안 전 규 격
■ Classifications:
- Type of protection against electrical shock: Class I
- Degree of protection against electrical shock (Patient connection): Type BF equipment
- Degree of protection against harmful ingress of water: Ordinary equipment
- Degree of safety of application in the presence of a flammable anesthetic material with air or
with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
- Mode of operation: Continuous operation
■ Electromechanical safety standards met:
IEC/EN 60601-1 Medical Electrical Equipment, Part 1General Requirements for Safety.
IEC/EN 60601-1-1 Safety requirements for medical electrical systems.
IEC/EN 60601-1-2 Electromagnetic compatibility -Requirements and tests.
IEC/EN 60601-2-37 Particular requirements for the safety of ultrasonic medical diagnostic and
monitoring equipment.
IEC 61157 Declaration of acoustic output parameters.
ISO 10993-1 Biological evaluation of medical devices.
UL 2601-1 Medical Electrical Equipment, Part 1 General Requirements for Safety.
CSA 22.2, 601.1 Medical Electrical Equipment, Part 1 General Requirements for Safety.