1 Introduction to the Accela UHPLC System
Overview
Thermo Scientific Accela UHPLC System Getting Connected Guide 3
The Accela autosampler draws a sample from a sample vial or well and meters it into an
isolated sample loop attached to the injection valve. The injection valve switches to the inject
position, which places the sample loop in line between the pump and column. Mobile phase
backflushes the sample from the sample loop into the LC column.
As compounds elute from the LC column, they pass through a short length of tubing into the
detector’s flow cell. In the Accela UV/Vis and Accela PDA detectors, each component in turn
absorbs light according to its molar absorptivity and concentration. The detectors measure the
reduction in light intensity due to the absorbance of each component, and the data system
reports the absorbance as a function of wavelength and time.
In the Accela RI Detector, which is a differential refractive index detector, eluting compounds
pass through the sample compartment of the detector’s flow cell. The RI detector measures
the deflection of a light beam caused by the refractive index difference between the contents of
the sample compartment and that of the reference compartment, which contains mobile
phase. The data system reports the signal in refractive index units (RIU) as a function of time.
When the Accela system is connected to a mass spectrometer, the liquid-phase mixture of
analyte components and mobile phase exits the LC column and travels through another
length of 0.005 in. ID, red tubing (supplied in the accessory kit for the MS detector) to the
mass spectrometer. The mass spectrometer’s ion source nebulizes and ionizes the liquid-phase
mixture, its ion optics transmit the ions into the mass analyzer, and the mass analyzer
separates the ions according to their mass-to-charge ratios (m/z). As the ions sequentially exit
the mass analyzer, they impinge upon a conversion dynode/electron multiplier detector,
which sends a signal to the data system computer. The data system reports the ion abundance
as a function of the m/z ratio of the ions.
Note Accela pumps shipped before March 2008 do not have a built-in dynamic mixer.
Call your local Thermo Fisher Scientific field service representative for information on
upgrading these pumps. Adding the dynamic mixer to the Accela Pump ensures mixing of
complex or relatively incompatible solvents and reduces the chromatographic baseline
noise for applications using proportioned mobile phases. The integrated dynamic mixer
also increases the accuracy and reproducibility of complex ternary and quaternary
gradients. The dynamic mixer adds 35 μL of gradient delay volume to the system.
The Accela 600 and Accela 1250 pumps ship with a placeholder module that connects the
flow path between the liquid displacement assembly and the autosampler. The placeholder
module is a block of stainless steel with an internal bore for the solvent flowpath and inlet
and outlet ports. The dynamic mixer is an add-on hardware option that provides solvent
mixing and physically replaces the placeholder module.
CAUTION The pressure rating for the flow cell of the Accela RI Detector is 517 kPA
(75 psi). When you connect two detectors in series, place the refractive index detector at
the end of the solvent path.