Endres+Hauser KA Proline 500 Short Instruction

Category
Measuring & layout tools
Type
Short Instruction
Products Solutions Services
Brief Operating Instructions
Proline 500
HART transmitter
with ultrasonic time-of-flight sensor
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 2 of 2: Transmitter
Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →  3
KA01475D/06/EN/02.24-00
71636524
2024-01-31
Proline 500 HART
2 Endress+Hauser
Order code:
Ext. ord. cd.:
Ser. no.:
www.endress.com/deviceviewer Endress+Hauser
Operations App
XXXXXXXXXXXX
XXXXX-XXXXXX
XXX.XXXX.XX
Serial number
1.
3.
2.
A0023555
Proline 500 HART Brief operating instructions Flowmeter
Endress+Hauser 3
Brief operating instructions Flowmeter
The device consists of a transmitter and a sensor.
The process of commissioning these two components is described in two separate manuals
which together form the Brief Operating Instructions for the flowmeter:
Brief Operating Instructions Part 1: Sensor
Brief Operating Instructions Part 2: Transmitter
Please refer to both parts of the Brief Operating Instructions when commissioning the device,
as the contents of the manuals complement one another:
Brief Operating Instructions Part 1: Sensor
The Sensor Brief Operating Instructions are aimed at specialists with responsibility for
installing the measuring device.
Incoming acceptance and product identification
Storage and transport
Mounting procedure
Brief Operating Instructions Part 2: Transmitter
The Transmitter Brief Operating Instructions are aimed at specialists with responsibility for
commissioning, configuring and parameterizing the measuring device (until the first
measured value).
Product description
Mounting procedure
Electrical connection
Operation options
System integration
• Commissioning
Diagnostic information
Additional device documentation
These Brief Operating Instructions are Brief Operating Instructions Part 2:
Transmitter.
The "Brief Operating Instructions Part 1: Sensor" are available via:
Internet: www.endress.com/deviceviewer
Smart phone/tablet: Endress+Hauser Operations App
Detailed information about the device can be found in the Operating Instructions and the
other documentation:
Internet: www.endress.com/deviceviewer
Smart phone/tablet: Endress+Hauser Operations App
Table of contents Proline 500 HART
4 Endress+Hauser
Table of contents
1 About this document .............................................................. 5
1.1 Symbols used ......................................................................... 5
2 Safety instructions ................................................................ 7
2.1 Requirements for the personnel ............................................................ 7
2.2 Intended use .......................................................................... 7
2.3 Workplace safety ...................................................................... 8
2.4 Operational safety ......................................................................8
2.5 Product safety ......................................................................... 8
2.6 IT security ............................................................................ 8
2.7 Device-specific IT security ................................................................ 8
3 Product description ............................................................... 9
4 Mounting procedure .............................................................. 9
4.1 Mounting the sensor .................................................................... 9
4.2 Mounting the transmitter ................................................................ 9
4.3 Cover locking ........................................................................ 16
4.4 Weather protection cover ............................................................... 17
5 Electrical connection ............................................................ 18
5.1 Electrical safety .......................................................................18
5.2 Connecting requirements ............................................................... 18
5.3 Connecting the measuring device ......................................................... 21
5.4 Ensuring potential equalization ........................................................... 28
5.5 Ensuring the degree of protection ......................................................... 28
5.6 Post-connection check ..................................................................28
6 Operation options ............................................................... 29
6.1 Overview of operation options ............................................................ 29
6.2 Structure and function of the operating menu ................................................ 30
6.3 Access to operating menu via local display ................................................... 31
6.4 Access to the operating menu via the operating tool ............................................ 34
6.5 Access to the operating menu via the web server .............................................. 34
7 System integration .............................................................. 35
8 Commissioning .................................................................. 35
8.1 Installation and function check ........................................................... 35
8.2 Setting the operating language ........................................................... 35
8.3 Configuring the measuring device ......................................................... 36
8.4 Protecting settings from unauthorized access ................................................. 37
9 Diagnostic information .......................................................... 37
Proline 500 HART About this document
Endress+Hauser 5
1 About this document
1.1 Symbols used
1.1.1 Safety symbols
DANGER
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTICE
This symbol contains information on procedures and other facts which do not result in
personal injury.
1.1.2 Symbols for certain types of information
Symbol Meaning Symbol Meaning
Permitted
Procedures, processes or actions that
are permitted.
Preferred
Procedures, processes or actions that
are preferred.
Forbidden
Procedures, processes or actions that
are forbidden.
Tip
Indicates additional information.
Reference to documentation
A
Reference to page
Reference to graphic
1.
,
2.
,
3.
Series of steps
Result of a step Visual inspection
1.1.3 Electrical symbols
Symbol Meaning Symbol Meaning
Direct current Alternating current
Direct current and alternating current Ground connection
A grounded terminal which, as far as
the operator is concerned, is grounded
via a grounding system.
About this document Proline 500 HART
6 Endress+Hauser
Symbol Meaning
Potential equalization connection (PE: protective earth)
Ground terminals that must be connected to ground prior to establishing any other connections.
The ground terminals are located on the interior and exterior of the device:
Interior ground terminal: potential equalization is connected to the supply network.
Exterior ground terminal: device is connected to the plant grounding system.
1.1.4 Communication-specific symbols
Symbol Meaning Symbol Meaning
Wireless Local Area Network
(WLAN)
Communication via a wireless, local
network.
Bluetooth
Wireless data transmission between
devices over a short distance.
LED
Light emitting diode is on.
LED
Light emitting diode is off.
LED
Light emitting diode is flashing.
1.1.5 Tool symbols
Symbol Meaning Symbol Meaning
Torx screwdriver Flat-blade screwdriver
Phillips head screwdriver Allen key
Open-ended wrench
1.1.6 Symbols in graphics
Symbol Meaning Symbol Meaning
1, 2, 3,... Item numbers
1.
,
2.
,
3.
Series of steps
A, B, C, ... Views A-A, B-B, C-C, ... Sections
-
Hazardous area
Safe area (non-hazardous area)
Flow direction
Proline 500 HART Safety instructions
Endress+Hauser 7
2 Safety instructions
2.1 Requirements for the personnel
The personnel must fulfill the following requirements for its tasks:
Trained, qualified specialists must have a relevant qualification for this specific function
and task.
Are authorized by the plant owner/operator.
Are familiar with federal/national regulations.
Before starting work, read and understand the instructions in the manual and
supplementary documentation as well as the certificates (depending on the application).
Follow instructions and comply with basic conditions.
2.2 Intended use
Application and media
The measuring device described in this manual is intended only for the flow measurement of
liquids.
Depending on the version ordered, the measuring device can also measure potentially
explosive, flammable, poisonous and oxidizing media.
Measuring devices for use in explosive atmospheres, in hygienic applications or where there is
a high risk of pressures, are labeled accordingly on the nameplate.
To ensure that the measuring device is in proper condition during the operation period:
Only use the measuring device in full compliance with the data on the nameplate and the
general conditions listed in the Operating Instructions and supplementary documentation.
Refer to the nameplate to check whether the ordered device can be operated for the
intended application in areas requiring specific approvals (e.g. explosion protection,
pressure equipment safety).
Use the measuring device only for media to which the process-wetted materials are
sufficiently resistant.
Keep within the specified pressure and temperature range.
Keep within the specified ambient temperature range.
Protect the measuring device permanently against corrosion from environmental
influences.
Incorrect use
Non-designated use can compromise safety. The manufacturer is not liable for damage caused
by improper or non-designated use.
Residual risks
LCAUTION
Risk of hot or cold burns! The use of media and electronics with high or low
temperatures can produce hot or cold surfaces on the device.
Mount suitable touch protection.
Use suitable protective equipment.
Safety instructions Proline 500 HART
8 Endress+Hauser
2.3 Workplace safety
When working on and with the device:
Wear the required personal protective equipment as per national regulations.
2.4 Operational safety
Damage to the device!
Operate the device in proper technical condition and fail-safe condition only.
The operator is responsible for the interference-free operation of the device.
2.5 Product safety
This measuring device is designed in accordance with good engineering practice to meet state-
of-the-art safety requirements, has been tested, and left the factory in a condition in which it
is safe to operate.
It meets general safety standards and legal requirements. It also complies with the EU
directives listed in the device-specific EU Declaration of Conformity. The manufacturer
confirms this by affixing the CE mark to the device..
2.6 IT security
Our warranty is valid only if the product is installed and used as described in the Operating
Instructions. The product is equipped with security mechanisms to protect it against any
inadvertent changes to the settings.
IT security measures, which provide additional protection for the product and associated data
transfer, must be implemented by the operators themselves in line with their security
standards.
2.7 Device-specific IT security
The device offers a range of specific functions to support protective measures on the operator's
side. These functions can be configured by the user and guarantee greater in-operation safety
if used correctly.
For detailed information on device-specific IT security, see the Operating Instructions for
the device.
2.7.1 Access via service interface (CDI-RJ45)
The device can be connected to a network via the service interface (CDI-RJ45). Device-specific
functions guarantee the secure operation of the device in a network.
The use of relevant industrial standards and guidelines that have been defined by national
and international safety committees, such as IEC/ISA62443 or the IEEE, is recommended.
This includes organizational security measures such as the assignment of access authorization
as well as technical measures such as network segmentation.
Transmitters with an Ex de approval may not be connected via the service interface (CDI-
RJ45)!
Proline 500 HART Product description
Endress+Hauser 9
3 Product description
The measuring system consists of a transmitter and two or one sensor sets.
The transmitter and sensor are mounted in physically separate locations. They are
interconnected via sensor cable(s).
1
2
2
A0041373
1 Transmitter with integrated ISEM
2 Sensor cable
For detailed information on the product description, see the Operating Instructions for
the device →  3
4 Mounting procedure
4.1 Mounting the sensor
For detailed information about mounting the sensor, see the Sensor Brief Operating
Instructions →  3
4.2 Mounting the transmitter
4.2.1 Mounting the transmitter housing
LCAUTION
Ambient temperature too high!
Danger of electronics overheating and housing deformation.
Do not exceed the permitted maximum ambient temperature .
If operating outdoors: Avoid direct sunlight and exposure to weathering, particularly in
warm climatic regions.
LCAUTION
Excessive force can damage the housing!
Avoid excessive mechanical stress.
Mounting procedure Proline 500 HART
10 Endress+Hauser
The transmitter can be mounted in the following ways:
Post mounting
Wall mounting
Wall mounting
Required tools
Drill with drill bit ⌀ 6.0 mm
20 (0.79)
100 (3.94)
100 (3.94)
! 8.6 (0.39)
! 18 (0.71)
! 10 (0.39)
A0029068
 1 Engineering unit mm (in)
Proline 500 HART Mounting procedure
Endress+Hauser 11
Pipe mounting
Required tools
Open-ended wrench AF 13
LWARNING
Order code for "Transmitter housing", option L "Cast, stainless": cast transmitters are
very heavy.
They are unstable if they are not mounted on a secure, fixed post.
Only mount the transmitter on a secure, fixed post on a stable surface.
1
4 x SW 13
! (! 0.79 to 2.75)20 70
A0029057
 2 Engineering unit mm (in)
Mounting procedure Proline 500 HART
12 Endress+Hauser
4.2.2 Turning the transmitter housing
To provide easier access to the connection compartment or display module, the transmitter
housing can be turned.
N
i
c
ht
u
n
t
e
r
a
r
e
ö
f
f
n
e
n
+E
ESC
1.
2.
3 mm
N
i
c
ht
u
n
t
e
r
a
r
e
ö
f
f
n
e
n
+E
ESC
3.
4 mm
4.
7 Nm (5.2 lbf ft)
A0029993
 3 Non-Ex housing
1. Depending on the device version: Loosen the securing clamp of the connection
compartment cover.
2. Unscrew the connection compartment cover.
3. Loosen the securing screw.
4. Turn the housing to the desired position.
5. Tighten the securing screw.
6. Screw on the connection compartment cover.
7. Depending on the device version: Attach the securing clamp of the connection
compartment cover.
Proline 500 HART Mounting procedure
Endress+Hauser 13
1.
2.
3.
4 mm 7 Nm (5.2 lbf !)
A
A
A0043150
 4 Ex housing
1. Loosen the fixing screws.
2. Turn the housing to the desired position.
3. Tighten the securing screws.
Mounting procedure Proline 500 HART
14 Endress+Hauser
4.2.3 Turning the display module
The display module can be turned to optimize display readability and operability.
N
i
c
ht
u
n
t
e
r
a
r
e
ö
f
f
n
e
n
+E
ESC
1.
2.
N
i
c
ht
u
n
t
e
r
a
r
e
ö
f
f
n
e
n
+E
ESC
3.
3 mm
+E
ESC
A0030035
1. Depending on the device version: Loosen the securing clamp of the connection
compartment cover.
2. Unscrew the connection compartment cover.
3. Turn the display module to the desired position: max. 8 × 45° in each direction.
4. Screw on the connection compartment cover.
5. Depending on the device version: Attach the securing clamp of the connection
compartment cover.
Proline 500 HART Mounting procedure
Endress+Hauser 15
4.2.4 Transmitter post-installation check
The post-installation check must always be performed after the following tasks:
Mounting the transmitter housing:
Post mounting
Wall mounting
Turning the transmitter housing
Turning the display module
Is the device undamaged (visual inspection)?
Turning the transmitter housing:
Is the securing screw firmly tightened?
Is the connection compartment cover screwed on tightly?
Is the securing clamp firmly tightened?
Turning the display module:
Is the connection compartment cover screwed on tightly?
Is the securing clamp firmly tightened?
Post and wall mounting:
Are the securing screws firmly tightened?
Mounting procedure Proline 500 HART
16 Endress+Hauser
4.3 Cover locking
NOTICE
Order code "Transmitter housing", option L "Cast, stainless": The covers of the
transmitter housing are provided with a borehole to lock the cover.
The cover can be locked using screws and a chain or cable provided by the customer on site.
The use of stainless steel chains or cables is recommended.
If a protective coating is applied, it is recommended to use a heat shrink tube to protect the
housing paint.
1
1
2
2
1
1
5 (0.2) min. 15 (0.6)
SW 2.5
M3
A0029799
1 Cover borehole for the securing screw
2 Securing screw to lock the cover
Proline 500 HART Mounting procedure
Endress+Hauser 17
4.4 Weather protection cover
146 (5.75)
48 (1.9)
12 (0.47)
280 (11.0) 255 (10.0)
134 (5.3) 30 (1.18)
A0029553
 5 Engineering unit mm (in)
Electrical connection Proline 500 HART
18 Endress+Hauser
5 Electrical connection
LWARNING
Live parts! Incorrect work performed on the electrical connections can result in an
electric shock.
Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the
device from the supply voltage.
In addition to the device fuse, include an overcurrent protection unit with max. 10 A in the
plant installation.
5.1 Electrical safety
In accordance with applicable national regulations.
5.2 Connecting requirements
5.2.1 Required tools
For cable entries: use appropriate tool
For securing clamp: Allen key 3 mm
Wire stripper
When using stranded cables: crimper for wire end ferrule
For removing cables from terminal: flat blade screwdriver ≤ 3 mm (0.12 in)
5.2.2 Requirements for connecting cable
The connecting cables provided by the customer must fulfill the following requirements.
Protective grounding cable for the outer ground terminal
Conductor cross-section < 2.1 mm2 (14 AWG)
The use of a cable lug enables the connection of larger cross-sections.
The grounding impedance must be less than 2 Ω.
Permitted temperature range
The installation guidelines that apply in the country of installation must be observed.
The cables must be suitable for the minimum and maximum temperatures to be expected.
Power supply cable (incl. conductor for the inner ground terminal)
Standard installation cable is sufficient.
Cable diameter
Cable glands supplied:
M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in)
Spring-loaded terminals: Suitable for strands and strands with ferrules.
Conductor cross-section 0.2 to 2.5 mm2 (24 to 12 AWG).
Proline 500 HART Electrical connection
Endress+Hauser 19
Signal cable
Current output 4 to 20 mA HART
A shielded cable is recommended. Observe grounding concept of the plant.
Current output 0/4 to 20 mA
Standard installation cable is sufficient
Pulse /frequency /switch output
Standard installation cable is sufficient
Double pulse output
Standard installation cable is sufficient
Relay output
Standard installation cable is sufficient.
Current input 0/4 to 20 mA
Standard installation cable is sufficient
Status input
Standard installation cable is sufficient
5.2.3 Connecting cable between the transmitter and sensor
Sensor cable for sensor - transmitter: Proline 500
Standard cable TPE: -40 to +80 °C (-40 to +176 °F)
TPE armored: -40 to +80 °C (-40 to +176 °F)
TPE halogen-free: -40 to +80 °C (-40 to +176 °F)
PTFE: -50 to +170 °C (-58 to +338 °F)
PTFE armored: -50 to +170 °C (-58 to +338 °F)
Cable length (max.) 30 m (90 ft)
Cable lengths (available for order) 5 m (15 ft), 10 m (30 ft), 15 m (45 ft), 30 m (90 ft)
Operating temperature Depends on the device version and how the cable is installed:
Standard version:
Cable - fixed installation 1): minimum –40 °C (–40 °F) or –50 °C (–58 °F)
Cable - movable installation: minimum –25 °C (–13 °F)
1) Compare details under the row "Standard cable"
Electrical connection Proline 500 HART
20 Endress+Hauser
5.2.4 Terminal assignment
Transmitter: supply voltage, input/outputs
The terminal assignment of the inputs and outputs depends on the individual order version of
the device. The device-specific terminal assignment is documented on an adhesive label in the
terminal cover.
Supply voltage Input/output 1 Input/output 2 Input/output 3
1 (+) 2 (–) 26 (+) 27 (–) 24 (+) 25 (–) 22 (+) 23 (–)
Device-specific terminal assignment: adhesive label in terminal cover.
Transmitter and sensor connection housing: connecting cable
The sensor and transmitter, which are mounted in separate locations, are interconnected by a
connecting cable. The cable is connected via the sensor connection housing and the
transmitter housing.
Terminal assignment and connection of the connecting cable .
5.2.5 Preparing the measuring device
Carry out the steps in the following order:
1. Mount the sensor and transmitter.
2. Sensor connection housing: Connect connecting cable.
3. Transmitter: Connect connecting cable.
4. Transmitter: Connect signal cable and cable for supply voltage.
NOTICE
Insufficient sealing of the housing!
Operational reliability of the measuring device could be compromised.
Use suitable cable glands corresponding to the degree of protection.
1. Remove dummy plug if present.
2. If the measuring device is supplied without cable glands:
Provide suitable cable gland for corresponding connecting cable.
3. If the measuring device is supplied with cable glands:
Observe requirements for connecting cables →  18.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Endres+Hauser KA Proline 500 Short Instruction

Category
Measuring & layout tools
Type
Short Instruction

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI