Samsung SA8000 SE User manual

Type
User manual
Chapter 4
Diagnosis Modes
Diagnosis Mode Types and Control ............................................................................... 2
Diagnosis Mode Type ............................................................................................................... 2
Changing Diagnosis Mode Format ........................................................................................... 2
Diagnosis Modes Optimization ................................................................................................. 4
Basic Mode..................................................................................................................... 6
2D Mode ................................................................................................................................... 6
M Mode ..................................................................................................................................... 9
Color Doppler Mode................................................................................................................ 11
Power Doppler Mode ..............................................................................................................14
Spectral Doppler Mode ...........................................................................................................15
Combined Mode ........................................................................................................... 20
B/C/PW Mode ......................................................................................................................... 20
B/PD/PW Mode....................................................................................................................... 20
B/C/CW Mode......................................................................................................................... 21
B/PD/CW Mode ...................................................................................................................... 22
B/C/M Mode............................................................................................................................ 22
Multi-Image Mode......................................................................................................... 23
Dual-2D Mode......................................................................................................................... 23
Dual-2D/Color Doppler Mode ................................................................................................. 23
Dual-2D/PD Mode................................................................................................................... 24
3D Mode....................................................................................................................... 25
3D ........................................................................................................................................... 25
Acquiring a 3D Image ............................................................................................................. 26
3D View................................................................................................................................... 26
4 - 2 SA8000 SE Operation manual
Diagnosis Mode Types and Control
Diagnosis Mode Type
This product supports a variety of diagnosis modes including Basic mode, Combined mode,
Multi-Image mode, and 3D mode.
Basic Mode
2D Mode, Color Doppler Mode, Power Doppler Mode, M Mode, PW Spectral Doppler
Mode, CW Spectral Doppler Mode.
Combined Mode
M Mode, PW Spectral Doppler Mode and CW Spectral Doppler Mode can be added in
Color Doppler Mode or Power Doppler Mode. Five types of Combined Mode are
supported.
Multi-Image Mode
In Dual Mode you can view two images side by side simultaneously and compare them.
3D Mode
3D images can be obtained. Provides the Freehand 3D option, which allows viewing of
three-dimensional images.
Changing Diagnosis Mode Format
The following section explains how to change the display format that is common to all diagnosis
modes.
Changing Display Format
Use the Flexible Soft button [4] Loop Format.
M Mode, PW Spectral Doppler Mode, or CW Spectral Doppler Mode provide two display
formats.
In M mode and PW Spectral Doppler Mode, Top Down Format sets the images at the top
and bottom of the screen. Side-by-Side Format sets the images side by side.
Chapter4. Diagnosis Modes 4 - 3
Changing Active Image
In Dual Mode, the control functions and screen menu configuration differ depending on the
active image. To change the Flexible soft menu of currently activated mode, press the
keyboard space bar.
For example, you can change the currently active image in PW Spectral Doppler Mode Æ
2D Mode Æ Color Doppler Mode order by pressing the space bar if 2D Mode is activated
in B/C/PW Mode.
However, the keyboard space bar is inactive in Freeze mode.
Changing Flexible Soft Menu
Use the keyboard space bar to move from Combined Mode to Flexible Soft Menu.
In B/C/D Mode, for example, you can change Flexible Soft Menu in PW Sepctral Doppler
Mode to Flexible Soft Menu in Color Doppler Mode by pressing the space bar.
Changing Activated Image Position in Multi-Image Mode
Use Set button or Dual button on the control panel.
For example, press Set button or Dual button in Dual-2D Mode. The image that is being
scanned is frozen in the left section of the screen, and the right image becomes active.
Freezing 2D Image in Spectral Doppler Mode
When 2D and PW Spectral Doppler images are showing simultaneously, press Set button
on the control panel to freeze the 2D image for closer observation of Doppler images.
Press Set button again to reactivate the image.
When Simultaneous Mode is off, the current image mode is shown as PW Only, CW Only
or 2D Only in the upper right portion of the screen. Activation of Simultaneous Mode is set
in the Misc. tab in the Setup window. If you set ‘Simultaneous Mode’ is ‘Off’, the item
‘Simultaneous’ on the PW Spectral Doppler mode will not be activated.
Image Depth Control
Raise/lower the Depth switch to decrease/increase the scanning depth of an image. The
maximum/minimum depth depends on the probe used.
Image Magnification
Use the HD Zoom switch on the control panel to save the magnified image. Raise the
switch, and the HD Zoom Box appears. Use the Change button and trackball to set the
position and size of the HD Zoom Box. Press the Change button to toggle between Zoom
Pos. and Zoom Size; the selection will be indicated on the left side of the screen.
4 - 4 SA8000 SE Operation manual
- Zoom Position: Press the Change button and the HD Zoom Box lines become solid
lines. Use the trackball to move the box to the desired position.
- Zoom Size: Press the Change button and the HD Zoom Box lines become dotted
lines. Use the trackball to change the size of the box.
After adjusting its position and size with the Trackball, press the Set button to enter HD
Zoom Mode. You can view the position of the magnified area through the Zoom
Navigation Box on the screen.
In HD Zoom mode, press the Z key on the keyboard to make the Zoom Navigation Box
disappear and Zoom Position deactivate. Press Z key again, and the Zoom Navigation
Box is shown again.
Adjust Depth switch or press Exit button to exit HD Zoom Mode.
Read Zoom
This function magnifies an image saved in the memory. Press the Zoom button on the
control panel to start Read Zoom mode.
In HD Zoom mode, it is available to press the Zoom button on the control panel to
start the Read Zoom mode. However, in Read Zoom mode, you cannot enter the HD
Zoom mode with the Zoom button.
Diagnosis Modes Optimization
For optimizing images, the following functions are common to all modes.
Gain Control
Use the D-Gain, C-Gain, or B-Gain dials on the control panel to adjust image brightness.
If you rotate the Gain dial clockwise, its value increases, and vice versa.
TGC Control
Ultrasound penetration gets weaker with depth. Six TGC slides are available for optimal
image output. Adjust Gain according to depth.
Move the TGC slide to the right(+) to increase Gain (brightening the image) and vice versa.
Frequency Change
Use the Frequency button on the control panel to obtain the optimal image by changing
the frequency. The meanings of the frequencies are shown below.
Chapter4. Diagnosis Modes 4 - 5
Focus Position Control
Raise/lower the Focus switch raise/lower the focusing point.
Harmonic Function
Press the Harmonic button on the control panel in 2D Mode. When the Harmonic function
is used, ‘HGen.’, ‘HRes.’ and ‘HPen.’ are displayed in the probe information section.
The Harmonic function is only activated with the specific probe.
Power Control
Power is the intensity of the acoustic output.
Use the Menu dial-button on the control panel to set the Power, in the left menu. Its value
ranges from 10 to 100.
4 - 6 SA8000 SE Operation manual
Basic Mode
2D Mode
This basic mode, also referred to as B Mode (Brightness mode), shows anatomy scan planes.
[Figure 4.1 2D Mode]
[Figure 4.2 2D Menu]
2D Mode Selection
Press the 2D button on the control panel.
2D Mode is the default mode and the 2D button is not a toggle button. Instead, it is used to
return to basic 2D Mode when in other image modes.
Chapter4. Diagnosis Modes 4 - 7
Changing 2D Image Format
Changing Image Direction
- Changing Left/ Right Orientation (Horizontal Orientation Direction)
Press the Flexible Soft button [5] Direction.
- Changing Up/down Orientation (Vertical Orientation. Apex)
Select Apex in the 2D menu using the Menu dial-button.
Changing View Area
Select View Area in the menu using the Menu dial-button on the control panel.
You can adjust View Area using the Size + and Size – buttons in the Flexible Soft Menu.
Widening/narrowing the View Area, decreases/increases the frame rate (FPS). If you want
to view other unseen parts, press the Pos.L and Pos.R buttons in Flexible Soft Menu to
move the View Area to the right or left.
[Figure 4.3 Flexible Soft Menu when changing View Area]
The changed view area value is displayed in the lower right-hand portion of the screen.
Pressing the Flexible Soft button [5] Return makes the value disappear.
M Line Display
Indicates where the image for M Mode or PW Spectral Doppler Mode appears in the 2D
image.
Pressing the Flexible Soft button [4] M Line makes the M Line appear or disappear.
2D Image Optimization
Frame Average Control
The Frame Average function is used to average frames together on the screen, thereby
reducing the speckles which make diagnosis difficult. Setting the Frame Average value as
high makes the display movements smoother but slower. (Though the actual frame rate
doesn’t drop.) Turn the function off in OB and cardiac examination where organs are
highly active.
Use the Menu dial-button on the control panel to press FA in the menu. The values range
from 0 to 9.
Dynamic Range Control
This function is used to change the range of contrast by adjusting the ratio of
4 - 8 SA8000 SE Operation manual
minimum/maximum input signal value ratio. Increase/decrease Dynamic Range to
smooth/roughen the image.
Use the Menu dial-button on the control panel to press DR in the left menu. Select a value.
Edge Enhance Control
This function is used to view more accurate images of organ or tissue boundaries.
Increase the Edge Enhance value to obtain clearer pictures of boundaries.
Use the Menu dial-button on the control panel to press EE in the menu. Select a value.
Reject Level Control
This function is used to eliminate noise or low level echoes for clearer signals.
Use the Menu dial-button on the control panel to press Reject in the left menu. Select a
value.
Focus Number Control
Use the Menu dial-button on the control panel to press Focus No in the menu. Or press
the Flexible Soft button [1] Focus No, and set the Focus Number with the Soft Menu
Focus + and Focus – buttons.
Frame Rate Control
The Frame Rate is the number of images generated per second. The Frame Rate should
be set to Fast when viewing fast moving organs. Select Fast, Normal, or Slow using the
Menu dial-button on the control panel.
Tissue Property Control
Select the desired ultrasound speed according to the patient’s tissue type. The Tissue
menu consists of Cystic, Adipose, Normal, and Solid.
Use the Menu dial-button on the control panel to select Tissue in the menu.
Trapezoid Function
This function is used to view the trapezoid-shaped 2D image when using the Linear probe.
Use the Menu dial-button on the control panel to select Trapezoidal in the menu.
When adjusting image depth, some depth values do not support the Trapezoid function.
When the Trapezoid function is on, the Zoom button is inactivated.
Chapter4. Diagnosis Modes 4 - 9
M Mode
In M Mode, you can observe the motion patterns of an area of anatomy - along a 2D image
reference line (M Line). This mode is especially useful for cardio-valvular observation.
[Figure 4.4 M Mode]
[Figure 4.5 M Mode Flexible Soft Menu]
M Mode Selection
Press the M button on the control panel. Press M button again. M Mode turns off and 2D Mode
turns on.
4 - 10 SA8000 SE Operation manual
Changing M Mode Format
Image Direction Change
Select Direction using the Menu dial-button on the control panel in M Mode.
M Line Control
Move the M Line to change the observation location. The M Line can be moved left and
right by using the Trackball on the control panel.
M Image Size Change
To adjust the size (Large, Normal or Small) of the 2D image and the M image, use the
Menu dial-button on the control panel to select M Size from the menu or select an M
Mode image size with the Flexible Soft button [1] M Size.
Sweep Speed Control
You can choose from 120Hz, 180Hz, 240Hz or 300Hz with the Speed button on the
control panel.
Loop Format Change
Select Top Down Format or Side-by-Side Format using the Flexible Soft button [4] Loop
Format. Top Down Format sets the 2D image and M image at the top and bottom of the
screen. Side-by-Side Format sets the images side by side.
NOTE
Refer to `2D Mode’ for more information about other items in the menu.
Chapter4. Diagnosis Modes 4 - 11
Color Doppler Mode
In this mode you can obtain a variety of information about blood flow within the 2D Image such
as the existence of blood flow, its average speed and its direction within the ROI (Region of
Interest).
[Figure 4.6 Color Doppler Mode]
[Figure 4.7 Color Doppler Mode Menu]
Color Doppler Mode Selection
Press the C button on the control panel.
Pressing C button again turns Color Doppler Mode off and 2D Mode on.
4 - 12 SA8000 SE Operation manual
Changing Color Image Format
Controlling Color Bar Baseline Control
The colors above the baseline represent the direction/speed of blood flow moving toward
the probe, and the colors below the baseline represent the direction/speed of blood flow
away from the probe.
Use the Baseline switch on the control panel. If you raise the switch, the Color Bar
Baseline is raised, and vice versa.
Baseline can be adjusted only when the Color Doppler Mode value is set to Velocity.
Color Bar Invert
Press the Invert button on the control panel to invert the color bar and the colors on the
image.
Changing Color Display Information
Select Color Mode in C-Menu using the Menu dial-button on the control panel.
Use the Menu dial-button to cycle through the available settings - Velocity, Power,
Variance, and Vel + Var.
Color Image Optimization
Changing ROI Box Position
The ROI is the area within the 2D image where color (blood flow) information is displayed
in Color Doppler Mode. Use the Trackball to change the ROI Box position.
Changing ROI Box Size
Enter the ROI Box Size Control Mode by rotating the Menu dial-button without opening the
pop-up menu or pressing the Change button on the control panel. Adjust the ROI Box size
with the Trackball. To return to the ROI Box Position Change Mode, rotate the Menu dial-
button without opening the pop-up menu or press Change button again.
Scale (PRF) Control
Use the Scale switch on the control panel. If you raise/lower the switch, PRF (Pulse
Repetition Frequency) increases/decreases so that the speed range of blood flow is
widened/narrowed.
Wall Filter Control
Wall filter is an electrical filter used to eliminate low-frequency Doppler signals caused by
the motions of vessel walls. Eliminate Doppler signals whose frequencies are lower than
the Cutoff Frequency by adjusting the Cutoff Frequency.
Chapter4. Diagnosis Modes 4 - 13
Use the Filter switch on the control panel. If you raise the Filter switch, the Cutoff
Frequency of the Wall Filter decreases, and vice versa.
Sensitivity Control
Select Sensitivity in the menu using the Menu dial-button on the control panel or use the
Flexible Soft button [1] Sensitivity to adjust the value (8~31).
To increase/decrease Sensitivity, raise/lower the Sensitivity value. Increasing the Color
Sensitivity decreases the frame rate, and vice versa.
Balance Control
This function is used to adjust the range of the color image displayed by comparing the
gray level of the 2D image with the Doppler signal value of the color image. If the Balance
value increases, the color image is seen even where the 2D image signal is strong (bright),
thereby increasing the color image range.
Select Balance using the Menu dial-button on the control panel and adjust the value
(1~16).
Steer Function
When Linear probe is selected, this function adjusts the ultrasound beam angle to prevent
the loss of color information caused by the ultrasound beam angle.
Pressing the Steer button on the control panel steers the ROI to the left, center and right -
in that order.
Display Function
Color Doppler Mode supports Color + B/W Image, Color Only, and B/W Only display.
Select Display in the menu using the Menu dial-button on the control panel. In Freeze,
use the Flexible Soft Button [5] to select the function.
TDI (Tissue Doppler Image) Function
This function is used to show the movements of the cardiac tissues. To use the TDI
function, use the Menu dial-button on the control panel to select Tissue Doppler from the
menu on the left-hand side of the screen.
It uses the color information of Color Doppler Mode or Power Doppler Mode. It also uses
the change of cardiac tissue of PW Spectral Doppler Mode as time goes by. It is only used
in the specific application (General, Cardiac, Pediatric) of Phased Array probe.
NOTE
For further information about other items in the menu, refer to ‘2D Mode’.
4 - 14 SA8000 SE Operation manual
Power Doppler Mode
Instead of the speed or direction of blood flow, Power Doppler Mode displays the overall
strength of blood flow, thereby showing information about the existence and amount of blood
flow in the sample area.
[Figure 4.8 Power Doppler Mode]
[Figure 4.9 Power Doppler Mode Menu]
Power Doppler Mode Selection
Press the PD button on the control panel. Pressing PD button again turns PD Mode off,
and 2D Mode on.
PD Mode Screen
Color Bar
Colors show the existence and amount of blood flow in PD Mode. The top of the color bar
is the brightest section, where the amount of blood flow is at its highest.
NOTE
For further information about other PD Image optimization, refer to ‘Color
Doppler Mode’.
Chapter4. Diagnosis Modes 4 - 15
Spectral Doppler Mode
Spectral Doppler Mode gives information on the speed/direction of blood flow at a specific site
in the form of a spectral trace and audio signal. It is divided into PW (Pulsed Wave) Spectral
Doppler Mode and CW (Continuous Wave) Spectral Doppler Mode.
PW Spectral Doppler Mode
PW Spectral Doppler Mode gives information on the speed of blood flow at a specific site in
the form of a spectral trace and audio signal. Since the pulse is transmitted being divided
according to time, distance (depth) information can be obtained. The mode is suitable for
measuring relatively slow blood flow such as that in abdominal and peripheral vessels. The
2D Mode image is shown simultaneously allowing the marking and adjustment of specific
sites within blood vessels.
Press the D button on the control panel. Press D button again to enter 2D Mode.
[Figure 4.10 PW Spectral Doppler Mode]
[Figure 4.11 PW Spectral Doppler Mode menu]
4 - 16 SA8000 SE Operation manual
CW Spectral Doppler Mode (Optional)
CW Spectral Doppler Mode gives information on the speed/direction of blood flow in the
form of a spectral trace and audio signal. The 2D Mode image is shown simultaneously
allowing the marking and adjustment of specific sites within blood vessels.
Press the CW button on the control panel. Press CW button again to enter 2D Mode.
[Figure 4.12 CW Spectral Doppler Mode]
[Figure 4.13 CW Spectral Doppler Mode menu]
Spectral Doppler Mode Screen
On the Spectral Doppler Mode image, the horizontal axis represents time, and the vertical
axis represents speed (or frequency).
Sample Volume
When Sample Volume is on the blood flow of 2D image, it represents Doppler Spectrum.
Its position is moved with the Trackball and displayed in the [email protected] mm format. The
Chapter4. Diagnosis Modes 4 - 17
format means that the Sample Volume of ‘xx.x’ mm size is located at a depth of ‘yy.y’ mm.
For example, [email protected] mm means that the Sample Volume of 2.0mm size is located at a
depth of 16.7mm.
NOTE
Sample Volume is available only in PW Spectral Doppler Mode.
Spectral Doppler Mode Format Change
Changing PW Spectral Doppler Image Size
To adjust the image position, the 2D image is shown together. The size of the 2D image
and PW Spectral Doppler image can be changed.
Select PW Size in the menu using the Menu dial-button on the control panel. Or press the
Flexible Soft button [1] PW Size to select the size (Large, Normal, or Small).
This function is activated in Simultaneous mode is on.
Activating Simultaneous Function
The PW Spectral Doppler image and 2D image can be displayed simultaneously or
separately. This function is not available in Steered CW Spectral Doppler Mode.
Select Simultaneous in the menu using the Menu dial-button on the control panel. If
Simultaneous is selected again when this mode is already on, the mode is deactivated.
Activate the Simultaneous function option by setting Simultaneous function as ‘Allow
2D/PW’ or ‘Allow 2D/C/PW’in the Misc. tab in the Setup window.
NB. Simultaneous function decreases Doppler PRF, thus decreasing the measurable
speed range.
Doppler Baseline Control
Raise/lower the Baseline switch on the control panel to raise/lower the Baseline.
Doppler Axis Invert
Press the Invert button on the control panel. The ‘+ velocity value’ and ‘- velocity value’,
shown at the Doppler axis, change positions.
Loop Format Change
Select Loop Format in the menu using the Menu dial-button on the control panel. Or
press Flexible Soft button [4] Loop Format to select Loop Format. ‘Hor’ means Top
Down Format, in which 2D image is set above the PW Spectral Doppler image, and ‘Ver’
means Side by Side Format, in which the two images are set side-by-side.
4 - 18 SA8000 SE Operation manual
Spectral Doppler Image Optimization
Sample Volume Position Control
Use the Trackball on the control panel.
Sample Volume Size Control
Adjust the Sample Volume size by pressing the Change button and using the Trackball
on the control panel. Press Change button again to return to the Sample Volume Position
Control screen. The icon showing the two Trackball functions (SV Pos/SV Size) is shown
for a second and disappears.
Sample Volume Angle Control
Use the Angle dial to adjust the angle for accurate measurement of speed.
Doppler Sound Volume Control
Use the Audio Slide Volume control on the control panel.
Average Control
The Average Control is used to increase spectrum quality.
Select Average in the menu using the Menu dial-button on the control panel. Select the
Average value (Low, Middle or High).
Spectrum Enhancement
The Spectrum Enhancement is used to compensate the signal of low gain Spectral
Doppler Spectrum. This function is available in Loop Only Mode or Simultaneous Function
and is not available in TDI function.
Select Spectrum Ehc. In the menu using the Menu dial-button on the control panel.
Select the value from 1 to 8.
HPRF (High PRF) Function
This function measures blood flow whose speed exceeds specified limits at a specified
depth. It doubles the original scale. This function is available only in PW Spectral Doppler
Mode.
NOTE
To use HPRF function, select ‘On’ in the Setting>System>Misc. tab.
- HPRF Activation
To activate HPRF, increase the Scale values at the required depth. The Phantom Gate
appears over the sample volume in D Line. Once HPRF is activated, the PRF doesn’t
increase even if the scale values are raised.
Chapter4. Diagnosis Modes 4 - 19
[Figure 4.14 HPRF]
- HPRF Exit
To exit HPRF, decrease Scale values by one notch while HPRF is activated. The
maximum PRF values in PW Spectral Doppler Mode are shown.
- Sample Volume Position Control
To move the Sample Volume position in the PW Only status, the system calculates PRF
values and the Phantom Gate position, and updates them on the PW Spectral Doppler
image. Exit HPRF when HPRF cannot be activated.
When Sample Volume is moved in the 2D Only status, the PRF values don’t change. In
PW Only status, the PRF values and Phantom Gate position are calculated and updated
on the PW Spectral Doppler image.
NOTE
HPRF is not activated in Simultaneous Mode. It is also not activated if
[PRF*2] exceeds 23KHz.
CAUTION
The Phantom Gate position can be located outside the 2D image area in
Zoom Mode.
Ensure that the Sample Volume and Phantom Gate are not positioned
over the area to be measured. If more than two SVs are located in the
vessels, all Doppler components will appear in the spectrum, causing
noise.
NOTE
For further information about Wall Filter, Dynamic Range and Steer Control,
refer to ‘2D Mode’ and ‘Color Doppler Mode’.
4 - 20 SA8000 SE Operation manual
Combined Mode
B/C/PW Mode
Color Doppler Mode and PW Spectral Doppler Mode are displayed simultaneously.
Press the D button on the control panel in Color Doppler Mode. In PW Spectral Doppler Mode
press the C button.
[Figure 4.15 B/C/PW Mode]
B/PD/PW Mode
PD Mode and PW Spectral Doppler Mode are displayed simultaneously.
Press the D button on the control panel in PD Mode. In PW Spectral Doppler Mode press the
PD button.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33

Samsung SA8000 SE User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI