Endres+Hauser Liquisys M COM223/253 Operating instructions

Category
Measuring & layout tools
Type
Operating instructions
BA199C/07/en/09.05
51500272
valid as of:
Software-version 2.40
Operating Instructions
Liquisys M COM223/253
Transmitter for Dissolved Oxygen
Endress+Hauser
Brief operating instructions
This explains how to use these Operating Instructions to commission your transmitter quickly and
safely:
Safety instructions
Page 5 ff.
Page 6 ff.
General safety instructions
Explanation of the warning symbols
You can find special instructions at the appropriate position in the chapter in question. The
positions are indicated with the icons Warning #, Caution " and Note !
â–¼
Installation
Page 10 ff.
Page 12 ff.
Here you can find information on installation conditions and the dimensions of the transmitter.
You can find information on how to install the transmitter on these pages.
â–¼
Wiring
Page 16 ff. Here you can find out how to connect sensors to the transmitter.
â–¼
Operation
Page 29 ff.
Page 33 ff.
Page 40 ff.
Page 67 ff.
The display and operating elements are described here.
The operating concept is explained here.
The system configuration is explained here.
You can find information on how to calibrate the sensor on these pages.
â–¼
Maintenance
Page 71 ff.
Page 75 ff.
Page 79 ff.
Page 86 ff.
Here you can find information on the maintenance of the entire measuring point.
The accessories which can be delivered for the transmitter are listed on the pages indicated.
Here you can find information on trouble-shooting.
Here you can find an overview of the spare parts which can be delivered as well as an overview
of the system.
â–¼
Technical data
Page 10 ff.
Page 93 ff.
Dimensions
Ambient and process conditions, weight, materials etc.
â–¼
Appendix
Page 98 ff. Here you can find the operating matrix
Liquisys M COM223/253
Endress+Hauser 3
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 5
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2 Installation, commissioning and operation . . . . . . . . 5
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
1.5 Notes on safety icons and symbols . . . . . . . . . . . . . . 6
2 Identification . . . . . . . . . . . . . . . . . . . . 7
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.1 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.1.2 Product structure . . . . . . . . . . . . . . . . . . . . 7
2.1.3 Additional functions of the Plus Package . . . 8
2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.3 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 8
3 Installation . . . . . . . . . . . . . . . . . . . . . . 9
3.1 Quick installation guide . . . . . . . . . . . . . . . . . . . . . . 9
3.1.1 Measuring system . . . . . . . . . . . . . . . . . . . . 9
3.2 Incoming acceptance, transport, storage . . . . . . . . . 10
3.3 Installation conditions . . . . . . . . . . . . . . . . . . . . . . 10
3.3.1 Field instrument . . . . . . . . . . . . . . . . . . . . 10
3.3.2 Panel-mounted instrument . . . . . . . . . . . . 11
3.4 Installation instructions . . . . . . . . . . . . . . . . . . . . . 12
3.4.1 Field instrument . . . . . . . . . . . . . . . . . . . . 12
3.4.2 Panel-mounted instrument . . . . . . . . . . . . 15
3.5 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 15
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 16
4.1 Electrical connection . . . . . . . . . . . . . . . . . . . . . . 16
4.1.1 Electrical connection variant 1
(DX/DS with COS41) . . . . . . . . . . . . . . . . 17
4.1.2 Measuring cable and sensor connection . . . 19
4.1.3 Electrical connection variant 2
(WX/WS with COS31/71 or COS61) . . . . 21
4.1.4 Measuring cable and sensor connection . . . 23
4.1.5 Electrical connection variant 2
(WX/WS with COS61) . . . . . . . . . . . . . . . 24
4.1.6 Measuring cable and sensor connection . . . 26
4.1.7 Alarm contact . . . . . . . . . . . . . . . . . . . . . . 28
4.2 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 28
5 Operation . . . . . . . . . . . . . . . . . . . . . . 29
5.1 Quick operation guide . . . . . . . . . . . . . . . . . . . . . . 29
5.2 Display and operating elements . . . . . . . . . . . . . . . 29
5.2.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.2.2 Operating elements . . . . . . . . . . . . . . . . . . 30
5.2.3 Key assignment . . . . . . . . . . . . . . . . . . . . . 31
5.3 Local Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.3.1 Automatic/manual mode . . . . . . . . . . . . . 33
5.3.2 Operating concept . . . . . . . . . . . . . . . . . . . 34
6 Commissioning . . . . . . . . . . . . . . . . . . 36
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.2 Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
6.3 Quick start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
6.4 System configuration . . . . . . . . . . . . . . . . . . . . . . . 40
6.4.1 Setup 1 (Oxygen) . . . . . . . . . . . . . . . . . . . . 40
6.4.2 Setup 2 (Salinity and temperature) . . . . . . . 41
6.4.3 Current input . . . . . . . . . . . . . . . . . . . . . . 41
6.4.4 Current outputs . . . . . . . . . . . . . . . . . . . . . 44
6.4.5 Monitoring functions . . . . . . . . . . . . . . . . . 49
6.4.6 Relay contact configuration . . . . . . . . . . . . 53
6.4.7 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
6.4.8 E+H Service . . . . . . . . . . . . . . . . . . . . . . . 66
6.4.9 Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.5 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
6.6 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
7 Maintenance . . . . . . . . . . . . . . . . . . . . 71
7.1 Maintenance of the entire measuring point . . . . . . . 71
7.1.1 Cleaning the transmitter . . . . . . . . . . . . . . 71
7.1.2 Test variant 1 (DX/DS with COS41) . . . . . 72
7.1.3 Test variant 2 (WX/WS with COS31 or
COS71) . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
7.1.4 Maintenance oxygen sensors . . . . . . . . . . . 73
7.1.5 Maintenance assembly . . . . . . . . . . . . . . . . 73
7.1.6 Connecting lines and junction boxes . . . . . 74
7.2 "Optoscope" service tool . . . . . . . . . . . . . . . . . . . . 74
8 Accessories . . . . . . . . . . . . . . . . . . . . . 75
8.1 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
8.2 Connection accessories . . . . . . . . . . . . . . . . . . . . . 75
8.3 Mounting accessories . . . . . . . . . . . . . . . . . . . . . . . 76
8.4 Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
8.5 Software and hardware add-ons . . . . . . . . . . . . . . . 78
8.6 Optoscope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
9 Trouble-shooting . . . . . . . . . . . . . . . . . 79
9.1 Trouble-shooting instructions . . . . . . . . . . . . . . . . . 79
9.2 System error messages . . . . . . . . . . . . . . . . . . . . . . 79
9.3 Process specific errors . . . . . . . . . . . . . . . . . . . . . . 81
9.4 Instrument specific errors . . . . . . . . . . . . . . . . . . . . 84
9.5 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
9.5.1 Dismantling the panel mounted instrument 86
9.5.2 Panel mounted instrument . . . . . . . . . . . . . 87
9.5.3 Dismantling the field instrument . . . . . . . . 89
9.5.4 Field instrument . . . . . . . . . . . . . . . . . . . . 90
9.5.5 Replacing the central module . . . . . . . . . . . 92
9.6 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
9.7 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
Liquisys M COM223/253
4Endress+Hauser
10 Technical data . . . . . . . . . . . . . . . . . . . 93
10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
10.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
10.4 Performance characteristics . . . . . . . . . . . . . . . . . . 95
10.5 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
10.6 Mechanical construction . . . . . . . . . . . . . . . . . . . . 96
10.7 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
11 Appendix. . . . . . . . . . . . . . . . . . . . . . . 98
Index. . . . . . . . . . . . . . . . . . . . . . . . . 110
Liquisys M COM223/253 Safety instructions
Endress+Hauser 5
1 Safety instructions
1.1 Designated use
Liquisys M is a transmitter for determining the oxygen concentration of a liquid medium.
The transmitter is particularly suited for use in the following areas:
• Sewage treatment plants
• Sewage treatment
•Drinking water
• Water conditioning and monitoring
• Surface water (rivers, lakes, oceans)
• Fish farming
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is, therefore, not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.
1.2 Installation, commissioning and operation
Please note the following items:
• Installation, electrical connection, commissioning, operation and maintenance of the measuring
system must only be carried out by trained technical personnel.
The technical personnel must be authorised for the specified activities by the system operator.
• Technical personnel must have read and understood these Operating Instructions and must
adhere to them.
• Before commissioning the entire measuring point, check all the connections for
correctness. Ensure that electrical cables and hose connections are not damaged.
• Do not operate damaged products and secure them against unintentional
commissioning. Mark the damaged product as being defective.
• Measuring point faults may only be rectified by authorised and specially trained
personnel.
• If faults can not be rectified, the products must be taken out of service and secured against
unintentional commissioning.
• Repairs not described in these Operating Instructions may only be carried out at the
manufacturer’s or by the service organisation.
1.3 Operational safety
The transmitter has been designed and tested according to the state of the art and left the factory in
perfect functioning order.
Relevant regulations and European standards have been met.
As the user, you are responsible for complying with the following safety conditions:
• Installation instructions
• Local prevailing standards and regulations.
Ex systems have an additional Ex documentation which is part of these Operating Instructions (see
also chapter "Scope of delivery").
EMC
This instrument has been tested for electromagnetic compatibility in industrial use according to
applicable European standards.
Protection against interference as specified above is valid only for an instrument
connected according to the instructions in these Operating Instructions.
Safety instructions Liquisys M COM223/253
6Endress+Hauser
1.4 Return
If the device requires repair, please send it cleaned to the sales centre responsible. Please use the
original packaging, if possible.
1.5 Notes on safety icons and symbols
Safety icons
Electrical symbols
#
Warning!
This symbol alerts you to hazards. They can cause serious damage to the instrument or to persons
if ignored.
"
Caution!
This symbol alerts you to possible faults which could arise from incorrect operation. They could
cause damage to the instrument if ignored.
!Note!
This symbol indicates important items of information.
%Direct Current (DC)
A terminal at which DC is applied or through which DC flows.
&Alternating Current (AC)
A terminal at which (sine-form) AC is applied or through which AC flows.
)Ground connecting
A terminal, which, from the user’s point of view, is already grounded using a grounding system.
*Protective earth terminal
A terminal which must be grounded before other connections may be set up.
bAlarm relay
Input
Output
DC voltage source
Temperature sensor
Liquisys M COM223/253 Identification
Endress+Hauser 7
2 Identification
2.1 Device designation
2.1.1 Nameplate
Compare the order code on the nameplate (on the transmitter) with the product structure (s.b.) and
your order.
The device version can be identified from the order code.
!Note!
The release codes for retrofitting the software for Chemoclean (to the left of the forward slash) or
the Plus Package (to the right of the forward slash) are listed under "Codes".
2.1.2 Product structure
C07-COM2x3xx-18-06-00-xx-001.eps
Fig. 1: COM253 nameplate (example)
C07-COM2x3xx-18-06-00-xx-002.eps
Fig. 2: COM223 nameplate (example)
Version
DX Sensor COS41 / 4 / 4HD, basic functions
DS Sensor COS41 / 4 / 4HD, with additional functions (Plus package)
WX Sensor COS31 / 61 / 71 / 3 / 3HD, basic functions
WS Sensor COS31 / 61 / 71 / 3 / 3HD, with additional functions (Plus package)
Power supply; approval
0230VAC
1115VAC
2 230 V AC; CSA Gen. Purp.
3 115 V AC; CSA Gen. Purp.
4 230 V AC; ATEX II 3G [EEx nAL] IIC
5100VAC
6 24 V AC/DC; ATEX II 3G [EEx nAL] IIC for COM223, EEx nA[L] IIC T4 for COM253
7 24 V AC/DC; CSA Gen. Purp.
824VAC/DC
Output
0 1 x 20 mA, dissolved oxygen
1 2 x 20 mA, dissolved oxygen and temperature/actuating variable
3PROFIBUS PA
4PROFIBUS DP
5 1 x 20 mA, dissolved oxygen with HART®
6 2 x 20 mA, dissolved oxygen with HART® and temperature/actuating variable
Additional contacts; analogue input
05 Not selected
10 2 x relay (limit/controller/timer)
15 4 x relay (limit/controller/Chemoclean)
16 4 x relay (limit/controller/timer)
20 2 x relay (limit/controller/timer); current input
25 4 x relay with cleaning (limit/controller/timer/Chemoclean); current input
26 4 x relay with timer (limit/controller/timer); current input
COM253-
complete order code
COM223-
131085-4D
order code
serial no
.
/
codes
COM 253-WS0116
4A234505G00
0 ... 20 mg/l
0 ... 200 %SAT
-10 … 60°C
0/4 … 20 mA
230 VAC 50/60 Hz 7.5 VA
-10 … +55°CIP 65
0/4…20mA
-
8732
prot. class ambient temp
.
output 1
output 2
mains
meas. range
temperature
3472
LIQUISYS M
diss. oxygen
Made in Germany, D-70839 Gerlingen
131085-4D
order code
serial no
.
/
codes
COM 223-WS0116
4A234505G00
0 ... 20 mg/l 0 ... 200 %SAT
-10 … 60°C
0/4 … 20 mA
230 VAC 50/60 Hz 7.5 VA
-10 … +55°CIP 54/ IP 30
0/4 … 20 mA
-
8732
3472
prot. class ambient temp.
output 1 output 2
mains
meas. range
temperature
LIQUISYS M
diss. oxygen
Made in Germany, D-70839 Gerlingen
Identification Liquisys M COM223/253
8Endress+Hauser
2.1.3 Additional functions of the Plus Package
• Current output table to cover large areas with varying resolution, fields O23x
• Process Check System (PCS): live check of the sensor, function group P
• Automatic cleaning function start, field F8
• At version DS: air pressure measurement
2.2 Scope of delivery
The delivery of the field instrument includes:
• 1 transmitter COM253
•1 plug-in screw terminal
• 1 cable gland Pg 7
• 1 cable gland Pg 16 reduced
• 2 cable glands Pg 13.5
• 1 operating instructions BA 199C/07/en
• versions with HART communication:
1 operating instructions Field Communication with HART, BA 208C/07/en
• versions with PROFIBUS communication:
1 operating instructions Field Communication with PROFIBUS PA/DP, BA 209C/07/en
• versions with explosion protection for hazardous area zone 2 (ATEX II 3G):
Safety instructions for use in explosion-hazardous areas, XA 194C/07/a3
The delivery of the panel mounted instrument includes:
• 1 transmitter COM223
• 1 set of plug-in screw terminals
• 2 tensioning screws
• 1 BNC-plug (solder-free)
• 1 operating instructions BA 199C/07/en
• versions with HART communication:
1 operating instructions Field Communication with HART, BA 208C/07/en
• versions with PROFIBUS communication:
1 operating instructions Field Communication with PROFIBUS PA/DP, BA 209C/07/en
• versions with explosion protection for hazardous area zone 2 (ATEX II 3G):
Safety instructions for use in explosion-hazardous areas, XA 194C/07/a3
If you have any questions, please contact your supplier or your sales centre responsible.
2.3 Certificates and approvals
Declaration of conformity
The product meets the legal requirements of the harmonised European standards.
The manufacturer confirms compliance with the standards by affixing the 4 symbol.
Explosion protection for Zone 2
Version Approval
COM253-..6... ATEX II 3G EEx nA[L] IIC T4
COM253-..4...
COM223-..4...
COM223-..6...
ATEX II 3G [EEx nAL] IIC
Liquisys M COM223/253 Installation
Endress+Hauser 9
3 Installation
3.1 Quick installation guide
#Warning!
If the measuring point or parts of the measuring point are in explosion-hazardous areas you have to
follow the "Safety instructions for electrical apparatus certified for use in explosion-hazardous
areas". These instructions (XA 194C/07/a3) are part of the scope of delivery.
Proceed as follows to completely install the measuring point:
• Install the transmitter (see "Installation instructions" section).
• If the sensor is not yet installed in the measuring point, install it (see Technical Information of the
sensor).
• Connect the sensor to the transmitter as illustrated in the "Electrical connection" section.
• Connect the transmitter as illustrated in the "Electrical connection" section.
• Commission the transmitter as explained in the "Commissioning" section.
3.1.1 Measuring system
A complete measuring system comprises:
Variant 1 (DX/DS with COS41)
• The transmitter Liquisys M COM223 or COM253 in version DX or DS
• An oxygen sensor COS41
• An immersion, flow or retractable assembly
Options: extension cable CMK, junction box VBM
Variant 2 (WX/WS with COS31, COS61 or COS71)
• The transmitter Liquisys M COM223 or COM253 in version WX or WS
• An oxygen sensor COS31, COS61 or COS71
• An immersion, flow or retractable assembly
Options: extension cable OMK, junction box VS
C07-COM2x3xx-14-06-00-xx-002.eps
Fig. 3: Complete measuring system Liquisys M COM223/253
1
2
3
4
Oxygen sensor
Immersion assembly CYA611
Universal hanging assembly holder CYH101
Weather protection cover CYY101
5
6
7
Liquisys M COM253
Liquisys M COM223
Retractable assembly COA451
CAL REL
REL1
REL1
ALARM
REL2
REL2
E
CAL REL
REL1
REL1
ALARM
REL2
REL2
E
1
2
35
6
7
4
Installation Liquisys M COM223/253
10 Endress+Hauser
3.2 Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging.
Keep the damaged packaging until the matter has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents.
Keep the damaged products until the matter has been settled.
• Check that the scope of delivery is complete and agrees with your order and the
shipping documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your sales centre responsible.
3.3 Installation conditions
3.3.1 Field instrument
C07-CxM253xx-06-06-00-en-001.EPS
Fig. 4: Field instrument
!Note!
There is a hole in the punching for the cable entry (connection of supply voltage). It serves as a
pressure balance during air freight dispatching. Make sure no moisture penetrates the inside of the
housing before the cable installation. The housing is completely air-tight after the cable installation.
CAL REL
REL 1
REL 1
ALARM
REL 2
REL 2
E
Pg 7
Pg 16
M 5
11/
0.43
Ø 6 /
Ø 0.24
115 / 4.53
157 / 6.18
247 / 9.72
70 / 2.76
70 / 2.76
154 / 6.06
170 / 6.69
mm / inch
Pg 13.5
Liquisys M COM223/253 Installation
Endress+Hauser 11
3.3.2 Panel-mounted instrument
C07-CxM223xx-06-06-00-en-001.EPS
Fig. 6: Panel-mounted instrument
C07-CxM253xx-11-06-00-xx-001.eps
Fig. 5: View into the field housing
1
2
3
4
Removable electronics box
Partition plate
Terminals
Fuse
1
2
3
4
90 / 3.54
139 / 5.47
96 / 3.78
6 /
0.24
mm / inch
92 / 3.62
149.5 / 5.89
Installation Liquisys M COM223/253
12 Endress+Hauser
3.4 Installation instructions
3.4.1 Field instrument
There are several ways of securing the field housing:
• Wall mounting with fixing screws
• Post mounting to cylindrical pipes
• Post mounting to square securing mast
!Note!
When mounting in the open air with unprotected exposure to weather conditions, a weather
protection cover (see Accessories) is required.
Transmitter wall mounting
C07-CxM253-11-06-00-en-002.eps
Fig. 7: Field instrument wall mounting
For wall mounting the transmitter, proceed as follows:
1. Drill the bores as shown in Fig. 7 .
2. Drive the two fixing screws through the securing bores (1) from the front.
3. Mount the transmitter on the wall as shown.
4. Cover the bores with plastic caps (2).
1
2
154 / 6.06
mm / inch
Ø 6 /
Ø 0.24
Liquisys M COM223/253 Installation
Endress+Hauser 13
Transmitter post mounting
!Note!
You require a post mounting kit to secure the field device to horizontal and vertical posts or pipes
(max. Ø 60 mm(2.36")). This can be acquired as an accessory (see "Accessories" section).
C07-CxM253xx-11-06-00-en-003.eps
Fig. 8: Post mounting field device to cylindrical pipes
For post mounting the transmitter, proceed as follows:
1. Guide the two securing screws (1) of the mounting kit through the openings of the securing
plate (3).
2. Screw the securing plate onto the transmitter using the four fixing screws (2).
3. Secure the retainer with the field device using the clip on the post or pipe.
1
2
3
1
1
3
Ømax. 60 /
2.36
Installation Liquisys M COM223/253
14 Endress+Hauser
You can also secure the field device to a square universal post in conjunction with the weather
protection cover. These can be acquired as accessories, see "Accessories" section.
C07-CxM253xx-11-06-00-xx-004.eps
Fig. 9: Mounting field device with universal posts and weather protection cover
For mounting the weather protection cover, proceed as follows:
1. Screw the weather protection cover with 2 screws (bores 1) to the upright post (bores 2).
2. Secure the field device to the weather protection cover. To do so, use the bores (3).
3
1
2
3
Liquisys M COM223/253 Installation
Endress+Hauser 15
3.4.2 Panel-mounted instrument
The panel-mounted instrument is secured with the clamping screws supplied (see →Fig. 10).
The necessary installation depth is approx. 165 mm (6.50").
C07-CxM223xx-11-06-00-en-001.eps
Fig. 10: Securing the panel-mounted instrument
1 Wall of the cabinet
2Seal
3 Clamping screws
* Required installation depth
3.5 Post-installation check
• After installation, check the transmitter for damage.
• Check whether the transmitter is protected against moisture and direct sunlight.
1
2
3
3
3
92 / 3.62
+0.5 +0.02
92 / 3.62
+0.5 +0.02
90 / 3.54
57 / 2.24
33 /
1.30
approx. 25 /
0.98
139 / 5.47
max. 45 /
1.77 6 /
0.24
96 / 3.78
92 /
3.62
approx. 165 / 6.50*
Wiring Liquisys M COM223/253
16 Endress+Hauser
4 Wiring
4.1 Electrical connection
#Warning!
• The electrical connection must only be carried out by authorised technical personnel.
• Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
• Ensure that there is no voltage at the power cable before beginning the connection work.
The electrical connection of the transmitter differs depending on the device version:
• If you are using the device version DX/DS (with COS41), please read the instructions in the
"Electrical connection variant 1" section.
• If you are using the device version WX/WS (with COS31, COS61 or COS71), please read the
instructions in the "Electrical connection variant 2" section.
Liquisys M COM223/253 Wiring
Endress+Hauser 17
4.1.1 Electrical connection variant 1 (DX/DS with COS41)
The wiring diagram depicted in Fig. 11 shows the connections of the transmitter with all the
options.
!Note!
The device is approved for protection class II and is generally operated without a protective ground
connection.
The circuits "C" and "F" are not galvanically isolated from each other.
C07-COM2x3xx-04-06-00-xx-001.eps
Fig. 11: Electrical connection of the transmitter in version DX or DS
A
B
C
D
E
F
Oxygen sensor COS41
Signal output 1 oxygen
Signal output 2 temperature/controller
Binary input 1 (Hold)
Binary input 2 (Chemoclean)
Aux. voltage output
G
H
I
J
K
L
M
Alarm (current-free contact position)
Relay 1 (current-free contact position)
Relay 2 (current-free contact position)
Relay 3 (current-free contact position)
Relay 4 (current-free contact position)
Current input 4 ... 20 mA
Power supply
* Aux. voltage output terminal 85/86 applicable
47
48
49
57
58
59
41
42
43
15 V
85
86
10-50 V
81
93
82
94
10-50 V
mA
31
32
mA
33
34
51
52
53
54
55
56
90
12
11
91
S
O2
O2
K
A
NTC
GN
RD
BN
23 +
24 -
mA
L+
L–N
L1
~ =
A
B
C
D
E
F
G
H
I
J
K
L
M
K
A
*
*
WH
Wiring Liquisys M COM223/253
18 Endress+Hauser
Field instrument connection version DX/DS
To connect the field instrument proceed as follows:
1. Open the front cover to get access to the terminals.
2. Cut out the marked PG gland hole of the housing. Mount a PG gland an guide the measuring
cable through the PG gland into the housing.
3. Connect the measuring cable in accordance with the connection diagram.
4. Fix the PG gland.
"Caution!
• Always protect plugs, terminals and cables against humidity.
• Terminals marked NC may not be wired.
• Unmarked terminals may not be wired.
C07-COM253xx-04-06-00-xx-002.eps
Fig. 12: Field instrument connection compartment sticker
!Note!
Please label the sensor terminal block with the sticker provided.
42
++
++++-
-
----
434154 55 56
NC
NC
NC NC
31
K11
12
32
A
3433
S
NC 938685 94 81 82 23 24 NC
51 52 53 57 58 59 47 48 49
O2
Temp.
(opt.)
+15V
10mA
RDGN
BN
REL 4SENSOR REL 3 REL 2 REL 1
131093-4F
COM 253
Digital 2
Digital 1
4..20mA
NC
Mains
Hilfsenergie
L1
+
N
-
(AC)
(DC)
Contacts:
max. 2A
AC: 250V/500VA
DC: 30V/60W
WH
Liquisys M COM223/253 Wiring
Endress+Hauser 19
Panel-mounted instrument connection
Connect the cables in accordance with the connection diagram.
C07-COM223xx-04-06-00-xx-001.eps
Fig. 13: Panel mounted instrument connection sticker
"Caution!
• Terminals marked NC may not be wired.
• Unmarked terminals may not be wired.
4.1.2 Measuring cable and sensor connection
You require screened special measuring cables to connect oxygen sensor COS41 to the transmitter.
To extend the measuring cable, use junction box and extension cable:
You can also use measuring cable CMK.
31 +
32 -
33 +
12
34 -
85 +
NC
NC
NC
NC
NC
Sensor
NC
86 -
93 + S
94 - 42
81 + 43
82 - 41
O2
Temp.
(opt.)
+15V
10mA
Digital 2
Digital 1
11
91
90 K
A
RD
BN
GN
131080-4C
Mains Contacts: max. 2A, AC:250V/500VA, DC:30V/60W
L1 +
AC DC
N-
REL1
REL2
REL3
REL4
4..20mA
24
23
49
COM
COM
COM
COM
-
+
59
54
55
56
52
51
53
58
57
48
47
131084-4C
WH
Sensor Sensor cablel Extension
COS41 CYK71 VBM junction box + CYK71 cable
Maximum cable length
COS41 50 m (164 ft) with CMK / CYK71 cable
Wiring Liquisys M COM223/253
20 Endress+Hauser
Structure of the measuring cable
!Note!
For further information on cables and junction boxes refer to chapter "Accessories".
Example of connecting an oxygen sensor
Connection of the oxygen sensor COS41:
C07-COM2x3xx-04-06-00-xx-002.eps
Fig. 15: Connection of COS41 with CYK71 cable
C07-COS41xxx-04-05-00-xx-001.eps
Fig. 14: Measuring cable CMK or CYK71
Terminal
S
12
90
91
11
Assignment
Outer screen
Active inner screen (NTC)
Cathode
Anode
NTC temperature sensor
BN
GN
WH
RD
S
12
90
91
11
K
A
S
CYK71
BN RD WH GN
Schirm /
Screen
11 12 90 91
O
2
KA
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112