Enraf-Nonius SmarTRAC User manual

Type
User manual
Dynatron SmarTRAC
User Manual
Copyright:
Dynatronics Corporation
7030 Park Centre Drive
Salt Lake City,
UT 84121.U.S.A.
Part number: 1671758-41
Date: April 11, 2016
3
TABLE OF CONTENTS
1 FOREWORD ......................................................................................................................................... 4
1.1 Intended Use ...................................................................................................................................... 4
1.2 Intended User/Operator ..................................................................................................................... 4
1.3 Essential Performance ....................................................................................................................... 4
2 DEVICE DESCRIPTION ........................................................................................................................ 5
2.1 Principles of Operation ....................................................................................................................... 5
2.2 Device Design .................................................................................................................................... 5
2.3 Touch Screen Control ........................................................................................................................ 6
2.4 Safety Features .................................................................................................................................. 6
2.5 Traction therapy programs ................................................................................................................. 7
2.6 Lumbar Traction ................................................................................................................................. 8
2.7 Cervical Traction (lying down)............................................................................................................ 9
3 PRECAUTIONARY INSTRUCTIONS ................................................................................................. 10
4 INDICATIONS ..................................................................................................................................... 13
4.1 Contra-indications ............................................................................................................................ 13
5 PACKAGE CONTENTS ...................................................................................................................... 14
6 INSTALLATION .................................................................................................................................. 15
6.1 Connection to power supply ............................................................................................................ 15
6.2 Disconnection from power supply .................................................................................................... 15
6.3 Attachment carabiner hook .............................................................................................................. 15
6.4 Mounting on the traction frame ........................................................................................................ 15
7 APPLICATION INFORMATION .......................................................................................................... 16
7.1 Operating sequence example .......................................................................................................... 16
7.2 Cord guidance .................................................................................................................................. 16
8 OPERATING INSTRUCTIONS ........................................................................................................... 17
8.1 Operator controls ............................................................................................................................. 17
8.2 Basic Operation ............................................................................................................................... 17
9 MAINTENANCE AND TROUBLESHOOTING ................................................................................... 25
9.1 Cleaning and Disinfection ................................................................................................................ 25
9.2 Warning Messages, Error Messages and Troubleshooting ............................................................. 25
9.3 Technical Maintenance .................................................................................................................... 26
9.4 End of life ......................................................................................................................................... 26
10 SPECIFICATIONS .............................................................................................................................. 27
10.1 Technical data .................................................................................................................................. 27
10.2 EMC details ...................................................................................................................................... 28
11 ORDERING INFORMATION ............................................................................................................... 30
12 PRODUCT LIABILITY ......................................................................................................................... 32
4
1 Foreword
This manual has been written for the owners and operators of the SmarTRAC. It contains general
instructions on operation, precautionary practices, maintenance and parts information. In order to
maximize the use, efficiency and lifespan of your unit, please read this manual thoroughly and become
familiar with the controls as well as the accessories before operating the unit.
1.1 Intended Use
The SmarTRAC Traction device, with its accessories, is intended to provide relief for the management of
pain and symptoms from a variety of pressures on muscular or skeletal structures. The SmarTRAC
Traction device may be used to treat pain and symptoms associated with the following conditions:
herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet
syndrome, acute facet problems, radicular pain, prolapsed discs, spinal root impingement, hypomobility,
degenerative joint disease, facet syndrome, compression fractures, joint pain, and discogenic pain.
1.2 Intended User/Operator
CAUTION: Federal law restricts this device to sale by or on the order of a healthcare provider.
The SmarTRAC is intended to be used, and shall only be used, by or under the supervision of
professional users in the field of physical medicine and rehabilitation.
1.3 Essential Performance
The essential performance of the device is to apply rhythmic, or according to a pattern, different dosed
forces on the spine or limbs of a patient, limited to 900N lumbar and 200 N cervical, without unacceptable
risk.
5
2 Device Description
2.1 Principles of Operation
The SmarTRAC traction device applies forces to the spine to alleviate pain caused by a variety of
pressures on muscular or skeletal structures. It exerts defined forces on the human body to alleviate pain.
It utilizes computer technology for programming of different kinds of traction. When a static force is
applied, the patient’s muscles relax. By increasing the force after relaxation of the muscles, the treatment
can be more effective.
2.2 Device Design
The SmarTRAC is a combination of computer technology, functional design and operational convenience.
The device is intended for use by professional users only, such as qualified personnel in physiotherapy,
rehabilitation and adjacent areas. The device is ideal for lumbar and cervical traction. Sequential
programming is possible to program combinations of traction which can be stored as favorites for later
use.
The device is designed in a plastic housing with a full color touch screen interface on which the
professional user can enter various parameters by a single touch. Parameters for traction include traction
force, base force, base hold time, treatment time, speed and others, depending on the selected traction
mode. Different settings can be configured, like language, brightness, end of treatment sounds, force unit
(N, Kg, lbs), and bodyweight unit (Kg, lbs).
The professional user must operate the device by selecting one or a combination of force protocols
categorized in 5 groups (Continuous, Intermittent, Cyclic, Progressive and Regressive), for the treatment
of the patient.
The device motor delivers traction to the patient via the cord and connected accessories at a force
between 15 and 900 Newton. There are Enraf-Nonius accessories suitable for lumbar or cervical traction
use with the SmarTRAC. The maximum protocol time is 150 minutes. If more than 150 minutes are set,
the device will not start, as a safety measure.
Several safety measures are running and controlling the device during operation. The patient stop is one
of these measures. It needs to be connected for the SmarTRAC to function. When the user presses the
patient stop, or if it becomes disconnected, the treatment will end and the force will be lowered by means
of the hardware design.
An optional remote can be used by the patient to decrease actual force during treatment.
The SmarTRAC is used in conjunction with a suitable treatment table. The treatment table is not part of
the SmarTRAC and must be purchased separately.
The original SmarTRAC was developed before 1982 and has proven to be a reliable therapeutic device
over the years.
6
2.3 Touch Screen Control
By means of a touch-screen interface, the device is easy to operate. A single touch can select four main
functions from the main touch screen:
Configuration
Clinical Protocols
Favorites (pre-programmed protocols)
Manual Operation
English is the default language. Other languages can be selected through the touch screen. Also,
brightness can be adjusted, an auditory tone signaling the end of treatment can be selected, and other
configurations are possible.
With one touch on the main screen “Clinical Protocol” can be selected. A new screen comes up, from
which different protocols can be directly selected. If necessary, additional information will be requested
before the selection can be made. With one touch “Favorites” can be selected from the main screen. It
will show the list of earlier stored manual protocols. When no “Clinical Protocol” or “Favorite” is requested,
a single click on “Manual Operation” from the main screen will start a wizard which leads the user through
all configurations that must be set up. This wizard is easy and fast to follow, and it does not skip any
necessary setting.
The SmarTRAC can be mounted on a treatment table where the patient lies down on their back. Popular
traction therapy programs can be selected and pre-programmed protocols can be used. Manually
programmed therapy programs can also be stored. Up to five types of therapy can be sequentially
programmed. The five types are as follows:
Continuous
Intermittent
Cyclic
Progressive
Regressive
Parameters can be adjusted during treatment. The device motor delivers traction force between 15
Newton and 900 Newton to the patient via the cord and other accessories. There are accessories
specifically designed for the SmarTRAC for use in lumbar or cervical traction.
2.4 Safety Features
To protect the patient from potential hazards, multiple mitigations are designed into the hardware and
software. There are electronic safeguards of force and mechanical safety for maximum force limitation.
Additional hardware mitigations are applied to protect the patient from a too high level of force. If no
software is running, or if no patient stop is connected, the hardware is designed to lower the force. There
is also a Patient Stop, which must be held by the patient during therapy. The patient can press it to stop
therapy immediately. If it is pressed, the force will lower to release the patient from the device. The
Patient Stop is controlled in the hardware, and any software will be overruled if it is pressed. A remote
control can be used as an option for decreasing the force delivered to the patient during treatment. As a
safety feature, the remote cannot increase the force.
7
2.5 Traction therapy programs
The therapy programs can be selected by the buttons corresponding with the graph. The buttons are
shown below the graph. See paragraph 8.2.1 for explanation about all the buttons.
Single Continuous traction
The force is increased to the Traction Force with the configured Speed and stays
there for the Traction Force Hold Time.
Depending on the configured Speed and the Traction Force the total treatment time is
calculated and shown in the treatment screen.
Single Intermittent traction
The force is increased to the Traction Force with the configured Speed and stays
there for the Traction Force Hold Time. Then it will lower the force towards the Base
Force with the same configure Speed and stays there for the Base Force Hold Time.
This sequence is Repeated as configured.
Depending on the configured Speed, the Traction Forces and number of repeats the
total treatment time is calculated and shown in the treatment screen.
Single Cyclic traction
The force is increased to the Traction Force with the configured Speed. Then it will be
lowered to the Base Force with the configured Ramp Down Time. When the Base
Force is reached, the force is increased again to the Traction Force, this time with the
configured Ramp Up Time. This sequence is Repeated as configured. After the last
repeat the force is lowered towards zero with the configured Speed again.
With the configured Speed, the Traction Forces , Ramp Up Time, Ramp Down Time
and number of repeats the total treatment time is calculated and shown in the
treatment screen.
Single Progressive Traction
The force is increased in a configured number of Steps. Each Step increases the
force partly, in such a way that all Steps together result in the programmed Traction
Force. Between each Step the force stays there for the Steps Hold Time. The force
between Steps is increased with the configured Speed. When the last Steps Hold
Time of the last Step has ended, the force is lowered towards zero (if no other traction
type is added).
With the configured Speed, the Traction Force and number of Steps, the total
treatment time is calculated and shown in the treatment screen.
Single Regressive Traction
The force is increased towards the Traction Force and then lowered in a configured
number of Steps. Each Step decreases the force partly, in such a way that all Steps
together result in a zero force value. Between each Step the force stays there for the
Steps Hold Time. The force between Steps is decreased with the configured Speed.
When the last Steps Hold Time of the last Step has ended, the force is zero.
With the configured Speed, the Traction Force and number of Steps the total
treatment time is calculated and shown in the treatment screen.
Note: Single Progressive Traction and Single Regressive Traction are not often used as single therapy
programs and mostly in combination with other traction types to start and finish a combined traction
treatment.
Note: Starting a combined traction protocol with regressive traction is not possible. It only lowers a
previous programmed Traction Force towards zero in a configured number of Steps or towards the
Traction Force value programmed in the following treatment type.
8
2.6 Lumbar Traction
Lumbar traction is used to apply traction on the back of a patient. The force limits are 15N to 900N. The
patient lies on a treatment table, such as the Ultra Traction Table, and is connected by hip and shoulder
belts (3445022 and 3445021). The patient lies down on the table and places legs on a flexi stool
(3013014) or a similar accessory to keep the legs at a 90 degree angle. The shoulder belt is connected to
the table and the hip belt is connected to the rope of the SmarTRAC. The Ultra Traction Table reduces
the friction of the patient’s hip to a minimum, so the force applied by the SmarTRAC is the force applied to
the back of the patient.
Patient connected for lumbar traction.
Layout of belts before patient is connected.
The hip belt is connected to the SmarTRAC
Shoulder belt is connected to treatment table
Optional accessories and table can be used with lumbar traction:
The SmarTRAC performs lumbar traction in conjunction with a suitable treatment table. A suitable
treatment table is one which is flat, provides a stable surface and is comfortable enough for the patient to
lie upon for duration of treatment. The hip and knees must be kept in the preferred angle of 90 degrees.
This can be done with a flexi stool which can be adjusted separately for each leg or with a similar
accessory to adjust the legs in the preferred angle.
Any suitable treatment table may be used. Recommended is the Ultra Traction Table which can be
ordered separately.
