  • Hello! I'm an AI assistant, and I've reviewed the user manual for the Rosemount OCX8800C Oxygen and Combustibles Analyzer. This document provides detailed instructions on installing, configuring, operating, and maintaining the device. It covers topics from safety instructions to calibration procedures and troubleshooting. The device is designed for hazardous areas and has multiple features like built-in calibrations, HART communications, and a blowback system. I'm here to help with your questions about the Rosemount OCX8800C.
  • What is the measuring range for oxygen?
    What is the measuring range for combustibles?
    What type of communication is standard with the device?
    What is the purpose of blowback system?
Manual
00809-0500-4881, Rev AB
April2023
Rosemount OCX8800C
Oxygen and Combustibles Analyzer for Hazardous Areas
Essential instructions
Read this page before proceeding!
Emerson designs, manufactures, and tests its products to meet many national and international standards. Because
these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure
they continue to operate within their normal specifications. You must adhere to the following instructions and
integrate them into your safety program when installing, using, and maintaining Emerson's Rosemount products.
Failure to follow the proper instructions may cause any one of the following situations to occur: loss of life, personal
injury, property damage, damage to this instrument, and warranty invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Follow all instructions during the installation, operation, and maintenance of this product.
To prevent personal injury, ensure that all components are in place prior to and during operation of this product.
Inform and educate your personnel in the proper installation, operation, and maintenance of the product.
Install equipment as specified in the installation instructions of the appropriate Quick Start Guide and per
applicable local and national codes. Connect all products to the proper electrical and pressure sources.
Ensure that all connections to pressure and electrical sources are secure prior to and during equipment operation.
If this reference manual is not the correct one, call 1-800-854-8257 or 949-757-8500 to request the correct
reference manual. Save this reference manual for future reference.
To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the
product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Emerson.
Unauthorized parts and procedures can affect the product's performance, place the safe operation of your
process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or
improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being
performed by qualified people, to prevent electrical shock and personal injury.
NOTICE
The information contained in this document is subject to change without notice.
NOTICE
If a Field Communicator is used with this unit, the software within the Field Communicator may require modification.
If a software modification is required, please contact your local Emerson Service Group or National Response Center
at 1-800-654-7768.
WARNING
Physical access
Unauthorized personnel may potentially cause significant damage to and/or misconfiguration of end users’
equipment. This could be intentional or unintentional and needs to be protected against.
Physical security is an important part of any security program and fundamental in protecting your system. Restrict
physical access by unauthorized personnel to protect end users’ assets. This is true for all systems used within the
facility.
Symbols
Earth (ground) terminal
2
Protective conducter terminal
Risk of electrical shock
Warning: Refer to Reference Manual.
Preface
The purpose of this manual is to provide information concerning the components, functions, installation, and
maintenance of the Rosemount OCX8800.
Emerson recommends that you thoroughly familiarize yourself with Description and specifications and Install before
installing your transmitter.
Description and specifications presents the basic principles of the transmitter along with its performance
characteristics and components. The remaining sections contain detailed procedures and information necessary to
install and service the transmitter.
Before contacting Emerson concerning any questions, first consult this manual. It describes most situations
encountered in your equipment's operation and details necessary action.
Definitions
The following definitions apply to warnings, cautions, and notes found throughout this publication.
WARNING
Highlights an operation or maintenance procedure, condition, statement, etc. that if not strictly observed, could result
in injury, death, or long-term health hazards of personnel.
CAUTION
Highlights an operation or maintenance procedure, practice, condition, statement, etc. that if not strictly observed,
could result in damage to or destruction of equipment or loss of effectiveness.
Note
Highlights an essential operating procedure, condition, or statement.
Notice
The contents of this publication are presented for informational purposes only, and while every effort has been made
to ensure their accuracy, they are not to be construed as warranties or guarantees, expressed or implied, regarding
the products or services described herein or their use or applicability. All sales are governed by Rosemount's
terms and conditions, which are available upon request. We reserve the right to modify or improve the designs
or specifications of such products at any time.
Rosemount does not assume responsibility for the selection, use, or maintenance of any product. Responsibility for
proper selection, use, and maintenance of any Rosemount product remains solely with the purchaser and end-user.
To the best of Rosemount's knowledge, the information herein is complete and accurate. Rosemount makes no
warranties, expressed or implied, including implied warranties of merchantability and fitness for a particular purpose,
with respect to this manual and, in no event, shall Rosemount be liable for any incidental, punitive, special, or
consequential damages including, but not limited to, loss of production, loss of profits, loss of revenue or use, and
costs incurred, including without limitation for capital, fuel and power, and claims of third parties.
Product names used herein are for manufacturer or supplier identification only and may be trademarks/registered
trademarks of these companies.
Rosemount and the Rosemount logo are registered trademarks of Rosemount. The Emerson logo is a trademark and
service mark of Emerson Electric Co.
3
4
Contents
Chapter 1  Description and specifications................................................................................................ 7
1.1 Component checklist...................................................................................................................7
1.2 System overview..........................................................................................................................9
1.3 Specifications............................................................................................................................. 19
1.4 Ordering information............................................................................................................... 22
Chapter 2  Install........................................................................................................................................23
2.1 Product safety............................................................................................................................23
2.2 Mechanical installation............................................................................................................. 24
2.3 Electrical installation................................................................................................................. 28
2.4 Pneumatic installation.............................................................................................................. 36
2.5 Initial start-up............................................................................................................................ 43
Chapter 3  Configuration and start-up....................................................................................................45
3.1 Verify installation.......................................................................................................................45
3.2 Initial power up..........................................................................................................................48
3.3 Setting test gas values..............................................................................................................48
3.4 Reset........................................................................................................................................... 49
3.5 Calibration solenoids................................................................................................................ 49
3.6 Blowback feature.......................................................................................................................50
3.7 Calibration verify feature..........................................................................................................51
3.8 Calibration tolerance feature...................................................................................................53
3.9 COe purge/zero feature........................................................................................................... 54
Chapter 4  Using the local operator interface (LOI).............................................................................. 57
4.1 Display orientation....................................................................................................................57
4.2 Local operator interface (LOI) controls...................................................................................58
4.3 Local operator interface (LOI) menu tree...............................................................................61
4.4 D/A trim procedures................................................................................................................. 66
Chapter 5  Using HART® communications.............................................................................................. 69
5.1 Overview.....................................................................................................................................69
5.2 Field Communicator signal connections................................................................................ 69
5.3 Field Communicator PC connections......................................................................................71
5.4 HART® menu tree...................................................................................................................... 72
5.5 D/A trim procedures................................................................................................................. 76
Chapter 6  Calibration............................................................................................................................... 79
6.1 Overview.....................................................................................................................................79
6.2 Fully automatic calibration.......................................................................................................79
6.3 Operator-initiated autocalibration..........................................................................................80
6.4 Manual calibration.................................................................................................................... 81
Chapter 7  Field Communicator............................................................................................................... 87
7.1 Overview.....................................................................................................................................87
7.2 Field Communicator connections............................................................................................87
Manual Contents
00809-0500-4881 April2023
Rosemount OCX8800C 5
7.3 Fieldbus menu tree................................................................................................................... 90
Chapter 8  FOUNDATION Fieldbus.........................................................................................................95
8.1 FOUNDATION Fieldbus technology........................................................................................... 95
8.2 Network communication..........................................................................................................98
8.3 Rosemount OCX8800 function blocks.................................................................................... 99
8.4 Resource block.........................................................................................................................100
8.5 Configure PlantWeb alert (PWA) simulation with the Field Communicator.................... 108
8.6 Transducer block.....................................................................................................................109
8.7 Transducer block enumerations........................................................................................... 116
8.8 Analog input (AI) function block............................................................................................123
8.9 Proportional/integral/derivative (PID) function block........................................................ 140
8.10 Arithmetic (ARTHM) function block.....................................................................................163
8.11 Advanced topics.................................................................................................................... 171
8.12 Input selector (ISEL) function block.................................................................................... 172
8.13 Operation with Emerson DeltaV........................................................................................179
Chapter 9  Troubleshoot..........................................................................................................................181
9.1 Overview...................................................................................................................................181
9.2 Diagnostic alarms....................................................................................................................183
9.3 Fault isolation.......................................................................................................................... 183
9.4 Alarm relay events.................................................................................................................. 200
Chapter 10  Maintenance and service..................................................................................................... 201
10.1 Overview.................................................................................................................................201
10.2 Removal and installation......................................................................................................201
10.3 Repair sensor housing..........................................................................................................227
10.4 Repair electronics housing.................................................................................................. 254
10.5 Replace tube fittings.............................................................................................................263
Chapter 11  Replacement parts................................................................................................................267
11.1 Sensor housing......................................................................................................................267
11.2 Electronics housing...............................................................................................................271
11.3 O2 cell and heater strut assembly.......................................................................................275
Appendix A  Safety data............................................................................................................................. 277
A.1 Safety instructions.................................................................................................................. 277
A.2 Safety data sheet for ceramic fiber products...................................................................... 279
A.3 High pressure gas cylinders.................................................................................................. 284
Appendix B  Product certifications........................................................................................................... 285
Appendix C  SPA with HART® alarm......................................................................................................... 287
C.1 Description...............................................................................................................................287
C.2 Install........................................................................................................................................ 289
C.3 Setup.........................................................................................................................................290
Appendix D  Returning material............................................................................................................... 297
Contents Manual
April2023 00809-0500-4881
6 Emerson.com/Rosemount
1  Description and specifications
1.1  Component checklist
A typical Rosemount OCX8800 oxygen/combustibles transmitter package should contain
the items shown in Figure 1-1.
Check the model number of your Rosemount OCX8800 against the transmitter features
and options, making sure options specified by this number are on or included with the
unit. Use this complete model number for any correspondence with Emerson.
Optional accessories provides a list of accessories for use with the Rosemount OCX8800.
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 7
Figure 1-1: Typical System Package
A. Quick Start Guide
B. Adapter plate with mounting hardware and gasket
C. Reference air and calibration set (optional)
D. Blowback hardware (optional)
E. Rosemount OCX8800 with remote electronics
F. Rosemount OCX8800 with integral electronics
Description and specifications Manual
April2023 00809-0500-4881
8 Emerson.com/Rosemount
1.2  System overview
1.2.1  Scope
This Reference Manual supplies details needed to install, start up, operate, and maintain
the Rosemount OCX8800.
Signal conditioning electronics outputs separate 4-20 mA signals representing oxygen (O2)
and combustibles (COe) values. You can access this information, plus additional details,
with the HART® Field Communicator or Emerson AMS software. The local operator interface
(LOI) also provides a communications interface with the electronics.
1.2.2  System description
Emerson has designed the Rosemount OCX8800 to measure oxygen and combustible
concentrations in flue gas temperatures up to 2600 °F (1427 °C). Electrical connections,
power, and communications are made through two ¾ national pipe thread (NPT) ports in
the flameproof electronics enclosure using fittings and cables provided by the customer.
Cable installation must meet NEC, IEC, and/or other applicable national or local codes for
Class I, Zone 1, Group IIB +H2 T3/T6 permanently mounted equipment. The transmitter is
close coupled to the process and requires minimum sample conditioning requirements.
The equipment measures oxygen percentage by reading the voltage developed across a
heated electrochemical cell, which consists of a small yttria-stabilized, zirconia disc. Both
sides of the disc are coated with porous metal electrodes. When operating at the proper
temperature, the following Nernst equation gives the millivolt output of the cell.
EMF = KT log10 (P1/P2) + C
where:
1. P2 is the partial pressure of the oxygen in the measured gas on one side of the cell.
2. P1 is the partial pressure of the oxygen in the reference air on the opposite side of
the cell.
3. T is the absolute temperature.
4. C is the cell constant.
5. K is an arithmetic constant.
Note
For best results, use clean, dry instrument air (20.95 percent oxygen) as the reference air.
When the cell is at operating temperature and there are unequal oxygen concentrations
across the cell, oxygen ions travel from the high oxygen partial pressure side to the
low oxygen partial pressure side of the cell. The resulting logarithmic output voltage is
approximately 50 mV per decade. The output is proportional to the inverse lograrithm
of the oxygen concentration. Therefore, the output signal increases as the oxygen
concentration of the sample decreases. This characteristic enables the transmitter to
provide exceptional sensitivity at low oxygen concentrations.
The Rosemount OCX8800 oxygen sensor measures net oxygen concentration in the
presence of all the products of combustion, including water vapor. Therefore, it may be
considered an analysis on a wet basis. In comparison with older methods, such as the
portable apparatus, which provides an analysis on a dry gas basis, the wet analysis will, in
general, indicate a lower percentage of oxygen.
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 9
The combustibles sensor is a catalytic sensor consisting of two resistance temperature
devices (RTDs). One RTD is the reference element covered with an inert coating. The other
RTD element is active, coated with a catalyst. As the sample gases flow by the sensor, the
combustible gases oxidize on the surface of the active element. The oxidation that occurs
produces heat and a temperature rise in the active element. The temperature difference
produces a resistance relationship between the two elements that is directly proportional
to the concentration of combustibles in the sample gases.
The catalyst is specifically designed to detect carbon monoxide (CO), but the sensor
responds to other combustible gases. The sensor is calibrated using CO; thus the output
should be expressed in terms of CO. However, as the sensor detects other combustible
gases, the output cannot just be labeled CO. The response of the sensor to other
combustible gases gives an output that is equivalent to the sensor detecting CO.
The term COe is used in this Manual to describe the sensor output. This term indicates that
the sensor is calibrated in terms of CO and that the sensor output is equivalent to CO but
not specific to CO.
Dilution air is provided to the COe sensor to ensure that there is adequate oxygen to fully
oxidize any combustible gases regardless of the concentration of oxygen in the process.
1.2.3  System configurations
Transmitters are available in four lengths, giving you the flexibility to use a penetration
appropriate to the size of the stack or duct. The length options are 18 in. (457 mm), 3 ft.
(0.91 m), 6 ft. (1.83 m), or 9 ft. (2.7 m). Probes are available in three material options: 316L
stainless steel, Inconel 600, and ceramic to accommodate higher temperatures.
The electronics are contained in a separate housing from the sensors. When the
transmitter is configured with the integral electronics option, the electronics and sensor
housings are mounted as a unit at the stack mounting flange. When the transmitter is
configured with the remote electronics option, the electronics housing is separated from
the sensor housing. The electronics housing may be mounted up to 150 ft. (45.7 m) from
the sensor housing.
The electronics control both sensor temperatures and provide output signals in one of two
ways:
1. Individual 4-20 mA isolated outputs that are proportional to the measured
oxygen and combustibles concentrations. The oxygen output also contains HART®
communication.
2. Single FOUNDATION Fieldbus output.
The electronics control both sensor temperatures and provide individual 4-20 mA isolated
outputs that are proportional to the measured oxygen and combustibles concentrations.
The power supply can accept voltages of 100 to 240 Vac and 50 to 60 Hz. The electronics
accepts millivolt signals generated by the sensors and produces the outputs to be used by
remotely connected devices. The outputs are isolated 4-20 mA linearized currents. Refer to
Configuration and start-up for specific instructions upon initial power up.
1.2.4  System features
1. The O2 cell output voltage and sensitivity increase as the oxygen concentration
decreases.
2. HART® or FOUNDATION Fieldbus communication is standard. To use this capability,
you must have either:
a. Field Communicator
Description and specifications Manual
April2023 00809-0500-4881
10 Emerson.com/Rosemount
b. Asset Management Solutions (AMS) software for the PC
3. Oxygen cell and heater/thermocouple assembly are field replaceable.
4. Electronics are automatically configured for line voltages from 100 to 240 Vac.
5. An operator can calibrate and diagnostically troubleshoot the transmitter in one of
two ways:
a. Local operator interface (LOI): The LOI is mounted to the end of the
electronics module and allows local communications with the electronics.
Refer to Using the local operator interface (LOI) for more information.
b. HART or FOUNDATION Fieldbus interface: The transmitter's output line
transmits a digital signal with the detected oxygen or combustible levels
encoded in a digital format. The HART output is superimposed on the oxygen
4-20 mA output line only. This information can be accessed through the
following:
Field Communicator: The handheld Field Communicator requires Device
Description (DD) software specific to the Rosemount OCX8800. The DD
software is supplied with many Field Communicators, but can also be
programmed into existing Field Communicators at most Emerson service
offices. Refer to Using HART® communications for additional information.
Personal computer (PC): The use of a personal computer requires AMS
software available from Emerson.
Selected distributed control systems: The use of distributed control
systems requires input/output (I/O) hardware and AMS software which
permit HART or FOUNDATION Fieldbus communications.
6. When the transmitter is configured without the LOI, you must calibrate and
diagnostically troubleshoot the transmitter using the HART or FOUNDATION Fieldbus
interface.
7. Optional blowback system: The blowback system periodically blows instrument
air back through the sample line filter and out the sample tube. This clears out
particulate and keeps the sample line filter from clogging.
1.2.5  System operation
Figure 1-2 shows the relationship between the components of the Rosemount OCX8800.
The sensor housing and probe mount to a duct or process wall so that the probe protrudes
into the flue gas stream. An air powered eductor continuously pulls samples of the process
flue gas through the probe to a chamber in front of the sensor housing where the sample
passes the O2 sensor and continues on the COe sensor. Dilution air is provided to the COe
sensor and reference air to the O2 sensor. After the gas sample flows past the O2 sensor
and through the COe sensor, it is drawn through the eductor where it mixes with the
eductor air and exits through exhaust back into the system.
The electronics housing contains the CPU and communication boards, which convert the
sensor inputs into 4-20 mA analog output signals. The CPU can also initiate and perform
calibrations. Solenoids can turn on and off three test gases and instrument air. A flow
meter between the electronics and sensor housings regulates test gas flow to the sensors.
Instrument air is separated into eductor air, reference air, and dilution air. The instrument
air solenoid does not allow air flow until the heaters are up to temperature. This minimizes
the amount of sampled process flue gas being pulled into the cold sensors causing
condensation.
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 11
Figure 1-2: System Operation Diagram
A. Sensor housing
B. COe combustibles sensor
C. Sample gas
D. Probe
E. O2 sensor
F. Eductor
G. Exhaust
H. Eductor air
I. Electronics housing
J. CPU
K. HART® board
L. Power supply
M. Optional test gas solenoids
N. Instrument air solenoid
O. Flow meter: 7 scfh
P. Eductor air
Q. Reference air
R. Low O2 test gas
S. High O2 test gas
T. CO test gas
U. Instrument air
V. Flow meter: 50 cc/min. (0.1 scfh)
W. Dilution air
Description and specifications Manual
April2023 00809-0500-4881
12 Emerson.com/Rosemount
1.2.6  Handling the transmitter
CAUTION
Equipment damage
The Rosemount OCX8800 is designed for industrial applications. The probe may contain
components made from ceramics, which are susceptible to shock when mishandled.
Treat each component of the system with care to avoid physical damage.
Only handle printed circuit boards and integrated circuits when adequate anti-static
precautions have been taken to prevent possible equipment damage.
1.2.7  System considerations
Prior to installing your transmitter, make sure you have all the components necessary to
make the system installation. Ensure that all components are properly integrated to make
the system functional.
After verifying that you have all the components, select mounting locations and
determine how each component will be placed in terms of available line voltage, ambient
temperatures, environmental considerations, convenience and serviceability. Figure 1-4
shows a typical system wiring for a system with integral electronics. Figure 1-5 shows
a typical system wiring for a system with remote electronics. Figure 1-6 shows the
dimensions for the optional sample tube support.
Figure 1-7 shows the dimensions for the optional in-situ filters. Figure 1-8 shows the
optional panel mounted blowback.
A source of instrument air is required at the transmitter for reference air, dilution air,
and eductor air. As the transmitter is equipped with an in-place calibration feature, make
provisions for connecting test gas tanks to the transmitter when calibrating it.
Note
The electronics module is designed to meet Type 4X and IP66, and the electronic
components are rated to temperatures up to 185 °F (85 °C).
Retain packaging in which the transmitter arrived from the factory in case any components
are to be shipped to another site. This packaging haws been designed to protect the
product.
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 13
Figure 1-3: Communications Connections and AMS Application - Rosemount OCX8800
with Integral Electronics
A. Rosemount OCX8800 with integral electronics
B. Instrument air
C. Three calibration lines by customer (300 ft. [91 m] maximum)
D. Signal output (twisted pairs)
E. Termination in control room
F. Field Communicator
G. Customer's laptop with AMS
H. AMS
Description and specifications Manual
April2023 00809-0500-4881
14 Emerson.com/Rosemount
Figure 1-4: Typical System Installation - Integral Electronics
A. Adapter plate
B. Signal outputs (twisted pairs)
C. Line voltage
D. Instrument air supply (reference gas)
E. High O2 test gas
F. Low O2 test gas
G. CO test gas
H. Gases
I. Duct
J. Stack
K. Test gas flow meter
L. Dilution air flow meter
M. Pressure regulator
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 15
Figure 1-5: Typical System Installation - Remote Electronics
A. Gases
B. Duct
C. Stack
D. Heater power cable (up to 150 ft. [46 m])
E. Signal cable (up to 150 ft. [46 m])
F. Test gas flow meter
G. Dilution air flow meter
H. 4-20 mA outputs (two twisted pairs)
I. Line voltage
J. Instrument air supply (reference gas)
K. Pressure regulator
L. High O2 test gas
M. Low O2 test gas
N. CO test gas
Description and specifications Manual
April2023 00809-0500-4881
16 Emerson.com/Rosemount
Figure 1-6: Optional Sample Tube Support
C. 0.75-in. (19 mm) diameter on 7.5-in. (190 mm) diameter B.C., eight places
D. 0.75-in. (19 mm) diameter on 4.75-in. (121 mm) diameter B.C., four places
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 17
Probe length in in. (mm) Length L in in. (mm)
18 in. (457.2 mm) 24.5 (621)
3 ft. (0.9 m) 42.5 (1078)
6 ft. (1.8 m) 78.5 (1993)
9 ft. (2.7 m) 114.5 (2907)
Figure 1-7: Optional In-Situ Filters
A. In-situ stainless steel or Hastelloy filter
B. In-situ high surface area stainless steel filter
Figure 1-8: Optional Panel Mounted Blowback and Calibration/Reference Air Set (19-
in. [482.6 mm] Rack or Wall Mount)
Description and specifications Manual
April2023 00809-0500-4881
18 Emerson.com/Rosemount
1.3  Specifications
Note
All static performance characteristics are with operating variables constant. Specifications
subject to change without notice.
1.3.1  Net O2 range
0-1 percent to 0-40 percent O2, fully field selectable
1.3.2  Combustibles
0-1000 ppm to 0-5 percent, fully field selectable
1.3.3  Accuracy
Oxygen
±0.75 percent of reading or 0.5 percent O2 (whichever is greater)
Combustibles
±50 ppm
1.3.4  System response to test gas
Oxygen
10 seconds T90
Combustibles
25 seconds T90
1.3.5  Temperature limits
Process
32 to 2600 °F (0 to 1427 °C)
Sensors housing
-40 to 212 °F (-40 to 100 °C) ambient
Electronics housing
-40 to 149 °F (-40 to 65 °C), ambient
-40 to 185 °F (-40 to 85 °C), internal operating temperature of electronics inside housing, as
read by HART®, FOUNDATION Fieldbus, or AMS software.
Manual Description and specifications
00809-0500-4881 April2023
Rosemount OCX8800C 19
Local operator interface
-40 to 158 °F (-40 to 70 °C), ambient
At temperatures above 158 °F (70 °C) inside instrument housing, the infrared keypad will
cease to function, but the Rosemount OCX8800 will continue to operate properly.
1.3.6  Nominal and approximate shipping weights
18-in. (457 mm) probe package
54 lb. (20 kg)
3-ft (0.91 m) probe package
55 lb. (20.5 kg)
6-ft (1.83 m) probe package
57 lb. (21 kg)
9-ft (2.74 m) probe package
59 lb. (22 kg)
1.3.7  Mounting
Flange
1.3.8  Materials
Probes
316L stainless steel: 1300 °F (704 °C)
Inconel 600: 1832 °F (1000 °C)
Ceramic: 2600 °F (1427 °C)
Enclosures
Low copper aluminum
1.3.9  Calibration
Semi-automatic or automatic
1.3.10  Calibration gas mixtures recommended (reference test
gas bottles kit #1A9919G04)
0.4 percent O2, balance N2
8 percent O2, balance N2
1000 ppm CO, balance air
Description and specifications Manual
April2023 00809-0500-4881
20 Emerson.com/Rosemount
/