2.4
Grounding
To protect operating personnel, the National Electrical
Manufacturer's Association
(NEMA)
recommends, and some local codes
require, instrument panels to be grounded.
module which, when used with the proper line cord and plugged
into an appropriate receptacle, grounds the instrument.
2.5
Installation
The Model 818 is equipped with a three conductor power entry
The Model 817 and Model 818 Cryopump Monitors are entirely
solid state and do not generate significant heat. They are
designed to be rack mounted in dead air space and should not be
subjected to temperatures outside of the specified operating
environment temperature range. Figure 2.1 illustrates the input
and output configurations for each instrument.
2.6
Repackaging for shipment
If the Model 817 or Model
818
appears to be operating incorrectly,
please discuss the problem with a factory representative before
returning the instrument. He may be able to suggest several field
tests which could avoid the unnecessary return of an instrument
to the factory when the malfunction is elsewhere. If these tests
determine that the fault
is
in the instrument, the representative
will provide shipping and labeling instructions for returning
it. In order to expedite the repair of the instrument, contact
the factory for a Returned Goods Authorization (RGA) number.
Include the instrument's model and serial numbers in all written
correspondence.
When returning an instrument, please attach a tag securely
to
the instrument itself (not on the shipping carton), clearly
stating
:
A.
Owner, address and phone number
B.
Instrument Model and Serial Numbers
C.
Malfunction Symptoms
D.
Description of External Connections and Cryostats
E.
Returned Goods Authorization Number
If
the original carton
is
available, repack the instrument in a
plastic bag, place it in the carton using original spacers to
protect protruding controls. Seal the carton with strong paper or
nylon tape. Affix shipping labels and
"FRAGILE"
warnings.
-6-