Hill-Rom Envella Air Fluidized Therapy Bed Operating instructions

Type
Operating instructions
Serial number V222EN and earlier labeled for US market only
Envella ® Air Fluidized
Therapy System
Instructions for Use
(Rental Only)
Product No. P0819
194344 REV 6
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) i
REVISION
© 2021 by Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
PATENTS / PATENT hillrom.com/patents
May be covered by one or more patents. See the above Internet address.
The Hill-Rom companies are the proprietors of European, US, and other patents and pending patent
applications.
No part of this text shall be reproduced or transmitted in any form or by any means, electronic or
mechanical, including photocopying, recording, or by any information or retrieval system without
written permission from Hill-Rom Services, Inc. (Hill-Rom).
The information in this manual is confidential and may not be disclosed to third parties without the
prior written consent of Hill-Rom.
The information in this manual is subject to change without notice. Hill-Rom makes no commitment to
update or keep current the information in this manual.
Hill-Rom reserves the right to make changes without notice in design, specifications, and models. The
only warranty Hill-Rom makes is the express written warranty extended on the sale or rental of its
products.
This manual (194344) was originally released and supplied in English. For a list of available translations,
contact Hill-Rom Technical Support.
Product images and labels are for illustrative purposes only. Actual product and label may vary.
Sixth Edition, 2021-12
First Printing, 2016
Clorox Healthcare® is a registered trademark of Clorox Professional Products Company.
Oxycide® is a registered trademark of EcoLab USA, Inc.
Oxivir® and Virex® are registered trademarks of DIVERSEY, Inc.
Wex-Cide™ is a trademark of Wexford Labs, Inc.
Hillrom™ and Quick View™ are trademarks of Hill-Rom Services, Inc.
Envella®, Graphical Caregiver Interface (GCI)®, and Point of Care® are registered trademarks of Hill-Rom
Services, Inc.
Replace this manual (194344) if it is damaged and/or can not be read.
For product support or to order additional printed copies of this manual (194344), call 800-638-2546,
contact your local Hill-Rom representative, or go to hillrom.com
NOTE:
A digital version of the user manual is available on the bed’s GCI.
Reference Documents
Envella® Air Fluidized Therapy System Service Manual (194345)
Manufactured by:
HILL-ROM, INC.
1069 STATE ROUTE 46 E
BATESVILLE, IN 47006-9167 USA
ii Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
NOTES:
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) iii
Table of Contents
Table of Contents
Revision. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Intended Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Air Fluidized Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Pressure Redistribution Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Graphical Caregiver Interface (GCI)® Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Symbols. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Document Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Product Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Caregiver Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Pendant (Hand) Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Graphical Caregiver Interface (GCI)®. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Other GCI Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Bed Frame Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
California Proposition 65 Statement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Quick View™ List of Features and Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Alarm Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Notification Priority. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Bed Temperature Too High / Communication Issue. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Power Loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Alerts and Information Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Audible Alerts and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Brake Not Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Visual Alerts and Indicators on the Caregiver Control Panels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Visual Indicator on the Upper Frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Bed Frame Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
CPR Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Use the CPR Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Cord Hooks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Brake and Steer Controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
iv Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
Table of Contents
Siderails. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Remove a Siderail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Install a Siderail. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Patient Restraint and Drainage Bag Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Patient Restraints. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Drainage Bag Holders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Equipment Rail . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Equipment Sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Night Light . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Line Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Support Surface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Air Fluidization Therapy Beads (A). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Filter Sheet (B). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Pressure Redistribution Therapy Cushions (C). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Clean Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Point of Care® Bed Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Lockout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Head Up and Down. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Bed Up and Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Bed Therapy (Air Fluidization). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Turn Bed Therapy On and Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Surface Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Transfer a Patient On to the Bed from a Bed, Stretcher, or other Flat Surface . . . . . . . . . . . . . 24
Transfer a Patient Out of the Bed to another Bed, Stretcher, or other Flat Surface . . . . . . . . 25
Side Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Transfer a Patient On to the Bed from a Standing or Chair Position . . . . . . . . . . . . . . . . . . . . . . 26
Transfer a Patient Out of the Bed to a Side-Sitting Position for Standing or
Placement into a Chair. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Foot Control for Bed Therapy (Air Fluidization) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Pendant (Hand Control) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Adjust the Head Elevation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
To Adjust the Comfort Settings for the Low Back, Upper Back, and Head Comfort. . . . . . . . 27
Graphical Caregiver Interface (GCI)® Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Unlock the GCI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Lock the GCI. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) v
Table of Contents
Use the GCI Menu Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Access the Help (?) Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Bed Exit Alert. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Turn the Bed Exit Alert On . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Turn the Bed Exit Alert Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Silence the Bed Exit Alert without Deactivating the System . . . . . . . . . . . . . . . . . . . . . . . . . 33
View Bed Exit Alert History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Bed Exit Alert System Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Head Angle Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Activate the Head Angle Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
When the Head Angle Alert Sounds. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Turn Off the Head Angle Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
View the Head Angle Alert History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Head Angle Alert Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Change the Alert Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Change the Alert Tone. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Scale . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Scale Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Weigh the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
View Weight History. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Weigh Patient Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Zero the Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Zero Scale Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Body Mass Index (BMI) Calculator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Adjust Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Add/Remove Items. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Bed Therapy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Adjust the Comfort Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
Adjust the Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
View Bed Therapy History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Bed Therapy—Air Fluidization and Temperature Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Bed Therapy System Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Miscellaneous Messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Turn Reminder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
vi Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
Table of Contents
Preferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Adjust Date/Time. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Electrocardiogram and Electroencephalogram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Position the Patient. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Adjust the Patient’s Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Turn the Patient for Side Lying with a Pillow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Turn the Patient for Linen Changes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Turn the Patient for Back, Wound, and Perineal Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Adjust the Female Patient’s Position for Catheterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Adjust the Patient’s Position for Bedpan Use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Transport the Bed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
After Transport . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Patient Helper (P0820). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
Foot Support Cushion (195108) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Impervious Sheet (P007012A). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
Infusion Support System (P158A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
IV Pole (P2217A) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Oxygen Tank Holder, Vertical (P276 and P27601) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Detachable parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Cleaning and Disinfecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Cleaning and Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
Prepare the Bed for Cleaning and Disinfecting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
STEP 1: Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
STEP 2: Disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Process the Bed after Patient Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Service Calls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Storage and Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
Intended Use
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) 1
INTENDED USE
The Envella® Air Fluidized Therapy System is intended for medical purposes to help treat or prevent
pressure injuries, to treat severe or extensive burns, or to aid in circulation. This bed is an ideal support
for patients who have advanced pressure injuries, flaps, grafts, or burns, and require frequent transfers
or variable head elevation, and any other conditions appropriate for air fluidized therapy. The bed also
can be used for intractable pain, extensive epidermal detachment, Stevens-Johnson Syndrome, purpura
fulminans, induce relaxation which may reduce need for sedative and pain medication, improve patient
outcome, and reduce wound healing time. The bed permits easy positioning and egress, thereby
enhancing the independence of patients.
INTRODUCTION
The Envella® Air Fluidized Therapy System is a patient management system that combines air fluidized
therapy and pressure redistribution technologies together on a low-height frame. The lower body
section provides air fluidized therapy, and the upper body section provides pressure redistribution.
The bed is most applicable in the long term care and acute care, medical, surgical, and critical care
settings, but it may be used in other clinical areas as well.
The bed may be used when continuous or variable head elevation and/or patient transfers and egress
may be required.
The bed is equipped with a scale intended to weigh the patient in the bed.
Before you operate the bed, make sure you read and understand in detail the contents of this manual. It
is important that you read and obey the aspects of safety contained in this manual.
Any reference to a side of the bed is from the patient’s view lying in the bed on his or her back.
AIR FLUIDIZED THERAPY
To support the patient’s buttocks, legs, and feet, the lower-body portion of the bed provides the
maximum pressure relief of air fluidized therapy. The bed’s air fluidization chamber contains silicone
coated microspheres (beads).
When the bed is turned on, room air is drawn into the base of the bed, filtered, heated, and channeled
into the bed through a porous diffuser. The air blows upward through the beads, and puts them into
motion, causing the beads to take on the properties of a fluid that the patient’s body can float in.
If the airflow is turned off, the beads settle into a mold around the body contours, creating a solid
support.
The bed supports these types of patients:
Patients in need of aggressive treatment for
multiple or advanced pressure injuries
Patients receiving treatment for flaps, grafts, burns,
or intractable pain
Patients at high risk for skin breakdown (to prevent
pressure injuries)
Symbols
2 Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
PRESSURE REDISTRIBUTION THERAPY
To support the patient’s head, shoulders, and back, the upper-body portion of the bed delivers pressure
redistribution therapy by controlling the escape of air and pressure distribution in the inflatable
cushions. The air is automatically adjusted for the patient’s weight, as well as changes in the patient’s
position and head elevation. The patient receives support according to the individual’s comfort at
therapeutic pressures.
GRAPHICAL CAREGIVER INTERFACE (GCI)® CONTROL
SYMBOLS
DOCUMENT SYMBOLS
This manual contains different typefaces and symbols to make the content easier to read and
understand:
Standard text—used for regular data.
Boldface text—emphasizes a word or phrase.
NOTE:—sets apart special data or important instruction clarification.
WARNING, CONTRAINDICATION, or CAUTION
A CONTRAINDICATION identifies situations or actions that may have an effect on patient
safety.
A WARNING identifies situations or actions that may have an effect on patient or user safety.
To ignore a warning could cause patient or user injury.
A CAUTION identifies special procedures or precautions that persons must obey to help
prevent equipment damage.
The GCI at the foot end of the bed constantly provides
information about bed features and the bed status.
Menus guide you through the bed functions. To access
the features, press the controls on the GCI. Indicator lights
show when a feature is on.
Some bed controls are available on the pendant (hand control), footboard, and sides of the bed.
Symbols
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) 3
PRODUCT SYMBOLS
Caregiver Control Panel
Pendant (Hand) Controls
Item or
Symbol Description Item or
Symbol Description
A Bed therapy (air fluidization) control
with indicator
FSide transfer control with indicator
B Head up and down controls with
lockout indicator
GService required indicator
C Head angle display (on the head-
end side controls only)
H Lockout control
D Bed up and down controls with
lockout indicator
I Bed not down indicator
ESurface transfer control with indica-
tor
Bed Unplugged Alarm indicator (on
the footboard controls only)
Item Description
A Head up and down controls
B Comfort adjust zone select control with indicators
C Comfort adjust pressure increase and decrease controls with indi-
cators
Symbols
4 Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
Graphical Caregiver Interface (GCI)®
Item Description Item Description
A Home screen indicator J Help (?) control
B Temperature Status indicators—
Green = the surface is at its
setpoint
Orange = the surface is
heating
Blue = the surface is cooling
K Head Angle Alert Status indicator
C Home Menu control L Head Angle display
D Alerts Menu control M Bed Height Status indicator
E Scale Menu control N Screen Lock control
F Bed Therapy Menu control O Preemptive Bed Exit Alert Silence control
G Reminder Menu control P Air Fluidization Status indicator
H Preferences Menu control Q Scale Zero Status indicator
I Down (and up) arrow control R Bed Exit Alert Status indicator
Other GCI Symbols
Symbol Description Symbol Description
Alert Volume control Patient Height Input control
Alert Tone control Increase and Decrease controls
Temperature Adjust controls Return control
Temperature Unit control Identifies that the alarm system is in the
audio off state.
Identifies an alarm condition
Symbols
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) 5
Bed Frame Labels
Symbol Description Symbol Description
CPR control label WARNING (yellow and black)
Patient left head rail (on the
siderail)
CAUTION (white and black)
Patient right head rail Identifies patient chest restraint
location
Patient left foot rail Identifies patient waist/wrist
restraint location
Patient right foot rail Identifies patient ankle restraint
and/or drainage bag location
Siderail release controls Brake pedal
Hip position locator Steer pedal
Safe working load for the
footboard rail
Before transport sequence
Detachable intermediate
siderail—left
After transport sequence
Detachable intermediate
siderail—right
Crush warning: must consult
accompanying documents
Detachable head siderail—
left
Identifies the bed power cord
Detachable head siderail—
right
Refer to the service manual for
fuse replacement and identifica-
tion
IV pole interference Protective earth
Do not use with Oxygen
Tents
ATTENTION: Consult accompany-
ing documents
Safety Information
6 Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
SAFETY INFORMATION
CONTRAINDICATION:
Use of active therapy surfaces with patients with unstabilized spinal cord injury could cause serious
injury to the patient.
Warning: no equipment stor-
age
Must consult accompanying doc-
uments
Type B applied part accord-
ing to IEC 60601-1
According to IEC 60529, Rating
for protection against fluid
ingress and identified as equip-
ment that is protected against
spraying and splashing water
Medical Equipment with
respect to Electric Shock,
Fire and Mechanical Hazards
only in accordance with
ANSI/AAMI ES60601-1,
IEC60601-1, IEC60601-2-52,
and IEC80601-2-35, and
CAN/CSA C22.2 No. 60601-1
Manufacturer
Patient minimum and maxi-
mum weight range
Manufacture date
Patient minimum and maxi-
mum weight range
Catalog number
Safe working load Serial number
Total bed weight (including
the safe working load)
Medical Bed for Adults
Symbol Description Symbol Description
Safety Information
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) 7
WARNING:
Obey all warnings throughout the manual and also the warnings below. Failure to do so could cause
injury and/or equipment damage:
Warning—Read and understand all warnings in this manual and on the bed itself prior to use
with a patient.
Warning—Follow the product manufacturer’s instructions.
Warning—Contact Hill-Rom if any of these occur:
The product does not respond to user controls as described in this manual.
“Notify Hill-Rom” shows on the GCI and an alert sounds.
There is an electrical or mechanical problem with the bed.
Air fluidization is sluggish or uneven, and not caused by blankets and pillows on the
bed or recent fluid ingress into the beads.
Warning—To help prevent the risk of hospital bed fires, make sure facility persons follow the
safety tips in the FDA Public Health Notification: Safety Tips for Preventing Hospital Bed Fires. (US
only)
Warning—The potential for electrical shock exists with electrical equipment. Failure to follow
facility protocols may cause death or serious injury.
Warning—Connect the power cord to a hospital grade power outlet only. (To avoid risk of
electric shock, this equipment must only be connected to a supply mains with protective earth.)
Warning—Do not connect the power cord to an extension cord or multiple outlet strip. There is
a risk of overheating and fire may cause injury or damage.
Warning—Incorrect use or handling of the power cord may cause damage to the power cord
which could cause electrical shock or other injury. If damage has occurred to the power cord or
any of its components, immediately remove the bed from service, and contact Hill-Rom.
Warning—Make sure the location of the power cord from the bed to its power source is such
that the cord will not cause a trip hazard.
Warning—The bed is intended for patients who are within the specified weight limit of 32 kg to
159 kg (70 lb to 350 lb). Use of the bed with a patient whose weight is outside of the specified
limits may cause loss of therapeutic value and/or damage to the bed.
Warning—This bed is not designed for use with oxygen tents and their accessories. Use oxygen
administering equipment of the nasal, mask, or ventilator type only.
Warning—To prevent a risk of explosion, do not use oxygen administering equipment in the
presence of flammable anesthetics.
Warning—Use detachable parts and accessories from Hill-Rom only. Do not modify or change
the bed system without approval from Hill-Rom.
Warning—Rental only product—there are no user replaceable service parts. For service part
replacement, contact Hill-Rom.
Warning—Make sure that there are no foreign objects in the patient zone of the bed.
Warning—A patient with continuous exposure to warm, dry air is subject to increased
evaporative water loss. Monitor the patient’s fluid status, and provide hydration therapy as
needed.
Warning—This product contains chemicals known to the State of California to cause cancer
and birth defects or other reproductive harm.
Warning—Make sure that the patient is in the correct position on the bed. Patient injury can
occur.
Safety Information
8 Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
CAUTION:
Caution—Do not store anything under the bed. To do so could cause equipment damage.
NOTES:
To remove power from the bed, unplug its power cord from the AC power outlet.
To make sure grounding is reliable, fully insert the plug into the power source.
Make sure the position of the bed is such that you can quickly, without obstruction, unplug the
power cord from the main power supply if necessary.
Rental only product—do not dispose of any parts of the bed. Contact Hill-Rom.
CALIFORNIA PROPOSITION 65 STATEMENT
WARNING:
Warning—This product can expose you to chemicals including Lead and Di (2-ethylhexyl) phthalate
(DEHP), which are known to the State of California to cause cancer and birth defects or other
reproductive harm. For more information go to www.P65Warnings.ca.gov.
Quick View™ List of Features and Controls
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) 9
QUICK VIEW™ LIST OF FEATURES AND CONTROLS
Item Feature Page Item Feature Page
A Graphical Caregiver Interface
(GCI)®
28 I Siderail release 15
B Caregiver control panel 21 J Patient restraint (the ankle
restraints can also be used as a
drainage bag holders
16
C Foot support cushion 21 K CPR control 13
D Support surface 18 L Night light 18
E Pendant 27 M Foot control and holder 27
F Line manager 18 N Brake and steer control 14
G Equipment socket 17 O Equipment rail 17
H Power cord hook 14 P Hip position locator 61
Alarm Conditions
10 Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
ALARM CONDITIONS
Different alarms are generated by the bed to notify the caregiver of potentially hazardous conditions.
The intended position of the caregiver to respond to alarms is standing in front of the GCI or control
panel that is on the footboard of the bed.
There are two alarm conditions: Bed Temperature Too High / Communication Issue is a medium
priority alarm, and Power Loss is a low priority alarm.
NOTIFICATION PRIORITY
Only one alarm or alert will show on the display at one time. Alarms are prioritized over alerts.
BED TEMPERATURE TOO HIGH / COMMUNICATION ISSUE
NOTE:
When you plug the bed into the power outlet, make sure that you hear a single beep. This lets you know
that the Bed Temperature Too High / Communication Issue alarm is operating correctly.
Priority Indication Condition Visual Warning Auditory
Warning Resolution
Medium
(1 second
inherent
alarm
delay)
Bed Tem-
perature Too
High / Com-
munication
Issue
The temperature
in the bead
section of the
bed exceeds
104°F (40°C).
Or
There is a
connection
issue with the
temperature
sensor.
Warning on the
GCI; the screen
has a yellow bor-
der that flashes.
Triple beep that
sounds repeat-
edly.
Follow the on-
screen instruc-
tions.
Low Power Loss The bed power
cord is
disconnected
from the AC
power outlet.
Or
A power outage
has occurred.
The Bed
Unplugged
Alarm indicator
is on and is yel-
low colored.
None Check that the
power cord is
plugged in.
Wait for power
return.
A triple beep will sound repeatedly and Bed Therapy will be
disabled if either of these conditions occur:
The temperature in the bead section of the bed
exceeds104°F (40°C).
There is a connection issue with the temperature
sensor.
If the alarm sounds, remove the patient from the bed, and
contact Hill-Rom.
Alerts and Information Indicators
Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6) 11
POWER LOSS
WARNING:
Warning—A patient should not remain on the bed with air fluidization not operating for more than
two hours.
A power loss can occur under these conditions:
The bed power cord has been disconnected from the AC power outlet.
A power outage has occurred.
A bed fuse has blown.
During a power loss condition, the bed can not operate. If the bed has been without power for more
than 2 hours (air fluidization has not operated), remove the patient from the bed.
When power is restored, the bed will resume operation at its previous state: if air fluidization was on, it
will turn on; if air fluidization was off, it will remain off.
ALERTS AND INFORMATION INDICATORS
Alerts and Information Indicators provide the caregiver with audible indicators and visual indicators. For
detailed information about the Bed Exit System and Head Angle alerts, see “Alerts” on page 30.
NOTE:
There must be AC power to the bed for the indicators to operate.
AUDIBLE ALERTS AND INDICATORS
A single beep will sound when an activity is successful or completed.
A triple beep will sound when there is an error or caregiver attention is needed.
A triple beep will sound every 10 minutes when there is a bed malfunction.
To turn off the alarm sound, press the Bed Therapy control
on one of the caregiver control panels. To turn the sound
on again, press the Bed Therapy control.
NOTES:
To make sure the Power Loss Alarm indicator operates correctly, do as follows:
1. Unplug the bed, and make sure that the indicator comes on.
2. Plug the bed in, and make sure that the indicator turns off.
The bed must be plugged into AC power to operate.
The Power Loss Alarm indicator stays on for approximately 10 minutes after the bed has stopped
receiving AC power.
Alerts and Information Indicators
12 Envella® Air Fluidized Therapy System Instructions for Use (194344 REV 6)
A triple beep will sound every 5 seconds during these situations:
When air fluidization is off because of a fan malfunction.
When a patient is in the bed and air fluidization has been turned off for more than 30 minutes.
The alert can be silenced and will sound again in 15 minutes if air fluidization is still off.
Brake Not Set
The Brake Not Set is an audible alert only. When the bed is plugged into AC power, and you release the
brakes, the alert will sound. To silence the alert, either unplug the bed (for transport) or set the brakes.
VISUAL ALERTS AND INDICATORS ON THE CAREGIVER CONTROL PANELS
Item Indicator
A Service Required—flashes when the bed detects a malfunction. Contact Hill-Rom for
assistance.
B Bed Not Down—flashes when the bed is not in its lowest position.
C Head Angle Display and Alert (side control panels only)—the display digitally shows the
approximate angle of the head section from 0° to +60° with respect to the floor. The Head
Angle Alert indicator is on when the Head Angle Alert is armed. The Head Angle display
has an accuracy of ± 4°. (The head angle and its alert status also show on the GCI.)
VISUAL INDICATOR ON THE UPPER FRAME
A Hip Position Locator label is on both sides of the upper
frame in between the attachment points for the head and
intermediate siderails. The label identifies the correct
location of the patient's hips while on the bed.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94

Hill-Rom Envella Air Fluidized Therapy Bed Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI