Amico NY 11788 Installation guide

Type
Installation guide

Amico NY 11788 is a horizontal console that provides medical gas and electrical access to patients in healthcare settings. It features adjustable lights for better visibility, a sleek design that blends seamlessly into modern healthcare environments, and easy installation with clear instructions and diagrams.

Amico NY 11788 is a horizontal console that provides medical gas and electrical access to patients in healthcare settings. It features adjustable lights for better visibility, a sleek design that blends seamlessly into modern healthcare environments, and easy installation with clear instructions and diagrams.

Installation Manual
Sapphire Series
Horizontal Console with Lights
1 Amico Corporation
Before Installation
Console installation should conform to local building codes.
The installation instructions provided are for the Amico standard products. Slight modifications to the
installations may be required. The product shipped to the job site may not be exactly as shown on the
installation drawings.
CAUTION
Please do not lift console by the plastic end caps.
1. Review a copy of the final approved shop drawing(s) and/or submittal package. These documents will provide you
with technical information specific to your installation, such as:
equipment types & quantities
room numbers and location
shop drawings of each unit type
wiring diagrams and rough-in locations
2. Locate the console unit from the shipping boxes that you wish to install. The contents of each shipping box are
listed on the shipping document(s).
www.amico.com 2
1. Using shop drawings, mark both devices center line AFF (above finished floor) on wall.
Installation
Devices Center Line AFF
2. Mark horizontal line for bottom of mounting bracket (to be installed later). This line is to be offset 4" (102 mm)
above the device’s center line.
Bottom of Mounting
Bracket Line AFF
4
Device Center Line AFF
*Unit shown for reference only
3 Amico Corporation
3. Using a stud locator, mark the location of any metal studs or blocking in the building wall that intersect the
console mounting location.
4. Mount the bracket(s) on the wall using hardware compliant to local building codes (supplied by others). Align
bracket horizontally according to the console mounting location. Refer to drawings. Bracket includes I.D. marks
for locating holes vertically, which are to be drilled on-site according to location of studs. Remember to align
bottom of bracket to “bottom of bracket” line marked in step 2.
Installation
Stud
Drywall
Stud Reference Line
Finished Floor
Bottom of Mounting
Bracket Line
Bottom of Mounting
Bracket Lines
Bracket
I.D. Mark for Holes
www.amico.com 4
Installation
5. Check rough-in drawing(s) for electrical and med gas rough-in locations. Remove any cover plates and devices,
then remove fascia panel(s) for access to rough-in locations.
6. Hang the console unit on the wall, over the mounting bracket, and center the unit with reference to bed center
line. Verify rough-in locations in wall and align with rough-in provisions on console.
7. Make all necessary electrical and medical gas connections as well as light connections, if required, at rough-in
locations.
Typical Rough-In
Fascia
Cover Plate
NOTE: End plate and cover removed in
graphic for clarity. Do not remove!
5 Amico Corporation
Installation
9. Install facia panel(s) back onto console, then install all devices and cover plates.
Mounting Locations
Back Plate
8. Secure console to the wall through back plate using hardware compliant to local building codes (provided by
others).
www.amico.com 6
Light Bulb Installation
1. The lens covers are snap fit into place. Remove the lens by carefully applying force at the fascia side and pulling
away.
2. Attach light bulbs (top and bottom).
3. Reattach the lens covers by lining up the cover with the groves in the extrusion and snapping the lens into place
at the fascia side.
www.amico.com
Amico Corporation | 85 Fulton Way, Richmond Hill, ON L4B 2N4, Canada
600 Prime Place, Hauppauge, NY 11788, USA
Toll Free Tel: 1.877.462.6426 | Tel: 905.764.0800 | Fax: 905.764.0862
Email: info@amico.com | www.amico.com
ACAIMSAPPHIREWITHLIGHTS 07.09.2021
C US LISTED
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Amico NY 11788 Installation guide

Type
Installation guide

Amico NY 11788 is a horizontal console that provides medical gas and electrical access to patients in healthcare settings. It features adjustable lights for better visibility, a sleek design that blends seamlessly into modern healthcare environments, and easy installation with clear instructions and diagrams.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI