4. Operation.
When first switching o, the monitor will display the previously set alarm level.
If 0.0 is displayed the alarm is disabled, and the monitor will then display the oxygen percentage.
If a value other than 0.0 is displayed, the alarm is enabled and the figure displayed is the alarm level.
4.1 Sampling the purge gas
There are two ways of sampling the purge:-
Flowing the exhaust from the job directly through the sample chamber
This is the best way, as you can then easily use the alarm facility to warn of purge loss, but beware, don’t
allow the situation where the tube to the monitor is the only means of escape for the purge gas - if this is
allowed to happen then the butt may become pressurized. Instead, if there is only one exhaust from the job,
put a “T” piece between the job and the monitor to allow any excess gas to vent to atmosphere.
Using the hand-bulb pump and capillary to sample the purge gas
This method is particularly useful for quick spot checks, or if the only access to the purge gas is via an open
butt, ie. tie-in welds. But don’t pump too hard or you may pump faster than the flow rate from the job and pull
in oxygen from the atmosphere, thereby getting an invalid reading.
4.2. The Alarm
The alarm, if enabled, works in the following way:-
1. When the oxygen level of the purge gas drops below the alarm level for the first time, the alarm will
sound continuously as long as the oxygen level remains below the alarm level.
2. Once you’re happy with the purge, press the “ON/OFF” button. This will kill the alarm and switch the
monitor into its second mode of operation. In this mode the alarm will not sound whilst the oxygen level
remains below the alarm level set, the alarm will only sound if the oxygen level rises above the alarm level
set. The monitor will now remain in this mode until it is switched off.
4.2.1. Enabling the alarm and setting the alarm level
1. With the “SET” button held down, press the “ON/OFF” button
The monitor will display “A0”
2. Press the “SET” button until the desired level is reached.
3. Press the “ON/OFF/ button
The monitor will now display the oxygen level and is ready for use.
4.3. Low Battery.
The Oxygen Monitor continuously checks its battery charge level. When the batteries are nearing exhaustion
“LO BAT” will be displayed. You can still use the Oxygen Monitor even though the battery level is getting
low - the Oxygen Monitor will continue to operate for approx. 1 hour after the low battery warning is given.
NB The Oxygen Monitor should not be re-charged before the low-battery
warning is displayed. If it is, battery life will be impaired
4.4. Re-charging.
Ensure that the Oxygen Monitor is switched off - plug the supplied charger into a 240v
outlet and connect to the Oxygen Monitor. A full charge can be achieved overnight.
NB The Oxygen Monitor should not be left on charge for more than 16 hours
or battery life will be impaired.