  • Hello! I am an AI chatbot trained to assist you with the Detecto SlimPRO Calibration Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
SlimPRO
Calibration
Instructions
1. Press and hold UNIT button (on the bottom of the scale).
2. Install the batteries. All display segments will turn on.
3. Press UNIT button 4 times quickly and hold UNIT button on the fourth press until the
scale shows the zero count.
4. Press UNIT button, and place the scale on the table or bench in 4 seconds.
5. The scale will show “0.0 kg” and then “65.0 kg”.
6. Put 65kg of calibrated test weight on the scale.
7. The scale will show “135.0 kg”.
8. Put 135kg of calibrated test weight on the scale.
9. The scale will show “PASS, and then power off automatically.
Technical Support: 1-866-254-8261 E-mail: tech@cardet.com
/