Mounting hardware inventory
Use these items to mount the wall system:
mounting rail bracket
accessory bin bracket
VESA mounting bracket (2 pieces)
self-adhesive foam strips (2 pieces)
• screws
Tools list
Use these tools to mount the wall system:
• #2 Phillips screwdriver
• level
• tape measure
• stud nder
• drill
• 1/8-inch (3.17 mm) diameter drill bit for “B” screw pilot holes
Mounting location
Mount the wall system to studs.
Mount the wall system within reach of the AC power outlet.
The power cord is 8 ft. (2.44 m) long.
Avoid brightly lit areas.
Blood pressure tubing is 5 ft. (1.52 m) long.
Position the wall system so that all instruments are accessible and in a location
that allows for ergonomic examinations.
CAUTION
Welch Allyn is not responsible for the integrity of any wall mounting interface. Welch Allyn recommends that you contact your Biomedical
Engineering Department or maintenance service to ensure professional installation, safety, and reliability of any mounting accessory.
CAUTION
Before mounting the wall system, consider the following
recommendations and wall system details to determine the best mounting location:
Sample room layout
1. Integrated Wall Panel
2. Examination table
x4
x2 x2 x4
x1
ABDEC
+
* Only available in 44WT
44WT 44XT
hillrom.com
Mat. No. 773560, 80028119 Ver.A
Revision date: 2021-03
© 2021 Welch Allyn, Inc. All rights reserved.
Hillrom Technical Support
hillrom.com/en-us/about-us/locations/
Welch Allyn, Inc. is a subsidiary of Hill-Rom Holdings, Inc.
77794-M4400 77794-2M4400
Welch Allyn®
Spot Vital Signs® 4400 Integrated Wall Panel
Assembly Instructions
2. Ax the mounting rail bracket to two studs at the selected height using two “B” screws.
3. Route the power cord through the channel in the back of the accessory bin bracket, then mount the bracket on the stud to the right at least 15 1/2 inches (39 cm)
below the mounting rail bracket using the two remaining “B” screws.
Mount the wall system
1. On the selected wall, nd and mark the studs, and choose the system height and corresponding height for the mounting rail bracket.
Recommendation: Place the mounting rail bracket 63 inches (1.6 m) from the oor.
x 4
B
2
3
CAUTION This drawing shows the physical relationships of the mounting brackets to each other and to the wall system after you complete the mounting
instructions. To avoid personal injury or equipment damage, do not place the wall system on the wall until you have completed all preliminary steps.
CAUTION Ensure that the upper “lip” of the bracket sticks out from the wall and that the bracket is level.
Stud
(reference)
15 1/2 in
(39 cm)
recommended
63 in
(1.6 m)
recommended
to floor
8 3/16 in
(20.8 cm)
reference
3 5/16 in
(8.4 cm)
reference
30 in
(76 cm)
reference
11 3/4 in
(29.8 cm)
reference
2
WARNING Ensure that the clips on the back of the wall system fully engage the mounting rail bracket.
The wall system should be level and ush to the wall.
WARNING Failure to install this security screw may result in personal injury and equipment damage.
x 1
A
4. Hang the wall system on the mounting rail bracket.
5. Secure the unit to the wall using the A screw.
3
x 2
D
1
2
3
4
4
x 2
C
1
2
3
4
1
2
3
5
x 4
E
1
2
1
2
3
* 44WT
6
3
Use these bolts to
adjust tilt tension.
1
2
7
2 h
Wall transformer for
otoscope/ophthalmoscope Spot Vital Signs 4400
KleenSpec®
dispenser
SpO2 cable management
1
2
1
2
3
8
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Hill-Rom Spot Vital Signs® 4400 Device User guide

Type
User guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI