Chapter4. Diagnosis Modes 4 - 3
▐ Changing Flexible Soft Menu
Use the Change Menu button to change the Flexible Soft Menu in Combined Mode.
In 2D/C/PW Mode, for example, you can change Flexible Soft Menu in PW Sepctral
Doppler Mode to Flexible Soft Menu in Color Doppler Mode by pressing the Change
Menu button.
▐ Changing Activated Image Position in Multi-Image Mode
Use Set button or Dual button on the control panel.
For example, press Set button or Dual button in Dual-2D Mode. The image that is being
scanned is frozen in the left section of the screen, and the right image becomes active.
▐ Freezing 2D Image in Spectral Doppler Mode
When 2D and PW Spectral Doppler images are showing simultaneously, press Update
button on the control panel to freeze the 2D image for closer observation of Doppler
images. Press Update button again to reactivate the image.
When Simultaneous Mode is off, the current image mode is shown as PW Only or 2D
Only in the upper right portion of the screen. Activation of Simultaneous Mode is set in the
Misc. tab in the Setup window. If you set ‘Simultaneous Mode’ is ‘Off’, the item
‘Simultaneous’ on the PW Spectral Doppler mode will not be activated.
▐ Image Depth Control
Raise/lower the Depth switch to decrease/increase the scanning depth of an image. The
maximum/minimum depth depends on the probe used.
Diagnosis Modes Optimization
For optimizing images, the following functions are common to all modes.
▐ Gain Control
Press the Gain dial-button on the control panel several times to select a specific mode.
After the LED of the mode is turned on, rotate the dial button to adjust the brightness of an
image. If you rotate the Gain dial clockwise, gain values increase and vice versa.
▐ TGC Control
Ultrasound penetration gets weaker with depth. Six TGC slides are available for optimal
image output. Adjust Gain according to depth.