Run Digital Accumeasure
Support Program
5
series Quick Start Guide
Start Digital Accumeasure Support
Program
FClick Scan for Units
FClick Connect
7001-0090 Rev 1.0
Fif you are using USB, you are now
setup to acquire data.
Fto switch to Ethernet connection
please proceed to the following steps.
Using Ethernet connection (make sure the USB
cable is still connected)
6
Default communication is via USB. To switch
to Ethernet, go to User Settings tab >
Communications > Mode Choose Ethernet
Confirm the
default
configurations are
as shown.
Connect the Digital Accumeasure to your PC via RJ45
Ethernet Cable
FClick start or button
FGo to Control Panel
For Windows 7 or 8
F Under Network and Internet
F Click Network and Sharing Center
FClick Local Area Connection
FClick Properties
FChoose Internet Protocol Version 4
(TCP/IPv4)
7
For Windows XP
FUnder Network Connections
FRight Click Local Area Connection
FClick Use the following IP address:
Network Credentials
IP address:
Subnet mask:
Default Gateway:
192 . 168 . 168 . 249
255 . 255 . 255 . 0
255 . 255 . 255 . 0
FClick Properties
FChoose Internet Protocol (TCP/IP)
FClick Properties
FGo back to Digital Accumeasure Support
Program
FClick Scan for Units. You should now see the
network connected Digital Accumeasure
FClick Connect
FIf not, you may need to reboot your PC after setup
FMake sure to SAVE your changes.
FPower off/Unplug Power from
the Digital Accumeasure.
FPower up/Plug in the Digital
Accumeasure.
FThe LED indicator will show
the communication has been
switched to Ethernet.