Ludlum Measurements 2200 Owner's manual

Type
Owner's manual

Ludlum Measurements 2200 is a self-contained counting instrument designed for operation with scintillation, proportional, or GM (Geiger Muller) detectors. It features a voltage-sensitive preamplifier, linear amplifier, electronic timer, and detector high-voltage power supply. A single-channel analyzer is also featured in this unit for use in gamma spectrum analysis. The unit can be powered by 95-250 Vac, 50-60 Hz, or by four “D” cell batteries. An RS-232 serial port is provided, which can be attached to a computer or optional printer.

Ludlum Measurements 2200 is a self-contained counting instrument designed for operation with scintillation, proportional, or GM (Geiger Muller) detectors. It features a voltage-sensitive preamplifier, linear amplifier, electronic timer, and detector high-voltage power supply. A single-channel analyzer is also featured in this unit for use in gamma spectrum analysis. The unit can be powered by 95-250 Vac, 50-60 Hz, or by four “D” cell batteries. An RS-232 serial port is provided, which can be attached to a computer or optional printer.

LUDLUM MODEL 2200
SCALER RATEMETER
November 2020
Serial No. 185785 and Succeeding
Serial Numbers
LUDLUM MODEL 2200
SCALER RATEMETER
November 2020
Serial No. 185785 and Succeeding
Serial Numbers
STATEMENT OF WARRANTY
Ludlum Measurements, Inc. warrants the products covered in this manual to be free
of defects due to workmanship, material, and design for a period of twelve months
from the date of delivery. The calibration of a product is warranted to be within its
specified accuracy limits at the time of shipment. In the event of instrument failure,
notify Ludlum Measurements to determine if repair, recalibration, or replacement is
required.
This warranty excludes the replacement of photomultiplier tubes, G-M and
proportional tubes, and scintillation crystals which are broken due to excessive
physical abuse or used for purposes other than intended.
There are no warranties, express or implied, including without limitation any implied
warranty of merchantability or fitness, which extend beyond the description of the
face there of. If the product does not perform as warranted herein, purchaser’s sole
remedy shall be repair or replacement, at the option of Ludlum Measurements. In no
event will Ludlum Measurements be liable for damages, lost revenue, lost wages, or
any other incidental or consequential damages, arising from the purchase, use, or
inability to use product.
RETURN OF GOODS TO MANUFACTURER
If equipment needs to be returned to Ludlum Measurements, Inc. for repair or calibration, please
send to the address below. All shipments should include documentation containing return shipping
address, customer name, telephone number, description of service requested, and all other necessary
information. Your cooperation will expedite the return of your equipment.
LUDLUM MEASUREMENTS, INC.
ATTN: REPAIR DEPARTMENT
501 OAK STREET
SWEETWATER, TX 79556
800-622-0828 325-235-5494
FAX 325-235-4672
Table of Contents
1. INTRODUCTION .................................................................................................................................... 1
2. SPECIFICATIONS ................................................................................................................................... 2
3. DESCRIPTION OF CONTROLS AND FUNCTIONS ........................................................................... 4
FRONT PANEL ............................................................................................................................................ 4
REAR PANEL ............................................................................................................................................. 5
INTERNAL CONTROLS ............................................................................................................................... 5
4. SAFETY CONSIDERATIONS AND WARNING MARKINGS ............................................................ 6
ENVIRONTMENTAL CONDITIONS FOR NORMAL USE ................................................................................. 6
CLEANING INSTRUCTIONS AND PRECAUTIONS .......................................................................................... 6
WARNING MARKINGS AND SYMBOLS ........................................................................................................ 7
REPLACEMENT OF FUSES ........................................................................................................................... 8
RETURN FOR REPAIR AND CALIBRATION ................................................................................................... 8
5. OPERATIONG PROCEDURES .............................................................................................................. 9
POWER ........................................................................................................................................................ 9
DISCR AND THRESHOLD SETTINGS ....................................................................................................... 9
OPERATING POINT .................................................................................................................................... 10
DETERMINING INSTRUMETN PLATEAU .................................................................................................... 11
OPERATING THE INSTRUMENT ................................................................................................................. 11
6. WINDOW OPERATION AND ENERGY CALIBRATION PROCEDURES ...................................... 12
GENERAL .................................................................................................................................................. 12
EQUIPMENT REQUIRED ............................................................................................................................ 12
PROCEDURE TO CALIBRATE THRESHOLD TO 10 KEV/TURN .................................................................... 12
PROCEDURE TO CALIBRATE THRESHOLD TO 100 KEV/TURN .................................................................. 13
7. DATA OUTPUT ..................................................................................................................................... 14
GENERAL .................................................................................................................................................. 14
SOFTWARE ................................................................................................................................................ 14
BUTTON FUNCTIONS ................................................................................................................................ 15
RS-232 PARAMETERS ............................................................................................................................... 16
8. RECYCLING .......................................................................................................................................... 17
9. PARTS LIST ........................................................................................................................................... 18
MAIN BOARD (DRAWING 167 X 413) ....................................................................................................... 18
AMPLIFIER BOARD (DRAWING 167 X 392) ............................................................................................... 20
SWITCH BOARD (DRAWING 167 X 273) ................................................................................................... 22
PUSHWHEEL BOARD (DRAWING 167 X 264) ............................................................................................ 22
DISPLAY BOARD (DRAWING 167 X 276) .................................................................................................. 22
WIRING DIAGRAM (DRAWING 167 X 418) ................................................................................................ 23
APPENDIX A CALIBRATION PROCEDURE ................................................................................... A-1
APPENDIX B ELECTRONICS CHECKOUT PROCEDURE ............................................................. B-1
APPENDIX C SCALER PRINTER OPTION ....................................................................................... C-1
PRINTER SPECIFICATIONS .................................................................................................................... C-1
CONFIGURATION ................................................................................................................................... C-1
APPENDIX D DRAWINGS AND DIAGRAMS ................................................................................. D-1
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 1 November 2020
Introduction
The Model 2200 Scaler Ratemeter is a self-contained counting instrument
designed for operation with scintillation, proportional, or GM (Geiger Muller)
detectors. Counting information is displayed both as a scaler count on the LED
(light-emitting diode) display and as an averaged count rate on the analog scaler.
The instrument can be powered by 95-250 Vac, 50-60 Hz, or by four “Dcell
batteries. The unit is complete with a voltage-sensitive preamplifier, linear
amplifier, electronic timer, and detector high-voltage power supply. A single-
channel analyzer is also featured in this unit for use in gamma spectrum analysis.
The analyzer may be switched on or off, allowing gross window counting. Ten-
turn potentiometers are supplied for threshold, window, and high-voltage
controls.
An RS-232 serial port is provided, which can be attached to a computer or
optional printer. A computer interface cable and software is provided to
start/stop and record counts.
Section
1
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 2 November 2020
Specifications
Meter: 1 mA, 6.4 cm (2.5 in.) scale, self-shielded DC mechanism
Ratemeter: four-decade, linear ratemeter with ranges of 0-500; 0-5000; 0-
50,000; and 0-500,000 cpm (counts per minute); 6.4 cm (2.5 in.) scale panel
meter; separate scales for battery check and high-voltage readout
Ratemeter Accuracy: within 10% of reference value
Scaler: a six-digit, LED readout
Scaler Accuracy: within 2% of reference value
Input Sensitivity: voltage-sensitive for use with all types of detectors
proportional, GM, scintillator
High Voltage: adjustable with a 10-turn control from 200-2500 V; supports
60 megohm scintillation loads to 1500 V
Time: time base is normally in minutes from 0-999 with multiples of X0.1
and X1. (Optional: Instrument can be switched to a seconds time base by
switching the internal dipswitch SW1 “2” to the ON position. In this case,
the count time may be set to 0-999 seconds with a multiplier switch of X1
and X10.)
Timer Accuracy: time base for the scaler is crystal-controlled and has a timer
accuracy of ±0.2% of the thumbswitch setting
Response Time: in fast position, 4 seconds; in slow position, 22 seconds, for
10-90% of full-scale deflection
Input Connection: series “C” coaxial connector; other connectors are
available upon request
Line Power: 95-250 Vac, 50-60 Hz single phase
Batteries: four “D” cells with typical life of 120 hours; self-contained in the
instrument
Section
2
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 3 November 2020
Temperature Range: -20 to 50 °C (-4 to 122 °F); may be certified to operate
from -40 to 65 °C (-40 to 150 °F)
Size: 21.6 x 12.7 x 21.6 cm (8.5 x 5 x 8.5 in.) (H x W x L), excluding handle
Weight: 3.4 kg (7.5 lb), including battery
Finish: beige powder coat with sub-surface printed front panel
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 4 November 2020
Description of
Controls and Functions
Front Panel
COUNT LAMP: a red LED, indicating that the scaler is in the count cycle.
COUNT SWITCH: resets and starts the scaler counting. The scaler turns off at the end
of the preset time.
COUNT TIME: Time base is normally in minutes from 0-999 with multiples of X0.1
and X1. It may be changed to seconds by an internal dipswitch.
MINUTES: a three-decade thumbwheel switch used for presetting count time.
Ratemeter Function Selector: a three-position switch labeled RATE, HV, and BAT.
The function of this switch is to allow the operator to have ratemeter readout on the
meter, HV readout on the meter, or battery check status on the meter.
THRESHOLD: A 10-turn potentiometer is used to set the basic pulse discrimination
point of the scaler. This control is linear in the range of 1.00 through 10.00. If set below
1.00, in addition to non-linearities, the system noise can cause errors.
WINDOW: a 10-turn potentiometer used for adjusting the window width. It is
calibrated with the THRESHOLD control so that one turn of the window control is
equivalent to one turn of the THRESHOLD control.
WINDOW ON-OFF: a toggle switch allowing insertion (ON) or removal (OFF) of the
window.
Detector Input Connection: a series “C coaxial connector; other connectors are
available upon request.
DISCR: a one-turn potentiometer for setting the range of the threshold potentiometer.
It is a non-indicating control, allowing a wide selection of operating points without
exceeding the linear ranges of the threshold/window circuits or detector voltage rating.
Maximum amplifier sensitivity occurs at maximum clockwise setting.
RANGE Selector Switch: a four-position switch providing range multipliers of X1,
X10, and X1K. With a scale of 0-500 cpm, the full range of the instrument is 0-500,00
cpm.
Section
3
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 5 November 2020
ZERO Switch: When depressed, this switch discharges the integrating capacitor, driving
the meter to zero.
F-S Response: a two-position toggle switch indicating fast or slow response. In “F
position, the meter will deflect from 10 to 90% of full scale in 4 seconds. In “S” potion,
the meter will deflect from 10 to 90% of full scale in 22 seconds.
HV: a 10-turn potentiometer control for adjusting HV from 200 V to 2500 V. It
provides a linear adjustment of the detector voltage supply. Changing the detector voltage
will cause the detector gain to change. A linear change in voltage will cause an exponential
change in detector gain. The instrument will support 100 megohm scintillation loads to
1500 V.
RATEMETER: a four-decade linear meter with ranges of 0-500, 0-5000, 0-50,000, and
0-500,000 cpm. Readout is on a 6.4 cm (2.5 in.) scale panel meter. Separate scales are for
battery check and high-voltage readout in conjunction with the Rate, HV, and Bat Switch.
Count Readout: a six-decade, LED readout indicating the count.
Rear Panel
RS-232 Connector: a 9-pin “Dconnector, allowing computer or printer output.
RS-232 Data Dipswitch: two-pole dipswitch located on the rear chassis
underneath the calibration cover; controls the RS-232 data. The top switch,
labeled “PC” and “PRNTR,” controls the type of data. In PC mode, data is bi-
directional, allowing the computer to start/stop counting. If the dipswitch is set
to PC, and the pushwheels are set to all zeroes, the instrument count time can
also be changed by the computer through the RS-232 port. In the PRNTR mode,
the scaler count is output at the end of each count only. If the dipswitch is set to
PRNTR, setting the pushwheels to 000 sets the count time to infinity, allowing
for very long count times. The bottom dipswitch, marked RECYCLE and
“OFF,” allows counting to recycle instead of requiring a manual count for each
cycle.
Input Power: a receptacle for 95-250 Vac, 50-60 Hz line cord.
Battery Compartment: holds fourD” cell batteries.
Internal Controls
Ratemeter Calibration Controls: four potentiometers used to calibrate the
ratemeter. Each range control is labeled respectively to the ratemeter range switch
on the front panel.
Time base Dipswitch: two-pole dipswitch located on the main circuit board. It
allows for selection of a seconds time base, variable from 0 to 999 seconds, with a
multiplier switch of X1 and X10.
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 6 November 2020
Safety Considerations and
Warning Markings
Environmental Conditions for Normal Use
Indoor use only
No maximum altitude
Temperature range of -20 to 50 °C (-4 to 122 °F)
Maximum relative humidity of less than 95% (non-condensing)
Mains supply voltage range 95-250 Vac
Maximum transient voltage of 1500 Vac
Installation Category (Overvoltage Category) II (as defined by IEC 1010-1)
Pollution Degree 2 (as defined by IEC 644)
Cleaning Instructions and Precautions
The Model 2200 may be cleaned externally with a damp cloth, using only water
as the wetting agent. Do not immerse the instrument in any liquid. Observe the
following precautions when cleaning:
1. Turn the instrument OFF and disconnect the instrument power cord.
2. Allow the instrument to sit for one minute before cleaning.
Section
4
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 7 November 2020
Warning Markings and Symbols
Caution!
The operator or responsible body is cautioned that the protection
provided by the equipment may be impaired if the equipment is
used in a manner not specified by Ludlum Measurements, Inc.
Caution!
Verify instrument voltage input rating before connecting to a power
converter. If the wrong power converter is used, the instrument
and/or power converter could be damaged.
The Model 2200 is marked with the following symbols:
DIRECT CURRENT (DC) (IEC 417, No. 5032) - designates an input
receptacle that accommodates a power cord intended for connection to DC
voltages. This symbol appears on the front panel.
PROTECTIVE CONDUCTOR TERMINAL (per IEC 417, No. 5019)
designates the central grounding point for the safety ground. This symbol is
visible inside the chassis.
ALTERNATING CURRENT (AC) (IEC 417, No. 5032) - designates an
input receptacle that accommodates a power cord intended for connection to
AC voltages. This symbol appears on the back panel.
CAUTION (per ISO 3864, No. B.3.1) designates hazardous live voltage
and risk of electric shock. During normal use, internal components are
hazardous live. This instrument must be isolated or disconnected from the
hazardous live voltage before accessing the internal components. This
symbol appears on the front panel. Note the following precautions:
Warning!
The operator is strongly cautioned to take the following precautions to
avoid contact with internal hazardous live parts that are accessible using
a tool:
1. Turn the instrument power OFF and disconnect the power cord.
2. Allow the instrument to sit for one minute before accessing internal
components.
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 8 November 2020
The crossed-out wheelie binsymbol notifies the consumer that the product
is not to be mixed with unsorted municipal waste when discarding; each material
must be separated. See section 8, Recycling for further information. Also
displayed on the side panel.
The “CE” mark is used to identify this instrument as being acceptable for use
within the European Union.
Replacement of Fuses
Warning!
For continued protection against risk of fire, replace only with fuses of
the specified type and current rating!
Return for Repair or Calibration
To return an instrument for repair or calibration, provide sufficient packing
material to prevent damage during shipment.
Every returned instrument must be accompanied by an Instrument Return
Form, which can be downloaded from the Ludlum website at
www.ludlums.com. Find the form by clicking the “Support” tab and selecting
“Repair and Calibration from the drop-down menu. Then choose the
appropriate Repair and Calibration division where you will find a link to the
form.
Note:
Ludlum Measurements, Inc. recommends recalibration at intervals
no greater than one year, assuming that regular operational checks
are performed. Check the appropriate local, state, and federal
regulations to determine required recalibration intervals.
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 9 November 2020
Operating Procedures
Power
Select either line or battery operation with the power switch.
Line Operation: Connect the instrument to line power of 95-250 Vac, 50-60
Hz. Turn the power switch to LINE and proceed to use the instrument.
Battery Operation: The battery lids are located at the rear of the instrument.
Use a coin or screwdriver to open the lids. Place 4 D” cells in the instrument
with the terminal “buttons” (positive ends) facing out. Turn the power switch to
BAT to operate. Check the battery condition by placing the RATE-HV-BAT
selector to BAT. (A downscale reading indicates battery reversal.)
Operational Check: Determine that the instrument is operational by pressing
COUNT. The count lamp should illuminate. (This action starts the scaler
count.)
DISCR and THRESHOLD Settings
The instrument DISCR control is calibrated at the factory so that 1.00 on the
threshold dial is equal to a 10 mV input sensitivity. The DISCR control should
not be changed unless a pulser is available to determine the threshold-versus-
millivolt relationship.
The threshold control should not be set below 1.00. Excessive counting may
occur at lower threshold settings.
Set DISCR and THRESHOLD for appropriate detector.
For scintillation detectors:
Set THRESHOLD on 1.00.
Connect scaler to pulser.
Set pulser pulse height at 10 mV.
Section
5
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 10 November 2020
Adjust DISCR until scaler meter reaches 75% of generated, incoming
count rate. The input sensitivity is now set for 10 mV with
THRESHOLD set for 1.00.