The patient’s legs must be supported to 90 degrees during lumbar traction treatment. The following
accessory is optional to provide leg support during lumbar traction:
- 3013014 Flexi stool grey, with separate height adjustment for each leg
The optional shoulder and hip belts are specified as articles:
- 3445022 Pelvic belt for lumbar traction
- 3445021 Thoracic belt for lumbar traction
These articles can be ordered separately and mentioned in this user manual chapter 11.
Document 0167336 explains how to mount the SmarTRAC onto the Ultra Traction Table. This document
is located on the CD-ROM.
9
2.7 Cervical Traction (lying down)
Cervical traction forces over 200 Newton are not allowed. By selecting a traction force higher than 200
Newton, a warning will be issued, exceeding this limit.
To ensure safe use, do NOT accept the limit warning when applying cervical traction to prevent
traction forces over 200 Newton for cervical traction.
A patient lies on a suitable treatment table, such as the Ultra Traction Table with a mounted traction
frame. On this frame an SmarTRAC is mounted and the cord of the SmarTRAC is connected to the
Saunders cervical traction unit. For an operating sequence example please refer to chapter 7.1.
With the Saunders cervical traction unit attached to your SmarTRAC it is possible to perform cervical
traction in the supine position. The patient can most easily relax in the supine position, which is very
important in traction therapy. This clinical unit pulls at the base of the occiput. Traction forces are directed
to the back of the head in this way, preventing compression of the Temporomandibular joint. It ensures
very effective cervical traction by eliminating friction between the patient and the treatment table. With the
SmarTRAC traction frame it is very easy to adjust the angle of pull.
The patient lies down on a suitable treatment table with the neck resting on the Saunders cervical
traction unit. Force is then applied on the neck of the patient. See picture above.
Optional accessories and table suitable for use with lying cervical traction:
The SmarTRAC applies supine cervical traction in conjunction with a suitable treatment table. A suitable
treatment table is one which is flat, provides a stable surface, is long enough for the patient to lie upon,
and be comfortable enough for the patient to lie explains upon for duration of treatment.
Any suitable treatment table may be used. The recommended Ultra Traction Table can be ordered
separately.
The optional cervical traction unit is specified as article:
- 1471808 Saunders cervical traction unit
This article can be ordered separately and mentioned in this user manual chapter 11.
Document 0167337 explains how to mount the Saunders cervical traction unit onto the Ultra Traction
Table. This document is located on the CD-ROM.
10
3 Precautionary Instructions
SYMBOLS:
Follow the instructions in the user manual
It is important that you read, understand and observe the precautionary and operating instructions.
TYPE B APPLIED PART
According IEC60601-1
Emergency stop
Connection for remote control
Fuse
Ratings are described in chapter 10.1
General warning sign
Warning is used to mean, “There is certain danger”
A Warning explains dangers that might result in personal injury or death.
General caution sign
Caution is used to mean, “There is certain danger”
A Caution explains hazards that could damage a product, including data loss.
General prohibition sign
Prohibition is used to mean, “You must not…”
General mandatory action sign
Mandatory action is used to mean, “You must…”
Caution: Finger can be stuck between rope and pulley.
End of life
Tranport values humidity
Tranport values Temperature
11
DO’S:
CAUTION: Federal law (USA) restricts this device to sale by, or on the order of, a physician or
licensed practitioner.
This device should be used only under the continued supervision of a physician or licensed
practitioner.
Make certain that the unit is electrically grounded by connecting only to a grounded electrical
service receptacle conforming to the applicable national and local electrical codes.
You should be aware of the contraindications.
The patient must be within sight of the practitioner at all times.
The patient must always hold the Patient Stop during treatment.
The Thoracic belt and Pelvic belt must be used across the clothes.
Read, understand and practice the precautionary and operating instructions. Know the limitations
and hazards associated with using any electrical stimulation device. Observe the precautionary
and operational stickers placed on the unit.
Before administering any treatment to a patient you should become acquainted with the operating
procedures for each mode of treatment available, as well as the indications, contra-indications,
warnings and precautions. Consult other resources for additional information regarding the
application of traction therapy.
DON'TS:
Do not use force greater than 200N for cervical traction.
Do not operate the unit in the vicinity (less than 2 meters distance) of a short-wave device.
The device should not be used in so-called wet rooms (hydrotherapy rooms).
This device is not suitable for use in the presence of flammable anesthetics mixtures with air,
oxygen, or nitrous oxide.
Do not expose the unit to direct sunlight, heat radiated from a heat radiator, excessive amounts of
dust, moisture, vibrations and mechanical shocks.
To ensure safe use, do NOT accept the limit warning when applying cervical traction to prevent
traction forces over 200 Newton for cervical traction.
WARNING:
This device should be kept out of the reach of children.
Use the SmarTRAC only in conjunction with the appropriate accessories.
Do not connect the unit to an electrical supply without first verifying that the power supply is the
correct voltage. Incorrect voltage may cause unit damage, malfunction, electrical shock, fire, or
personal injury.
Should fray or damage to the Nylon Cord or knot occur, immediately stop use of the unit and
contact qualified technical personnel. Do not attempt to repair the cord or knot yourself.
The patient stop switch should be handed to the patient and its use should be explained. Make
sure the stop switch can be used by the patient during the whole treatment.
In the event of a loss of power to the unit, traction tension should only be released by having the
patient move towards the traction device to release the tension on the rope. Once the tension on
the rope has been released, the mechanical connection to the motor is loosened and the rope
can be pulled out.
12
CAUTION:
This unit should be transported and stored at temperatures between -10 and +70° C (14° and
158° F) with a relative humidity ranging from 10-90%, non-condensing and an atmospheric
pressure between 500 and 1060 hPa.
This unit should be operated between 10 and 40° C (50° and 104° F) with a relative humidity
ranging from 10-90%, non-condensing and an atmospheric pressure between 800 and 1060 hPa.
In the case of ingress of liquids, unplug the unit from the power supply and have it checked by an
authorized person (see the paragraph on maintenance).
Use care on patients who do not tolerate supine positions, they may not be good candidates for
traction therapy.
The unit should be routinely checked before each use to determine all controls function normally.
Handle the unit with care. Inappropriate handling of the unit may adversely affect its
characteristics.
Before each use, inspect the Nylon Cord for wear. Prolonged wear on the cord will cause it to
break, which may cause sudden release of traction pressure on a patient.
Test the Patient Safety Stop before each use for proper operation.
Do not use sharp objects such as a pencil point or ballpoint pen to operate the buttons on the
touch screen base as damage may occur
Do not use the traction unit near devices such as X-ray units or diathermy units. These units may
emit high frequency noise that may affect the operation of the unit.
The device may only be opened or repaired by a qualified technician.
The device should not be used when placed immediately next to or stacked on top of other
devices. If operation is necessary when immediately next to or stacked on top of other devices,
the device should be monitored to ensure it is operating as intended in this arrangement.
13
4 Indications
The SmarTRAC Traction device, with its accessories, is intended to provide relief for the management of
pain and symptoms from a variety of pressures on muscular or skeletal structures. The SmarTRAC
Traction device may be used to treat pain and symptoms associated with the following conditions:
herniated discs, bulging or protruding intervertebral discs, degenerative disc disease, posterior facet
syndrome, acute facet problems, radicular pain, prolapsed discs, spinal root impingement, hypomobility,
degenerative joint disease, facet syndrome, compression fractures, joint pain, and discogenic pain.
4.1 Contra-indications
Traction is contraindicated for the following:
Any condition for which movement is contraindicated
Cardiac or pulmonary problems
Discitis
Primary or metastatic bone tumor
Spinal cord tumor
Osteoporosis
Untreated hypertension
Severe anxiety or Claustrophobia
Clinical signs of myelopathy
Pregnant
Aortic aneurysms
Recent fractures
Inadequate expertise of the practitioner providing the treatment
Vertebral basilar artery insufficiency
Rheumatoid arthritis and other connective tissue disorders
Midline herniated nucleus pulposus
Acute torticollis
Restrictive lung disease or other respiratory disorders
Do not use traction on patients with structural disease (e.g. osteomyelitis, spinal caries, and
ankylosing spondylitis).
Do not use traction on patients with joint instability, hypermobility, or spinal fracture.
Do not use traction on patients with severe cardiovascular disease, vascular compromise, aortic
aneurysm or severe respiratory disease.
Do not use traction on patients with acute sprains, strains, or inflammation that could be
aggravated by traction.
Stop traction if there is an increase in radiated pain to the extremities.
Do not use lumbar traction on patients with hiatal or abdominal hernias.
14
5 Package Contents
Device
1671902 SmarTRAC
Standard accessories SmarTRAC:
1671801 Patient stop switch
1471801 Fixing knobs set of 4
1471806 Nylon cord 170 cm with carabiner hook
1671759 Information booklet SmarTRAC
1671760 CD-ROM
3444672 Power cord UL-CSA 115V straight socket black or grey (for USA)
The CD-ROM (1671760) includes the following documents:
1671758 Operating manual SmarTRAC
0167337 Saunders Clinical Traction Device Assembly Instructions
0167336 Information sheet mounting Ultra traction frame
1671780 Information sheet Traction Belts
1671759 Information booklet SmarTRAC
15
6 Installation
Please see the diagram in section 8 for numbered parts of the device.
6.1 Connection to power supply
Insert the power cable into socket [5] and connect it to a wall socket.
Do not place the device in a location where the power cord could be tripped over or pulled out
during treatment.
Do not attempt to use the device if it is not properly grounded. Make certain that the device is
electrically grounded by connecting it only to a grounded electrical service receptacle
conformable with the applicable national and local electrical codes regarding medical
environments.
Turn on the device with push button [4]
The device will initialize and perform a self-test. This may take a while.
At the end of the self-test the device enters the Home menu and is ready for use.
6.2 Disconnection from power supply
Turn off the device with push button [4]
6.3 Attachment carabiner hook
The cord must be attached to the carabiner hook by means of a knot capable of withstanding
considerable traction, without coming undone. A suitable knot is shown in the figure below.
6.4 Mounting on the traction frame
The SmarTRAC may be mounted on different Enraf-Nonius traction benches by way of a traction frame.
Moreover, the device may be mounted on the mobile traction frame for cervical traction. For mounting
instructions, please refer to the documentation supplied with the traction frames.
16
7 Application Information
7.1 Operating sequence example
Operating sequence depends upon the therapy and accessories employed.
7.1.1 Lumbar traction operating example
Requirements:
- SmarTRAC and suitable traction table
- Pelvic belt
- Thoracic belt
- Flexi stool
- Fixation clamps (2x)
a. Connect the SmarTRAC to the traction table as described in document 0167336. This is the
information sheet for mounting an SmarTRAC to a traction table and is located on the CD-ROM.
b. Install the belts for lumbar traction as described in document 1671780. This is the information
sheet for adjusting the traction belts and is located on the CD-ROM.
c. Hand over the patient stop switch to the patient and explain its operation.
d. Start the treatment.
e. If necessary, adjust the parameters during treatment.
f. After treatment, the patient should have sufficient time to recover.
7.1.2 Cervical traction operating example (lying)
For cervical traction, forces over 200 Newton are not allowed. By selecting a traction force higher than
200 Newton, a warning will be issued about exceeding this limit.
To ensure safe use, do NOT accept the limit warning when applying cervical traction to prevent
traction forces over 200 Newton for cervical traction.
Requirements:
- SmarTRAC and suitable traction table
- Saunders cervical traction unit
a. Connect the SmarTRAC to the traction table as described in document 0167336. This is the
information sheet for mounting an SmarTRAC to a traction table and is located on the CD-ROM.
b. Install the Saunders cervical traction unit as described in document 0167337. This is the
information sheet for adjusting the device and is located on the CD-ROM.
c. Hand over the patient stop switch to the patient and explain its operation.
d. Start the treatment.
e. If necessary, adjust the parameters during treatment.
f. After treatment, the patient should have sufficient time to recover.
