7
1
LIFE2000® VENTILATOR INTRODUCTION
WARNING:
• For any accessories, read the label and accompanying document(s) before use.
• Use only approved accessories and replacement parts with the ventilator. If unauthorized accessories or
replacement parts are used with the ventilator, the ventilator may be damaged and performance may be
degraded.
• Do not connect the ventilator components or accessories to any other equipment that is not described in
this
Instructions for Use
.
• Adding attachments or other components and/or sub-assemblies to the ventilator breathing system can
cause an increase in expiratory resistance at the patient connection
• Adding humidification or nebulization can increase the resistance of the breathing circuit. The operator of
the ventilator needs to monitor the breathing system for increased resistance and blockage.
• Ventilator accuracy can be aected by the gas added by use of a nebulizer.
• Adding attachments or other components or sub-assemblies to the ventilator can change the pressure
gradient across the ventilator system and can aect ventilator performance.
• To ensure accuracy of oxygen administration and to monitor for the presence of contamination (incorrect gas
connected), use an external oxygen monitor to verify the oxygen concentration in the delivered gas.
• To monitor minute volume, use an external exhaled volume monitor.
• Before beginning ventilation therapy, verify that there is an adequate supply of source gas for the intended
duration of the therapy. Otherwise, the patient may not receive appropriate therapy.
• Use only an approved oxygen or air hose with the ventilator. If an unauthorized oxygen or air hose is used
with the ventilator, the ventilator may be damaged.
• Use the ventilator only with approved medical compressed gas (oxygen or dry air). Use with non-approved
sources of gas may cause the ventilator to malfunction and the patient may not receive appropriate
respiratory therapy.
• If the ventilator is not used with a regulator capable of 41 PSI to 87 PSI (nominal 50 PSI) with greater than 40
LPM capability, patients may not receive appropriate respiratory therapy.
• To prevent risk of cross-contamination, clean and disinfect the ventilator before using it on a new patient,
and use a new Breathe Pillows Entrainment Interface.
• The Breathe Pillows Entrainment Interface is designed for single-patient use. To prevent risk of cross-
contamination, use a new Breathe Pillows Entrainment Interface.
• Do not subject Breathe Pillows Entrainment Interface to heat sterilization, hot water pasteurization,
autoclaving, radiation sterilization, ethylene oxide gas sterilization, or attempt to clean them in a dishwasher
or microwave oven. Doing any of these may damage the interface and impair gas delivery.
• Properly secure the Breathe Pillows Entrainment Interface to the face and route tubing around the ears to
avoid strangulation.
• Do not cover the ventilator, touch screen, speaker, or backup alarm buzzer with tape or any other object.
Covering the ventilator or any of its parts might cause diculty in hearing alarms and might aect ventilator
performance.
• Reducing the alarm loudness level to lower than the ambient sound level will impede alarm condition
recognition.
• If upgrading software from version 05.11.00 to 05.12.00 re-evaluate the ventilator settings if PEEP is applied
• If upgrading a patient ventilator from ventilator REF MS-01-0001 to ventilator REF MS-01-0118 re-evaluate
vetntilator settings if PEEP is applied.