Proceed to use instrument with THRESHOLD setting at 1.00 or
more.
For Geiger Mueller (GM) detectors:
Set the THRESHOLD and the pulser as in the first three steps for
scintillation detectors.
Disconnect pulser and set the THRESHOLD on 3.00 and proceed
to use the instrument. The input sensitivity is approximately 30 mV.
At settings less than this, double-pulsing may occur. Some GM tubes
may require even higher settings of 7.00 or 8.00.
For proportional detectors:
Set the THRESHOLD on 1.00.
Connect the scaler to the pulser.
Set the pulser pulse height at 2 mV.
Adjust the DISCR until the scaler meter reaches 75% of the
generated incoming count rate. The input sensitivity is now set for 2
mV with the THRESHOLD SET ON 1.00.
Proceed to use instrument with THRESHOLD setting at 1.00 or
more.
Operating Point
The instrument and detector operating point is established by setting the
probe voltage (HV) and instrument sensitivity (DISCR and THRESHOLD).
Efficiency, background, and noise for a given detector system are fixed by
the physical makeup of the detector and rarely vary from unit to unit.
However, the selection of the operating point makes a marked difference in
the apparent contribution of these three sources of count.
In the singular case of the GM detector, a minimum operating voltage is
required to establish the GM operating region. (At lower voltages, the
detector operates as a very insensitive proportional counter.) This detector is
not capable of energy discrimination (pulse-height discrimination).
For gain sensitive detectors (proportional or scintillation), the most
straightforward method of selecting the operating point is to develop a
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 11 November 2020
graph, relating count rate to system gain. This relationship is commonly
referred to as a plateau or instrument plateau curve. System gain may be
changed by adjusting detector high voltage (HV), THRESHOLD control or
DISCR (amplifier gain).
Determining Instrument Plateau
Set window ON-OFF at OFF.
Set the time multiplier toggle switch to X0.1 and the MINUTES
thumbwheel switch to 001. This gives a 0.1 minute count.
Set THRESHOLD control at 1.0.
With the detector shielded from the source, adjust the high-voltage control
by 100-volt increments and take a plot of HV-versus-background coutn rate
until the detector voltage rating is reached. Return the HV control to zero.
Note: If the detector voltage is reached before any background counts are
detected, turn the DISCR clockwise to increase the sensitivity.
Expose the detector to a source and again make a plot of count-versus-
voltage.
Plot both sets of data and select the operating point to correspond with the
maximum source count and minimum background count. Avoid areas of
very fast count rate changes with small changes in detector voltage. The
optimum operating point for low background detectors is just above the
inflection point (or break-over point) of the plateau curve. If the background
count is irrelevant, shift the operating point to the plateau center for greater
stability.
Operating the Instrument
Ensure that all settings and adjustments have been properly made. Proceed
to use the instrument.
Model 2200 Scaler Ratemeter
Ludlum Measurements, Inc. Page 12 November 2020
Window Operation and Energy
Calibration Procedures
General
The Model 2200 is calibrated at the factory so that one turn on the
THRESHOLD controls is equal to one turn on the window dial.
Equipment Required
Detectors capable of energy discrimination. Examples used are the
Ludlum Models 44-3 and 44-2.
Known gamma radiation sources. Typical sources are 137Cs (662 keV)
and 241Am (60 keV).
Procedures to Calibrate
Threshold to 10 keV/Turn
Initial Model 2200 control settings
Window ON-OFF switch at OFF
Ratemeter RANGE selector switch at X10
THRESHOLD dial at 5.50
WINDOW dial at 1.00
HV dial at 0.00
Turn-on
Connect the Model 44-3 detector to the instrument.
Turn the power switch to the appropriate power supply.
Operation and Calibration
Expose 241Am to the detector.
Increase the HV setting until the count from the source just starts to
count.
Section
6
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62

Ludlum Measurements 2200 Owner's manual

Type
Owner's manual

Ludlum Measurements 2200 is a self-contained counting instrument designed for operation with scintillation, proportional, or GM (Geiger Muller) detectors. It features a voltage-sensitive preamplifier, linear amplifier, electronic timer, and detector high-voltage power supply. A single-channel analyzer is also featured in this unit for use in gamma spectrum analysis. The unit can be powered by 95-250 Vac, 50-60 Hz, or by four “D” cell batteries. An RS-232 serial port is provided, which can be attached to a computer or optional printer.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI