0000CF04465 Iss. 1 12/16
Alarm Action
TITRATION NOT CONFIRMED
The infusion rate has been
changed, but has not been
confirmed and 2 minutes has
expired without any operation.
• Press the c button to silence the alarm,
then press the CANCEL softkey to clear
this message.
• Check infusion rate and confirm by pressing
the b button or press the h button to
revert to the previous rate.
• Pressing QUIT will cancel the titration and
keep the original rate..
VTBI DONE
The pre-set Volume To Be Infused
is complete.
• Prepare next infusion, if required.
AC POWER FAIL
AC Power has been disconnected
and the pump is operating on
battery power.
• Reconnect AC power supply or press the c
button to silence the alarm and continue
with battery operation.
• The alarm will automatically cancel if the AC
power supply is reconnected.
ADD DRUG NOT COMPLETE
The ADD DRUG operation has not
been completed.
• Press CANCEL softkey and complete ADD
DRUG option.
Error Code and Message
The alarm system has detected an
internal malfunction.
• Note the malfunction code.
• Remove pump from service for examination
by a qualified service engineer.
ATTENTION (with “3 Beeps”)
Pump left on hold for 2 minutes
without starting operation.
• Review pump setup.
• Start infusion or turn o pump.
*Alaris™ CC Syringe Pump and the Alaris™ CC Guardrails Syringe Pump only
Display Action
DOSE WOULD EXCEED
Infusion rate entered exceeds a Guardrails
soft limit.
• Check the infusion setting.
• To confirm OVERRIDE LIMIT?
press YES.
• To deny OVERRIDE LIMIT?
press NO.
DOSE UNDER
Infusion rate entered is under a Guardrails
soft limit.
Alarms continued