easily change the voltage and current settings.
Output off timer function
To prevent battery overcharge and similar problems, this
function automatically turns the output off after a preset amount
of time has passed while the output is on.
CC priority mode
Compared to general switching power supply units, the PLR
power supply unit is better able to reduce current overshoot,
thanks to our original current-overshoot inhibit circuit. This
circuit produces a load that enables the unit to operate at a
constant current while the output is on.
Sequence function
Sequence programs can be written to the unit from a computer
via the interface boards (PLR-RS, PLR-GU and PLR-LU).
Sequence programs can be executed by performing panel or
computer operations. You can also perform sequence operations
with a maximum of 1000 steps by using a program written by the
computer. The minimum step unit is 50ms.
Remote sensing function
This function uses the remote sensing terminal to compensate for
voltage drop caused by the wires.
Protective functions
The PLR power supply unit has protective functions against
overvoltage, under voltage and overcurrent for the primary side,
and overvoltage, overcurrent, remote sensing (terminal open),
and internal heat for the secondary side. The unit is also
equipped with OVP (over-voltage protection), UVP (under-
voltage protection), and OCP (over-current protection) for the
load. The setting values for these protective functions are
changeable.
Master-slave operation
The PLR power supply unit can perform master-slave operation.
Master-slave operation can be performed with a maximum of 3
units (of the same voltage model) connected in parallel, and with