wtw Turb 350 IR Operating instructions

Type
Operating instructions
Turb 350 IR
ba41114e02 04/2003
43
Operating manual
Portable Turbidimeter
44
Status at time of
going to press
The use of advanced technology and the high quality stan-
dard of our instruments are the result of a continuous devel-
opment. This may result in differences between this
operating manual and your instrument. Also, we cannot
guarantee that there are absolutely no errors in this manual.
Therefore, we are sure you will understand that we cannot
accept any legal claims resulting from the data, figures or
descriptions.
Copyright
© Weilheim 2000, WTW GmbH & Co. KG
Reproduction in whole - or even in part - is prohibited without
the express written permission of WTW GmbH & Co. KG,
Weilheim.
Printed in Germany.
List of contents
45
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
1.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
1.2 Keypad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . .52
2.2 General safety instructions . . . . . . . . . . . . . . . .52
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . 55
3.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . .55
3.2 Initial commissioning . . . . . . . . . . . . . . . . . . . . .55
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
4.1 Instructions for operating . . . . . . . . . . . . . . . . . .57
4.1.1 Marking and aligning cuvettes . . . . . . . .57
4.1.2 Venting the sample . . . . . . . . . . . . . . . . .59
4.2 Measuring turbidity . . . . . . . . . . . . . . . . . . . . . .60
4.3 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63
4.3.1 Basic information on calibration . . . . . . .63
4.3.2 Calibration procedures . . . . . . . . . . . . . .63
4.3.3 Preparing the calibration . . . . . . . . . . . .65
4.3.4 Four-point calibration . . . . . . . . . . . . . . .65
4.3.5 Partial calibration . . . . . . . . . . . . . . . . . .67
5 Maintenance, cleaning, disposal . . . . . . . . . . 71
5.1 Changing the batteries . . . . . . . . . . . . . . . . . . .71
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
5.2.1 Cleaning the measuring instrument . . . .73
5.2.2 Cleaning the cuvettes . . . . . . . . . . . . . . .73
5.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .74
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . 75
7 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . 77
8 Accessories, Options . . . . . . . . . . . . . . . . . . . 79
9 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
List of contents
46
Overview
47
1Overview
The Turb 350 IR enables rapid and reliable turbidity mea-
surements to be carried out on individual samples. This
complies with the measuring method used by ISO 7027 /
DIN EN 27027.
1
2
3
Cuvette
Alignment ring
Mark
4
5
Display
Keypad
TURB 350 IR
CAL
4283
NTU
1
2
3
4
5
Overview
48
1.1 Display
The LCD display has the following display elements:
1 Calibration symbol - instrument in the Calibration
mode
2 - Number of the selected calibration standard (when
the CAL display appears), or:
- Error number (flashes when the error symbol
appears)
3 Error symbol - flashes when an error occurs
4 Battery symbol - flashes if the battery needs to be
changed
5 Measuring unit NTU
6 Main line
...8
NTU
CAL ERR
13
2
4
123 4
5
6
Overview
49
1.2 Keypad
The Turb 350 IR has a keypad with the following 5 keys:
Key Function
Switch measuring instrument on/off
Increase a value;
Select settings
Start measurement
Reduce value;
Select settings
Call up Calibration mode
CAL
Overview
50
Safety
51
2 Safety
This operating manual contains basic instructions that you
must follow during the commissioning, operation and main-
tenance of the measuring instrument. Consequently, all re-
sponsible personnel must read this operating manual before
working with the measuring instrument.
The operating manual must always be available within the
vicinity of the measuring instrument.
Target group The measuring instrument was developed for measure-
ments on site and in the laboratory. We assume that, due to
their professional training and experience, operators are fa-
miliar with the safety measures that are necessary when
handling chemicals.
Symbols
used
Warning
indicates instructions that have to be followed to prevent
damage to your instrument.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. Appli-
cation reports.
Safety
52
2.1 Authorized use
Authorized use consists exclusively of the turbidity mea-
surement of individual samples on site or in a laboratory en-
vironment.
The technical specifications given in the T
ECHNICAL DATA
chapter must be observed. Only the operation and running
of the measuring instrument according to the instructions
given in this operating manual is authorized. Any other use
is considered unauthorized.
2.2 General safety instructions
This instrument left the factory in a technically safe and se-
cure condition.
Function and
operating safety
The smooth functioning and operational safety of the mea-
suring instrument can only be guaranteed if the generally
applicable safety measures and the specific safety instruc-
tions that appear in this operating manual are followed dur-
ing operation.
The smooth functioning and operational safety of the mea-
suring instrument can only be guaranteed under the climatic
conditions specified in the chapter, T
ECHNICAL DATA.
If the instrument was transported from a cold environment to
a warm environment, the formation of condensate can im-
pair the functioning of the instrument. In this event, wait until
the temperature of the instrument reaches room tempera-
ture before putting the instrument back into operation.
Safety
53
Safe operation If safe operation is no longer possible, the instrument must
be taken out of service and secured against inadvertent op-
eration!
Safe operation is no longer possible if the measuring instru-
ment
l has been damaged in transport
l has been stored under adverse conditions for a lengthy
period of time
l is visibly damaged
l no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the in-
strument.
Obligations of the
operator
The purchaser of this measuring instrument must ensure
that the following laws and guidelines are observed when
using dangerous substances:
l EEC directives for protective labor legislation
l National protective labor legislation
l Safety regulations
l Safety datasheets of the chemical manufacturers.
Safety
54
Commissioning
55
3 Commissioning
3.1 Scope of delivery
l Turb 350 IR portable turbidity measuring instrument
l Operating manual
l 4 calibration standards (0.02 / 10.0 / 100 / 1000 NTU)
l Two empty cuvettes
l 4 alignment rings
l Cleaning cloths
l Carrying case with short operating manual
l 4 alkali-manganese batteries, type AAA/Micro
3.2 Initial commissioning
Inserting
the batteries
Insert the batteries that are included in the scope of delivery
into the instrument as follows:
1
2
3
4
Commissioning
56
The instrument is ready for operation.
Note
The measuring instrument was calibrated and tested by the
manufacturer before delivery. Consequently, you can begin
measuring immediately. However, we recommend perform-
ing a calibration again following the initial commissioning
(see section 4.3 C
ALIBRATION). Thus, you will quickly be-
come familiar with operating the measuring instrument and
the calibration procedure.
1 Place the instrument with the operating panel face
down on a soft surface.
2 Undo the two screws (1).
3 Open the lid of the battery compartment (2).
4 Insert the batteries (3). While doing so, check that the
poles are correct according to the diagram in the bat-
tery compartment.
5 Replace the lid of the battery compartment.
6 Attach the lid with the two screws. At the same time,
make sure that the sealing gasket around the battery
compartment (4) forms a tight seal with the lid.
Operation
57
4Operation
4.1 Instructions for operating
4.1.1 Marking and aligning cuvettes
Even completely clean quality cuvettes exhibit tiny direction-
al differences in their light transmittance. Therefore, each
cuvette should be marked - both the measuring cuvettes as
well as the cuvettes with the calibration standards (refer to
section 2130 of the "Standard Methods for the Examination
of Water and Wastewater", 19th edition). Consequently,
each cuvette can always be inserted in the correct position
and you can achieve precise and reproducible measuring
results.
Clean the cuvette The cuvette must be absolutely clean (see section
5.2.2 C
LEANING THE CUVETTES).
Marking a cuvette This is how to mark a cuvette (empty or full):
1 Switch on the turbidity measuring instrument by
pressing the
key. Let the turbidity measuring in-
strument warm up for at least 75 s.
2 The cuvette must be closed with the black light pro-
tection cap. Make sure that the outside of the cuvette
is clean, dry, and free of fingerprints.
3 Insert the cuvette in the cuvette shaft.
4 Slowly rotate the cuvette by one complete rotation
(by 360 °) in small steps and hold the
key de-
pressed while doing so. After each step, wait for a
moment until the display is stable. Watch the display
of the turbidimeter while you rotate the cuvette.
5 Turn the cuvette back to the position with the lowest
displayed value.
Operation
58
Note
Always use the cuvette and the light protection cap together;
only in this way is the cuvette marked.
Aligning a marked
cuvette
A marked cuvette is aligned as follows:
Aligning an
unmarked cuvette
An unmarked cuvette is aligned as follows:
6 With the cuvette in this position, place an alignment
ring on the cuvette so that the arrow on the alignment
ring points to the mark on the enclosure.
7 Leave the marker ring on the light protection cap of
the cuvette. This cuvette is now permanently marked.
1 Insert the cuvette in the cuvette shaft.
2 Align the marked cuvette so that the arrow on the
alignment ring points to the marking on the enclo-
sure.
1 Insert the cuvette in the cuvette shaft.
2 Slowly rotate the cuvette by one complete rotation
(by 360 °) in small steps and hold the
key de-
pressed while doing so. After each step, wait for a
moment until the display is stable. Watch the display
of the turbidimeter while you rotate the cuvette.
3 Turn the cuvette back to the position with the lowest
displayed value.
4 Leave go of the
key.
Operation
59
4.1.2 Venting the sample
Air bubbles in the sample affect the measuring result to a
massive extent because they have an extensive scattering
effect on the incident light. Larger air bubbles cause sudden
changes in the measured values whereas smaller air bub-
bles are recorded by the instrument as turbidity. Therefore,
avoid or remove air bubbles:
Avoiding or remov-
ing air bubbles
l During sampling, ensure all movement is kept to a mini-
mum
l If necessary, vent the sample (ultrasonic baths, heating or
adding a surface-active substance to reduce surface ten-
sion)
Operation
60
4.2 Measuring turbidity
Warning
Never pour liquid directly into the cuvette shaft. Always use
a cuvette for the measurement. The measuring instrument
only measures precisely when the cuvette is closed with the
black light protective cap (WTW cuvettes).
Note
The outside of the cuvette always has to be clean, dry, and
free of fingerprints. Clean the cuvette before starting to mea-
sure (see section 5.2.2 C
LEANING THE CUVETTES). Only hold
the cuvettes by the top or by the black light protection cap.
Measuring This is how to measure the turbidity of a sample using the
Turb 350 IR:
1 Switch on the turbidity measuring instrument by
pressing the
key. Let the turbidity measuring in-
strument warm up for at least 75 s. During this time,
continue with the steps 2 to 5.
2 Rinse out a clean cuvette with the sample to be mea-
sured:
Pour approximately 10 ml sample into the cuvette.
Close the cuvette and rotate it several times before
throwing the sample away.
3 Repeat the rinsing procedure twice more.
4 Fill the cuvette with the sample to be measured (ap-
prox. 15 ml). Close the cuvette with the black light
protection cap.
5 Make sure that the outside of the cuvette is clean,
dry, and free of fingerprints.
6 Insert the cuvette in the cuvette shaft so that it clicks
into place.
7 Align the cuvette (see section 4.1.1 MARKING AND ALI-
GNING CUVETTES).
Operation
61
Display when
measuring range is
exceeded
If the measured value lies outside the measuring range of
the Turb 350 IR, the display shows "
-Or-
-Or--Or-
-Or-
" (Overrange):
8 Press the
key. As long as the measured value
is being recorded, the display flashes.
9 Read off the measured value when the measurement
is complete and the display no longer flashes:
10 Repeat steps 2 to 9 for further samples.
CAL
ERR
13
2
4
4 2 8 3
NTU
CAL
ERR
13
2
4
- O r -
NTU
Operation
62
Display when mea-
suring range is
switched over or
due to air bubbles
The measuring instrument may briefly display "
----
--------
----
":
This can be caused by the following:
l The measuring instrument carries out an automatic
switchover of the measuring range. Wait until the instru-
ment has completed the automatic switchover of the mea-
suring range.
l There are air bubbles in the sample. Vent the sample if
necessary.
CAL
ERR
13
2
4
- - - -
NTU
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42

wtw Turb 350 IR Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI