Buchi Pure C-805 User manual

Type
User manual
Pure Chromatography Instrument C-805
Operation Manual
11594236 | E en
Imprint
Product Identification:
Operation Manual (Original) Pure Chromatography Instrument C-805
11594236
Publication date: 09.2023
Version E
BÜCHI Labortechnik AG
Meierseggstrasse 40
Postfach
CH-9230 Flawil 1
BUCHI reserves the right to make changes to the manual as deemed necessary in the light of
experience, especially with respect to structure, illustrations and technical details.
This manual is copyrighted. Information from it may neither be reproduced, distributed, or used for
competitive purposes, nor made available to third parties. The manufacture of any component with the aid
of this manual without prior written agreement is also prohibited.
Büchi Labortechnik AG Table of contents
Operation Manual Pure Chromatography Instrument C-805 iii
Table of contents
1 About this document......................................................................................................................... 6
1.1 Warning notices in this document........................................................................................................ 6
1.2 Symbols............................................................................................................................................... 6
1.2.1 Warning symbols ...................................................................................................................6
1.2.2 Mark-ups and symbols...........................................................................................................6
1.3 Trademarks.......................................................................................................................................... 7
1.4 Connected devices .............................................................................................................................. 7
2 Safety .................................................................................................................................................. 8
2.1 Proper use ........................................................................................................................................... 8
2.2 Use other than intended ...................................................................................................................... 8
2.3 Staff qualification ................................................................................................................................. 8
2.4 Residual risks ...................................................................................................................................... 8
2.4.1 Faults during operation ..........................................................................................................9
2.4.2 Malware infection due to connections with other devices or network ....................................9
2.4.3 Damage to the internal memory due to incorrect shutting down of the instrument................9
2.5 Personal protective equipment ............................................................................................................ 9
2.6 Modifications........................................................................................................................................ 9
3 Product description......................................................................................................................... 10
3.1 Description of function ....................................................................................................................... 10
3.2 Configuration ..................................................................................................................................... 11
3.2.1 Front view ............................................................................................................................11
3.2.2 Rear view .............................................................................................................................11
3.2.3 Connections on gradient valve.............................................................................................13
3.3 Type plate.......................................................................................................................................... 13
3.4 Scope of delivery ............................................................................................................................... 14
3.5 Technical data ................................................................................................................................... 14
3.5.1 Pure Chromatography Instrument C-805.............................................................................14
3.5.2 Ambient conditions...............................................................................................................15
3.5.3 Material ................................................................................................................................15
4 Transport and storage..................................................................................................................... 16
4.1 Transport ........................................................................................................................................... 16
4.2 Storage .............................................................................................................................................. 16
4.3 Lifting the instrument ......................................................................................................................... 17
5 Installation........................................................................................................................................ 18
5.1 Installation site................................................................................................................................... 18
5.2 Before installation .............................................................................................................................. 18
5.3 Securing against earthquakes ........................................................................................................... 18
5.4 Establishing electrical connections.................................................................................................... 19
5.5 Establishing solvent connection......................................................................................................... 19
5.6 Assigning solvent to solvent lines...................................................................................................... 20
5.7 Assembling the waste bottle.............................................................................................................. 20
5.8 Assembling the solvent bottle platform (option)................................................................................. 21
5.9 Assembling the cartridge holder ........................................................................................................ 22
Table of contents Büchi Labortechnik AG
iv Operation Manual Pure Chromatography Instrument C-805
6 Operation.......................................................................................................................................... 24
6.1 Control panel ..................................................................................................................................... 24
6.1.1 Layout of control panel.........................................................................................................24
6.1.2 Enter value...........................................................................................................................25
6.1.3 Menu bar..............................................................................................................................25
6.1.4 Conditions panel ..................................................................................................................28
6.1.5 Collection panel ...................................................................................................................29
6.1.6 Gradient panel .....................................................................................................................30
6.1.7 Run panel.............................................................................................................................31
6.1.8 Solvent selection panel ........................................................................................................32
6.1.9 Detector selection panel ......................................................................................................32
6.1.10 Slope detection panel ..........................................................................................................33
6.2 Editing a method................................................................................................................................ 33
6.2.1 Selecting a cartridge (Flash mode) ......................................................................................33
6.2.2 Selecting a solvent...............................................................................................................35
6.2.3 Editing the gradient ..............................................................................................................35
6.2.4 Editing the wavelength for the UV detector..........................................................................38
6.2.5 Editing the fraction collection criteria ...................................................................................39
6.2.6 Editing the fraction collection time .......................................................................................40
6.2.7 Editing the vapor sensor sensitivity......................................................................................41
6.2.8 Saving a method ..................................................................................................................42
6.3 Editing a solvent ................................................................................................................................ 42
6.3.1 Adding a new solvent...........................................................................................................42
6.3.2 Deleting a solvent ................................................................................................................43
6.4 Tasks during a separation ................................................................................................................. 44
6.4.1 Injecting a sample into the flash system ..............................................................................44
6.4.2 Installing and removing a cartridge ......................................................................................45
6.4.3 Inserting the fraction collection tray .....................................................................................46
6.4.4 Selecting an existing method ...............................................................................................46
6.4.5 Selecting values on the dialog box Sample Loading ...........................................................47
6.5 Performing a Flash separation using a method................................................................................. 47
6.5.1 Preparing the instrument......................................................................................................47
6.5.2 Starting Flash separation using a method ...........................................................................47
6.5.3 Changings during a separation ............................................................................................48
6.5.4 Ending a Flash separation ...................................................................................................48
6.5.5 Shutting down the instrument ..............................................................................................48
6.6 Performing a Flash separation manually ........................................................................................... 48
6.6.1 Preparing the instrument......................................................................................................48
6.6.2 Starting Flash separation manually......................................................................................49
6.6.3 Changings during a separation ............................................................................................49
6.6.4 Ending a Flash separation ...................................................................................................50
6.6.5 Shutting down the instrument ..............................................................................................50
6.7 Identifying fractions............................................................................................................................ 50
6.7.1 Identifying fractions by peak ................................................................................................50
6.7.2 Identifying fractions per vial .................................................................................................51
6.8 Importing and exporting data............................................................................................................. 51
6.8.1 Printing a run report .............................................................................................................51
6.8.2 Sending data to USB ...........................................................................................................52
6.8.3 Sending PDF to USB ...........................................................................................................52
7 Cleaning and servicing ................................................................................................................... 53
7.1 Regular maintenance work................................................................................................................ 53
7.2 Carrying out a data backup................................................................................................................ 54
7.3 Calibrating the display ....................................................................................................................... 54
7.4 Cleaning the check valve................................................................................................................... 54
7.5 Cleaning the instrument..................................................................................................................... 55
Büchi Labortechnik AG Table of contents
Operation Manual Pure Chromatography Instrument C-805 v
8 Help with faults ................................................................................................................................ 57
8.1 Faults, possible causes and solutions (general)................................................................................ 57
8.2 Faults, possible causes and solutions (cartridge).............................................................................. 57
8.3 Faults, possible causes and solutions (solvent delivery)................................................................... 57
8.4 Faults, possible causes and solutions (fraction collection)................................................................ 58
8.5 Faults, possible causes and solutions (detection) ............................................................................. 59
8.6 Error messages ................................................................................................................................. 59
8.7 Resetting the Fraction collector arm.................................................................................................. 63
9 Taking out of service and disposal................................................................................................ 64
9.1 Taking out of service.......................................................................................................................... 64
9.2 Disposal............................................................................................................................................. 64
9.3 Returning the instrument ................................................................................................................... 64
10 Appendix .......................................................................................................................................... 65
10.1 Spare parts and accessories ............................................................................................................. 65
10.1.1 Accessories..........................................................................................................................65
10.1.2 Spare parts ..........................................................................................................................66
10.1.3 Tools and adapter kits..........................................................................................................66
10.1.4 Tubes ...................................................................................................................................66
10.1.5 Wear parts ...........................................................................................................................67
10.2 Folder locations ................................................................................................................................. 67
10.3 Finding out separation conditions with the navigator......................................................................... 67
10.3.1 Open the navigator ..............................................................................................................67
10.3.2 TLC Silica.............................................................................................................................68
10.3.3 LC-C18.................................................................................................................................69
10.3.4 LC-Transfer..........................................................................................................................70
1 | About this document Büchi Labortechnik AG
6/72 Operation Manual Pure Chromatography Instrument C-805
1 About this document
This operation manual is applicable for all variants of the instrument.
Read this operation manual before operating the instrument and follow the
instructions to ensure safe and trouble-free operation.
Keep this operation manual for later use and pass it on to any subsequent user or
owner.
BÜCHI Labortechnik AG accepts no liability for damage, faults and malfunctions
resulting from not following this operation manual.
If you have any questions after reading this operation manual:
uContact BÜCHI Labortechnik AG Customer Service.
https://www.buchi.com/contact
1.1 Warning notices in this document
Warning notices warn you of dangers that can occur when handling the device. There
are four danger levels, each identifiable by the signal word used.
Signal word Meaning
DANGER Indicates a danger with a high level of risk which could result in
death or serious injury if not prevented.
WARNING Indicates a danger with a medium level of risk which could result in
death or serious injury if not prevented.
CAUTION Indicates a danger with a low level of risk which could result in mi-
nor or medium-severity injury if not prevented.
NOTICE Indicates a danger that could result in damage to property.
1.2 Symbols
The following symbols are displayed in this operation manual or on the device:
1.2.1 Warning symbols
Symbol Meaning
General warning
Dangerous electrical voltage
Flammable substances
Laser class 1
1.2.2 Mark-ups and symbols
NOTE
This symbol draws attention to useful and important information.
RThis character draws attention to a requirement that must be met before the
instructions below are carried out.
uThis character indicates an instruction that must be carried out by the user.
ðThis character indicates the result of a correctly carried out instruction.
Büchi Labortechnik AG About this document | 1
Operation Manual Pure Chromatography Instrument C-805 7/72
Mark-up Explanation
Window
Software Windows are marked-up like this.
Tab
Tabs are marked-up like this.
Dialog
Dialogs are marked-up like this.
[Button] Buttons are marked-up like this.
[Field names] Field names are marked-up like this.
[Menu / Menu item] Menus or menu items are marked-up like this.
Status Status is marked-up like this.
Signal Signals are marked-up like this.
1.3 Trademarks
Product names and registered or unregistered trademarks that are used in this
document are used only for identification and remain the property of the owner in
each case.
1.4 Connected devices
In addition to these operating instructions, follow the instructions and specifications in
the documentation for the connected devices.
2 | Safety Büchi Labortechnik AG
8/72 Operation Manual Pure Chromatography Instrument C-805
2 Safety
2.1 Proper use
The instrument is designed and built for laboratories.
The instrument can be used for the following tasks:
Purification
Separation of one or more compounds from a mixture
2.2 Use other than intended
Use of any kind other than that described in Chapter2.1 "Proper use", page8 and
any application that does not comply with the technical specifications (See
Chapter3.5 "Technical data", page14) constitutes use other than intended.
In particular, the following applications are not permissible:
Use of the instrument in areas which require explosion-safe instruments.
Use of the instrument with solvents containing peroxides.
Use the instrument for production purposes.
2.3 Staff qualification
Unqualified persons are unable to identify risks and are therefore exposed to greater
dangers.
The device may only be operated by suitably qualified laboratory staff.
These operating instructions are aimed at the following target groups:
Users
Users are persons that meet the following criteria:
They have been instructed in the use of the device.
They are familiar with the contents of these operating instructions and the
applicable safety regulations and apply them.
They are able on the basis of their training or professional experience to assess
the risks associated with the use of the device.
Operator
The operator (generally the laboratory manager) is responsible for the following
aspects:
The device must be correctly installed, commissioned, operated and serviced.
Only suitably qualified staff may be assigned the task of performing the operations
described in these operating instructions.
The staff must comply with the local applicable requirements and regulations for
safe and hazard-conscious working practices.
Safety-related incidents that occur while using the device should be reported to the
manufacturer ([email protected]).
BUCHI service technicians
Service technicians authorized by BUCHI have attended special training courses and
are authorized by BÜCHI Labortechnik AG to carry out special servicing and repair
measures.
2.4 Residual risks
The device has been developed and manufactured using the latest technological
advances. Nevertheless, risks to persons, property or the environment can arise if the
device is used incorrectly.
Büchi Labortechnik AG Safety | 2
Operation Manual Pure Chromatography Instrument C-805 9/72
Appropriate warnings in this manual serve to alert the user to these residual dangers.
2.4.1 Faults during operation
If a device is damaged, sharp edges, moving parts or exposed electrical wires can
cause injuries.
uRegularly check device for visible damage.
uIf faults occur, switch off the device immediately, unplug the power cord and
inform the operator.
uDo not continue to use devices that are damaged.
2.4.2 Malware infection due to connections with other devices or
network
Connections with other devices or a network can cause a malware infection to the
instrument.
uInstall antivirus software and firewall on the instrument before connecting it to
other devices or network.
2.4.3 Damage to the internal memory due to incorrect shutting
down of the instrument
Incorrect shutting down of the instrument can cause damage to the internal memory.
uShut down the instrument as described. See Chapter6 "Operation", page24
2.5 Personal protective equipment
Depending on the application, hazards due to heat and/or corrosive chemicals may
arise.
uAlways wear appropriate personal protective equipment such as safety goggles,
protective clothing and gloves.
uMake sure that the personal protective equipment meets the requirements of the
safety data sheets for all chemicals used.
2.6 Modifications
Unauthorized modifications can effect safety and lead to accidents.
uUse only genuine BUCHI accessories, spare parts and consumables.
uCarry out technical changes only with prior written approval from BUCHI.
uOnly allow changes to be made by BUCHI service technicians.
BUCHI accepts no liability for damage, faults and malfunctions resulting from
unauthorized modifications.
3 | Product description Büchi Labortechnik AG
10/72 Operation Manual Pure Chromatography Instrument C-805
3 Product description
3.1 Description of function
The instrument is a purification device designed to purify complex samples by flash
chromatography.
Flash chromatography has the ability to separate gram size samples in short
period.
The instrument allows:
2 different solvents
Injection of liquid or solid sample
Separation on a cartridge
Identifying the compounds by UV
Collecting the desired fractions
Instrument schematic:
Gradient valve
UV
detector
Fraction
collection
Solvent 2
Solvent 1
Pump
Sample
introduction
Cartridge
Büchi Labortechnik AG Product description | 3
Operation Manual Pure Chromatography Instrument C-805 11/72
3.2 Configuration
3.2.1 Front view
1
2 3 4
5
6
7
Fig.1: Front view
1 Control panel 2 RFID reader
3 On/Off switch 4 USB Port
5 Cartridge holder 6 Flash pump
7 Fraction collection bay
3.2.2 Rear view
NOTE
All electrical connections are not limited energy.
3 | Product description Büchi Labortechnik AG
12/72 Operation Manual Pure Chromatography Instrument C-805
9
8
7
6
1
2
3
4
5
Fig.2: Rear view
1 Ventilation slot 2 Gradient valve
(See Chapter3.2.3 "Connections on
gradient valve", page13 )
3 Signal connection
(for external air supply)
4 LAN port
5 Power supply connection 6 Fuse
7 On/Off master switch 8 USB ports
9 Type plate
(See Chapter3.3 "Type plate",
page13
Büchi Labortechnik AG Product description | 3
Operation Manual Pure Chromatography Instrument C-805 13/72
3.2.3 Connections on gradient valve
3
2
1
Fig.3: Connections on the gradient valve
1 Waste line 2 Solvent line 2
3 Solvent line 1
3.3 Type plate
The type plate identifies the instrument. The type plate is located at the rear of the
instrument.
Buchi Labortechnik AG
9230 Flawil / Switzerland
Type: xxx
SN: 10xxxxxxxx
Volt: xxx-xxx VAC
Frequ.: xx Hz
Power: xxxx W
Built: 20xx
Made in Switzerland
6
5
4
1
3
2
7
8
910 9
Fig.4: Type Plate
1 Company name and address 2 Instrument name
3 Serial number 4 Input voltage range
5 Frequency 6 Power consumption maximum
7 Year of manufacture 8 Product code
9 Approvals 10 Symbol for "Do not dispose of as
household waste"
3 | Product description Büchi Labortechnik AG
14/72 Operation Manual Pure Chromatography Instrument C-805
3.4 Scope of delivery
NOTE
The scope of delivery depends of the configuration of the purchase order.
Accessories are delivered as per the purchase order, order confirmation, and
delivery note.
3.5 Technical data
3.5.1 Pure Chromatography Instrument C-805
Pure C-805
Dimension
(W x D x H)
365 x 570 x 680 mm
Weight 24 kg
Clearance on each side of the instrument 200 mm
Power consumption 350 W
Connection voltage 100 - 240 V ± 10 %
Frequency 50 - 60 Hz
Fuse 4 A
Overvoltage category II
Pollution degree 2
IP Code IP 20
Solvents 2
USB port 3
LAN port
(RJ45)
1
RFID reader
(racks)
Yes
RFID reader
(cartridges)
Yes
Fraction collector bay closed
Lighted fraction collector bay Yes
Vapor sensor Yes
Pump Flash Mode
Pure C-805
Gradient binary
Pressure range 0 - 50 bar
Flow rate 0 - 250 mL/min
Flow rate accuracy < 2 %
Functional principle self-priming
Specification 3 pistons
radial arranged
UV Detector
Pure C-805
UV wavelength range 200 - 400 nm
Büchi Labortechnik AG Product description | 3
Operation Manual Pure Chromatography Instrument C-805 15/72
Pure C-805
Light sources Deuterium
Lifetime lamp 2000 hours
Detector DAD
3.5.2 Ambient conditions
For indoor use only.
Max. altitude above sea level 2000 m
Ambient temperature 5‒40°C (25°C)
No maximum performance above 25°C
Maximum relative humidity 80% non-condensing, for temperatures up
to 31°C
Storage temperature max. 45 °C
3.5.3 Material
Component Materials of construction
Housing Powder coated steel 1.4301
Fraction collection bay PMMA / PET
Pump head PEEK
Pump cover PP
Pump excenter housing Aluminum
Tubings FEP
Fitting POM
Valve screw fitting POM
Ferrule ETFE
Cone ring POM
Radial seal PTFE
Pistons Ceramic
4 | Transport and storage Büchi Labortechnik AG
16/72 Operation Manual Pure Chromatography Instrument C-805
4 Transport and storage
4.1 Transport
NOTICE
Risk of breakage due to incorrect transportation
uMake sure that all parts of the device are safely packed in such a way as to
prevent breakage, ideally in the original box.
uAvoid sharp movements during transit.
uAfter transportation, check the device for damage.
uDamage that has occurred in transit should be reported to the carrier.
uKeep packing for future transportation.
4.2 Storage
uMake sure that the ambient conditions are complied with (see Chapter3.5
"Technical data", page14).
uMake sure a clean solvent like ethanol or isopropanol is in the pump.
uWherever possible, store the device in its original packaging.
uAfter storage, check the device, all seals and tubing for damage and replace if
necessary.
Büchi Labortechnik AG Transport and storage | 4
Operation Manual Pure Chromatography Instrument C-805 17/72
4.3 Lifting the instrument
WARNING
Danger due to incorrect transportation
The possible consequences are crushing injuries, cuts and breakages.
uThe instrument should be transported by two persons at the same time.
uLift the instrument at the points indicated.
uLift the instrument – this requires two persons each lifting at one of the points
indicated on the bottom of the instrument.
5 | Installation Büchi Labortechnik AG
18/72 Operation Manual Pure Chromatography Instrument C-805
5 Installation
5.1 Installation site
The installation site must meet the following requirements:
The installation site has a firm, level surface.
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter3.5 "Technical data", page14
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site has a own mains outlet socket for the instrument.
The installation site has enough space that cables / tubes can be routed safely.
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site is not exposed to increased electromagnetic emissions.
Electromagnetic fields in the frequency range between 200 to 300 MHz can cause
the instrument to operate incorrectly.
The installation site meets the requirements of the safety data sheets for all
solvents and samples used.
5.2 Before installation
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
uClimatize the instrument after transportation.
5.3 Securing against earthquakes
The instrument has an earthquake fixing point to protect the device against falling.
1
Fig.5: Securing against earthquakes
1 Fixing point
uTie the lashing mount to a fixed point using strong cord or a wire.
Büchi Labortechnik AG Installation | 5
Operation Manual Pure Chromatography Instrument C-805 19/72
5.4 Establishing electrical connections
NOTICE
Risk of instrument damage because of not suitable power supply cables.
Not suitable power supply cables can cause bad performance or an instrument
damage
uUse only BUCHI power supply cables.
Precondition:
RThe electrical installation is as specified on the
type plate.
RThe electrical installation is equipped with a
proper grounding system.
RThe electrical installation is equipped with
suitable fuses and electrical safety features.
RThe installation site is as specified in the
technical date. See Chapter3.5 "Technical
data", page14
uConnect the power supply cable to the
connection on the instrument. See Chapter3.2
"Configuration", page11
uConnect the mains plug to an own mains outlet
socket.
5.5 Establishing solvent connection
NOTICE
Solvent bottles on top of the instrument.
Solvent bottles on top of the instrument can cause property damages.
uLocate the solvent bottles next to the instrument.
uUse the optional solvent bottle platform.
Precondition:
RMake sure that the instrument is not connected
to the power supply.
uAttach all solvent lines to the instrument.
Connections see Chapter3.2 "Configuration",
page11
uPut the other end of the solvent line into the
solvent bottle.
uAssign the solvent to the solvent lines. See
Chapter5.6 "Assigning solvent to solvent lines",
page20
5 | Installation Büchi Labortechnik AG
20/72 Operation Manual Pure Chromatography Instrument C-805
5.6 Assigning solvent to solvent lines
Navigation path
Tools Solvent Loading
Precondition:
RThe solvent bottle is connected to the instrument. See Establishing solvent
connection
RThe solvent you wish to use is part of the solvent library. See Chapter6.3 "Editing
a solvent", page42
uNavigate to the
Solvent Loading
dialog according to the navigation path.
ðThe display shows the dialog box
Solvent Loading
.
uTap the drop-down list next to Line 1.
ðThe display shows a drop-down list with selectable solvents.
uSelect the solvent which is connected to solvent line 1.
ðThe solvent for Line 1 is assigned.
ðThe drop-down list closes.
uRepeat the solvent selection for each line.
uActivate the checkbox next to the line you wish to prime.
uTap the button [Auto Prime].
uWait till priming finished.
uTap the button [Close].
ðAll lines are assigned with solvents.
ðThe dialog box closes.
5.7 Assembling the waste bottle
NOTICE
Waste bottle on top of the instrument
Waste bottle on top of the instrument can cause property damages.
uMake sure that there is a decline between the connection on the instrument and
the waste bottle.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72

Buchi Pure C-805 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI