CONTROL SOLUTIONS VFC400 User guide

Type
User guide

The CONTROL SOLUTIONS VFC400 is a precise and reliable data logger designed to monitor and record temperature data in various applications. It features a user-friendly interface with buttons for easy operation. The VFC400 allows you to review minimum and maximum temperatures, place inspection marks in the log, and start, stop, or clear recordings. It also has a start/clear/stop button to initiate or stop data logging and exit day summaries. With its ability to store a significant amount of data, the VFC400 ensures you have ample記錄 for analysis and reporting.

The CONTROL SOLUTIONS VFC400 is a precise and reliable data logger designed to monitor and record temperature data in various applications. It features a user-friendly interface with buttons for easy operation. The VFC400 allows you to review minimum and maximum temperatures, place inspection marks in the log, and start, stop, or clear recordings. It also has a start/clear/stop button to initiate or stop data logging and exit day summaries. With its ability to store a significant amount of data, the VFC400 ensures you have ample記錄 for analysis and reporting.

VFC400 Download of Data
Downloading data from your the VFC400
Control Solutions, Inc.
35851 Industrial Way, Suite D
St. Helens, OR 97051
503-410-5996
www.vfcdataloggers.com
Downloading data
Buttons Actions
Overview Review/Mark button
Press to enter REVIEW mode and then press
again to advance through the min/max
temperatures. This action also places an
inspection mark in the log if RECORDING is
active. This validates your twice daily
temperature logs.
Start/Clear/Stop button
Press to start recording
Press to Stop recording
Press to exit day summary
Downloading data
The first thing we want to do
before we download data from
the VFC400 is stop the data logger
from logging temperatures.
Downloading data
Stopping the data logger
When it is time to download the data from the data logger
we suggest stopping the logger. To accomplish this you
hold down the Start/Clear/Stop button for 1-2 seconds.
The word STOPPED will appear on the screen as shown on
the top left picture. The REC symbol will disappear after 1-
2 seconds and you take your finger off the button quickly.
The logger will stop and the screen will only say STOPPED
and show the number of days collected.
Downlaoding data
At your PC
open up the
Control
Solutions
VTMC
software by
double clicking
on the Icon.
VFC400 Data Logger Configuration
Plug your Docking Station into the USB port of your PC
Insert data logger into the Docking Station firmly
Docking Station
Note: Make sure to download
your data from your data
logger prior to configuring/re-
configuring as this will erase
any and all data stored on your
data logger.
Downloading data
After you insert the logger into the docking station you will see the
following message:
A few seconds later the chart/data will pop-up on the screen and then the
following message:
Click OK in the box and
remove the data logger
from the docking station.
Downloading data
Starting The Logger
The logger is now ready to start.
In Ready mode the recorder displays the time
and the READY icon.
To Start: Press and hold the Start/Clear/Stop
button. First, STARTING is displayed in addition to
READY. Then READY disappears.
Recorder in READY mode
Downlaoding data
Starting The Logger
Once READY disappears, immediately release the button.
STARTING will also disappear, and the symbol RECwill be
shown. The VFC 400 is now recording temperature data.
You can see that there are no alarms because there is a
mark in the upper left corner and the audible alarm is not
beeping.
Thank you for your business
Control Solutions, Inc.
www.vfcdataloggers.com
503-410-5996
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

CONTROL SOLUTIONS VFC400 User guide

Type
User guide

The CONTROL SOLUTIONS VFC400 is a precise and reliable data logger designed to monitor and record temperature data in various applications. It features a user-friendly interface with buttons for easy operation. The VFC400 allows you to review minimum and maximum temperatures, place inspection marks in the log, and start, stop, or clear recordings. It also has a start/clear/stop button to initiate or stop data logging and exit day summaries. With its ability to store a significant amount of data, the VFC400 ensures you have ample記錄 for analysis and reporting.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI