Promax 7350370080540 Syringe Pump User manual

Type
User manual
ProMAX
Syringe Pump
User Manual
Syringe Pump for Veterinary Use
Table of Contents
Preface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Application Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Applicable Object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Use Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Paraphrase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description of Revision of User Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Safety Information & Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Dialogue Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Application Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Expected Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Expected Working Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Suitable object . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Working Principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Structure and Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Structure and Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Product Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Appearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Operation Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Display Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Title Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Typical Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Typical Interface Icon Paraphrase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Unpacking and Checking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Install the Covetrus® ProMAX Syringe Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Use Preparation & Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Use Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Operation Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Basic Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Operation Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Infusion Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Equipment Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Starting and Self-test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Install Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Remove Air Bubble . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Change the Rate During Infusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Bolus Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Remove the Syringe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Power OFF or Standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Table of Contents
Set Infusion Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Introduction to Infusion Parameters Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Infusion Mode Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Rate Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Time Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Body Weight Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Drug Library Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Sequence Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
System Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Drug Library . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
KVO Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Bolus Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Commonly used syringe brand . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
DPS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Occlusion Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Pressure Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Pump Idle Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Finish Pre-Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Micro Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Reset Total Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Date & Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Sound . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Screen Lock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Night Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Battery Capacity Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Nurse Call (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Nurse Call Alarm Level (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Net Work (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Factory Default . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Serial Number (SN) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Software Version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Other Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Patient Information (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Prescription (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
History Entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Last Therapies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Anti-bolus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Alarm Prompt and Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Introduction to Alarm Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Multilevel Alarm Rules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Alarm Treatment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Fault Analysis and Solution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Table of Contents
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Cleaning, Disinfecting, and Sterilizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Periodical Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Check the Appearance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Performance Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Maintenance Plan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Normal Repair Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Maintenance for Long-Term Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Equipment Components/Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Production Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Recycling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Check the Battery Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Replacing the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
After-Sale Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Appendix Alarm and Solution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
1
Preface
Application Scope
Applicable to the Covetrus® ProMAX Syringe Pump.
This user manual describes the product's most complete configuration. For more detailed information,
please contact Covetrus.
Applicable Object
It applies to professionally trained veterinarians and maintenance technicians of this equipment.
Use Instructions
The Covetrus ProMax Syringe Pump user manual covers the basic information on the safety and
effectiveness of the product for guiding the operator to correctly install, test, operate, use, and maintain
the product. Please read this manual thoroughly before use.
Covetrus is responsible for the reliability and performance of the equipment only if all the following
conditions are met:
1. Use the equipment according to this user manual
2. The equipment can only be disassembled, assembled, replaced, tested, improved, and repaired by
Triumph Medical
3. All components and accessories, as well as consumables for repair, are provided by Covetrus
4. Relevant electric devices meet the international standard IEC/EN 60601-1 and this user manual
Paraphrase
Mechanical button
Touch button
( ) Information
- Inapplicable
Accordant
Operation steps
Bolus: Infuse large volume of liquid in a short time.
KVO: Keep vein open and prevent blood back to the syringe extension line.
Anti-bolus: Motor automatically reverses while the syringe extension line and needle have high pressure.
IrDA: Infrared Communication
Warning /Attention: can possibly cause physical injury or death if the cautions covered in the Warning
are not obeyed.
Caution: can possibly cause physical injury if the cautions are not obeyed.
Note: if user fails to follow the supplementary or prompt information on the operating instructions, it may
cause physical damage to the equipment if it is not obeyed.
Accessories: the optional components which are necessary and (or) suitable for use with the equipment
to achieve the expected purpose, provide convenience for attaining the expected purpose, improve the
expected purpose, or increase the additional functions of the equipment.
Description of Revision of User Manual
The Covetrus® ProMAX Syringe Pump User Manual will be revised subject to product improvement,
laws updating, or instructions improvement based on the preconditions of meeting related laws and
regulations. All revisions will be documented in the new version.
2
Safety Instructions
Safety Information and Warnings
Before using, please check the equipment, connecting wire, and accessories to ensure they can work
normally and safely. If there's anything abnormal, immediately stop working and contact Covetrus®.
Additionally, the adhesion or intrusion of fluid/drug may cause the equipment to fault and malfunction.
Therefore, please clean the equipment after use, and store it correctly.
It is not allowed to store or use the equipment in an environment with active chemical gas (including gas
for disinfecting) and a moist environment since it may influence the inside components of the Covetrus®
ProMAX Syringe Pump and may possibly cause performance drop or damage the inside components.
If the syringe extension line is twisted, the filter or needle is obstructed, or there is blood in the needle which
may obstruct the syringe, the pressure in the syringe will rise. Removing such occlusion may cause a "bolus
infusion" (temporary excess infusion) to the patient. The correct method is to tightly hold or clamp the
extension line near the puncturing position, then loosen the syringe, solve the reason for occlusion, and
restart the infusion. If the infusion is restarted when the occlusion reason exists, then it may cause a
persistent occlusion alarm. This will cause the pressure in the syringe to rise and may break or cut OFF the
connection or hurt the patient.
This equipment has an occlusion detection function for detecting and alarming when the syringe needle
deviates from the position in the vein or the needle is not correctly punctured in the vein. However, it only
alarms when the occlusion pressure has reached a certain numerical value. At this point, the puncturing
site may have become reddish, swelling, or bleeding. Additionally, it is possible that the device doesn't
alarm for an extended period if the actual occlusion pressure is lower than the alarm threshold value;
therefore, please periodically check the puncturing site. If there's any abnormal phenomenon for the
puncturing site, please take suitable measures, such as puncturing again.
Tightly fix this equipment on the infusion stand and ensure its stability of the infusion stand. Be careful
when moving the infusion stand and the equipment to avoid the equipment dropping and the infusion
stand falling or knocking the surrounding objects.
It is not allowed to place and use the equipment in an environment with anesthetic and other inflammable
or explosive articles to avoid fire or explosion.
The operator shall guarantee that the inputted infusion parameters of this equipment are the same as the
medical advice before starting the infusion.
Only those sterile hypodermic syringes for single use and other medical components that meet the local
laws and regulations and the requirements covered in this user manual can be adopted. It is suggested to
adopt the syringe with the same brand as defaulted in this equipment. Infusion accuracy is not ensured if
an unsuitable syringe is adopted. Please use the syringe and the extension line with luer locks, or the
extension line might pop off, causing injury to the patient.
To avoid the risk of electric shock, this equipment must only be connected to a supply main with protective
ground earth.
Disassembling or refitting this equipment or
using it for other purposes except normal
infusion is prohibited.
No one is allowed to repair this equipment
except Triumph Medical.
This equipment must be operated by trained
professional medical care personnel.
This equipment does not apply to blood
transfusion.
Please do not depend only on the alarm system during use; please periodically check it to avoid an
accident.
3
Cautions
Dialogue Windows
Dialogue window mainly includes operation select, operation confirm, and information.
Before its first use after purchase, or if the equipment is not used for an extended period, please charge the
equipment with an AC Power Supply. If not fully charged, the equipment may not work with the built-in
battery power supply under a power failure.
Under general conditions, please use an AC power supply as much as possible since it can prolong the
service life of the battery to a certain degree. When using an AC power supply, ensure that the grounding
wire is reliably connected to the ground, and only the AC power cord attached to this equipment shall be
used. Please pay attention to the plug position of the AC power cord to ensure that you can disconnect it at
any time. The built-in battery can only be used as an assistant power supply when the AC power supply
can't reliably connect with the ground and is not under normal conditions (power failure or in-transport
infusion).
Before connecting this equipment to the power supply, please keep the power socket and plug dry. The
power voltage and frequency must meet the requirements listed on the equipment label or in this user
manual.
The equipment is equipped with an audible and visual alarm system, the red and yellow alarm indicators will light on
by turn, and the speaker makes the "beep" sound.
Please use the fluid/drug after it has reached or nearly reached room temperature. When the fluid/drug is
used at a low temperature, the air which is dissolved in the fluid/drug may cause more air bubbles.
It is not allowed to press and operate the button with a sharp object (such as a pencil tip and nail). Otherwise, it may
possibly cause early damage to the button or surface lm.
