)
L
INEARIZATION
Keypad entered coefficients for linearizing 1, 2 or (all) 3
ranges
P
OWER
R
EQUIREMENTS
115/230 VAC ±10%, 50/60 Hz, 350 Watts
S
PECIFICATIONS
- S
AMPLE
S
AMPLE
C
ELL
12.0 inches (305 mm) long, 110 cc volume
M
ATERIALS IN
C
ONTACT WITH
S
AMPLE
W
INDOWS
Suprasil II
C
ELLS
Pyrex
T
UBING
FEP Teflon
F
ITTINGS
316 Stainless Steel
O-R
INGS
Viton-A
S
AMPLE
P
RESSURE
Maximum 10 psig (69 kPa)
1. Performance specifications based on ambient temperature shifts of less than 20° Fahrenheit (11° Celsius) per hour.
P
REFACE
January 2000 Rosemount Analytical
748460-A
P-5
S
PECIFICATIONS
- P
HYSICAL
E
NCLOSURE
General purpose for installation in weather-protected area.
Optional purge kit per Type Z, ANSI/NFPA 496-1993
1
D
IMENSIONS
8.7 x 19 x 24 inches (221 x 483 x 610 mm) H x W x D
W
EIGHT
65 lbs. (30 kg)
S
PECIFICATIONS
- O
PTIONS
A
LARM
2
Two single point, field programmable high or low,
deadband up to 20% of fullscale
A
LARM
R
ELAY
C
ONTACTS
Two Form C contact rated 3A, 125/250 VAC or 5A, 30 VDC
(resistive)
C
ALIBRATION
G
AS
C
ONTROL
Two front panel actuated contact closures
R
ELAY
O
UTPUTS
Two Form C contact rated 3A, 125/250 VAC or 5A, 30 VDC
(resistive)
A
UTO
Z
ERO
/S
PAN
Four form C contact closures, rated 3A, 125/250 VAC or
5A, 30 VDC (resistive), field programmable frequency and
duration of closure
R
ELAY
O
UTPUTS
Two form A contact closures for indication of insufficient
zero and span adjustment, rated (resistive load):
Max. switching power: 10 Watts
Max. switching voltage: 30 VDC
Max. switching current: 0.5 A
R
EMOTE
I
NPUT
/O
UTPUT
Three remotely changeable ranges with positive
identification.
R
ANGE
C
HANGE
Binary or decimal, field selectable.
A
UTO
Z
ERO
/S
PAN
Auto Cal request and status.
R
ELAY
O
UTPUTS
Eight form A contact rated (resistive load):
Max. switching power: 10 Watts
Max. switching voltage: 30 VDC
Max. switching current: 0.5 A
I
NPUTS
Eight optical couplers
I
NPUT
R
ANGE
+5 VDC to +24 VDC
1 When installed with user-supplied components, meets requirements for Class I, Division 2 locations per National
Electrical Code (ANSI/NFPA 70) for analyzers sampling nonflammable gases. Analyzers sampling flammable gases
must be protected by a continuous dilution purge system in accordance with Standard ANSI/NFPA 496-1993, Chapter 6.
Consult factory for recommendations.
2 Fail-safe jumper configuration.
P
REFACE
Rosemount Analytical January 2000
748460-A
P-6
C
USTOMER
S
ERVICE
, T
ECHNICAL
A
SSISTANCE AND
F
IELD
S
ERVICE
For order administration, replacement parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
R
ETURNING
P
ARTS TO THE
F
ACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when
you call: Model Number, Serial Number, and Purchase Order Number or Sales Order
Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive
or other hazardous material or used in such a hazardous environment, the user must
attach an appropriate Material Safety Data Sheet (MSDS) or a written certification that
the material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
USA
T
RAINING
A comprehensive Factory Training Program of operator and service classes is
available. For a copy of the Current Operator and Service Training Schedule contact
the Technical Services Department at:
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
D
OCUMENTATION
The following Model 890 Analyzer instruction materials are available. Contact
Customer Service or the local representative to order.
748460 Instruction Manual (this document)
P
REFACE
January 2000 Rosemount Analytical
748460-A
P-7
C
OMPLIANCES
M
ODEL
890 SO
2
A
NALYZER
The Model 890 SO
2
Analyzer is intended for sampling only non-hazardous gases in
non-hazardous locations. When equipped with the optional Type Z Purge Kit (PN
624446), this analyzer is approved for use in Class I, Division 2, Groups B, C, and D
hazardous locations and use indoor non-hazardous locations when sampling
flammable gases.
Rosemount Analytical has satisfied all obligations from the European Legislation to
harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May
1993).
This product satisfies all obligations of all relevant standards of the EMC framework in
Australia and New Zealand.
M
ODEL
890 Cl
2
A
NALYZER
The Model 890 Cl
2
Analyzer is intended for sampling only non-hazardous gases in
non-hazardous locations. When equipped with the optional Type Z Purge Kit (PN
624446), this analyzer is approved for use in Class I, Division 2, Groups B, C, and D
hazardous locations and use indoor non-hazardous locations when sampling
flammable gases.
APPROVED
FM
NAMUR
N96
97-C209
P
REFACE
Rosemount Analytical January 2000
748460-A
P-8
N
OTES
I
NTRODUCTION
Rosemount Analytical
748460-A
January 2000
1-1
1.1 GENERAL DESCRIPTION
The Model 890 Ultraviolet Analyzer is designed to determine continuously the
concentration of the component of interest in a flowing gaseous mixture. The analyzer
is capable of measurement in the 50 to 5,000 ppm range for SO
2
and 100 to 5,000
ppm for Cl
2
.
F
IGURE
1-1. M
ODEL
890 A
NALYZER
O
PTICAL
B
ENCH
The ultraviolet source emits a pulsed (30 Hz) beam of energy. This energy is split by a
beam splitter, each beam being directed to pairs of detectors before and after the
sample cell.
One of the unique features of the Model 890 is the use of spectrally selective,
“Transflectance”© mirrors. These mirrors isolate the sample and reference spectral
passbands for the detectors. They reflect energy below a wavelength region and
I
NTRODUCTION
Rosemount Analytical
748460-A
January 2000
1-2
transmit the remaining, higher wavelengths, all with much lower energy loss than the
more commonly used bandpass interference filters.
Four detectors are used in this system, two before the sample cell (sample before [Sb]
and reference before [Rb]) and two after (sample after [Sa] and reference after [Ra]).
Sb and Sa receive energy in the specific wavelength regions depending on the
application (265 to 310 nm for SO
2
, 310 to 355 nm for Cl
2
), Rb and Ra in the 310 to 355
nm region for SO
2
and 355 to 400 nm for Cl
2
.
These four detectors measure the component of interest (COI) concentration and
correct for NO2 interference and UV lamp fluctuations. The difference between
detector determinations is the COI concentration, following this formula:
COI = [f(Rb)-Sb]-[f(Ra)-Sa]
where:
Ra, Rb, Sa, Sb = signals from those detectors so identified
f = attenuation factor for the reference signal, adjusted to compensate for NO2
interference.
The sample gas is introduced to the sample cell, and the COI absorbs ultraviolet
energy in proportion to the concentration in the gas. The difference between the
signals of the detectors located at both ends of the sample cell determines the
concentration of COI in the sample.
F
IGURE
1-2. M
ODEL
890 O
PTICAL
B
ENCH
R
a
DETECTORS
MIRRORS
TRANSFLECTANCE©
MIRRORS
DETECTORS
MIRROR
TRANSFLECTANCE©
MIRRORS
SAMPLE CELL
UV LAMP
BEAM SPLITTER
S
a
S
b
R
b
I
NTRODUCTION
January 2000 Rosemount Analytical
748460-A
1-3
Additionally, the adjacent (non-COI-absorbing) reference wavelengths are used as a
baseline for measurement and correction of sample interferent components,
particularly NO2.
Readout is on a 16-character, LED-backlighted liquid crystal display. COI
concentration data is presented in parts per million, percent of composition, or percent
of fullscale. Additionally, 0 to +5 VDC output for a potentiometric (voltage) recorder
and 0 to 20 mA or 4 to 20 mA isolated current output (maximum load 700 ohms) are
provided as standard.
A case heater with fan assembly maintains proper operating temperature.
L
INEARIZATION
A linearizer, based on a fourth-order polynomial, is incorporated in the electronic
circuitry. By turning the linearizer ON and entering the correct coefficients, an output
linear with concentration is obtained.
1.2 AVAILABLE OPTIONS
Operation of the Model 890 can be enhanced with the choice of several options:
D
UAL
A
LARMS
(
STANDARD AND FAIL
-
SAFE
)
User-set dual alarms are available with configurable HI/LO designations and
deadband.
A
UTO
Z
ERO
/S
PAN
An Automatic Zero/Span Option is available for unattended calibration of all three
ranges.
C
ALIBRATION
G
AS
C
ONTROL
A Calibration Gas Control Option allows two solenoids to be remotely actuated from
the front panel, enabling one-man calibration without leaving the analyzer.
R
EMOTE
R
ANGE
I/O
An optional remote range input/output is available.
A
IR
P
URGE
K
IT
Air purge kit, when installed with user-supplied components, meets Type Z
requirements of standard ANSI/NFPA 496-1993 for installation in Class I, Division 2
locations as defined in the National Electrical Code (ANSI/NFPA 70) when sampling
I
NTRODUCTION
Rosemount Analytical
748460-A
January 2000
1-4
nonflammable gases. If the analyzer is used to sample a flammable gas, it must be
protected by a continuous dilution purge system per standard ANSI/NFPA 496-1993,
Chapter 6, or IEC publication 79-2-1983, Section Three. (Consult factory for further
information.)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96

Rosemount 890 UV Analyzer-Rev A Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual