Invacare COMPRESSOR is a medical device designed to efficiently and conveniently compress oxygen into compatible cylinders. This allows users to refill their own oxygen supply, providing greater independence and flexibility in managing their oxygen therapy. The compressor is compatible with specific Invacare oxygen concentrators, enabling users to seamlessly integrate it into their home oxygen system.

With its user-friendly design, the Invacare COMPRESSOR simplifies the oxygen refilling process. By following the clear instructions provided in the user manual, users can safely connect the compressor to their concentrator and oxygen cylinder. The compressor's intuitive controls and indicator lights provide real-time feedback on the filling status, ensuring proper operation.

Invacare COMPRESSOR

Invacare COMPRESSOR is a medical device designed to efficiently and conveniently compress oxygen into compatible cylinders. This allows users to refill their own oxygen supply, providing greater independence and flexibility in managing their oxygen therapy. The compressor is compatible with specific Invacare oxygen concentrators, enabling users to seamlessly integrate it into their home oxygen system.

With its user-friendly design, the Invacare COMPRESSOR simplifies the oxygen refilling process. By following the clear instructions provided in the user manual, users can safely connect the compressor to their concentrator and oxygen cylinder. The compressor's intuitive controls and indicator lights provide real-time feedback on the filling status, ensuring proper operation.

User: Before using the product, read this manual
and save for future reference.
Dealer:
This manual must be given to the end user.
For more information regarding
Invacare products, parts, and services,
please visit www.invacare.com
User Manual
HomeFill
®
Compressor
WARNING
HomeFill
®
Compressor 2 Part No 1100873
WARNING
DO NOT use this product or any available optional
equipment without first completely reading and
understanding these instructions and any additional
instructional material such as owner’s manuals, service
manuals or instruction sheets supplied with this product
or optional equipment. If you are unable to understand
the warnings, cautions or instructions, contact a
healthcare professional, dealer or technical personnel
before attempting to use this equipment - otherwise,
injury or damage may occur.
NOTICE
Theinformationcontainedinthisdocumentissubjectto
changewithoutnotice.
ACCESSORIES WARNING
Invacare products are specifically designed and
manufactured for use in conjunction with Invacare
accessories. Accessories designed by other manufacturers
have not been tested by Invacare and are not
recommended for use with Invacare products.
NOTE:Updatedversionsofthismanualareavailableon
www.invacare.com.
TABLE OF CONTENTS
Part No 1100873 3 HomeFill
®
Compressor
SPECIAL NOTES ............................................... 5
TYPICAL PRODUCT PARAMETERS ................... 7
LABEL LOCATION ............................................ 9
DEALER INFORMATION ................................... 9
HOMEFILL INFORMATION .............................. 10
SECTION 1—IMPORTANT SAFETY
INSTRUCTIONS ............................................. 11
Operating Information .....................................................................11
Maintenance........................................................................................13
Double-Insulated Products .............................................................13
Disposal of Equipment and Accessories ......................................14
Radio Frequency Interference........................................................14
SECTION 2—TECHNICAL DESCRIPTION .......... 15
SECTION 3—INITIAL SETUP ............................ 16
Initial Set Up of Compressor..........................................................16
Compressor Initial Setup Checklist..........................................16
Choosing a location......................................................................16
Transporting the Compressor...................................................17
Installing the Power Cord...........................................................17
Connecting the Compressor to the Concentrator..............18
SECTION 4—CHECKLIST AND CYLINDER INSPEC-
TION ............................................................. 20
Compressor Operation Checklist.................................................20
Cylinder Prefill Inspection ...............................................................20
The Hydrostatic Testing Date...................................................22
External Examination ...................................................................22
SECTION 5—COMPRESSOR OPERATION .......... 23
TABLE OF CONTENTS
HomeFill
®
Compressor 4 Part No 1100873
Connecting/Disconnecting the Cylinder to/from the
Compressor ..............................................................................................23
Connecting the Cylinder to the Compressor........................23
Disconnecting the Cylinder from the Compressor.............. 26
Turning the Compressor On .........................................................28
SECTION 6—INDICATOR LIGHTS .................... 30
SECTION 7—TROUBLESHOOTING ................... 31
SECTION 8—MAINTENANCE .......................... 33
Cleaning the Compressor Filter ....................................................33
Cabinet ................................................................................................33
LIMITED WARRANTY .................................... 35
SPECIAL NOTES
Part No 1100873 5 HomeFill
®
Compressor
SPECIAL NOTES
Checkallpartsforshippingdamage.Incaseofdamage,DO
NOTuse.ContactInvacareforfurtherinstruction.
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothefollowingtablefordefinitions
ofthesignalwords.
ACCESSORIES
Therearemanydifferenttypesofoxygentubing,cannulas
andmasksthatcanbeusedwiththisdevice.Youshould
contactyourlocalhomecareproviderforrecommendations
onwhichofthesedeviceswillbebestforyou.Theyshould
alsogiveyouadviceontheproperusage,maintenance,and
cleaning.
CAUTION
“Caution:Federallawrestrictsthisdevicetosaleorrentalby
oronorderofaphysician,oranyotherpractitionerlicensed
bythelawoftheStateinwhichhe/shepracticestouseor
ordertheuseofthisdevice.”
SIGNAL WORD MEANING
DANGER Danger indicates an imminently hazardous
situation which, if not avoided, will result in
death or serious injury.
WARNING Warning indicates a potentially hazardous
situation which, if not avoided, could result
in death or serious injury.
CAUTION Caution indicates a potentially hazardous
situation which, if not avoided, may result in
property damage or minor injury or both.
SPECIAL NOTES
HomeFill
®
Compressor 6 Part No 1100873
DANGER
DO NOT SMOKE while using this device. Keep all
matches, lit cigarettes or other sources of ignition out of
the room in which this product is located and away from
where oxygen is being delivered.
NO SMOKING signs should be prominently displayed.
Textiles and other materials that normally would not
burn are easily ignited and burn with great intensity in
oxygen enriched air. Failure to observe this warning can
result in severe fire, property damage and cause physical
injury or death.
TYPICAL PRODUCT PARAMETERS
Part No 1100873 7 HomeFill
®
Compressor
TYPICAL PRODUCT PARAMETERS
TYPICAL PRODUCT PARAMETERS
- COMPRESSOR
Standard Listing: Double Insulated Product
ETL Certified as Complying with UL60601-1,
First Edition
ETL Certified Complying with CAN/CSA
C22.2 No. 601-1 M:90 (Canada)
EMC IEC 60601-1-2
Alternating Current
Type BF equipment
Unit running
Unit not running
Attention - Consider Accompanying
Documents
DO NOT smoke
Class II, Double Insulated
Not Protected
Electrical Requirements:
120 VAC ± 10% (108 VAC/132 VAC),
60 Hz
Rated Current Input:
2.0 AMP
Operating Temperature:
50 to 95° F (10 to 35° C) @ 20 to 60%
non-condensing humidity
Storage Temperature: -10 to 150° F (-23 to 65° C) @ 15 to
95% humidity
IPX0
TYPICAL PRODUCT PARAMETERS
HomeFill
®
Compressor 8 Part No 1100873
Input Pressure Required: 14 - 21 psi
Input Flow Required: 2 L/min
Input O
2
% Required: >90% O
2
Width: 20 ¼ inches
Height: 15 inches
Depth: 16 inches
Weight: 33 lbs
Shipping Weight: 39 lbs
No AP/APG:
Not suitable for use in the presence of a
flammable anaesthetic mixture.
Electrical:
No extension cords.
Placement:
No closer than 7.5 cm (3 in) from any
wall, furniture, draperies, or furniture to
assure sufficient air flow. Avoid deep pile
carpets and heaters, radiators or hot air
registers. Floor location only.
Tubing:
2 m (7 ft) cannula with a maximum 15 m
(50 ft) of Crush-Proof Tubing (DO NOT
pinch).
Time of Operation:
Up to 24 hours per day.
Recommended Storage and
Shipping Temperature:
-29°C to 65°C (-20°F to 150°F) at
15-95% relative humidity.
Environment:
Smoke and soot-free. No confined
spaces (Example: No closets).
Minimum Operating Time:
30 Minutes
LABEL LOCATION
Part No 1100873 9 HomeFill
®
Compressor
LABEL LOCATION
DEALER INFORMATION
NOTE:Invacarerecommendsleavingafullcylinderofoxygenwith
thepatientaftersettinguptheHomeFillcompressor.
BeforeleavingaHomeFillCompressorwithauser,the
followingchecklistMUSTbecompleted:
Makesurethepressuretestoncylinderhasnotexpired.
Instructtheuseronthesafeoperationoftheconcentrator
andreviewallwarnings.
InstructtheuseronthesafeoperationoftheHomeFill
compressorandreviewallwarnings.
Leaveafullcylinderofoxygenwiththeuser.
Leaveacopyofallwarningsandfillinginstructionswith
theuser.
WARNING
Keep all connections free of oil and grease to avoid violent
ignition. See Owner’s Manual before cleaning.
P/N 1075220 Rev C - 10/04
WARNING
DO NOT REMOVE UNIT COVER
This unit contains moving and high pressure
components. Manufacturer Service Only.
P/N 1075221 Rev B- 10/04
WARNING
Keep all connections free of oil and grease.
See Owner’s Manual before cleaning. Never oil or lubricate
coupling.
Cylinders filled on this unit MUST be labeled “Contents of
this cylinder are the product of an oxygen concentrator.”
P/N 1075238 Rev C - 11/17/06
HOMEFILL INFORMATION
HomeFill
®
Compressor 10 Part No 1100873
HOMEFILL INFORMATION
HomeFillisanaccessorydevicetotheoxygenconcentrator
thatenablesthepatienttorefillcustomportableoxygen
cylindersforpersonaluse.
NOTICE
TheHomeFillcompressorisforusewithspeciallydesigned
InvacarePlatinum™5,10,XLandPerfecto
2
HomeFill
compatibleoxygenconcentrators.
Theoxygenconcentratorcanbesettothefollowing
maximumflowrateswhenoperatingwithHomeFill
compressor:
Platinum5,XLandPerfecto
2
:3L/minorless.
Platinum10:5L/minorless.
TodetermineifyouroxygenconcentratorisHomeFill
compatible,examinethelowerrightcorneroftherearofthe
oxygenconcentratorforanoutletfitting.Refertothe
illustrationbelow.
TOP VIEW
Concentrator
Power Cord
Outlet Fitting
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS
Part No 1100873 11 HomeFill
®
Compressor
SECTION 1—IMPORTANT SAFETY
INSTRUCTIONS
WARNING
Thissectioncontainsimportantinformationforthesafe
operationanduseofthisproduct.
DANGER
Risk of electric shock. DO NOT disassemble. Refer
servicing to qualified service personnel. No user
serviceable parts.
TO REDUCE THE RISK OF BURNS, ELECTROCUTION,
FIRE, OR INJURY TO PERSONS
NEVER use near any type of flame or flammable/explosive
substances, vapors or atmosphere.
Operating Information
DONOTallowoilfromyourhandsorothersourcestocome
intocontactwiththecompressorcouplingorbottlefittings.
Thesesubstancesmaybecomeflammableinthepresenceof
oxygenandcauseinjury.Avoidtouchingthebottlefitting
wheneverpossible.Exampleofitemstoavoidarehand
creamsorlotions,cookingoils,suntanoilsorsunblocksand
similaroilyproducts.Ifyouusethesetypesofproducts,wash
yourhandspriortousingtheHomeFillunit.Ifthebottle
fittingshouldcomeincontactwithsuchsubstancesitshould
becleanedoffwithadampclothpriortoconnectingittothe
compressor.
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS
HomeFill
®
Compressor 12 Part No 1100873
NEVERoilorlubricatethecompressorcouplerorbottle
fittings.NEVERuseanypenetratingoilorlubricantsuchas
WD40
®
or3in1oil
®
.Ifconnectiondifficultyisexperienced
verifythatthebottlefittingisbeinginsertedstraightdown
intothecompressorcoupler.Ifcontinueddifficultyis
experiencedcontactyoursupplierforassistance.
Beforemovingorrepositioningeitherthecompressoror
concentrator,ALWAYSpushthepowerswitchtotheOff(O)
position,disconnecttheACpowercordandtheinterconnect
hosebetweenthecompressorandconcentrator.Failuretodo
somayresultindamagetotheunitorpersonalinjury.
ChildrenshouldALWAYSbesupervisedaroundthe
HomeFillunit.WhenusedwiththeHomeFilltableorsimilar
elevatedsupportsurfaces,achildcouldpulloveranysuch
tablecausingtheunittofall.Failuretoobservethiswarning
mayresultinpersonalinjuryorproductdamage.
Useextremecarewhenhandlingandfillinganoxygen
cylinder.Fulloxygencylindersareunderpressureandcan
becomeaprojectileifdroppedormishandled.
DONOTstorefilledoxygencylindersnearafurnace,hot
watertank,inthetrunkofacarorotherhightemperature
area.Storageinareassuchasthiscanresultinburstingofthe
cylinderorfire.
Theuseofthisdeviceislimitedtotheoxygenpatient.
CylindersMUSTbeusedONLYbytheoxygenpatientandare
nottobedistributedtoanyotherindividualforanypurpose.
DONOTuseoutdoors.
Themaximumpatientflowwhilefillingacylinderwiththe
IOH200is3L/minforPlatinum5,XLandPerfecto
2
concentratorsand5L/minforPlatinum10concentrators.
Patientswithahigherprescribedflowshouldnotfilla
cylinderwhilereceivingsupplementaloxygentreatment
fromthePlatinumConcentrator.
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS
Part No 1100873 13 HomeFill
®
Compressor
NEVERblocktheairopeningsofthe productorplaceitona
softsurface,suchasabedorcouch,wheretheairopeningmay
beblocked.Keeptheopeningsfreefromlint,hairandthelik e.
Ifthecompressorhasadamagedcordorplug,ifitisnot
workingproperly, ifithasbeendroppedordamaged,or
submersedinwater,callaqualifiedtechnicianfor
examinationandrepair.
IfanyleakageoftheoxygencylinderisdetectedDONOT
attempttousethecylinder.Turntheflowselectorknobonthe
regulatortoOff.Ifleakagepersists,placethecylinder
outdoorsandnotifyyourHomeHealthCareProvideror
ServiceRepresentativeofthiscondition.
TheHomeFillisequippedwithahighpressurereliefvalveto
ensuretheuserssafety.Whenactivated,thissafetyfeature
willmakeanextremelyloudnoise.Ifthisnoiseoccurs,turn
theunitOffandcontactyourHomeHealthCareProvideror
ServiceRepresentative.
Unplugunitfromoutletwhennotinuseandbeforecleaning.
Maintenance
DONOTdisassemble.Theunitisprovidedwithdouble
insulationandhasnoserviceableparts.Ifanyserviceis
required,contactInvacareTe chnicalService.
Double-Insulated Products
Inadoubleinsulatedproduct,twosystemsofinsulationare
providedinsteadofgrounding.Nogroundingmeansis
requiredonadoubleinsulatedproduct,norisameansfor
groundingtobeaddedtotheproduct.Adoubleinsulated
productismarkedwiththewords“DOUBLEINSULATION”
or“DOUBLEINSULATED”orthesymbol().
SECTION 1—IMPORTANT SAFETY INSTRUCTIONS
HomeFill
®
Compressor 14 Part No 1100873
DONOTunplugthepowercordbypullingonthecord.To
unplug,grasptheplugnotthecord,otherwisedamagemay
occur.
Disposal of Equipment and Accessories
Followlocalgoverningordinanceandrecyclingplans
regardingdisposalofdevicecomponents.
Radio Frequency Interference
Thisequipmenthasbeentestedandfoundtocomplywith
EMClimitsspecifiedbyIEC/EN6060112.Theselimitsare
designedtoprovideareasonableprotectionagainst
electromagneticinterferenceinatypicalmedicalinstallation.
Otherdevicesmayexperienceinterferencefromeventhelow
leve lsofelectromagneticemissionspermittedbytheabove
standards.TodetermineiftheemissionsfromthePerfecto
2
is
causingtheinterference,turnthePerfecto
2
Off.Ifthe
interferencewiththeotherdevice(s)stops,thenthePerfecto
2
iscausingtheinterference.Insuchrarecases,interference
maybereducedorcorrectedbyoneofthefollowing
measures:
•Reposition,relocate,orincreasetheseparation
betweentheequipment.
Connecttheequipmentintoanoutletonacircuit
differentfromthattowhichtheotherdevice(s)is
connected.
Avoidcreationofanysparknearmedicaloxygenequipment.
Thisincludessparksfromstaticelectricitycreatedbyanytype
offriction.
SAVE THESE INSTRUCTIONS
SECTION 2—TECHNICAL DESCRIPTION
Part No 1100873 15 HomeFill
®
Compressor
SECTION 2—TECHNICAL
DESCRIPTION
TheHomeFillOxygenSystem(IOH200)isan
electromechanicalprescriptionsystemforuseinthehome.
Itsintendedfunctionanduseistosupplypressurizedoxygen
tofillInvacareproprietarygascylindersforambulatoryuse
whilethepatientreceivestherapeuticoxygenfroman
accessoryoxygenconcentrator.TheIOH200gassystemis
designedforusewithanInvacarePlatinumoxygen
concentratorwithanintegralHF2accessoryportthatcan
supplyagasconcentrationgreaterthan90%O
2
around2
L/min.whilesupplyingapatienttherapyflowrateupto2.5
L/min.forPlatinum5and5L/min.forPlatinum9.
TheHomeFillOxygensystemconsistsofIOH200gas
compressor,connectiontubing,andtheInvacaresupplied
portablegascylinderswiththeintegralflowregulatorand
interlockingfillfitting.
TheIOH200compressorconsistsofamultistagecompressor,
internalreservoir,controlelectronicswithanintegraloxygen
concentrationsensor,interlockingfillfitting,inletfitting,and
drivemotor.Thecompressoroperationiscontrolledbythe
electronicsassembly.Themotor/compressorisallowedtofill
acylinderonlywhentheoxygenconcentrationoftheinletgas
streamisgreaterthan90%.Themotor/compressorfill
operationiscompletedwhenthegasinacylinderreachesthe
maximumfillpressureof2000PSI(13786kPa).
TheIOH200compressorcanfillcylindersforportableuseby
ambulatorypatients.TheInvacaresuppliedportablegas
cylindershaveaninterlockingfillfittingthatmatesonlywith
theInvacareIOH200compressorassembly.Thegascylinder
alsocontainsanintegralflowregulatorwithfixed,selectable
flowsettings.
SECTION 3—INITIAL SETUP
HomeFill
®
Compressor 16 Part No 1100873
SECTION 3—INITIAL SETUP
NOTE:DealersOnly‐RefertoDe alerInformationonpage 9befor e
performingthispr ocedure.
Initial Set Up of Compressor
Compressor Initial Setup Checklist
BeforeoperatingtheHomeFillCompressorforthefirsttime,
completethefollowingchecklist:
Choosealocationforthecompressor.
Transportthecompressortothedesiredlocation.
Connectthepow ercordtothecompressor.
Connectthecompressortotheconcentrator.
Choosing a location
Thecompressorshouldbeatleastthreeinches aw ayfromwalls,
draperiesorfurnituretoensuresufficientairflow.
Avoidplacingthecompressornexttoheaters,radiatorsorhotair
registers.
DONOTuseinacloset.
Thecompressorshouldsitonahardflatstablesurface,suchasa
tableortheReadyRac k.
WARNING
NEVERus enearanytypeofflameorflammable/explosive
substances,vaporsoratmosphere.
SECTION 3—INITIAL SETUP
Part No 1100873 17 HomeFill
®
Compressor
Transporting the Compressor
WARNING
NEVERtransportacompressorwithacylinderconnectedtoor
storedonthecompressor ,otherwiseinjuryordamagecanoccur.
CAUTION
Whentransportingthecompressor ,becarefulnottobumpthe
compressorortheconnectors intoobstacles.Otherwise,damage
tothecompressormayoccur.
NOTE:Forthisprocedur e, refertoFIGURE3.1.
1. Therearetw omethodsto
safelytransportthe
compressor:
•OneHand‐Graspthe
handgripatthe
bottomrearofthe
compressor.
•TwoHands‐Graspthe
handgripsonthe sides
ofcompressor.
FIGURE 3.1 Transporting the
Compressor
Installing the Power Cord
NOTE:Forthisprocedur e, refertoFIGURE3.2onpage 18.
1. Chooseaproperlocation. RefertoChoosingalocationon
page 16.
2. MakesurethepowerswitchisintheOff(O)position
3. Plugthepowercordintotheelectricconnectorinthebackof
thecompressor.
4. Plugtheoppositeendofthepowercordintothewalloutlet.
Hand Grip
Hand Grip
Hand Grip
SECTION 3—INITIAL SETUP
HomeFill
®
Compressor 18 Part No 1100873
FIGURE 3.2 Installing the Power Cord
Connecting the Compressor to the Concentrator
NOTE:Forthisprocedur e, refertoFIGURE3.3onpage 19.
1. Insertoneendoftheinterconnecthoseintotheoutletfitting
onthebackoftheconcentrator(Detail“A ) .
2. Inserttheoppositeendoftheinterconnecthoseintotheinlet
fittingonthesideofthecompressor(Detail“B”).
NOTE:Themetaltabsontheconcentratoroutletfittingandthe
compre ssorinletfittingwillpopoutwithanaudible“click”whenthe
endoftheinter connecthoseisproperlyinstalled(Detail“C”).
Wall
Outlet
Power
Cord
Compressor
(REAR)
Electrical
Connection
NOTE:Thisproductmaycomewithapolarizedplug(onebladeis
widerthantheother).Asasafetyfeature,thisplugwillfitina
polarizedoutletonlyoneway.Iftheplugdoesnotfitfullyinthe
outlet,reversetheplug.Ifitstilldoesnotfit,contactaqualified
electrician.DONOTattempttodefeatthissafetyfeature.
SECTION 3—INITIAL SETUP
Part No 1100873 19 HomeFill
®
Compressor
FIGURE 3.3 Connecting the Compressor to the Concentrator
Insert End of
Interconnect Hose into
Inlet/Outlet Fitting
Inlet/Outlet Fitting
Metal
Tab
Interconnect Hose
Compressor
Inlet Fitting
Power Cord
Outlet Fitting
CONCENTRATOR TOP VIEW
DETAIL “A”
DETAIL “C”
DETAIL “B”
SECTION 4—CHECKLIST AND CYLINDER INSPECTION
HomeFill
®
Compressor 20 Part No 1100873
SECTION 4—CHECKLIST AND
CYLINDER INSPECTION
Compressor Operation Checklist
EachtimetheHomeFillCompressorisusedtofillacylinder,
completethefollowingchecklist:
EnsuretheconcentratorhasbeenOnforatleastthirty
minutes.RefertotheconcentratorOwnersManual.
Performtheprefillinspectiononthecylinder.
Connectthecylindertothecompressor.
PushthecompressorpowerswitchtotheOn()position.
Examinetheindicatorlightsonthecontrolpanel.
Disconnectandremovethefullcylinder.
PushthecompressorpowerswitchtotheOff(O)position.
Iffillinganothercylinder,repeatthischecklist.
Cylinder Prefill Inspection
WARNING
ONLYusecylindersthathavetheInvacareHomeFill
connectionandHomeFilllabel.
AllcylindersMUSTbeinspectedbeforeattemptingtofill.
Otherwise,injuryordamagemayoccur.
NOTE:Forthisprocedure,refertoFIGURE4.1onpage 21.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Invacare COMPRESSOR User manual

Type
User manual
This manual is also suitable for

Invacare COMPRESSOR

Invacare COMPRESSOR is a medical device designed to efficiently and conveniently compress oxygen into compatible cylinders. This allows users to refill their own oxygen supply, providing greater independence and flexibility in managing their oxygen therapy. The compressor is compatible with specific Invacare oxygen concentrators, enabling users to seamlessly integrate it into their home oxygen system.

With its user-friendly design, the Invacare COMPRESSOR simplifies the oxygen refilling process. By following the clear instructions provided in the user manual, users can safely connect the compressor to their concentrator and oxygen cylinder. The compressor's intuitive controls and indicator lights provide real-time feedback on the filling status, ensuring proper operation.