Emerson 00809-0200-4377, 752 Remote Indicator-Option Y0040, Rosemount 752 User manual

  • Hello! I am an AI chatbot trained to assist you with the Emerson 00809-0200-4377 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Manual Supplement
00809-0200-4377, Rev AA
December 2004
Rosemount 752 Remote Indicator
www.rosemount.com
Option Y0040 for the
Rosemount 752 Remote Indicator
SPECIAL CONSIDERATIONS FOR THE ROSEMOUNT 752
ORDERED WITH THE Y0040 OPTION
Important
Rosemount Inc. has verified that the Rosemount 752 Remote Indicator
ordered with a LCD meter will not be damaged and will remain intact at
temperatures down to –60 °F (–51 °C). However, the LCD meter will not
operate below the specified operating temperature of -4 °F (-20 °C).
Any hazardous location approvals ordered with this transmitter are void if the transmitter is
operating outside of the ambient temperature limits specified for that certification.
Approved ambient temperature limits are specified in the following locations:
Enclosure approval label
Hazardous locations installation drawings
“Approvals” appendix of the reference manual
Emerson Process Management
© 2004 Rosemount Inc. All rights reserved.
Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc.
All other marks are the property of their respective owners.
¢00809-0200-4378¨¤
Rosemount Inc.
8200 Market Boulevard
Chanhassen, MN 55317 USA
T (U.S.) 1-800-999-9307
T (International) (952) 906-8888
F (952) 949-7001
www.rosemount.com
Manual Supplement
00809-0200-4377, Rev AA
December 2004
Rosemount 752 Remote Indicator
/