8 Service instructions RZ.5K.A1.02 © Danfoss 9/1998 AKS 41
7. Push the programming button and keep it down until the green LED goes out which will
happen after about 30 seconds. (It is important that the button is released not later than
three seconds after the green LED goes out).
8. Wait 10 seconds.
The new calibration has now been saved in the memory. Remove the shorting strap from
the two pins and check that the transmitter’s output signal is 20 mA. The shorting strap is
stored by placing it over one of the pins.
N.B.
You may calibrate the lower limit of the measuring range (4 mA point) without at the same
time calibrating the upper limit (20 mA) and vice versa. However, be aware of the following:
When you calibrate the lower limit of the measuring range, it is in actual fact the zero-point of
the built-in signal converter you are adjusting.
When you calibrate the upper limit of the measuring range, it is in actual fact the width of the
measuring range (span) of the built-in signal converter you are adjusting.
Consequently, if the lower limit of the measuring range (4 mA point) is moved, the upper limit
(20 mA point) will automatically be moved a similar distance.
Calibration of plant (all refrigerants)
If the transmitter is to be calibrated in the plant it may however be difficult to get the rod right
out of the liquid, and most often impossible to have the liquid reach the upper end of the rod. It
is furthermore a problem to ascertain the liquid level, when the level signal from the liquid
level transmitter cannot be used. If you are lucky, you may have a short gauge glass to help
you, but they hardly ever cover the lower and uppermost parts of the rod.
In short, it is impossible in practice to carry out a calibration of the
extreme points
of the
measuring range, if calibration is to be made with the transmitter mounted in the system.
But a calibration does not either necessarily have to take place at the extreme points of the
measuring range (corresponding to the 4 mA and the 20 mA points). The relation between
signal current and liquid level can be shown as a straight line - the so called liquid level
transmitter working line. In order for you to draw a straight line you must know at least two
points traversed by the line, but these two points need not be the end points of the line.
The calibration function has been designed to take advantage of this possibility. Instead of
calibrating the lower working point to 4 mA, it may be calibrated to 6, 8 or 10 mA, and instead
of calibrating the upper working point to 20 mA, it may be calibrated to 12, 14, 16 or 18 mA.
In practice it works, as follows. (It is assumed here that the transmitter has been put in
calibration mode by short-circuiting the two pins with the supplied strap, as described earlier,
and that a voltmeter has been connected to output terminals 2 and 3):
When the programming button is kept down, the 4-20 mA signal will be increased in 2 mA
steps, where each step is maintained for four seconds. If the programming button is released
at a certain step, the actual liquid level will be calibrated to the value the current had reached.
If the programming button is released when the current has reached 4, 6, 8 or 10 mA, this will
be understood as a calibration of the lower point of the working line.
If you wait releasing the programming button until the current has reached 12, 14, 16, 18 or 20
mA, this will be understood as a calibration of the upper point of the working line.
Signal current
terminals 2 - 3