Page 7
Chapter 1 PhD
2
Description
1.1 PhD
2
capabilities
The PhD
2
is a gas detector with numerous features that
can be modified to meet user requirements. This chapter
discusses some of those features.
1.1.1 Methods of sampling
The PhD
2
may be used as either a "Diffusion" or
"Sample-Draw" type monitoring device.
In normal operation, the PhD
2
detector is worn on the
belt, used with its shoulder strap, or held by hand. Once
turned on, the PhD
2
monitors continuously. The
atmosphere being measured gets to the sensors by
diffusing through vents in the sensor compartment cover.
Normal air movements are enough to carry the sample to
the sensors. The sensors react immediately to changes
in the concentrations of the gases being measured. This
type of "diffusion" operation monitors only the
atmosphere that immediately surrounds the detector.
It is possible to use the PhD
2
to sample remote locations
by using a sample-draw kit. Two sample-drawing kits are
available. In each case the sample is drawn in through a
probe assembly, and sucked through a length of hose
back to the instrument. One type of kit uses a hand-
operated squeeze-bulb to draw the sample through the
hose, the other uses a battery-operated continuous
mechanical pump. A hand-aspirated sample-draw kit is
included as an accessory with every PhD
2
. Use of the
sample draw kits is covered in section 2.4.
1.1.2 Multi-sensor capability
The PhD
2
can be configured to simultaneously monitor
one, two, three, or four gases. Sensors can be added,
deleted, changed, or replaced in the field. The PhD
2
microprocessor circuitry eliminates the need for manual
switch setting and other laborious set-up procedures. (It
is still necessary to verify the accuracy of the PhD
2
by calibration with known concentration test gas
whenever a change is made to the sensors installed
in the instrument.)
The PhD
2
design uses highly specific, electrochemical
toxic sensors that have been designed to minimize the
effects of common interfering gases. These sensors
provide accurate, dependable readings for many common
toxic gases. Currently available toxic sensors for use in
the PhD
2
include hydrogen sulfide (H2S), carbon
monoxide (CO), ammonia (NH3), chlorine (Cl2), sulfur
dioxide (SO2), nitric oxide (NO), nitrogen dioxide (NO2),
hydrogen chloride (HCl), and hydrogen cyanide (HCN).
In addition to sensors designed to measure specific toxic
hazards, Biosystems also offers the TOX 1 broad range
electrochemical sensor for situations requiring use of a
single sensor to monitor for multiple toxic hazards.
Different measurement units are used depending on the
gas being measured. Common examples include:
Concentration of ... Is indicated in
Oxygen (O
2
) Percentage of air by volume
Combustible gas Percentage of the lower
explosive limit (LEL)
Carbon Monoxide (CO) Parts per million in air
Hydrogen Sulfide (H
2
S) Parts per million in air
Table 1
Sensor configuration procedures are discussed in
greater detail in section 5.3.
1.1.3 Calibration
The PhD
2
detector has been designed for easy
calibration without the use of manually adjusted
calibration pots.
The accuracy of the PhD2 should
be checked periodically with known concentration
calibration gas. Failure to check accuracy can lead
to inaccurate and potentially dangerous readings.
Calibration is a two step procedure. In the first step the
PhD
2
is taken to an area where the atmosphere is fresh
and a "zero" adjustment is made automatically on-
demand by pressing a button.
The second step of the calibration procedure is the
sensor response or "span" calibration. In this step the
accuracy of the PhD
2
sensors is established by exposing
them to known concentration calibration gas and noting
the response. If there is a deviation from the expected
response the instrument may be adjusted immediately by
pressing buttons on the instrument key pad, or the
instrument may be taken out of service and returned to
another location for adjustment.
Use of these push-buttons in span adjustment
procedures is reserved for authorized personnel.
Calibration procedures are discussed in detail in
Chapter 3.
1.1.4 Alarm logic
PhD
2
gas alarms are user adjustable and may be set
anywhere within the range of the sensor channel. When
an alarm set point is exceeded a loud (100 dBA at six
inches) audible alarm sounds, and an individual bright
red LED alarm light for each affected sensor flashes.
PhD
2
gas alarms are normally of the self-resetting type.
When readings drop back below the pre-set alarm levels,
the instrument reverts back to normal operation, and the
visual and audible alarms cease.
It is possible, if desired, to set PhD
2
gas alarms so that
they "latch." In the latched condition, once an alarm
occurs both visual and audible alarms continue to sound
even after the atmospheric hazard has cleared. The