GasWatch 3 for CO(-H2) Chapter 3: Measuring Mode • 15
3. If CAL.RMDR is set to ON (factory setting) and a calibration is due, the screen that
appears next depends on how CAL.EXPD is set in User Mode (see pg.52). The three
possible screens are described below. If a calibration is not due, the instrument displays
the number of days left until a calibration is due.
CAL.EXPD set to
CONFIRM (factory setting)
CAL.EXPD set to
CANT.USE
CAL.EXPD set to
NONE
LCD
Sound Buzzer sounds double pulsing
tone
Buzzer sounds double pulsing
tone
None
Action • Option A, Perform calibration
:
Press and release POWER/
MODE to enter User Mode
and perform a calibration. The
instrument takes you straight
to the calibration start screen
in User Mode’s GAS CAL\A-
CAL(E-CAL) item. See pg.41
for calibration instructions.
NOTE: If Password Protection
is set ton On (factory setting is
Off) using the GW-3 Config Pro-
gram, you will be prompted for a
password to enter User Mode.
If the calibration is success-
ful, the screen above will not
appear again until the unit is
due for calibration. If the cali-
bration is not successful, the
screen above will again appear
in the startup sequence.
• Option B, Bypass message
: To
continue without performing a
calibration, press and release
AIR.
The GasWatch 3 cannot be used
until a successful calibration is
performed. Press and release
POWER/MODE to enter User
Mode and perform a calibration.
The instrument takes you straight
to the calibration start screen in
User Mode’s GAS CAL/A-CAL
(E-CAL) item. If you don’t press
POWER/MODE, the instrument
automatically goes to the calibra-
tion start screen after 6 seconds.
See pg.41 for calibration instruc-
tions.
NOTE: If Password Protection is
set ton On (factory setting is Off)
using the GW-3 Config Program,
you will be prompted for a pass-
word to enter User Mode.
If the calibration is successful,
the screen above will not appear
again until the unit is due for cali-
bration. If the calibration is not
successful, the screen above will
again appear in the startup
sequence.
• Option A, Perform calibration
:
If you want to enter User
Mode and perform a
calibration, press and release
POWER/MODE. The
instrument takes you straight
to the calibration start screen
in User Mode’s GAS CAL/A-
CAL (E-CAL) item.
• Option B, Bypass message
:
To continue without
performing a calibration, wait
a few seconds for the
instrument to continue with its
startup sequence.