7.2 Cord guidance
Make sure the cord is guided over the pulley as shown in the left drawing.
17
8 Operating Instructions
8.1 Operator controls
[1] Cord
[2] Pulley
[3] Display with touch screen
[4] Power switch
0 Device disconnected from power supply
1 Device connected to power supply
[5] Power input and power fuses
[6] Type number plate
Provides information of the device like type and serial number, as well as data on voltage and
maximum current consumption.
[7] Connection patient stop switch
[8] Connection remote control
8.2 Basic Operation
Switch on the device as described in paragraph 6
18
8.2.1 Navigation
The following buttons can be displayed on the navigation bar.
Button
Significance
Back, return to previous screen.
Continue, continue to next screen.
Home, return to Home screen.
Page number / number of pages in multi-page menu screens or treatment step number
/ number of treatment steps in sequential protocols.
Arrow up. Scrolls through pages.
Arrow down. Scrolls through pages.
Store therapy parameters or a programmed sequential protocol as favorite.
Delete.
Pause / Interrupt treatment. Power will be reduced to zero.
Start / Continue treatment.
Accept selected option.
Stop. Power reduces till zero.
Emergency stop. Power reduces till zero at maximum speed.
System settings.
Remote control is connected.
Static traction type selection button
Intermittent traction type selection button
Progressive traction type selection button
Regressive traction type selection button
Cyclic traction type selection button
Add traction type button. When the first traction type is selected, a following traction
type can be added by this button to configure a combined traction treatment.
19
Please select a menu item by pressing the relevant button to navigate to the next screen. You can return
to the previous screen by pressing the back arrow at the top of the screen. You can return to the Home
menu anywhere during navigation, by pressing the home button.
8.2.2 Home
The Home menu provides access to the following
functions:
Protocols
Favorites
Manual operation
System settings
(The button for system settings is located at the top right-hand side
of the navigation bar)
8.2.3 Protocols
Selecting protocols
After selecting the Protocols menu in the main menu,
a list will appear with preprogrammed protocols.
Scroll through this list using the arrows on the
navigation bar. Select the required protocol by
pressing the relevant button.
Protocol information
For more information about a protocol, press the info
button on the left hand side of the protocol.
8.2.4 Favorites
Retrieving favorites
After selecting the Favorites in the main menu, a list
will appear with stored favorites.
By selecting the required favorite, the treatment
screen will appear showing the main parameters of the
favorite. If required, each parameter can be adjusted
here, see 8.2.6.
(A program that requires use of the remote control will be preceded
by the remote control symbol.
Removing favorite
To remove a favorite, first press the recycle bin and
then the favorite you want to remove.
20
8.2.5 Manual Operation
Form of therapy
After selecting Manual Operation in the main menu,
the possible forms of therapy appear.
Depending on the selected form of therapy, you will be
prompted to enter the required parameters.
See paragraph 8.2.1 for button explanation.
See paragraph Error! Reference source not found.
or traction type information.
Setting of parameters
Use the arrow keys on the screen to adjust the
parameter.
Use the tick on the navigation bar to confirm the set
value.
The screen for setting the next parameter appears.
You can use the back key on the navigation bar to
return to the previous parameter.
Setting with body weight
You have the option to set the body weight and a
percentage to determine the required power.
To do this, press the "Setting with body weight" key at
the bottom of the screen.
Set the patient's body weight. In the next screen you
can insert the percentage of the body weight.
(The desired tractive power will be calculated for you. The absolute
value will be shown when you are in the treatment screen)
Speed
The speed with which the tractive power changes can
be set very accurately using the arrow keys.
Moreover you can use the three pictograms at the
bottom of the screen: slow, medium and fast.
When all required parameters have been set, you will
arrive at the treatment screen.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Enraf-Nonius SmarTRAC User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI