Mettler Toledo HX204 Halogen Moisture Analyzer Operating instructions

Type
Operating instructions
Ex
c
e
llence
Plus
Moisture Analyzer
Excellence Plus HX204
Reference Manual
30019565F1/30/2020 12:46 PM- Schema ST4 PDF engine -Layout by Victor Mahler
Table of Contents
1 Introduction 5
2 Safety Information 6
2.1 Definition of signal warnings and symbols...................................................................... 6
2.2 Product specific safety information................................................................................. 6
3 Design and Function 9
3.1 Overview..................................................................................................................... 9
3.1.1 Overview drying unit...................................................................................... 9
3.1.2 Overview terminal ......................................................................................... 10
3.1.3 Overview type plates...................................................................................... 10
3.2 Compliance Information................................................................................................ 11
3.3 Function description..................................................................................................... 11
4 Installation and Putting into Operation 13
4.1 Selecting the location ................................................................................................... 13
4.2 Unpacking .................................................................................................................. 13
4.3 Scope of delivery ......................................................................................................... 13
4.4 Connecting the instrument............................................................................................. 14
4.5 Opening and closing sample chamber ........................................................................... 14
4.6 Setup wizard ............................................................................................................... 15
4.7 Setting up the instrument .............................................................................................. 15
4.8 Leveling the instrument................................................................................................. 16
4.9 Anti-theft device ........................................................................................................... 18
4.10 Setting reading angle of terminal.................................................................................... 18
4.11 Setting date and time.................................................................................................... 18
4.12 Adjusting after setup..................................................................................................... 18
4.13 Transporting, packaging and storing.............................................................................. 19
4.13.1 Transporting the Moisture Analyzer.................................................................. 19
4.13.2 Putting into operation after transport................................................................ 19
4.13.3 Packaging and storing................................................................................... 19
4.14 Installing USB device driver........................................................................................... 20
5 Operating Elements and Interfaces 21
5.1 Operation keys ............................................................................................................ 21
5.2 Electrical interfaces ...................................................................................................... 21
6 Operation 23
6.1 Operation orientation.................................................................................................... 23
6.2 User interface .............................................................................................................. 23
6.2.1 Operation Keys ............................................................................................. 23
6.2.2 Screen navigation ......................................................................................... 23
6.2.3 Basic elements on the touch screen ................................................................ 24
6.2.4 Icons and buttons ......................................................................................... 24
6.2.4.1 Status message icons.............................................................................. 24
6.2.4.2 Title bar icons......................................................................................... 25
6.2.4.3 Action bar buttons ................................................................................... 25
6.2.5 Home Screen................................................................................................ 25
6.2.6 Work Screen................................................................................................. 26
6.2.7 Input dialogs ................................................................................................ 27
6.2.7.1 Entering characters and numbers.............................................................. 27
6.2.7.2 Entering numerical values ........................................................................ 28
6.2.7.3 Changing date and time........................................................................... 28
6.2.8 Lists and tables ............................................................................................ 29
6.3 Performing a Simple Measurement................................................................................. 30
Table of Contents 1Moisture Analyzer
7 Software Description 34
7.1 Method Definition......................................................................................................... 34
7.1.1 Method parameters ....................................................................................... 34
7.1.1.1 Main measurement parameters................................................................. 34
7.1.1.2 Result and value handling........................................................................ 44
7.1.1.3 Workflow handling .................................................................................. 47
7.1.1.4 General method properties........................................................................ 49
7.1.2 Creating a method......................................................................................... 50
7.1.2.1 Creating a method with the Method Wizard ................................................ 50
7.1.2.2 Creating a method manually..................................................................... 51
7.1.2.3 Creating a method by consulting the method library .................................... 51
7.1.3 Editing a method........................................................................................... 52
7.1.4 Testing a method .......................................................................................... 53
7.1.4.1 Temperature assistant.............................................................................. 53
7.1.4.2 Test measurement................................................................................... 54
7.1.4.3 Testing with AutoMet................................................................................ 55
7.1.5 Preparing a method for QuickPredict ............................................................... 55
7.1.5.1 Prerequisites to use QuickPredict for a method............................................ 56
7.1.5.2 Activating QuickPredict............................................................................. 57
7.1.5.3 Establishing the prediction knowledge ....................................................... 59
7.2 Measurement .............................................................................................................. 62
7.2.1 Performing a measurement ............................................................................ 62
7.2.2 Optimum sample preparation ......................................................................... 64
7.2.3 Working with weighing-in aid ......................................................................... 65
7.2.4 Working with QuickPredict ............................................................................. 65
7.2.4.1 Performing measurements with QuickPredict .............................................. 65
7.2.4.2 Handling QuickPredict results ................................................................... 68
7.2.4.3 Working with uncertainty thresholds .......................................................... 69
7.2.5 Working with shortcuts .................................................................................. 69
7.3 Results ....................................................................................................................... 70
7.3.1 Graphical evaluations of measurement results.................................................. 70
7.3.2 Method versioning......................................................................................... 73
7.3.3 Exporting results ........................................................................................... 73
7.3.3.1 Exporting results to a USB storage device................................................... 74
7.3.3.2 Exporting results with EasyDirect Moisture.................................................. 74
7.3.3.3 Exporting results via FTP server................................................................. 74
7.3.3.4 Export settings ........................................................................................ 75
7.4 Test/Adjust .................................................................................................................. 76
7.4.1 Tests ........................................................................................................... 77
7.4.1.1 Weight test - internal................................................................................ 77
7.4.1.2 Weight test - external............................................................................... 77
7.4.1.3 Weight test - Repeatability external............................................................ 78
7.4.1.4 Temperature test...................................................................................... 79
7.4.1.5 SmartCal test .......................................................................................... 80
7.4.1.6 Measures after failed test.......................................................................... 82
7.4.2 Adjustments ................................................................................................. 82
7.4.2.1 Fully automatic adjustment (FACT)............................................................ 82
7.4.2.2 Weight adjustment - internal..................................................................... 82
7.4.2.3 Weight adjustment - external .................................................................... 83
7.4.2.4 Temperature adjustment........................................................................... 83
7.4.3 History......................................................................................................... 85
7.5 Settings ...................................................................................................................... 86
7.5.1 User preferences ........................................................................................... 87
7.5.1.1 Languages ............................................................................................. 87
7.5.1.2 Screen .................................................................................................. 88
7.5.1.3 Sound.................................................................................................... 89
7.5.1.4 Password............................................................................................... 89
Table of Contents2 Moisture Analyzer
7.5.2 Instrument settings ........................................................................................ 90
7.5.2.1 Regional settings..................................................................................... 90
7.5.2.2 Peripherals............................................................................................. 90
7.5.2.3 Default settings ....................................................................................... 92
7.5.2.4 Company information .............................................................................. 93
7.5.2.5 Touch screen adjustment ......................................................................... 93
7.5.2.6 Network interfaces................................................................................... 94
7.5.3 Applications settings...................................................................................... 98
7.5.3.1 Identifications management...................................................................... 98
7.5.3.2 Print and export management................................................................... 99
7.5.3.3 Method selection..................................................................................... 102
7.5.4 Quality management ..................................................................................... 103
7.5.4.1 Test / Adjustment settings......................................................................... 103
7.5.4.2 QuickPredict ........................................................................................... 107
7.5.5 User management......................................................................................... 108
7.5.5.1 Groups .................................................................................................. 108
7.5.5.2 Users..................................................................................................... 109
7.5.5.3 Account policies...................................................................................... 110
7.5.6 System and data management ....................................................................... 111
7.5.6.1 Result management................................................................................. 111
7.5.6.2 Export / Import........................................................................................ 111
7.5.6.3 Back up / Restore.................................................................................... 113
7.5.6.4 Export event logs..................................................................................... 113
7.5.6.5 License manager..................................................................................... 114
7.5.6.6 Reset ..................................................................................................... 114
7.5.6.7 Update................................................................................................... 115
7.5.7 Help and tutorials.......................................................................................... 115
7.5.7.1 Instrument tutorial ................................................................................... 115
8 Maintenance 116
8.1 Maintenance table........................................................................................................ 116
8.2 Cleaning..................................................................................................................... 116
8.2.1 Sample chamber........................................................................................... 117
8.2.2 Heating module ............................................................................................ 117
8.2.3 Fan grill ....................................................................................................... 120
8.2.4 Putting into operation after cleaning ................................................................ 120
8.3 Replacing the Dust filter................................................................................................ 120
8.4 Replacing the power line fuse........................................................................................ 120
8.5 Updating the software................................................................................................... 121
9 Troubleshooting 122
9.1 Malfunctions shown on the display................................................................................ 122
9.2 Malfunctions when operating the instrument.................................................................... 123
10 Accessories and Spare Parts 127
10.1 Accessories................................................................................................................. 127
10.2 Spare parts ................................................................................................................. 131
11 Technical Data 132
11.1 General data ............................................................................................................... 132
11.2 Model-specific data...................................................................................................... 133
11.3 Dimensions................................................................................................................. 134
11.3.1 Drying unit ................................................................................................... 134
11.3.2 Terminal ...................................................................................................... 135
11.4 Interfaces specification ................................................................................................. 135
11.4.1 RS232C....................................................................................................... 135
11.4.2 USB host...................................................................................................... 136
11.4.3 USB device................................................................................................... 136
Table of Contents 3Moisture Analyzer
12 Disposal 137
Index 139
Table of Contents4 Moisture Analyzer
Introduction 5Moisture Analyzer
1 Introduction
Thank you for purchasing a Halogen Moisture Analyzer from METTLER TOLEDO.
Your Moisture Analyzer is fast, reliable and offers a high level of operating convenience and useful functions to
facilitate determination of the moisture content of your samples.
The Moisture Analyzer conforms with all common standards and directives. It supports stipulations, work
techniques and result records as demanded by all international quality assurance systems, e.g. GLP (Good
Laboratory Practice), GMP (Good Manufacturing Practice). The instrument has a CE declaration of conformity
and METTLER TOLEDO as the manufacturer has been awarded ISO 9001 and ISO 14001 certification. This
provides you with the assurance that your capital investment is protected in the long term by a high product
quality and a comprehensive service package (repairs, maintenance, servicing, adjustment service).
EULA
The software in this product is licensed under the METTLER TOLEDO End
User License Agreement (EULA) for Software.
When using this product you agree to the terms of the EULA.
u www.mt.com/EULA
Finding more information
u www.mt.com/moisture
u www.mt.com/hxhs
Search for software
u http://www.mt.com/moisture-software
Software version
This document is based on the software version V2.40 / V3.10.
Search for documents u www.mt.com/library
For further questions, please contact your authorized METTLER TOLEDO dealer or service representative.
u www.mt.com/contact
Conventions and symbols
Key and button designations are indicated by a picture or bold text (e.g. Save).
Refers to an external document.
Note
For useful information about the product.
Elements of instructions
In this manual, step-by-step instructions are presented as follows. The action steps are numbered and can
contain prerequisites, intermediate results and results, as shown in the example. Sequences with less than two
steps are not numbered.
§
Prerequisites that must be fulfilled before the individual steps can be executed.
1 Step 1
ð Intermediate result
2 Step 2
ð Result
Safety Information6 Moisture Analyzer
2 Safety Information
Two documents named "User Manual" and "Reference Manual" are available for this instrument.
The User Manual is printed and delivered with the instrument.
The electronic Reference Manual contains a full description of the instrument and its use.
Keep both documents for future reference.
Include both documents if you transfer the instrument to other parties.
Only use the instrument according to the User Manual and the Reference Manual. If you do not use the
instrument according to these documents or if the instrument is modified, the safety of the instrument may be
impaired and Mettler-Toledo GmbH assumes no liability.
2.1 Definition of signal warnings and symbols
Safety notes contain important information on safety issues. Ignoring the safety notes may lead to personal
injury, damage to the instrument, malfunctions and false results. Safety notes are marked with the following
signal words and warning symbols:
Signal words
WARNING
A hazardous situation with medium risk, possibly resulting in death or severe injury if
not avoided.
CAUTION
A hazardous situation with low risk, resulting in minor or moderate injury if not avoided.
NOTICE
A hazardous situation with low risk, resulting in damage to the instrument, other
material damage, malfunctions and erroneous results, or loss of data.
Warning symbols
General hazard: read the User Manual or the Reference Manual for information about the hazards
and the resulting measures.
Hot surface Notice
2.2 Product specific safety information
Intended use
This instrument is designed to be used by trained staff. The instrument is intended for determining the weight
loss during drying of samples.
Any other type of use and operation beyond the limits of use stated by Mettler-Toledo GmbH without consent
from Mettler-Toledo GmbH is considered as not intended.
Moisture determination applications must be optimized and validated by the user according to local
regulations. Application-specific data provided by METTLER TOLEDO is intended for guidance only.
Responsibilities of the instrument owner
The instrument owner is the person holding the legal title to the instrument and who uses the instrument or
authorizes any person to use it, or the person who is deemed by law to be the operator of the instrument. The
instrument owner is responsible for the safety of all users of the instrument and third parties.
METTLER TOLEDO assumes that the instrument owner trains users to safely use the instrument in their
workplace and deal with potential hazards. METTLER TOLEDO assumes that the instrument owner provides the
necessary protective gear.
Safety Information 7Moisture Analyzer
Personal protective equipment
Insulated gloves Lab coat
Goggles
Safety notes
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured and the instrument
can be damaged.
1 Check that the voltage printed on the instrument is the same as your local power supply
voltage. If this is not the case, under no circumstances connect the instrument to the
power supply, but contact a METTLER TOLEDO representative.
2 Only use the three-core power cable with equipment grounding conductor supplied by
METTLER TOLEDO to connect your instrument.
3 Only connect it to a three-pin power socket with earthing contact.
4 Only use standardized extension cables with equipment grounding conductor for
operating the instrument.
5 Make sure that the power plug is accessible at all times.
6 Arrange the cables so that they cannot be damaged or interfere with the operation.
7 Keep all electrical cables and connections away from liquids.
WARNING
Injury or death due to toxic or corrosive substances
Heating up toxic or corrosive substances, e.g., acids, can result in toxic or corrosive vapors
that can cause injuries if they come in contact with the skin or the eyes or if they are
inhaled.
1 When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
2 Set up the instrument in a well-ventilated location.
3 When using dry substances that form toxic gases, place the instrument in a fume hood.
WARNING
Death or serious injuries due to flammable solvents
Flammable solvents in the vicinity of the instrument can ignite and lead to fire and
explosions.
1 Keep flammable solvents away from the instrument.
2 When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
Safety Information8 Moisture Analyzer
CAUTION
Burns due to hot surfaces
During operation, parts of the instrument can reach temperatures that can cause burns if
touched.
1 Do not touch the area marked with the warning symbol.
2 Ensure sufficient free space around the instrument to avoid heat accumulation and
overheating (approx. 1 m free space above the heating module).
3 Never cover, tape or clog the vent over the sample chamber. Do not tamper with the vent
in any other way.
4 Exercise caution when removing a sample. The sample itself, the sample chamber, the
draft shield and the sample pan may be very hot.
5 Do not open the heating module during operation. Always let it cool down completely
before opening.
6 Do not modify the heating module in any way.
NOTICE
Damage to the instrument due to corrosive substances and vapors
Corrosive substances and corrosive vapors can damage the instrument.
1 When using chemicals and solvents, comply with the instructions of the manufacturer
and the general laboratory safety rules.
2 Ensure that the instrument parts touching your sample substance can not get altered by
it.
3 Wipe off any condensation of corrosive vapours after an operation.
4 Work with small samples.
NOTICE
Damage to the instrument due to the use of unsuitable parts
Using unsuitable parts with the instrument can damage the instrument or cause it to
malfunction.
Only use parts from METTLER TOLEDO that are intended to be used with your
instrument.
For related information concerning the QuickPredict function, see [Prerequisites to use QuickPredict for a
method}Page56].
Design and Function 9Moisture Analyzer
3 Design and Function
In order to work with your Moisture Analyzer, you need to familiarize yourself with the instrument layout, drying
unit, graphical user interface of the touch screen as well as the terminal.
3.1 Overview
3.1.1 Overview drying unit
Legend drying unit
1
Heating module (with thermal overload
protection)
2
Handles for opening the sample chamber
3
Power contacts for halogen lamp
4
Draft shield element
5
Sample chamber
6
Level indicator / Level sensor
7
Leveling screw
8
Sample pan handler
9
Sample pan holder
10
Contacts for temperature kit (optional)
11
Temperature sensor
12
Inspection window and vent
13
2 Aux ports (n.a.)
14
Socket for terminal connection cable
15
Fastening for terminal stand
16
Fan
17
RS232C (for production only)
18
Dust filter (optional)
19
Kensington slot for anti-theft purposes
20
Power supply socket
21
Power line fuse (plus replacement fuse)
Design and Function10 Moisture Analyzer
3.1.2 Overview terminal
Legend terminal
1
Touch screen
2
USB Host port 1
3
Foot
4
USB Host port 2
5
USB Device port
6
RS232C port
7
Ethernet RJ45 port
8
Socket for terminal connection cable
9
SD Memory Card slot
3.1.3 Overview type plates
Drying unit type plate
The type plate of the drying unit is located on the side of the drying unit and contains the following information:
5
1
4
6
7
2 3
1
Model designation
2
Readability
3
Maximum capacity
4
Manufacturer
5
Power supply
6
Year of manufacture
7
Serial number (SNR)
Design and Function 11Moisture Analyzer
Terminal type plate
The type plate of the terminal is located on the terminal and contains the following information:
1
2
1
Model designation
2
Serial number (SNR)
3.2 Compliance Information
European Union
The instrument complies with the directives and standards listed on the EU Declaration of Conformity.
United States of America
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful inter-
ference when the equipment is operated in a commercial environment. This equipment generates, uses, and
can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual,
may cause harmful interference to radio communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be required to correct the interference at his own
expense.
Changes or modifications not expressly approved by the party responsible for compliance could void the user's
authority to operate the equipment.
The FCC Supplier Declaration of Conformity is available online.
u http://www.mt.com/ComplianceSearch
3.3 Function description
METTLER TOLEDO Moisture Analyzers work on the thermogravimetric principle. At the start of the measurement
the Moisture Analyzer determines the weight of the sample, the sample is then quickly heated by the absorption
of the emitted infrared radiation. During the drying process the instrument continually measures the weight of
the sample and displays the weight reduction, up to the final result.
120 °
C
With halogen heating technology, maximum heating power is quickly reached and allows use of high temper-
atures. Uniform heating of the sample material ensures repeatability of the drying results and makes it possible
to use small amounts of sample.
The METTLER TOLEDO Moisture portfolio comprises a range of moisture analyzers that differ from each other in
both hardware and software.
The following features are available for the HX204 model:
Accurate results for all sample types and repeatable moisture results thanks to 0.001% moisture content
readability.
Design and Function12 Moisture Analyzer
Increased safety and process control: management of multiple users with individual access rights and
automatic lock-out option. See [User management}Page108]
Increased efficiency and adherence to SOPs and regulations with Zero results option.
Built-in testing methods to ensure performance: FACT, SmartCal and tests for the heating and weighing unit.
See [Test/Adjust}Page76]
Reduced measurement time and quick decision making with the embedded QuickPredict
TM
function. See
[Working with QuickPredict}Page65]
Integrated method development support. See [Creating a method}Page50]
Quick and easy cleaning. See [Cleaning}Page116]
Network connectivity. See [Electrical interfaces}Page21] and [Network interfaces}Page94]
Level control. See [Leveling the instrument}Page16]
Simple system integration and automation.
Tailor-made reporting. See [Exporting results}Page73]
ID management. See [Identifications management}Page98]
Control limits. See [Control limits}Page44]
Installation and Putting into Operation 13Moisture Analyzer
4 Installation and Putting into Operation
4.1 Selecting the location
Requirements of the location
Place indoors on stable
table
Ensure sufficient spacing Level the instrument Provide adequate lighting
Avoid direct sunlight Avoid vibrations Avoid strong drafts Avoid temperature fluctu-
ations
Sufficient distance for Moisture Analyzers: > 15 cm next to the instrument, > 1 m above the heating module.
4.2 Unpacking
Note
Store all parts of the packaging. This packaging guarantees the best possible protection for the transport of your
instrument.
After receiving the instrument, proceed as follows:
1 Open the package and remove drying unit, terminal and the accessories.
2 Remove the packaging from the instrument.
3 Check the Moisture Analyzer for transport damage and complete scope of delivery.
4 In the event of damage or missing accessories, immediately inform a METTLER TOLEDO representative.
4.3 Scope of delivery
Moisture Analyzer Documentation Accessories
1 Drying unit
1 Draft shield
1 Power cable
1 Terminal cable
1 Sample handler
1 Sample pan holder
1 User Manual
1 Application brochure «Guide
to Moisture Analysis»
1 Production certificate
1 Voucher for E-learning "Proper
Moisture Determination"
EU-countries: 1 CE declaration
of conformity
80 aluminium sample pans
3 Specimen samples (glass
fibre filter)
1 SmartCal sampler
Specific countries: 1 Wifi dongle
Installation and Putting into Operation14 Moisture Analyzer
4.4 Connecting the instrument
WARNING
Death or serious injury due to electric shock
Contact with parts that carry a live current can lead to injury and death. If the instrument
cannot be shut down in an emergency situation, people can be injured and the instrument
can be damaged.
1 Check that the voltage printed on the instrument is the same as your local power supply
voltage. If this is not the case, under no circumstances connect the instrument to the
power supply, but contact a METTLER TOLEDO representative.
2 Only use the three-core power cable with equipment grounding conductor supplied by
METTLER TOLEDO to connect your instrument.
3 Only connect it to a three-pin power socket with earthing contact.
4 Only use standardized extension cables with equipment grounding conductor for
operating the instrument.
5 Make sure that the power plug is accessible at all times.
6 Arrange the cables so that they cannot be damaged or interfere with the operation.
7 Keep all electrical cables and connections away from liquids.
Two different versions of drying units with country-specific power cable are available:
110VAC or 230VAC
Connecting the terminal to the drying unit
§
Drying unit and terminal are at the final location.
1 Connect one end of the supplied terminal cable to the system connection socket on the terminal.
2 Screw the connector firmly.
3 Connect the other end to the system connection socket on the drying unit.
4 Screw the connector firmly.
Connecting the drying unit to the power supply
§
The terminal is connected to the drying unit.
1 Connect one end of the power cable to the power supply socket on the drying unit.
2 Connect the other end to the power line outlet.
See also
2 Overview terminal}Page10
2 Overview drying unit}Page9
4.5 Opening and closing sample chamber
NOTICE
Damage to the instrument due to manual opening
Opening the sample chamber manually can damage the opening function and hinges of the
instrument.
1 In ordinary situations, always use the key [
] to open and close the sample chamber
2 Only use the opening handles on the heating module in case of an emergency.
Regular opening
In ordinary situations, open and close the sample chamber by pressing the key [
].
Installation and Putting into Operation 15Moisture Analyzer
Emergency opening
In case of an emergency:
1 Grab the handles on the heating module with both hands and pull the module firmly upwards to open the
sample chamber.
2 After a manual emergency opening, always reboot the system and check for damage to the instrument or
opening function.
4.6 Setup wizard
When you switch on your Moisture Analyzer for the first time or when a factory reset was performed, the Setup
Wizard appears. The Setup Wizard is a step-by-step instruction, that helps to define the basic functions
(Language and Date format) and to install the sample chamber of your device. The buttons in the action bar
can be used for navigation.
It must be successfully finished. If it is not finished, the device is not operational and the function will reappear
when the device is switched on the next time.
After the application Setup Wizard is successfully finished, the settings can be changed via Settings of the user
home screen.
Note
The application Setup Wizard appears automatically when the instrument was switched on for the first time (or
when a factory reset was performed). It must be successfully finished. If it is not finished, the device is not
operational and the function will reappear when the device is switched on the next time.
After the application Setup Wizard is successfully finished, the settings can be changed via Settings of the user
home screen.
Note
The application Instrument tutorial starts automatically after the application Setup Wizard has succesfully
finished. This option can be skipped by inactivating the checkbox Proceed to the Instrument Tutorial. If you
want to redo the application Instrument tutorial, it can be found in the section Settings.
4.7 Setting up the instrument
§
Terminal and drying unit are connected.
§
The instrument is connected to the power supply.
1 Switch on the Instrument by pressing [
].
2 Press [
].
ð The motorized automatic sample chamber opens.
3 Place the draft shield element. Only one position is possible.
Installation and Putting into Operation16 Moisture Analyzer
4 Insert the sample pan holder carefully. Check that the
sample pan holder is hooked correctly.
5 Insert the sample pan handler.
4.8 Leveling the instrument
Exact horizontal positioning and stable installation are prerequisites for repeatable and accurate results. To
compensate for small irregularities or inclinations (±2%) at the location, the instrument must be leveled.
Note
For leveling the drying unit use the function Leveling guide that can be found on Settings > Help and tutorials
> Instrument tutorial > 1. Leveling the instrument.
Installation and Putting into Operation 17Moisture Analyzer
For exact horizontal positioning, the drying unit has a level indicator
(level) and 2 leveling screws. When the air bubble in the level
indicator is exactly in the center, the instrument is standing perfectly
horizontally. To level it, proceed as follows:
1 Position your Moisture Analyzer at the selected location.
2 Turn the two leveling feet until the air bubble is in the center of the
level indicator.
L = left foot
R = right foot
Air bubble at 12
o'clock:
turn both feet clockwise.
Air bubble at 3
o'clock:
turn left foot clockwise, right foot counter-
clockwise.
Air bubble at 6
o'clock:
turn both feet counterclockwise.
Air bubble at 9
o'clock:
turn left foot counterclockwise, right foot
clockwise.
Leveling the drying unit with optional stand for terminal mounted.
§
The drying unit is positioned at the selected location.
1 To loose the clamps for the safety feet of the stand, turn them
outwards.
2 Level the drying unit as previously described.
3 To lock the clamps for the safety feet, turn them inwards as far as
they will go.
Level sensor
The built-in level sensor constantly monitors the correct horizontal alignment of the drying unit. When the level
is incorrect, the status icon
appears in the display, see Status Message Icons.
Note
The level sensor cannot be switched off.
Installation and Putting into Operation18 Moisture Analyzer
4.9 Anti-theft device
For protection against theft, your drying unit is fitted
with a Kensington security slot for an anti-theft device.
To find a suitable anti-theft cable see [Accessories
and Spare Parts}Page127].
4.10 Setting reading angle of terminal
To change the reading angle, fold out the two tilt feet.
4.11 Setting date and time
Navigation: Home > Settings > Instrument settings > Regional Settings
When you put your new instrument into operation for the first time, you should enter the current date and time.
These settings are retained even if you disconnect your instrument from the power supply.
Note
Various languages are available for the dialog with your instrument and can be selected in the menu:
Navigation: Settings > User preferences.
Set the current date
§
Regional Settings is selected.
1 Tap Date.
2 Set day, month and year.
3 Confirm with [Set date].
Set the current time
§
Regional Settings is selected.
1 Tap Time.
2 Set Hours and minutes.
3 Confirm with [Set time].
4.12 Adjusting after setup
To obtain accurate measuring results, it is necessary to adjust the integrated balance as well as the heating
module.
Adjusting is necessary:
before the instrument is used for the first time.
at regular intervals.
after a change of location.
The following adjustment options are possible:
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146

Mettler Toledo HX204 Halogen Moisture Analyzer Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI