4.0 Warnings and Special Precautions
Breath coming from the lungs must not be contaminated
with smoke particles or "mouth alcohol" from a recent
drink, regurgitation, mouthwashes, breath sprays or
medications; which would cause false high readings.
CAUTION: If a subject has been smoking, wait 3 minutes
before conducting a test.
CAUTION: Ensure that a test subject has taken nothing
by mouth for at least 15 minutes or rinses the mouth
cavity with water prior to a breath test.
5.0 Error Codes
The microcontroller of the ALERT J4X.ec automatically
conducts a system diagnostic check when the unit is
turned "On". In addition it monitors the status of several
functions during the test procedure. If any faults are de-
tected an "Err" code appears on the display and after 10
seconds the unit powers down.
Err 1 Low battery indication. Test should be discontin-
ued and unit should be placed on charge. Alternatively,
when using the AC power adapter it signifies a supply
voltage below 3.0 volts.
Err 2 AC power adapter voltage too high. Return the AC
power adapter to ACS for service.
Err 3 Lithium battery voltage low. Return the ALERT
J4X.ec to ACS for service (battery replacement).
Err 4 Baseline signal too high. Return the ALERT J4X.ec
to ACS for service (sensor may need replacement).
Err 5 Baseline signal too low. Return the ALERT J4X.ec
to ACS for service (sensor may need replacement).
Err 6 Sensor heater open. Return the ALERT J4X.ec to
ACS for service.
Err 7 Test memory log full. Connect ALERT J4X.ec to PC
and run program to read and clear log.
Err 8 Calibration range low. (Sensor may need replace-
ment.)
Err 9 Calibration range high.
Err 10 Calibration recall. It has been 12 months or 1,000
tests since the last calibration. Perform calibration.
Err 11 Pressure signal too low.
Err 12 Pressure signal to high.
Er 13 Verification test result is too low.
Er 14 Verification result is too high.
Er 15 Verification tests were low and calibration lockout
is initated.
Er 16 Verification tests were high and calibration lockout
is initiated.
Er 17 The maximum number of days since last calibra-
tion has been exceeded.
Er 18 The maximum number of tests since last calibra-
tion has been exceeded.
6.0 Time Between Tests
The ALERT J4X.ec unit uses an electro-chemical (fuel cell)
sensor. While being very accurate, specific to alcohol and
power efficient this type of sensor requires a period of time
to recover from exposure to ethanol. The following
guidelines should be observed to ensure accurate tests are
obtained.
The following chart provides the recommended waiting
time before another test is conducted after a positive
alcohol test result is obtained.