Under the condition of low flow rate infusion, please pay special attention to an occlusion. The lower the
infusion flow rate, the longer the time of detecting occlusion, which may cause a long-time infusion stop
during this period.
If the equipment is damaged from dropping or impact, please stop using it immediately, and contact
Covetrus®. The equipment's inside components may be damaged even if the appearance is not damaged
and abnormality does not occur when working.
This equipment can't be used in locations with radiological installation, magnetic resonance equipment, or places with
high-pressure oxygen therapy.
Please keep the equipment away from the AC Power socket for a certain distance to avoid uid/drug splashing or
dropping in the socket. Otherwise, it may possibly cause a short circuit.
4
Symbols
Not all of the below symbols exist on the equipment you have purchased.
Lot Number
Attention, consult
accompanying documents
Non-Ionizing Radiation
Environment-friendly use period
(20 years)
Date of Manufacture
Defibrillation Proof
Type CF Applied Part
Serial Number
Unlock
Class I Equipment
Alternating Current
Input and output
Handle with care
Manufacturer
Please refer to the User Manual
Ingress Protection
(prevent solid objects larger than 2.5
mm in diameter and water intrusion
from splashing in all directions)
Lock
Marks Description Marks Description
IP34
SN
5
Overview
Application Scope
Expected Purpose
This product is used for constant-speed intravenous infusion in a hospital.
Expected Working Environment
Including but not limited to the operation room and surgery table in an animal hospital/or pet clinic.
Suitable object
Companion animals
Contraindications
No
Working Principle
It is controlled to move into a linear motion from a microcontroller-based system that drives a step motor,
allowing a wide range of pumping rates configured to the inside diameter of the loaded syringe. The
syringe plunger is driven from a lead screw and drive-nut mechanism, infusing the drugs into the patient.
Structure and Performance
Structure and Performance
The Covetrus® ProMAX Syringe Pump comprises the main unit and built-in battery. This equipment
provides several infusion modes, such as rate mode, time mode, body weight mode, drug library mode,
and sequence mode. It also has functions such as history records, drug library, Anti-bolus, alarm, etc.
Product Specification
Electric Protection Type Class I
Electric Protection Level Defibrillation proof type CF applied part
Ingress Protection IP34 (Prevent solid objects larger than 2.5 mm in diameter and water intrusion
from splashing in all directions)
Working mode Continuous
Classification Portable equipment, non-portable syringe pump
Safety Classification
Compatible Syringes 3 ml, 5 ml, 10 ml, 20 ml, 30 ml, 50/60 ml
System Accuracy ≥1ml/h, ±2% 1ml/h, ±5%
Infusion Rate Syringe size 3 ml: (0.1-100) ml/h
Syringe size 5 ml: (0.1-150) ml/h
Syringe size 10 ml: (0.1-300) ml/h
Syringe size 20 ml: (0.1-600) ml/h
Syringe size 30 ml: (0.1-900) ml/h
Syringe size 50 ml: (0.1-1500) ml/h
Specification Parameters
6
Bolus Rate Syringe size 3 ml: (0.1-100) ml/h
Syringe size 5 ml: (0.1-150) ml/h
Syringe size 10 ml: (0.1-300) ml/h
Syringe size 20 ml: (0.1-600) ml/h
Syringe size 30 ml: (0.1-900) ml/h
Syringe size 50 ml: (0.1-1500) ml/h
Bolus preset value Min: 0.1 ml Max: max rate of accordingly loaded syringe size
KVO Rate 0-5.00 ml/h
Micro Mode Setting Range Syringe size 3 ml: (100) ml/h
Syringe size 5 ml: (100-150) ml/h
Syringe size 10 ml: (100-300) ml/h
Syringe size 20 ml: (100-600) ml/h
Syringe size 30 ml: (100-900) ml/h
Syringe size 50 ml: (100-1500) ml/h
Minimum Flow Rate Increment 0.01 ml/hour
VTBI 0-9999.99 ml, minimum step is 0.01 ml
Total Volume Infused 0.01-9999.99 ml, minimum step is 0.01 ml
Time Range 1 minute - 99 hours 59 minutes
Fuse Type slow fuse 2A 250V
Dimensions 242.5 (w) x 111 (d) x 126.5 (h) mm
Weight 1.7 kg, 3.75 lbs
Specification Parameters (continued)
AC Power Supply 100-240V 50/60Hz
Input Power 50VA
DC Power Supply DC 12V
Battery Specifications Specification: 7.4V 2500mAh
Charging time: single battery less than 2.5 hours, two batteries less than 5 hours
(under OFF status)
Working time: single battery over 5.5 hours, two batteries over 11 hours
(after completely charging battery/batteries, when the environment temperature
is 25°C or 77° F and flow rate is 5 ml/h constantly)
Power Supply
7
Non-AP/APG type equipment Do not use it in an environment with inflammable anesthetic gas mixed with air,
and inflammable anesthetic gas mixed with oxygen or nitrous oxide
Operating 1. Temperature: 5–40°C, 41–104°F
2. Humidity: 15-95%, Non-Condensable
3. Atmospheric Pressure: 57-106kPa
Transport and Storage 1. Temperature: -20–55°C, -4–131°F
2. Humidity: 10-95%, Non-Condensable
3. Atmospheric Pressure: 50-106 kPa
Environment
Alarm Signal Sound Pressure Level When the sound is set at lowest level, alarm signal sound pressure level ≥50dB(A)
When the sound is set at highest level, alarm signal sound pressure level ≤80dB(A)
Alarm Information VTBI near end, Syringe near empty, VTBI infused, Syringe empty, Occlusion
(Pressure high), Battery nearly empty, Battery empty, no battery inserted,
No power supply, check syringe, Pump Idle Alarm, Standby time expired,
KVO finished
Alarm
Main Safety Standards IEC 60601-1 : 2005+A1 : 2012
Medical Electrical Equipment – Part 1 :
General Requirements for Basic Safety and Essential Performance
IEC60601-2-24 : 2012
Medical Electrical Equipment – Part 2-24 :
Requirements for the Safety of Syringe Pumps and Controllers
IEC60601-1-82006+A12012
Medical Electrical Equipment – Part 1-8 :
General Requirements for Basic Safety and Essential Performance –
Collateral Standard: General Requirements, Tests, and Guidance for Alarm
Systems in Medical Electrical Equipment and Medical Electrical Systems
EN60601-1-2 : 2007+AC : 2010
Medical Electrical Equipment – Part 1-2 :
General Requirements for Basic Safety and Essential Performance –
Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
Safety Standard
8
Appearance
Front View
Part Number Name
Handle Control the Covetrus® ProMAX Syringe Pump
push-pull sliding box and clip
Slider Box
Pressure Sensor Detect the pressure of the syringe
Syringe Clip Clamp the syringe plunger
Leadscrew
Syringe Fixture Lever Pull forward, turn 90° right or left,
and install the syringe into the slot.
Keep the extension line straight and neat
9
Operation Panel
Part Number Name
Touch Screen
Power Pump power switch, short pressing the power button
to enter the shutdown setting interface, the user can
set shutdown, standby (time), or cancel.
Long pressing the power button until the screen turns off.
AC Indicate Light When connected to the AC power supply,
the AC indicator light is on.
Alarm Indicator Light While pump alarms, the indicator light glitter, different
level, different frequencies and color, for more
information, please refer to page 21
Running Lights
Start/Stop
Bolus/Purge
Home Enter system home page
10
Display Screen
The display screen interface layout composes of a title bar and typical interface
Title Bar
The title bar displays real-time information and is not touchable, and the left upper corner displays the
name of the current editing parameter.
Typical Interface
During pre-infusion and infusion, the standard interface will display the following: main interface,
working interface, alarm interface, prompt interface, control panel, parameters setting, input method,
standby interface, etc.
Syringe Apparatus Syringe apparatus indication icon
Indication Icon
Lock Screen Unlock state icon is
Indication Icon
WIFI Indicates WIFI connection state
Indication Icon
Battery Charging Display the current battery charging state
Indication Icon
Battery Status Battery life percentage is displayed on the left side of the icon
Indication Icon Battery icon may show the following states:
Icon Paraphrase Description
11
Typical Interface Icon Paraphrase
Input Method Interface
The input method interface comprises the title bar, input box, and editing box.
Title bar: displays the name of the current editing parameter.
Input box: real-time display of the input content.
Editing box: It composes of the main button area and function button area.
The main button area comprises the numerical value, letters, and icons. Click it continuously to change
the sequence.
The function button area composes of a clear button, backspace button, , , and Shift.
Start Click this icon to start infusion
Stop Click this icon to stop infusion
Bolus/ 1. During infusion, it is aBolusfunction–click it to start fast infusion
Purge 2. Before infusion starts, it is aPurgefunction–click it to exhaust air
from the syringe
Home Click this icon, return to the main interface
Icon Paraphrase Description
Backspace Button Click it to backspace and delete
Shift Button Click it to switch the capital and lowercase English letters
Cancel Button Click it to cancel editing and exit
OK Button Click it to save editing and exit
Icon Paraphrase Description
12
Rear View
Part Number Name
USB Port Port only for software upgrade
DC Input Port External 12V DC Power Supply
Handle
A/C Adapter Port External 100-240V 50/60Hz AC Power Supply
Pole Clamp Using for fixing the equipment on the infusion stand
Loudspeaker
IrDA Using for communicating with infusion docking station
(Optional)
Latch for stackable function
Slider Box
13
Installation
Unpacking and Checking
Please check the appearance before unpacking; if broken, please contact Covetrus®.
Please carefully open the package to avoid damaging the equipment and relevant accessories.
After unpacking, please check the objects according to the packaging list. If there are insufficient or
damaged accessories, please contact Covetrus.
Please keep relevant accessories, warranty card, and User Manual.
Please keep the packing case and packing materials for future transportation or storage.
Installation
Install the Covetrus® ProMAX Syringe Pump
Rotate the Pole Clamp screw (knob) and unscrew it to leave space
Lock the Pole Clamp on the infusion stand, adjust the position of the Covetrus® ProMAX Syringe
Pump, and tighten the Pole Clamp to fix the ProMAX Syringe Pump on the infusion stand (as shown
in the drawing below). Hold the ProMAX Syringe Pump when tightening the fixing clamp; carefully
let go of it after tightening to avoid it from falling
The Pole Clamp supports the vertical pole at the default state. To adjust the Pole Clamp direction,
please remove the bolt from the Pole Clamp, take out the Pole Clamp and, adjust the direction, then
tighten the bolt
Please put the packing materials out of reach of children. Please obey local laws and regulations when
disposing of items.
The Covetrus® ProMAX Syringe Pump shall be installed per the instructions of this user manual
All devices that connect with this equipment must pass the designated IEC standards (for example, IEC60950
information technology equipment safety and IEC60601-1 medical electric device safety) certification, and all
devices must be connected according to the correct version of the IEC60601-1-1 system. The technician who
connects to additional devices with the equipment interface is responsible for meeting the IEC60601-1-1 standard.
Please contac tCovetrus® for any questions.
When connecting this equipment with other electric devices to form a combination for special
circumstances, if the combination can't be confirmed dangerous or not, please contact Covetrus or an
electrical expert of the hospital to ensure the necessary safety of all devices in the combination won't
be destroyed
This equipment must be used and stored in an environment aligned guidelines from this manual
14
Use Preparation & Cautions
Use Preparation
If the Covetrus® ProMAX Syringe Pump is new, reused after storing for an extended period, or reused after
repair, please check and ensure the following before use:
The equipment's appearance is clean and in good condition without cracks and leakage
The moving components are smooth and working. The buttons are working
The touch screen can be operated smoothly and effectively
The power cord is installed tightly and won't be easily damaged when pulling
Set and check the system time to ensure the history records will be correctly recorded
In case only the built-in battery is adopted for supplying power, please charge it fully before using,
and check the battery state frequently
Carefully read the Warnings, Cautions, and Operation Steps in this User Manual
Operation Cautions
Avoid direct sunlight, high temperature, and high humidity
The equipment shall be put at a position less than 0.65 m (both up and down) to the heart of the patient
The parameters can only be set or changed by trained and professional personnel
Avoid using faulty equipment to avoid medical negligence, which may hurt the health and even the life
of the patient
Infusion accuracy or abnormal work of the equipment may be compromised if the working environment
temperature exceeds the designated range
The viscosity and specific gravity of infusion fluid will influence infusion accuracy
15
Basic Operation
Operation Flow
Mount the Covetrus® ProMAX Syringe Pump on the IV stand: refer to page 13
Power on: press two seconds, power on equipment; refer to below
Install syringe: refer to below
Confirm syringe brand and size: Select syringe brand
Remove an air bubble in the line: refer to page 16
Select infusion mode: Select infusion modes according to requirement
Set infusion parameters: set infusion parameters according to requirement
Connect infusion line with the animal
Start infusion: press , start infusion
Finish infusion: refer to page 17
Remove syringe: refer to page 17
Power off or Standby: refer to page 17
Infusion Operation
Equipment Installation
Mount the device on the infusion stand according to page 13, connect with AC power supply, and
check the AC indicator lights. The battery will start to charge once AC power is connected.
Starting and Self-test
Press two seconds, then power on equipment
After power is on, the system will automatically check the motor, sensor, battery, memorizer,
CPU communication, and alarm indicator.
After passing the self-test, Covetrus® ProMAX Syringe Pump enters into rate mode interface.
Install Syringe
Hold the clutch (Handle) and pull the slider (arm) to the right side
Pull the Syringe Fixture Lever (barrel clamp), and turn 90° right or left
Insert the syringe flange into the slot, turn the Syringe Fixture Lever 90° spring back to tighten the
syringe, and confirm the correct syringe brand/size
Hold the clutch (Handle) and push leftward—release after it touches the plunger firmly
Put the extension line of the syringe into the extension line hook
• ClickSetting Commonly used Syringe brandt o choose syringe brand if not selected
If the self-test fails, the Covetrus ProMAX Syringe Pump cannot be operated properly or is damaged, it
cannot be used for animal infusion; please contact Covetrus®.
Only use syringe brands defaulted in this syringe pump
Ensure syringe brand and size on the display screen are in accordant with the one in use
Contact Covetrus® if you are unsure of the correct syringe Manufacturer and size–selecting the
wrong Manufacturer or size can compromise accuracy and possibly cause injury to the animal.
Check to ensure no air bubbles are in the syringe
Make sure the syringe is correctly installed. Otherwise, accuracy will not be assured and may harm
the animal
16
Set Infusion Parameters
Refer to page 18
Remove Air bubble
Under the parameters setting interface, Press theBolusbutton and hold, or touch the purge icon
on the product panel, to eliminate the air bubbles in the line.
The purge total volume is not calculated in the Total Volume Infused.
Start Infusion
Connect syringe extension line to the patient, confirm infusion parameters, PressStartbutton ,
click yes in the pop-up prompt interface, and start infusion.
Change the Rate During Infusion
Under the running interface, click the rate number in the Touch Screen orStop ,
enter parameters setting interface, and reset the target infusion rate.
Note: Only rate mode, time mode, and body weight mode support online rate modification
function during infusion without pressingStop .
Bolus Application
During operation, the bolus functions have two operation modes: Manual bolus and Automatic bolus.
Manual bolus: press and hold the Bolus button on the product panel, and the pump will work
at the max flow rate of the current syringe size or set the max bolus rate under the setting interface.
(For syringe flow rate range, please check in the setting interface and refer to page 5, release the
button, and Covetrus® ProMAX Syringe Pump will return to the previous infusion rate
Automatic bolus: Under the running interface, click the Bolus on the Touch Screen, set two
parameters among bolus infusion volume, rate, and time, and click Start ; it will beep at every 1 ml
infused. After the bolus infusion is finished, the equipment will return to the previous infusion rate
Before purging the air, double-check to confirm the infusion line is not connected with the animal
Purge rate is the max rate of the syringe size. The purge will automatically stop when the volume is
≥ 5 ml
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34

Promax 7350370080540 Syringe Pump User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI