  • Hello! I am an AI chatbot trained to assist you with the RKI Instruments GX-2009 Reference guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
1. Turning the GX-2009 On
a.PressandbrieyholddownthePOWERMODEbutton.The
backlightwillturnonandactivateallLCDsegments.Release
thebuttonwhenyouhearabeep.
i.IfCalC-LIMITisdisplayed,unitispastdueforcalibration.Pressand
releasethePOWERbuttontoconrmandcontinueuse.
ii. IfFAILC-LIMITisdisplayed,unitispastdueforcalibrationandcannotbe
usedwithoutcalibrating.
b.Theinstrumentwillgothroughabriefwarmupshowingdate,
numberofdaystonextcalibration,timeofday,batterypower,
alarmlatchsetting,fullscalevaluesforallchannels,warning
alarmsettingandhighalarmsettingforallchannels,STEL
alarmsetpointforH2SandCOandTWAalarmsetpointfor
H2SandCOchannels.
c.Ifthebatterylevelisloworthebatteryiconisashingandthe
unit’sbuzzersounds,rechargetheGX-2009beforeuse.
2. Measuring Mode Screen
a.Afterwarm-uptheGX-2009willdisplaythecurrentO2,H2S,
COand%LELgasreadings.
b.Ifthereadingsarenotdisplayingfreshairvalues,itis
necessarytoperformaFreshAiradjustment.
3. Performing a Fresh Air Adjustment
a.Findafresh-airenvironmentfreeoftoxicorcombustible
gasesandofnormaloxygencontent(20.9%).
b.PressandholdtheAIRbuttonforaboutthreeseconds.The
displaywillindicateAirCALHOLDAIR.Continuetoholduntil
ADJandRELEASEaredisplayed.ReleasetheAIRbutton.
TheGX-2009willsettofreshairvalues,0%LELCH4,20.9%
O2,0.0ppmH2Sand0ppmCO.
4. Performing a Breath Test
a.PerformingabreathtestquicklyteststheGX-2009oxygen
sensor,audible,visualandvibratoryalarmsforproper
operation.Performingabreathtestisnotasubstitutefor
calibrationorperformingabumptestonallsensors.
b.CuphandsoverO2sensoropeningandgentlyexhaleover
thesensor.
c.Theoxygensensoroutputwilldropbelowthealarmpointof
19.5%andactivatealarms.
d.Ifalarmsaresettolatching,pressthePOWERMODEbutton
toresetalarmsonceoxygenreadingisabove19.5%.
i. Note:YoumayalsoseeslightresponseontheCOgassensorwhen
performingabreathtest.
Quick Reference Guide For Model GX-2009
Turn on and adjust the GX-2009 gas monitor in a known fresh air area.
Quick Reference Guide For Model GX-2009
PN71-0160RKRevP1
RKI Instruments, Inc. 33248 Central Ave, Union City, CA 94587
Phone(510)441-5656•(800)754-5165•www.rkiinstruments.com
5. Sensor Failure
a.IftheGX-2009experiencesasensorfailureduringstartup,the
LCDwillindicateFAILandindicatethesensorthathasfailed.
Example:FAILCOppmSENSOR.Tocontinueoperation
pressandreleasethePOWERMODEbuttontoacknowledge
thefailure.Gasreadingsforthatsensorwillbereplacedby
dashes.
b.Itisnecessarytoreplacesensorandcalibrateassoonas
possible.
6. Display Mode
a.PressandreleasethePOWERMODEbuttontoaccessPEAK
readings.
i. ThiswilldisplaythehighestLEL,COandH2SandlowestO2readings
encounteredwhiletheinstrumenthasbeeninoperation.
ii. ToResetPEAKreadings,pressandholdtheAIRbuttonuntilCLEAR
RELEASEisdisplayed.
b.PressthePOWERMODEbuttonagaintomovetotheSTEL
screen.
i. STELisa15minuteaverageexposureforCOandH2S.
c.PressthePOWERMODEbuttonagaintomovetotheTWA
screen
i. TWAisan8houraverageexposureforCOandH2S.
d.PressthePOWERMODEbuttonagaintodisplaythefullscale
readingsofeachchannel.
e.PressingthePOWERMODEbuttonagainwillreturnyoutothe
normalmeasuringscreen.
7. Resetting Alarms
a.IftheGX-2009alarmsaresetuptolatch,pressandrelease
thePOWERMODEbuttontoresetalarmsoncealarm
conditionhascleared.
b.IftheGX-2009alarmsaresetupforself-resetting,thealarms
willautomaticallyresetoncethealarmconditionhascleared.
8. Turning off the GX-2009
a.PressandholdthePOWERMODEbutton.
b.TURNOFFwillbedisplayedandtheinstrumentwillbeepthree
timesbeforeshuttingdown.
9. Consult the operator’s manual for further
information.
081003/1000
/